JAVAC D-TEK 3 Operation and maintenance manual

JAV-1283
D-TEK®3
LEAK DETECTOR
Safety Instructions & Operation Manual
EN

JAVAC Table of Contents
074-710-P21-B JAVAC D-TEK 3 Operating Manual 2
Table of Contents
1 EU Declaration of Conformity ......................................................................................................................3
2 UKCA Declaration of Conformity .................................................................................................................4
3 Cautions and Warnings .................................................................................................................................5
4 Specifications................................................................................................................................................. 6
5 D-TEK 3 ...........................................................................................................................................................8
6 Charging the Battery....................................................................................................................................10
7 Turning On the Instrument and Preparing for Use ...................................................................................11
8 Pinpoint Mode...............................................................................................................................................11
9 Manual Zero Mode........................................................................................................................................12
10 Earbuds and Volume Control......................................................................................................................12
11 Removing and Installing the Lithium Ion Battery......................................................................................13
12 Removing and Installing the Sensor ..........................................................................................................14
13 Replacing Filters ..........................................................................................................................................15
14 Extra-long Probe ..........................................................................................................................................16
15 Needle Probe Extension ..............................................................................................................................16
16 Optional Sensors..........................................................................................................................................17
17 Cleaning and Storage ..................................................................................................................................18
18 Replacement Parts and Accessories .........................................................................................................18
19 Troubleshooting Guide................................................................................................................................19
20 Warranty and Liability-Limitation ...............................................................................................................20

1 | EU Declaration of Conformity JAVAC
3 / 21074-710-P21-B JAVAC D-TEK 3 Operating Manual
1 EU Declaration of Conformity
EU DECLARATION OF CONFORMITY
This declaration is issued under the sole responsibility of the manufacturer INFICON. The object of the declaration is
to certify that this equipment, designed and manufactured by:
INFICON Inc.
Two Technology Place
East Syracuse, NY 13057
USA
is in conformity with the relevant Community harmonization legislation. It has been constructed in accordance with
good engineering practice in safety matters in force in the Community and does not endanger the safety of persons,
domestic animals or property when properly installed and maintained and used in applications for which it was made.
D-TEK 3 Refrigerant Leak Detector
721-20x-Gxx
2014/35/EU
2014/30/EU
Equipment Description:
Model Number:
Applicable Directives:
Applicable Standards:
Safety:
2011/65/EU
2006/66/EC
EN 61010-1:2010
(Applicable to all Group numbers)
Low Voltage Directive (LVD)
General Electromagnetic Compatibility (EMC)
as amended by 2015/863/EU RoHS
as amended by 2013/56/EU Battery Directive
Part 1: General Requirements: Safety
requirements for electrical equipment for
measurement, control, and laboratory use.
EN 62133:2013 Safety requirements for portable sealed
secondary cells and for batteries made from
them, for use in portable applications. CB Test
Cert DK-73443-UL
UL 2054 UL Standard for Safety Household and
Commercial Batteries Cert 20180518-MH29443
UL 60950-1 & CAN/CSA-C22.2 No. 60950-1-07
UL Standard for Safety Information Technology
Equipment –Safety –Part 1: General
Requirements Cert 20180518-MH294
UN 38.3 UN Manual of Tests and Criteria, Part III, sub-
section 38.3. Safe Transport of Li-Ion
Rechargeable Battery
Emissions: EN 61326-1:2013 Edition 2.0 (Radiated, Conducted & Harmonic
Emissions) (EMC- Measurement, Control &
Laboratory Equipment)
CISPR 11/EN 55011:2009 (+A1:2010)
Immunity: EN 61326-1:2013
Emission standard for industrial, scientific, and
medical (ISM) radio RF equipment, Class A
Edition 2.0 (EMC – Measurement, Control &
Laboratory Equipment) Immunity per Table A.1 –
Portable Test and Measurement Equipment
RoHS
CE Implementation Date:
Compliant
April 20, 2020
EU Authorized Representative
INFICON GmbH
50968 Köln, Bonner Str. 498
Authorized Representative:
Brian King
INFICON
General Manager – Service Tools
Two Technology Place
East Syracuse, NY USA 13057
ANY QUESTIONS RELATIVE TO THIS DECLARATION OR TO THE SAFETY OF INFICON'S PRODUCTS SHOULD BE DIRECTED, IN WRITING, TO THE AUTHORIZED
REPRESENTATIVE AT THE ABOVE ADDRESS.

