isweek MF5700 User manual

Gas Mass Flow Meter
VC.3
Model MF5700

MF5700UserManual1|Page
Gas Mass Flow Meter
withMEMSthermalsensingtechnology
MF5700Series
User Manual
DocumentNo. 11‐2021‐MF2EN
Issuedate: 2021.11
Revision: VC.3

MF5700UserManual2|Page
Attention!
Usewithcaution!
Pleasecarefullyreadthismanualpriortooperatingthisproduct.
Donotopenormodifyanyhardwarewhichmayleadtoirrecoverable
damage.
Donotusethisproductifyoususpectanymalfunctionsordefection.
Donotusethisproductforcorrosivemediaorinastrongvibration
environment.
Usethisproductaccordingtothespecifiedparameters.
Onlythetrainedorqualifiedpersonnelshallbeallowedtoperform
productservices.
Becautiousfortheelectricalsafety,evenitoperatesatalowvoltage,
anyelectricalshockmightleadtosomeunexpecteddamages.
Thegastobemeasuredshouldbecleanandfreeofparticles.Donotapply
thismeterforliquidmedium.
Donotapplyforanyunknownornon‐specifiedgasesthatmay
damagetheproduct.
ForwirelessdatawithNB‐IoT,pleasebesurethenetworkisavailable.

MF5700UserManual3|Page
TableofContents
1.Overview..................................................................................................5
2.Receipt/unpackoftheproducts................................................................6
3.Knowingtheproducts...............................................................................7
3.1Productdescription...............................................................................................7
3.2LCDdescription.....................................................................................................8
3.3Poweranddatacabledescription...........................................................................9
3.4Mechanicaldimensions..........................................................................................9
4.Installation..............................................................................................11
5.OperationandMENUdescription..............................................................14
5.1Checktheproductspecifications...........................................................................14
5.2Checktheleakage...............................................................................................14
5.3Powerthemeteranddigitaldataconnection.........................................................14
5.4Wirelessdatacommunication...............................................................................15
5.5MeterMENUdescription......................................................................................15
5.5.1Startingthemeasurement............................................................................................16
5.5.2MENUentrywithaverifiedpassword............................................................................16
5.5.3SettheRS485Modbusaddress......................................................................................17
5.5.4SettheRS485communicationbaudrate........................................................................17
5.5.5Resetorcalibratetheoffset..........................................................................................17
5.5.6Gasconversionfactor(GCF)fordifferentgasmeasurement.............................................18
5.5.7SettheResponsetime..................................................................................................18
5.5.8Selectthedisplaymode................................................................................................19
5.5.9Selectthedisplaylanguage...........................................................................................19
5.5.10Open/closethevalve.....................................................................................................19
5.5.11Setanalarm:upperinstantflowratelimit.....................................................................20
5.5.12Setanalarm:lowerinstantflowratelimit.....................................................................20
5.5.13Setanalarm:accumulatedflowrateortotalizerlimit.....................................................21
5.5.14Changethedefaultpassword........................................................................................21

MF5700UserManual4|Page
5.5.15Resettheaccumulatedortotalizedflowrate..................................................................22
5.5.16ExittheMENU..............................................................................................................22
5.5.17Sleepingmode.............................................................................................................22
5.5.18Wakeupthemeter.......................................................................................................22
5.5.19MENUkeysequenceforthesettings..............................................................................23
5.6RS485Modbuscommunicationprotocol...............................................................24
5.6.1Hardwareconnection....................................................................................................24
5.6.2Communicationparameters..........................................................................................24
5.6.3Frame..........................................................................................................................25
5.6.4Functioncodes.............................................................................................................25
5.6.5Registers......................................................................................................................25
6.Productselectionandorderinformation...................................................29
6.1Productselection................................................................................................29
6.2Ordercontactandcustomersupport.....................................................................29
7.Technicalspecifications............................................................................30
8.Technicalnotesfortheproductperformance.............................................31
8.1Measurementprinciple........................................................................................31
8.2Precautionsforthebestperformanceoftheproduct..............................................31
8.2.1Comparisonwithathirdpartyreferencemeter...............................................................31
8.2.2Particlecontaminationandfluidiccleanness..................................................................32
8.2.3Applytoadifferentgasmedium....................................................................................32
9.Troubleshooting......................................................................................33
10.WarrantyandLiability..............................................................................34
11.Servicecontact........................................................................................36
AppendixI:Productevaluationkit..................................................................37
AppendixII:Documenthistory.......................................................................38

