Kane KANE-EGD User manual

KANE-EGD
Stock No: MAN00344 Rev: 1.01022 OCTOBER 2022
© Kane International Ltd
Leak Detector’s
KANE-RGD
G

CONTENTS Page No.
ANALYSER OVERVIEW 3-4
INSTRUMENT FEATURES AND KEYPAD 5-7
SPECIFICATIONS 8
END OF LIFE DISPOSAL 9
EU DECLARATION OF CONFORMITY 10
SERVICE - CALIBRATE - RECERTIFY 11-14
KANE ASSET MANAGER (KAM)
COLD WEATHER PRECAUTIONS 15
2 KANE-EGD/RGD MANUAL

OVERVIEW
INTRODUCTION
Please read these instructions fully before using the KANE-EGD/RGD.
The KANE-EGD/RGD is powered by two AA batteries. Take care to ensure the
batteries are inserted correctly. When replacing batteries please dispose of used
batteries in accordance with local regulations.
The KANE-EGD/RGD will enable the operator to pin-point small gas leaks, It does
not MEASURE gas levels.
Do not use the KANE-EGD/RGD as a personal safety monitor.
Take extreme care when using the KANE-EGD/RGD, especially when gas is
detected. If in any doubt, open doors and windows to ventilate rooms under test,
ensure people are not at risk and call the Emergency Service Provider.
THE LEAK DETECTOR OPERATION:
Warm up:
Switch the KANE-EGD/RGD ON in fresh air, (press and hold the on/off button).
The tip light will briefly illuminate. The top (red) LED will illuminate and the tic rate
will be once a second.
The LEDs will then cycle for approximately 20 seconds to indicate the warm-up.
When the warm up is complete the bottom (green) LED will be illuminated and the
KANE-EGD/RGD tic rate will be once every 2 seconds.
Normal operation:
The Green battery status LED will remain illuminated if the batteries have sufficient
power to run the KANE-EGD/RGD. If the Green LED is flashing the batteries in the
KANE-EGD/RGD must be replaced.
In fresh air the unit will tic once every two seconds.
If gas is detected the tic rate will increase, the Yellow LED’s will illuminate.
The Red LED’s will illuminate at higher levels of gas and the tic rate will increase to
become almost continuous. Rotate the thumb wheel anti clockwise to reduce the tic
rate, resetting the KANE-EGD/RGD to the new background level.
KANE-EGD/RDG MANUAL 3

Move the KANE-EGD/RGD tip into higher concentrations of gas indicated by the
increased tic rate and illumination of the yellow and red LED’s until the leak is
found.
The tip light is activated by pressing the light/mute button, to turn the light off press
the button once more.
To mute the tic sound press and hold the light/mute button, to reinstate the tic
sound press and hold the mute button again.
Tips:
If the initial tic rate does not settle to once every two seconds, switch off and back
on to allow additional re-setting time.
The KANE-EGD/RGD may be reset at any time by switching off and back on.
Some leak detection fluids can cause a (temporary) false indication.
Use the KANE-EGD/RGD first!
The KANE-EGD/RGD is intended for use in sheltered areas or indoors as severe
draughts may cause (temporary) false indication.
Avoid contamination by water or dirt and do not allow any reactive chemicals etc.
to come into contact with the sensor.
4 KANE-EGD/RGD MANUAL

KANE-EGD
KANE-EGD/RGD MANUAL 5
Power On / Off Torch
and Mute
Torch and Sensing
Element
Display
Sensitivity
Adjustment

KANE-RGD
6 KANE-EGD/RGD MANUAL
Power On / Off
Display
Sensitivity
Adjustment
Torch and
Mute Alarm
Torch and Sensing
Element