JAVAC UKCA Declaration of Conformity | 2
074-710-P21-B JAVAC D-TEK 3 Operating Manual 4 / 21
2 UKCA Declaration of Conformity
UK DECLARATION
OF
CONFORMITY
This declaration is issued under the sole responsibility of the manufacturer INFICON. The object of the declaration is
to certify that this equipment, designed and manufactured by:
INFICON Inc.
Two Technology Place
East Syracuse, NY 13057
USA
is in conformity with the requirements regarding safety, health and relevant provisions of the relevant legislation by
design, type and the versions, which are brought into circulation by us. It has been constructed in accordance with
good engineering practice in safety matters in force in the Community and does not endanger the safety of persons,
domestic animals or property when properly installed and maintained and used in applications for which it was made.
D-TEK 3 Refrigerant Leak DetectorEquipment Description:
Model Number:
Applicable Directives:
Applicable Standards:
Safety:
721-20x-Gxx
S.I. 2016 No. 1101
S.I. 2016 No. 1091
S.I. 2012 No. 3032
S.I. 2008 No. 2164
EN 61010-1:2010
EN 62133:2013
(Applicable to all Group numbers)
LVD
EMC
RoHS
Battery Directive
Part 1: General Requirements: Safety
requirements for electrical equipment for
measurement, control, and laboratory use.
Safety requirements for portable sealed
secondary cells and for batteries made from
them, for use in portable applications. CB Test
Emissions: EN 61326-1:2013 Edition 2.0 (Radiated, Conducted & Harmonic
Emissions) (EMC- Measurement, Control &
Laboratory Equipment)
CISPR 11/EN 55011:2009 (+A1:2010)
Immunity: EN 61326-1:2013
Emission standard for industrial, scientific, and
medical (ISM) radio RF equipment, Class A
Edition 2.0 (EMC –Measurement, Control &
Laboratory Equipment) Immunity per Table A.1 –
Portable Test and Measurement Equipment
RoHS Compliant
UKCA Implementation Date: 27 July 2021
EC Authorized Representative
INFICON GmbH
50968 Köln, Bonner Str. 498
Authorized Representative:
Brian King
INFICON
Business Line Manager –Service Tools
Two Technology Place
East Syracuse, NY USA 13057
ANY QUESTIONS RELATIVE TO THIS DECLARATION OR TO THE SAFETY OF INFICON'S PRODUCTS SHOULD BE DIRECTED, IN WRITING, TO THE AUTHORIZED
REPRESENTATIVE AT THE ABOVE ADDRESS.

3 | Cautions and Warnings JAVAC
5 / 21074-710-P21-B JAVAC D-TEK 3 Operating Manual
3 Cautions and Warnings
Cautions:
• Only use a certified charger/cord with an output of 5 V (dc) ±5%, 1 A ±5%.
• Keep the device out of extremely high or low temperature locations.
• Do not expose the battery to liquid.
• Do not use the device if you notice any damage to the battery.
• Do not disassemble or modify the battery.
• Handle and dispose of the battery per local regulations.
• If the recharging operation fails to complete, even when the specified recharging
time has elapsed, immediately stop further recharging.
• Do not leave the battery unattended while charging.
• Unplug the charger when the battery is fully charged.
• Improper use or disposal of lithium ion batteries can cause a fire.
• High RF environments may cause a false alarm.
WARNING
This symbol alerts the user to the presence of important operating and
maintenance (servicing) instructions.
WARNING
Exposure to high concentrations of CO2 or refrigerants is dangerous and
can be life-threatening.
The instrument is not for use in toxic or hazardous environments. It is not a personal
protection or life-saving device. Always exercise extreme caution in potentially toxic or
hazardous environments.
WARNING
This product is not intrinsically safe and should not be used in the
presence of explosive fumes, explosive dust, or other explosive
chemicals. Use in an environment with flammable refrigerant
concentration approaching the LEL (lower explosive limit) could cause
an explosion or fire resulting in serious injury, death, or damage to
property.