MF5700UserManual5|Page
1. Overview
Allcontactinformationcanbefoundattheendofthismanual.
ThismanualprovidesessentialinformationfortheoperationoftheMF5700seriesofgasmassflow
metersgeneral‐purposegasmetrologyapplications.Theproductperformance,maintenance,and
trouble‐shootingaswellastheinformationforproductorder,technicalsupport,andrepairarealso
included.
TheMF5700massflowmeterisoneofthebest‐sellingmassflowmeterproductsonmarket,dueto
itsperformanceandcost‐effectivedesignenabledwiththecompany’sproprietaryMEMS(micro‐
electro‐mechanicalsystems)thermalsensingtechnology.Thecurrentmodelisthe3rdgenerationof
theproductshavingsignificantlyupgradedfeaturessinceitsfirstreleasein2010.Theupgrade
includesmuch‐improvedaccuracy,anadditionalmodeltocoverflowrangeandchannelsize
compatibility;significantlyextendedmeasurementdynamicrange,wirelessdataoptions,andvalve
options.Theupgradealsoensuresback‐compatibilityforourexistingcustomers.Siargois
committedtocontinuingproductinnovationwhileofferingthebestvaluetoourcustomers.
Thisserialofproductsisdesignedforgeneralpurposemassflowsensingandcontrolpurposes.The
wettedmaterialsoftheproductsarecompatiblewithmostofthecommongasessuchasoxygen,
nitrogen,air,argon,carbondioxide,etc.Siargoalsooffersawidespectrumofoff‐the‐shelfand
customizedflow,pressure,temperature/humidity,andgasconcentrationsensingproductsand
providesintegrationandturn‐keysolutionstoourcustomers.Pleasecontactthemanufacturerfor
additionalinformation.

MF5700UserManual6|Page
2. Receipt/unpackoftheproducts
Uponreceiptoftheproducts,pleasecheckthepackingboxbeforethedismantlementofthepacking
materials.Ensurenodamagesduringshipping.Ifanyabnormalityisobserved,pleasecontactand
notifythecarrierwhoshippedtheproductandinformthedistributorsorsalesrepresentativesifthe
orderisnotplaceddirectlywiththemanufacturer,otherwise,themanufacturershouldbeinformed
aswell.Foranyfurtheractions,pleaserefertothereturnandrepairsectioninthismanual.
Ifthepackingboxisintact,proceedtoopenthepackingbox,andyoushallfindtheproduct(either
themeterorthemeterwiththevalvepertheactualorder).Thepoweradapterand/ordatacableas
shownbelowmayalsobefoundaccordingtoyouractualorder.
or
FlowmeterFlowsensor+valveAdapterdatacable
Figure2.1:MF5700flowmeterandaccessories
Pleasecheckimmediatelyfortheintegrityoftheproductaswellasthepoweranddatacable,ifany
abnormalisidentified,pleasenotifythedistributor/salesrepresentativeormanufacturerassoonas
youcan.Ifanydefectsareconfirmed,anexchangeshallbearrangedimmediatelyviatheoriginal
saleschannel.(Note:theLCDscreenshallnotbelighteduntilthebatteryisinstalledorthepower
cableispluggedin).Thisusermanualshallalsoeitherbeincludedinthepackingboxorviaanonline
requestforanelectronicversion.Inmostcases,thismanualshallbemadeavailabletothecustomer
beforetheactualorder.
Themeterisdesignedwithlowpowerthatcanbeoperatedwith4AAbatteries.However,ifyou
orderedtheproductwiththecontrolvalve,itisrecommendedtohaveexternalpowersincethevalve
requireslargepowerconsumption.Apoweradaptermusthaveasafetycertificationforusewiththis
meter.Theexternalpower/dataconnectorisatype‐CUSB.Seethedetailedrequirementsinthe
operationsection.