KANE-EGD/RGD MANUAL 7
Battery
Apartment
Serial
Number

SPECIFICATIONS
8 KANE-EDG/RDG MANUAL
Power Supply: 2 x AA alkaline batteries
Sensor: Solid state semiconductor
Sensitivity
KANE-EGD <50ppm methane >20ppm H2
KANE-RGD 10g/yr based on R134a
Indicators: Audible: Variable tic rate
Visual: Multi-colored LEDs
Warm-up: 20 seconds
Response Time: <1 second
Duty Cycle: Continuous
Battery Life: >12 hours typical use
Dimensions: Instrument = 180mm long Probe = 430 mm long
Weight: Approx. 250gm
Operational Range: 0-40°C, 10% to 90% RH non-condensing
Gases detected include
KANE-EGD
Acetone, Alcohol, Ammonia, Benzene, Butane, Ethylene,
Gasoline – Petrol, Hydrogen, Industrial Solvents, Jet Fuel,
Lacquer, Thinners, Methane, Naphtha, Natural Gas,
Propane, Toluene
KANE-RGD CFC, HFC, HCFC, R11, R22, R134a, R290, R1234yf

END OF LIFE DISPOSAL
The Waste Electrical or Electronic Equipment (WEEE) Directive requires
countries in the EU to maximise collection and environmentally responsible
processing of these items.
Products are now labelled with a crossed out wheeled bin symbol to remind
you that they can be recycled.
Please Note: Batteries used in this instrument should be disposed of in
accordance with current legislation and local guidelines.
KANE-EDG/RDG MANUAL 9

EU DECLARATON OF CONFORMITY
This declaration of conformity is issued under the sole responsibility of the
manufacturer:-
Kane International Ltd.
Kane House, 11 Bessemer Road, Welwyn Garden City, Hertfordshire, AL10 1GF, UK.
Tel: + 1707 375550 Web: www.kane.co.uk
The KANE-EGD and KANE-RGD are in conformity with the relevant Union harmonization
The following harmonised standards and technical specifications have been applied:
EMC (UK & EU)
EN61326-1:2013, EN61326-2-2:2013
SAFETY (UK & EU)
EN61010-1:2010
ROHS (UK & EU)
IEC62321-2:2013, IEC62321-1:2013, IEC62321-3-1:2013, IEC62321-5:2013,
IEC62321-4:2013, IEC62321-7-2:2017, IEC62321-7-1:2015, IEC62321-6:2015,
IEC 62321-8:2015
Signed for on behalf of:- Kane International Ltd.
01. July 2022
Paul Morrison
Engineering Manager
10 KANE-EGD/RGD MANUAL
UK Directive
The Electromagnetic Compatibility Regulations 2016 (EMC)
The Restriction of the Use of Certain Hazardous Substances in Electrical and
Electronic Equipment Regulations 2012 (RoHS)
Electrical Equipment (Safety) Regulations 2016
EU Directive Title
201430EU Electromagnetic Compatibility (EMC)
201165EU Restriction of the use of certain hazardous substances in
electrical and electronic equipment (EMC)
2014/35 Low Voltage Directive (LVD)

More info at: www.kane.co.uk/service-and-re-calibration
10210118
Use all 8 digits
Welcome to
KANE CARE
www.kane.co.uk www.kanetest.ie
www.kane.eu www.kane-care.com
KANE CARE is our award winning promise to never let you down
or your money back
KANE CARE applies to any KANE Analyser Service & Recertification
booked online via www.kane.co.uk & www.kanetest.ie - some
exclusions apply
For KANE CARE International Service & Recertification
Please register your analyser on our website to download your instruction manual
PLEASE READ ALL SAFETY WARNINGS IN THE MANUAL
Use our website to manage your analyser, buy spares & other
KANE products
Find FAQs on our website or YouTube Channel
Any questions? Call our Customer Service Team 7am - 5pm any
normal weekday 0800 059 0800
KANE-EDG/RDG MANUAL 11