JAVAC Specifications | 4
074-710-P21-B JAVAC D-TEK 3 Operating Manual 6 / 21
4 Specifications
Usage Indoor/outdoor
Sensor type Infrared
Compatible refrigerants
• Refrigerant sensor
(standard, PN 724-701-G1)
All CFCs, HCFCs, HFCs, HFOs, and blends
(including A2Ls)
• CO2 sensor (PN 724-701-G2) R744 (carbon dioxide)
• Flammable refrigerant (HC) sensor (PN
724-701-G3)
R290, R600a, R441a (see safety warnings)
Sensitivity (tested to EN14624 standard)1
• R134a 0.04 oz./yr (1g/yr)
• R1234yf 0.02 oz./yr (0.5g/yr)
• R290 0.2 oz./yr (5g/yr)
• R744 (CO2) 0.07 oz./yr (2g/yr)
• Ammonia 0.1 oz./yr (3g/yr)
Battery type Lithium ion
Charging input type Micro USB
Charging time (starting at 0%) Approximately 3 hours
Battery life Approximately 9 hours
Sensor life Up to 2000 hrs
Input voltage 5 V (dc) ±5%
Input current 1 A ±5%
Warm-up period 45–90 seconds
Temperature ranges and humidity
• Storage -20–60°C (-4–140°F)
• Operating2-20–50°C (-4–122°F)
• Charging 0–45°C (32–113°F)
• Humidity 95% RH NC maximum
Altitude 2000 m (6500 ft.)
Pollution degree 2
Overvoltage category 2
Weight (with battery; not including carrying case
or accessories)
0.97 lb. (0.44 kg)

4 | Specifications JAVAC
074-710-P21-B JAVAC D-TEK 3 Operating Manual
1 To achieve optimal performance and the specified sensitivities, it is recommended to
allow D-TEK 3 to run for 15 minutes prior to use.
2 Use in temperatures below 0°C (32°F) should be limited. Extended warm-up time is
recommended before use in low temperature environments.
Specification table in accordance with EN 14624:2020
R134a R1234yf R744
(CO2)
Ammonia
(R717)
R290
Minimum sensitivity,
fixed (static)
1 g/yr 0.5 g/yr 2 g/yr 3 g/yr 5 g/yr
Minimum sensitivity,
moving (dynamic)
1 g/yr 1 g/yr 2 g/yr 3 g/yr 5 g/yr
Minimum response/
detection time
<1 s <1 s <1 s <1 s <1 s
Recovery time for x g/yr
exposure3
7.6 s 6.4 s 13 s 4 s 4 s
Minimum sensitivity in
contaminated
environment
>2 g/yr 1 g/yr > 4g/yr 3 g/yr 7 g/yr
Calibration frequency Check annually with calibrated leak standard
3 x = R134a: 32 g/yr
R1234yf: 35 g/yr
R744: >50 g/yr
R717: 35 g/yr
R290: 30 g/yr
Patents (Pending) Application #10 2018 208 826.8
7 / 21

JAVAC D-TEK 3 | 5
074-710-P21-B JAVAC D-TEK 3 Operating Manual
5 D-TEK 3
MODE button SENS/ZERO button
leak indicators
sensitivity indicators
power indicator MODE indicator
battery door
eject button
8 / 21

5 | D-TEK 3 JAVAC
074-710-P21-B JAVAC D-TEK 3 Operating Manual
power button micro USB power input
volume button headphone jack
battery door
eject button
9 / 21

JAVAC Charging the Battery | 6
074-710-P21-B JAVAC D-TEK 3 Operating Manual 10 / 21
6 Charging the Battery
D-TEK 3 uses a rechargeable lithium ion battery that comes partially charged. JAVAC
recommends charging the battery before use. Using the supplied charger or charging
cradle accessory, a dead battery can be charged to 80% in approximately 2 hours and
100% in approximately 3 hours. A full charge typically lasts about 9 hours of operation,
depending on the operating temperature. The power indicator indicates the remaining
battery percentage.
Color % Charge
Green >30%
Orange 10–29%
Red <10%
D-TEK 3 can be used while charging.