MF5700UserManual7|Page
3. Knowingtheproducts
3.1 Productdescription
Figure3.1:MF5700partsdescription
Figure3.2:MF5700withvalve
Meterhead
(rotatable)
Meterbody
LCDdisplay
MENUkeys
BSPP
connector
Flow
direction
Battery
chamber
cover
Dataand
powerport
Installation
screws
ON/OFF
solenoidvalve

MF5700UserManual8|Page
3.2 LCDdescription
TheLCDprovidesallinformationthattheproduct
measures.Somesymbolsarereservedforfuture
upgrades,andwillnotbelightedduringthe
operation.Thefollowingtabledetailsthemeaning
ofeachofthesymbols.
Figure3.3:LCDsymbolillustration.
Table3.1:Symbolsdescription
ACC
Thetoprow.ThedefaultdisplaysthetotalizedflowrateinL(liters),
m3(cubicmeters),orSCF(standardcubicfeet).Italsodisplays
pressure(ifkPalightsup)ortemperature(if°Clightsup)*.
FlowThe2ndrow.DisplaysinstantflowrateinL/min(Litersperminute)or
SCFH(standardcubicfeetperhour)orm3/h(cubicmetersperhour).
Batterystatus.
Valvestatus.ThecurrentproductprovidesanON/OFFcontrolvalve.
LoRawirelessprotocolisenabledwhenlightsup.
NB‐IoTwirelessprotocolisenabledwhenlightsup.
WIFIwirelessprotocolisenabledwhenlightsup.
BluetoothLE4.2wirelessprotocolisenabledwhenlightsup.
ReservedforfutureupgradeofGPRSinclusion.
*Note:Boththepressureandtemperaturesensorscanbeintegratedwiththecurrentproduct
butnotwiththedefaultmodels.Pleasecontactthemanufacturerforfurther
information.

MF5700UserManual9|Page
3.3 Poweranddatacabledescription
Table3.2:MF5700wireassignment.
Figure3.4:MF5700cable(USB‐C‐100)
3.4 Mechanicaldimensions
Figure3.5:MF5700meterdimensions
Wire Color Definition
1RedPowersupply(5~24Vdc)
2BlackGND,ground
3GreenRS485A
4WhiteRS485B

MF5700UserManual10|Page
Figure3.6:MF5700meterwithcontrolvalvesdimensions
Table3.3:Mechanicaldimensionsofthemeters/meterswithcontrolvalves
Connector,GL1L2LV1LV2W1W2H1H2H3
MF5706BSPP¼”613495.512334251272812.5
MF5708BSPP3/8”814511013445301343515.5
MF5712BSPP½”905011514345301414219.5

MF5700UserManual11|Page
4. Installation
Donotopenoralteranypartoftheproductwhichwouldleadtomalfunctionandirrecoverable
damage.Itwillalsoforfeitthetermsofthewarrantyandcauseliability.
Theproductatthetimeofshipmentisfullyinspectedforitsqualityandmeetsallsafetyrequirements.
Additionalsafetymeasuresduringtheinstallationshouldbeapplied.Thisincludes,butisnotlimited
totheleakageverificationprocedures,standardEDS(electrostaticdischarge)precautions,DC
voltageprecautions.Othertaskssuchascalibration,partreplacement,repair,andmaintenance
mustonlybeperformedbytrainedpersonnel.Uponrequest,themanufacturerwillprovide
necessarytechnicalsupportand/ortrainingofthepersonnel.
Theproductispreferabletobeinstalledhorizontally.Flowdirectionshouldbealignedwiththearrow
markonthemeterbody.Iftheflowingfluidmayhaveparticlesordebris,afilterisstrongly
recommendedtobeinstalledupstreamofthemeter.
Pleasefollowthefollowingstepstocompletetheinstallation:
Figure4.1:MF5700meterinstallation
a)Uponopeningthepackage,theproduct'sphysicalintegrityshouldbeinspectedtoensurenovisual
damage.
b)Beforeinstallationoftheproduct,pleaseensurethatthepipedebrisorparticlesoranyother
foreignmaterialsarecompletedremoved.
c)Cautionsduringinstallation:
(i)Itispreferablytofirstinstalltheinletendofthemeterandthentheoutletendofthemeter;
Toensurethemeasurementaccuracy,anupstreamstraightpipeoflengthnolessthan10DN
andadownstreamstraightpipeoflengthnolessthan5DNshouldbeinplace.Pleasereferto
thefollowingrecommendedinstallationconfiguration.

