Irradian DA221-Cos Instruction Manual

DA221A-Cos
DA221B2-Cos
DA221B3-Cos
DA221AB-Cos
Handbook Ref No. IH006/Issue A
File Ref: DA221-v2a.doc


1. Introduction
2. Specification
3. Spectral Response
4. Cosine Angular Response
5. Circuits & Connections
6. Outline Drawings
7. Care and Maintenance
8. Environmental Care, Recycling and Disposal


DA221-Cos Waterproof UV Detector Irradian Limited
1
WATERPROOF UV DETECTORS WITH PRE-AMPLIFIERS
Irradian's DA221-Cos range of data logger compatible ultra-violet
detectors comprise of an ultra-violet sensitive photodiode detector with
filter glasses and a transimpedance amplifier, providing a voltage output.
Each detector is sealed in a waterproof enclosure and operates from a
single rail power supply.
Models in the DA221-Cos series include a UVA detector with peak
response at 365nm, a UVB2 and narrower bandwidth UVB3 detector with
peak responses at 311nm and a broad band UVAB detector with a peak
at 352nm.
Key design features include:
•Good cosine response
•All aluminium body with weather resistant anodising
•High stability photodiodes
•Internal calibration resistor, adjustable for annual recalibration
Four models of sealed UV detectors are available:
UVA: Model DA221A-Cos
UVB: Model DA221B2-Cos
Model DA221B3-Cos
UVAB Model DA221AB-Cos

Irradian Limited DA221-Cos Waterproof UV Detector
2
Models Region Figure λ
peak
FWHM
A-Cos UVA 1 365 ± 2nm 35 ± 2nm
B2-Cos UVB2 2 311 ± 2nm 19 ± 2nm
B3-Cos UVB3 3 311 ± 2nm 14 ± 2nm
AB-Cos UVAB 4 352 ± 5nm 84 ± 5nm
Cosine Corrected Input: Within ±5 % upto 70°. Refer to Figure 5.
Sensitivity*: 10mV per W.m
-2
*Other ranges available on request
.
Calibration: ± 7.5% Traceable to NPL standards.
Linearity: <1% Error
Operating Temperature Range: -10 to +60 °C
Storage Temperature Range: -20 to +60 °C
Temperature Dependence: Sensitivity <0.1% / °C (typical)
Amplifier DC offset Refer to Figure 6.
Input Supply: 7 to 15V DC
Current Consumption: <2mA @12V
Warm Up Period: 10 seconds
Cable: 5 meter screened multicore.
(other lengths available)
Mounting: Two M4 tapped holes on a 43mm PCD.
Housing: Black anodised aluminium, also
available in stainless steel.
Enclosure Rating: Waterproof to 10 meters depth.
Weight: 230g (excluding cable)

DA221-Cos Waterproof UV Detector Irradian Limited
3
DA221A-Cos
Relative Spectral Response
Figure 1
DA221B2-Cos
Relative Spectral Response
Figure 2
0
20
40
60
80
100
320
330
340
350
360
370
380
390
400
410
Wavelength (nm)
R
e
s
p
o
n
s
e
%
0
20
40
60
80
100
270
280
290
300
310
320
330
340
350
360
R
e
s
p
o
n
s
e
%
Wavelength (nm)

Irradian Limited DA221-Cos Waterproof UV Detector
4
(continued):
DA221B3-Cos
Relative Spectral Response
Figure 3
DA221AB-Cos
Relative Spectral Response
Figure 4
0
20
40
60
80
100
270
280
290
300
310
320
330
340
350
360
R
e
s
p
o
n
s
e
%
Wavelength (nm)
0
20
40
60
80
100
240
260
280
300
320
340
360
380
400
R
e
s
p
o
n
s
e
%
Wavelength (nm)

DA221-Cos Waterproof UV Detector Irradian Limited
5
COSINE CORRECTED INPUT
Angular Response: Accurately cosine corrected to Lambert's
Cosine Law.
Maximum error is less than ±5% from true
response to 70°from normal incidence.
Typical DA221-Cos
Cosine Angular Response
Figure 5
True Cosine
60°
30°
DA221 Series

Irradian Limited DA221-Cos Waterproof UV Detector
6
!
! !
!
Amplifier DC Offset
-8
-6
-4
-2
0
2
4
6
8
10
-10 0 10 20 30 40 50 60
O
f
f
s
e
t
µ
V
Temperature °C
Figure 6
DA221-Cos Wiring
Supply (+)
------ Red
Supply (-)
------ Yellow
Output (+)
------ Blue
Output (-)
------ Green
NOTE: Do not link Output (-) to Supply (-) on data logger.
Do not connect cable screen to data logger.
+
-
+V
O/P
7 - 15V
Supply
+5V
-5V
0V
0V
Zero
Gain / Cal.
Reg.
Invertor

DA221-Cos Waterproof UV Detector Irradian Limited
7
"
""
"
DA221-Cos Series Detector
DA221-Cos Series Schematic
Cosine Diffuser
Photodiode
Filter Glasses
Amplifier
50mm
Cable Gland
50mm

Irradian Limited DA221-Cos Waterproof UV Detector
8
#
##
#
LOOKING AFTER THE DETECTOR
1. The DA221-Cos detector should be cleaned using a moist cloth with
detergent, solvent or alcohol.
2. The diffuser on the detector should be kept clean at all times.
3. The detector is a precision instrument, protect it from shocks.
4. Avoid supporting the detector by the cable.

DA221-Cos Waterproof UV Detector Irradian Limited
9
$
$$
$ % &
% & % &
% &
WASTE ELECTRICAL AND ELECTRONIC EQUIPMENT
The purpose of the European Commission WEEE directive (Waste
Electrical and Electronic Equipment; 2002/96/EC) is to ensure that
electrical and electronic products are recycled using the best treatments,
recovery and recycling techniques that are currently available. This is so
that high health standards and a lasting environmental protection can be
achieved and maintained.
This product has been designed and manufactured using high quality
materials and components, many of which can be recycled and reused.
Please remember to observe the local regulations that govern both the
disposal of the packaging materials accompanying this product and any
used batteries.
DO NOT DISPOSE OF THIS PRODUCT IN YOUR
GENERAL WASTE BIN.
Please inform yourself about your local WEEE collection
system which is available for electrical and electronic
products that are marked with the symbol shown here.
When disposing of this meter, please use one of the following options:
1. Use your local designated WEEE collection facilities to dispose of the
complete product (including cables, detectors, filters & accessories).
2. Return the complete product back to Irradian, marking it clearly as
intended for WEEE disposal.

This page left blank

This page left blank

'(
'( '(
'(
9 Elphinstone Road Telephone: +44 (0)1875 898-083
Tranent Facsimile: +44 (0)1875 616-528
Scotland EH33 2LG Web: www.irradian.co.uk
This manual suits for next models
4
Other Irradian Security Sensor manuals
Popular Security Sensor manuals by other brands

Suresafe
Suresafe SH-055DV user manual

S+S Regeltechnik
S+S Regeltechnik HYGRASGARD AFTF-EtherCAT P Operating Instructions, Mounting & Installation

System Sensor
System Sensor B114LP Installation and maintenance instructions

Industrial Scientific
Industrial Scientific iBrid MX6-K1230201 Ordering guide

Benning
Benning VT 1 operating manual

Toyama
Toyama T 76 252 Technical manual