7 | Turning On the Instrument and Preparing for Use JAVAC
074-710-P21-B JAVAC D-TEK 3 Operating Manual
7 Turning On the Instrument and
Preparing for Use
If the unit does not turn on, the battery is low and needs to be charged. D-TEK 3 can
be used while charging.
1Long press the power button (located on the left side of the body of the
instrument) to turn D-TEK 3 On or Off.
ðD-TEK 3 begins a variable-time warm-up for 45 to 90 seconds. When warm-
up is complete, D-TEK 3 is ready to use.
2To switch modes, press the MODE button. This toggles between Pinpoint and
Manual Zero modes.
D-TEK 3 always starts in the last mode that was used.
8 Pinpoint Mode
1Place the tip of D-TEK 3 as close as possible to the suspected leak (do not
block the air flow).
2Slowly move the probe past each possible leak point.
ðIf a leak is detected, D-TEK 3 alarms and the leak indicators illuminate.
3When a leak is identified, pull the probe away from the leak for a few seconds
and then recheck the spot to verify the leak.
In Pinpoint mode, D-TEK 3 automatically zeros to the background refrigerant
concentration and only alarms again with a higher concentration of refrigerant. When
this occurs, either continue looking for a higher concentration of refrigerant or move
the probe to an area of lower concentration for a few seconds to reset the zero point.
Press the SENS/ZERO button to switch the sensitivity setting. When working with a
large leak, it can be easier to pinpoint the leak location using a lower sensitivity
setting. The current sensitivity is displayed by the sensitivity (SENS) indicator.
11 / 21

9 | Manual Zero Mode
JAVAC
074-710-P21-B JAVAC D-TEK 3 Operating Manual
9 Manual Zero Mode
Manual Zero mode operates similar to Pinpoint mode, but is identified by the MODE
indicator slowly flashing. Manual Zero mode allows the user to manually zero to the
background refrigerant by pressing the SENS/ZERO button. Once the new zero point
is set, D-TEK 3 will not alarm unless a higher concentration of refrigerant is detected.
Manual Zero mode beeps faster at the zero point than other modes. If the
concentration is lower than the current zero point, the beeping slows. This allows the
user to know if they are moving away from the leak by listening to a change in the
beep rate.
There is no sensitivity setting in Manual Zero mode.
Manual Zero mode requires an extra warm-up time of up to 15 minutes for optimal
performance.
10 Earbuds and Volume Control
A headphone jack is located on the right side of D-TEK 3 for listening to the alarm
sounds through headphones or earbuds.
WARNING
If attempting to use headphones not supplied by JAVAC, be sure to test
them carefully to avoid hearing damage.
A volume button is located next to the headphone jack. Press the volume button to
toggle from 100% volume, to 50% volume, and to mute. The volume defaults to 100%
at startup. When headphones or earbuds are plugged in, the volume toggles between
100% volume and mute.
12 / 21

11 | Removing and Installing the Lithium Ion Battery JAVAC
13 / 21 074-710-P21-B JAVAC D-TEK 3 Operating Manual
11 Removing and Installing the Lithium Ion
Battery
1Press the eject button on the back of D-TEK 3 and remove the battery door.
eject button
battery release tab
2Remove the battery by moving the battery release tab to the side until the
battery begins to eject. Slide the battery out.
3Re-insert the battery by first aligning the battery with the rails.
4Gently push the battery along the rails until the battery release tab engages.
ðDo not force the battery. If the battery does not slide freely, check the
alignment and try again.
5Reinstall the battery door.

JAVAC Removing and Installing the Sensor | 12
074-710-P21-B JAVAC D-TEK 3 Operating Manual 14 / 21
12 Removing and Installing the Sensor
D-TEK 3 uses a cartridge style sensor that is quick and easy to remove and replace in
the field. In addition to the standard sensor, JAVAC offers a CO2 specific sensor for
use in refrigeration and air conditioning applications. See Replacement Parts and
Accessories [}23].
eject button
sensor release
tab
To replace the sensor:
1Press the eject button on the back of D-TEK 3 and remove the battery door.
2Grasp the sensor release tab and gently pull it out.
3Align the new sensor with the rails.
4Gently push the sensor along the rails until it is fully inserted.
ðDo not force the sensor. If it does not slide freely, check the alignment and try
again.
5Reinstall the battery door.