MF5700UserManual12|Page
(a)90‐degreeelboworT‐piece(b)2x90‐degreeelbow
Figure4.2:MF5700meterinstallation
(ii)Iftheupstreamordownstreampipesizeisdifferentfromthatoftheproduct,thesizeofthe
installationlinepipediameter(s)shouldbelargerthantheflowchannel(pipe)sizeofthemeters
tobeinstalled.Forsometypicalsituations,pleasefollowtheinstallationrecommendation
detailedinthefollowingsketches.
(c)2x90‐degreeelbow,3D(d)Pipesize‐reduction
(e)Pipesizeexpansion(f)Controlvalveatupstreamordownstream
Figure4.3:MF5700meterinstallation
(iii)Duringinstallation,pleasemakesurenoforeignmaterials(suchaswater,oil,dirt,particles,
etc.)enterintotheinstallationpipeline.

MF5700UserManual13|Page
d)Install4‐AAbatteriesintothebatterychamber.Incasethattheexternalpowersupplyisdesired,
connectelectricalwiresperthewiredefinitioninTable3.2.Pleasebesureofthepowersupply
range(i.e.,5~24VDC)andpowersupplypolarization.Ifanadapterotherthantheonesupplied
bythemanufacturer,makesuretheadaptermeetsindustrialstandardsandhasallsafety
certifications.
e)Forthecommunicationwireconnection,pleasefollowthedescriptioninTable3.2andmakesure
thatthewiresarecorrectlyconnectedtotheproperportsonyourdatadevice/equipment.Please
makesurethedatacablemeetsindustrialstandardswithpropershielding.
f)Oncethebatteryisinstalled,ortheexternalpowerissuccessfullyconnected,theLCDshouldbe
lightedupwiththeproperinformationdisplayedworkscorrectly.
g)Slowlyopenthevalve(s)ifany,upstreamordownstreamorbothofthepipeline,andthemeter
shouldthenstarttomeasuretheflowinthepipeline.
Note:becausethemeterhasalargedynamicalmeasurementrange,itcouldbenormalifyousee
thesmallinstantflowratebeforeyouopenthevalveastherecouldbesomeleakage.
However,makesurethemeterreadsnullwhenthereisnoflowpresentinthepipeline.
h)Thiswillconcludetheinstallation.
Cautions
a)Don'talteranypartsoftheproduct.
b)Ensuretheelectricalconnectionisproperlydonepertheinstructions.
c)Makesurenomechanicalstressesintheconnections.
d)Thestrongelectromagneticinterferencesourcesclosebyoranymechanicalshocksatthe
pipelinemayalsocreatemalfunctioningoftheproduct.
e)Slowlyopen/closevalvestopreventabruptpulseflowimpact.

MF5700UserManual14|Page
5. OperationandMENUdescription
5.1 Checktheproductspecifications
Beforestartingtousethisproduct,checktheproductspecificationsthatcanbefoundinthismanual
orthebasicinformationlocatedonthebackpaneloftheproduct.
ThedetailedproducttechnicalspecificationscanbefoundinSection7.Foraspecificapplication,
thepressureratingmustnotbehigherthanthesystempressuretobemeasured,andtheflowrange
shouldalsobewithinthespecifiedones.Inmostcases,theuseofahighfull‐scalerangedmeterfor
theverylowflowratemeasurementoftenresultsinerroneousdata.Thegastobemeasuredmust
alsobeconsistentwiththatspecifiedbytheproduct.Beparticularlycautiousaboutthesupplied
voltageindicatedinthespecification.Ahighervoltagemayleadtoirrecoverabledamage,anda
lowervoltagewillnotpowertheproductforanydesiredfunctions.
Forthebestperformanceoftheproduct,itisadvisedthatthegastobemeasuredmustbecleanand
freeofparticlesorotherforeignmaterials.
5.2 Checktheleakage
Checkgasleakagebeforeanymeasurement.Ifitisneeded,thepressurizednitrogenoraircanbe
usedfortheleakagecheck.
5.3 Powerthemeteranddigitaldataconnection
AlthoughthisproductcomplieswiththeCE‐requiredEMCregulations,italsorequirestheproductto
beusedaccordingtothestandardelectricaldevicepractice.Forgeneral‐purposeusage,themeter
canbepoweredby4AAbatteries.Besuretoselectthebatterieswithsafetyproofandwithinthe
expirationdate.Donotusebatterieswithoutaknownmanufacturerandexpirationdate.Formodels
withcontrolvalvesorforextensiveusage,itisrecommendedtohavethemeterpoweredwith
externalDCpoweroranAC‐DCadapter.Makesurethesupplyvoltageiswithintherangeofthe
specifiedonesinSection7.Beforeconnectingtheproductwiththeexternalpower,makesurethe
standardelectricaldeviceprecautionssuchasEDS(electrostaticdischarge)andDCvoltageare
observed.Excessiveelectrostaticdischargemaydamagetheproduct.