12 KANE-EDG/RDG MANUAL
UK & Ireland
Same Day Despatch
KANE CARE Service & Recertification
*Excludes KANE-EGA, AUTO600, ‘9 series’ analyser & UKAS certificates - Ireland 2 day turnaround
Register your analyser on
www.kane.co.uk or
www.kanetest.ie
Book & pay to Service & Recertify
via your dashboard
Select FREEPOST for tracked
carriage - UK & Ireland mainland only
Your analyser will be despatched
the same day we receive it…
OR YOUR MONEY BACK*

KANE-EDG/RDG MANUAL 13
International
All Inclusive Price
KANE CARE Service & Recertification
*Excludes local tax & duty
email: sales@kane.co.uk with
analyser name & serial number
Pay our all inclusive
KANE CARE quote
Send your analyser & accessories
We will return free of charge* after
KANE CARE Service & Recertification
with 1 year warranty if under 10 years old

14 KANE-EDG/RDG MANUAL
Customer Service
0800 059 0800
+44 1707 375550
10 Year warranty with annual Service & Recertification
20 Year all inclusive pricing
Includes accessory repair or replacement as required including probe,
charger, carry bag & printer
Free tracked return
Full test certification & repair report
KANE CARE International applies to any KANE Analyser Service & Recertification
More info at: www.kane.co.uk/service-and-re-certification
________________________________________________________________________________________________
KANE CARE International Service & Recertification
KANE CARE is our promise to never let you down or your money
10 Year warranty with annual Service & Recertification
Same Day Service & Recertification guaranteed
Free next day tracked delivery to KANE
Free pre 9am next day return & Saturday am
Theft protection - 50% off when replacing your stolen KANE analyser
KANE CARE applies to any KANE FGA Service & Recertification booked online
via www.kane.co.uk & wwwkanetest.ie - some exclusions apply
More info at: www.kane.co.uk/service-and-re-certification
________________________________________________________________________________________________
KANE CARE UK & Ireland Service & Recertification
KANE CARE is our promise to never let you down or your money back
www.kane.co.uk www.kanetest.ie
www.kane.eu www.kane-care.com
Follow us on...
Email: [email protected]
KANE and the Kane logo (registered in the UK and elsewhere) are trademarks of Kane Group Limited, used under licence
All other marks are trademarks of their respective holders. © 2022 Kane International Limited
Trustpilot

COLD WEATHER PRECAUTIONS
It is important you keep your product in a warm place overnight.
Electronic devices that become really cold, by being left in a vehicle
overnight suffer when taken into a warm room the next morning.
Condensation may form which can affect the product’s performance & cause
permanent damage
Electrochemical sensors can be affected by condensation or water.
If you think that your product is affected by condensation or water ingress,
it may be possible to rectify the problem yourself. Simple leave the product
running in a warm place for a few hours (use mains adapter/battery charger
if needed).
If you still experience problems please call our Customer Service team on
0800 059 0800 7am to 5pm any normal weekday.
KANE-EDG/RDG MANUAL 15

THIS PRODUCT CONFIRMS WITH THE FOLLOWING
PLEASE RECYCLE
MADE IN CHINA
16 KANE-EDG/RDG MANUAL

Thank you for buying this analyser.
Before use, please register on our website
www.kane.co.uk
Scan QR code to directly to
register your product online.
Kane International Ltd
Kane House, 11 Bessemer Road
Welwyn Garden City
Hertfordshire
AL7 1GF, UK
email: sales@kane.co.uk
telephone: 0800 059 0800
KANE-EDG/RDG MANUAL 17
This manual suits for next models
1
Table of contents
Popular Security Sensor manuals by other brands

EDWARDS SYSTEMS TECHNOLOGY
EDWARDS SYSTEMS TECHNOLOGY CRC Technical reference manual

DSC
DSC pg9945 installation instructions

LUDLUM
LUDLUM 70 System manual

System Sensor
System Sensor JTWB-BCD-5151EIS Installation and maintenance instructions

Rielta
Rielta FOTON-SH1 installation guide

C.P. Electronics
C.P. Electronics MWS1A-IP installation guide