13 | Replacing Filters JAVAC
15 / 21 074-710-P21-B JAVAC D-TEK 3 Operating Manual
13 Replacing Filters
D-TEK 3 uses a hydrophobic filter cartridge that allows air and refrigerants to pass
through while filtering out water, dirt, and oil. Examine the white cloth to determine if
the filter needs to be changed. If the cloth appears discolored, install a new filter.
Changing the filter is also an easy troubleshooting step if you suspect your leak
detector is not properly detecting refrigerants. A clogged air filter can limit the sample
air flow.
probe tip
filter
Exposing the filter to water or oil can block air flow. If this occurs, remove the filter with
the unit turned off and the probe facing down to avoid getting contaminants in the
probe and install a new filter. If the filter is wet, it can be reused once it dries.
CAUTION
Never use the instrument without a probe tip and filter.
To replace the filter:
1Unscrew the probe tip and remove the filter.
2Insert the new filter into the probe.
3Screw the probe tip on. Do not overtighten.

JAVAC Extra-long Probe | 14
074-710-P21-B JAVAC D-TEK 3 Operating Manual 16 / 21
14 Extra-long Probe
The optional extra-long replacement probe can be used for leak checking in hard-to-
reach areas.
To install the extra-long probe:
1Unscrew the standard probe from the D-TEK 3 body using a 10 mm wrench.
2Screw on the extra-long probe to approximately 35 in·lb (4 N·m).
Do not overtighten.
3Unscrew the probe tip from the standard probe and remove the filter (or use a
new one).
4Insert the filter into the extra-long probe.
5Screw the probe tip onto the extra-long probe. Do not overtighten.
CAUTION
Always remove the probe with a 10 mm wrench at the base of the probe.
Unscrewing the probe in any way other than the specified method can cause damage
to the probe.
15 Needle Probe Extension
The optional needle probe extension accessory allows for leak checking in tight spots
and insulation. This probe is pointed to easily puncture insulation and fit into small
areas.
To install the needle probe extension:
1Unscrew the probe tip. Leave the filter in place.
2Insert the new filter into the probe tip.
3Screw the probe tip on. Do not overtighten.

JAVAC Optional Sensors | 16
074-710-P21-B JAVAC D-TEK 3 Operating Manual 17 / 21
16 Optional Sensors
Optional sensors are available for use in CO2 (PN 724-701-G2) or flammable
refrigerant (PN 724-701-G3) applications. To use the CO2 or flammable refrigerant
sensor, remove the standard sensor and install the new sensor following the
instructions in Removing and Installing the Sensor [}17]. D-TEK 3 automatically
recognizes the sensor type and the MODE indicator illuminates the appropriate color
to indicate the sensor type the entire time the sensor is installed. Green indicates CO2,
red indicates flammable refrigerant, and orange indicates the standard refrigerant
sensor is installed.
If D-TEK 3 does not automatically recognize your flammable refrigerant sensor,
contact JAVAC to arrange a software update.
The flammable refrigerant sensor does not detect hydrogen, including hydrogen
forming gas.
While searching for CO2 leaks, it is recommended to wear a respirator or mask to
avoid exhaling CO2 toward the probe.
WARNING
Exposure to high concentrations of CO2 or refrigerants is dangerous and
can be life-threatening.
The instrument is not for use in toxic or hazardous environments. It is not a personal
protection or life-saving device. Always exercise extreme caution in potentially toxic or
hazardous environments.
WARNING
This product is not intrinsically safe and should not be used in the
presence of explosive fumes, explosive dust, or other explosive
chemicals. Use in an environment with flammable refrigerant
concentration approaching the LEL (lower explosive limit) could cause
an explosion or fire resulting in serious injury, death, or damage to
property.