MF5700UserManual15|Page
ThepoweranddataareconnectedviaastandardType‐CUSBport.Themanufacturer‐supplied
Type‐CUSBcablehasthescrewlockingfixture.Itisrecommendedsuchafeaturetoensurethecable
isproperlyengagedandwillnotbeaccidentallygotunplugged.
Half‐duplexRS485Modbusisusedfordigitaldatacommunication.Makesurethewiresareproperly
connectedtothereceiverside.
5.4 Wirelessdatacommunication
Theproducthasoptionalwirelessdatatransmission,includingBluetoothLE4.2,LoRa,WIFI,andNB‐
IoTwirelessdata.
AnapplicationAPPnamed“SiargoSmartMeter2.0”canbedownloadedfromeitherAppleStorefor
smartdevicesbasedontheiOSsystemorfromGooglePlayStoreforsmartdevicesbasedonthe
Androidsystem.
TheAPPprovidesthebasicfunctionstoaccessthemeter/sensordata,plotaninstantflowratechart,
calculatetotalizedoraccumulatedflowvolumeataspecifictimedefinedbytheuser.Itallowsthe
usertosendthelocaldatainstantlytoaremotedestinationwhenthewiredconnectionisnotan
option.Also,theinterfaceprovidesthereadoutoftheproductinformationaswellastechnical
supportinformation.
ForthedetailedprotocolsofLoRa,WIFI,andNB‐IoT,pleasecontactthemanufacturerforfurther
information.
5.5 MeterMENUdescription
Themeterhasafront3‐keyboardfortheusertosetthedesiredfunctions,accessdata,andcheck
forthestatus.TheMenukey(M)isatthecentralpositionthatallowstheusertoselectafunction
andconfirmationorotherrelatedactionsthatwillbedetailedbelow.Twokeys(“Up”and“Down”)
toselectthemenuandsub‐menu.
U
p
MENUDown

MF5700UserManual16|Page
5.5.1 Startingthemeasurement
Oncethepowerissuppliedandnoabnormalissuesareobserved,
themeterisreadytoperformthemeasurements.Thedefault
displayisforthemassflowmeasurementhavingtwonumerical
linesontheLCD.Theupperlineistheaccumulatedortotalized
flowrate,andthelowerlineistheinstantflowrate.Ifthebattery
symbolturnsinto,proceedtochangethebatteryorplugin
theexternalpowercord.
5.5.2 MENUentrywithaverifiedpassword
Attheflowmeasurement(main)display,pressthecentral“M”MENUkey,itwillenterintothe
passwordsettingandverificationMENU.Thedefaultpasswordis“11111”.Ifthepasswordisincorrect,
thedisplaywillreturntothemaindisplay.
Toenteranewpassword,pressthe“Up”or“Down”keytochangethedigitthatflashes,andpress
the“M”keytoconfirm.Repeatthisprocessforall5digitsandthemeterwillenterintothemenu
settinginterfaces/screen.
Subsequently,theMENUallowstheuserto:
SetModbusaddressF2‐Addr
Changecommunicationbaudrate F3‐bPS
Resetorcalibratetheoffset F11‐oFFST
Enterthegasconversionfactor(GCF) F12‐GCF
Changetheresponsetime F16‐rESPS
Changethedisplaymode F38‐dISP
Changethedisplaylanguage F39‐LAnG
Open/Closethevalve F43‐VALVE
Setupperflowratelimitalarm F51‐ALM‐H
Setlowerflowratelimitalarm F52‐ALM‐L
Settotalizedoraccumulatedflowratealarm F53‐ALM‐A
Changethedefaultpassword F91‐PASS
Clearorresetthetotalizeroraccumulatedflowrate F92‐CLr‐A
ExitfromtheMENUF99‐qUIT
Note:Duringthisprocess,themeterwillcontinuetomeasuretheflowwithoutbeinginterrupted.