17 | Cleaning and Storage JAVAC
18 / 21
074-710-P21-B JAVAC D-TEK 3 Operating Manual
17 Cleaning and Storage
D-TEK 3 can be cleaned with mild detergent or isopropyl alcohol. Care should be
taken to prevent cleaner from entering the instrument. Do not clean with gasoline,
acetone, or other aggressive solvents as they may damage the plastic or display.
18 Replacement Parts and Accessories
Earbuds 721-607-G1
12 V (dc) car charger 721-605-G1
AC wall charger (includes plugs for
multiple regions)
721-606-G1
Lithium ion battery 721-702-G1
Battery charging cradle 721-610-G1
Battery/charging cradle combination 721-604-G1
Standard sensor (detects CFCs, HCFCs,
HFCs, HFOs, and blends (including
A2Ls))
724-701-G1
CO2 sensor 724-701-G2
Flammable refrigerant (HC) sensor 724-701-G3
Filter cartridges (quantity, 5) 712-707-G1
Replacement probe cap 712-705-G1
Needle probe extension 721-612-G1
Extra-long probe 721-611-G1
TEK-Check R134a test leak 703-080-G10

19 | Troubleshooting Guide JAVAC
19 / 21 074-710-P21-B JAVAC D-TEK 3 Operating Manual
19 Troubleshooting Guide
Problem Cause Remedy
The power indicator quickly
flashes red.
A battery error has occurred.
This can be caused by a failed
battery or by the battery being
improperly installed, or having a
poor connection.
Remove and reinstall the
battery. If the problem is not
fixed, replace the battery. Refer
to Removing and Installing the
Lithium Ion Battery [}16].
The battery is above or below
the ideal temperature range and
may not charge properly.
Allow the battery to return to
normal temperature.
All leak indicators flash. A sensor error has occurred.
This can be caused by a failed
sensor, or by the sensor being
improperly installed, or having a
poor connection.
Remove and reinstall the
sensor. If the problem is not
fixed, replace the sensor. Refer
to Removing and Installing the
Sensor [}17].
The indicators do not turn on
after long-pressing the power
button.
The battery level is critically low. Charge the battery or plug the
unit into a charger.
The unit turns on, but does not
detect refrigerant.
The unit has not completed
warm-up.
Wait for the warm-up to
complete. This takes 45 to 90
seconds.
The filter is clogged, restricting
the air flow.
Replace the filter cartridge.
Refer to Replacing Filters [}18].
The pump has failed. Listen for the pump sound. If the
pump is not making a sound and
the battery has a proper charge,
contact JAVAC.
The sensitivity is set too low
(Pinpoint mode only).
Verify the sensitivity level. For
very small leaks, High or Super
should be used.
The incorrect sensor is installed. Verify that the correct sensor is
being used (refrigerant sensor
or CO2 sensor).
The unit alarms in clean air. The incorrect sensor is installed. Verify that the refrigerant sensor
is installed instead of the CO2
sensor.
The pump is not making a
sound.
The pump has failed. If the battery has a proper
charge, contact JAVAC.

JAVAC Warranty and Liability-Limitation | 20
074-710-P21-B JAVAC D-TEK 3 Operating Manual 20 / 21
20 Warranty and Liability-Limitation
JAVAC warrants your D-TEK 3 Refrigerant Leak Detector to be free from defects of
materials or workmanship for one or two years (depending on region) from the date of
purchase. JAVAC does not warrant items that deteriorate under normal use, including
batteries, sensors, and filters. In addition, JAVAC does not warrant any instrument that
has been subjected to misuse, negligence, or accident, or has been repaired or
altered by anyone other than JAVAC. JAVAC liability is limited to instruments returned
to JAVAC, transportation prepaid, not later than thirty (30) days after the warranty
period expires, and which JAVAC judges to have malfunctioned because of defective
materials or workmanship. JAVAC liability is limited to, at its option, repairing or
replacing the defective instrument or part. This warranty is in lieu of all other
warranties, express or implied, whether of MERCHANTABILITY or of FITNESS FOR
A PARTICULAR PURPOSE or otherwise. All such other warranties are expressly
disclaimed. JAVAC shall have no liability in excess of the price paid to JAVAC for the
instrument plus return transportation charges prepaid. JAVAC shall have no liability for
any incidental or consequential damages. All such liabilities are EXCLUDED.
This manual suits for next models
1
Table of contents
Other JAVAC Security Sensor manuals