MF5700UserManual17|Page
5.5.3 SettheRS485Modbusaddress
AfterthepasswordisverifiedandenteredintotheMENUsettings,
pressthe“Up”or“Down”keyuntilthescreenshowstheF2‐Addras
indicatedtotheleft.
TheModbusaddresshas3digits,whichcanbeanynumberbetween
001to255.Pressthe“M”keytoenterintothechangeaddressscreen.
Pressthe“Up”or“Down”keytochangetheflashingdigits,andthen
pressthe“M”keytoconfirm.Aftertheaddressisset,thedisplaywill
returntoF2‐Addr,whichindicatesthetaskiscompleted.Pressthe
“Up”or“Down”keytoselectF99‐qUITandthe“M”keytoexitthe
MENUandreturntotheMainDisplayscreen.
5.5.4 SettheRS485communicationbaudrate
Followingtheabove‐mentionedsteps,attheMENUsettingscreen,
usethe“Up”or“Down”keytoselectF3‐bPSandthenpressthe
“M”keytosettheRS485communicationbaudrate.
Thereare6baudratesselectable,dependingonyoursystem
requirements:4800,9600,19200,38400,57600,and115200.The
defaultbaudrateis9600.Usethe“Up”or“Down”keytoselectthe
desiredoneandpressthe“M”keytoconfirm.Thedisplaywillthen
returntotheF3‐bPSscreen,whichindicatesthetaskiscompleted.
Usethe“Up”or“Down”keytoselectF99‐qUITandthe“M”keytoexittheMENUandreturntothe
MainDisplayscreen.
5.5.5 Resetorcalibratetheoffset
Afteracertaintimeofusage,themeter’soffset(zeroflowrate)
mightormightnothaveasmallshift.Orwhenyouapplythemeter
fordifferentgases,theoffsetmightbeshifted.Toensure
measurementaccuracy,itisnecessarytoresetorcalibratetheoffset.
Followingtheabove‐mentionedsteps,attheMENUsettingscreen,
usethe“Up”or“Down”keytoselectF11‐oFFST.Beforeperforming
thetask,makesurethereisabsolutelynoflowintheflowchannel,
otherwise,itwillcreateevenbiggererroneousmeasurementresults.

MF5700UserManual18|Page
Pressthe“M”keytoconfirmthetask,anditwillopenthesub‐MENUaskingyoutoconfirm.Usethe
“Up”or“Down”keytoselectthedesiredoneandpressthe“M”keytoconfirm.Thedisplaywillthen
returntotheF11‐oFFSTscreen,whichindicatesthetaskiscompleted.Usethe“Up”or“Down”key
toselectF99‐qUITandthe“M”keytoexittheMENUandreturntotheMainDisplayscreen.
5.5.6 Gasconversionfactor(GCF)fordifferentgasmeasurement
Forthegeneralpurposeoftheapplication,agasconversionfactor
(GCF)canbeappliedtometerthegasdifferentfromthedefaultone
ortheoneusedforcalibration.TheGCFisdeterminedbythe
thermalcalorimetricsensingprincipleaswellasthemeterfluidic
dynamicdesignandthecontrolcircuitry.Contactthemanufacturer
toobtainthevaluescorrespondingtothecorrectmodels.
TheGCFforairis1000.
Note:Ifthemeterduringpurchaseisorderedforaspecialrealgascalibration,contactthe
manufacturerbeforefurtherproceeding.
Followingtheabove‐mentionedsteps,attheMENUsettingscreen,usethe“Up”or“Down”keyto
selectF12‐GCF.Pressthe“M”keytoconfirm,anditwillopenthesub‐MENUshowingthecurrent
gasconversionvalue.Usethe“Up”or“Down”andthe“M”confirmingkeytoinputthedesiredvalue,
andpressthe“M”keyagaintocompletethetask.ThedisplaywillthenreturntotheF12‐GCFscreen,
whichindicatesthetaskiscompleted.Usethe“Up”or“Down”keytoselectF99‐qUITandthe“M”
keytoexittheMENUandreturntotheMainDisplayscreen.
5.5.7 SettheResponsetime
Followingtheabove‐mentionedsteps,attheMENUsettingscreen,
usethe“Up”or“Down”keytoselectF16‐rESPSandthenpress
the“M”keytosettheresponsetime.
Thereare6responsetimeselectable:125,250,500,1000,2000,and
5000,unitismsec.Thedefaultresponsetimeis125msec.Usethe
“Up”or“Down”keytoselectthedesiredoneandpressthe“M”key
toconfirm.ThedisplaywillthenreturntotheF16‐rESPSscreen,
whichindicatesthetaskiscompleted.Usethe“Up”or“Down”key
toselectF99‐qUITandthe“M”keytoexittheMENUandreturntotheMainDisplayscreen.

MF5700UserManual19|Page
5.5.8 Selectthedisplaymode
Thisfunctionistosetthedisplaymodes,cycledisplaymode,or
fixeddisplaymode.Ifonelikestoswitchbetweenthesetwodisplay
modes,followingtheabove‐mentionedsteps,attheMENUsetting
screen,usethe“Up”or“Down”keytoselectF38‐dISP.Pressthe
“M”keytoconfirm,anditwillopenthesub‐MENUshowingthe
currentdisplaymode.Usethe“Up”or“Down”andthe“M”
confirmingkeytoselectthedesiredone,andpressthe“M”key
againtocompletethetask.ThedisplaywillthenreturntotheF38‐
dISPscreen,whichindicatesthetaskiscompleted.Usethe“Up”or“Down”keytoselectF99‐qUIT
andthe“M”keytoexittheMENUandreturntotheMainDisplayscreen.
5.5.9 Selectthedisplaylanguage
Thisfunctionisreservedforfutureupgrades.Thecurrentmodelsare
configuredtobeEnglishorChinesebasedontheshipping
geographicregions.However,ifonelikestoswitchbetweenthese
twodefaultlanguages,followingtheabove‐mentionedsteps,atthe
MENUsettingscreen,usethe“Up”or“Down”keytoselectF39‐
LAnG.Pressthe“M”keytoconfirm,anditwillopenthesub‐MENU
showingthecurrentlanguage.Usethe“Up”or“Down”andthe“M”
confirmingkeytoselectthedesiredone,andpressthe“M”key
againtocompletethetask.ThedisplaywillthenreturntotheF39‐
LAnGscreen,whichindicatesthetaskiscompleted.Usethe“Up”or“Down”keytoselectF99‐qUIT
andthe“M”keytoexittheMENUandreturntotheMainDisplayscreen.
5.5.10 Open/closethevalve
Thisfunctionistoopenorclosethevalve.Ifonelikestoopenorclose
thevalve,followingtheabove‐mentionedsteps,attheMENU
settingscreen,usethe“Up”or“Down”keytoselectF43‐VALVE.
Pressthe“M”keytoconfirm,anditwillopenthesub‐MENUshowing
thecurrentvalvestatus.Usethe“Up”or“Down”andthe“M”
confirmingkeytoselectthedesiredone,andpressthe“M”keyagain
tocompletethetask.ThedisplaywillthenreturntotheF43‐VALVE
screen,whichindicatesthetaskiscompleted.Usethe“Up”or
“Down”keytoselectF99‐qUITandthe“M”keytoexittheMENU
andreturntotheMainDisplayscreen.
Table of contents
Other isweek Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Fluke Calibration
Fluke Calibration RUSKA 7615 user manual

Fluke
Fluke Datapaq Furnace Tracker user manual

BRECOflex
BRECOflex SM5 quick start guide

Bilanciai
Bilanciai B-TEK Scales T503W Quick start manual

Vega Absolute
Vega Absolute VEGA MT X CAN user manual

Water Analytics
Water Analytics AquaMetrix AM-ODO-TX Installation and operation manual