
7
14
8
9
10
12
13
7.1 Zero point calibration
▸ Hold up the refractometer against a sufficiently bright light source and look through
the eyepiece [5], holding the rubber eyeshade [4]closely against your eye/glasses.
▸ Rotate the eyepiece [5]to adjust it for your eyesight, until you can see the scale
sharply.
▸ Open the prism cover [1].
▸ Thoroughly clean the prism [2]and the underside of the prism cover [1]using a so
cloth or soft paper (with alcohol if necessary)and wipe dry.
▸ Now apply a few drops of the calibration liquid [10]onto the prism surface [2].
▸ Close the prism cover [1]. The amount of fluid should be sufficient to moisten most
of the prism surface. There should not be any air bubbles between the measuring
prism [2]and prism cover [1].
▸ Wait about 30 seconds to allow the temperatures of the fluid and prism to
equalise.
▸ Look through the eyepiece [4]while pointing the refractometer‘s prism surface [2]
at a bright light source.
▸ Through the eyepiece [4], you will see a bright and a blue field. The boundary line
between them shows the measured value on the scale which is also visible
through the eyepiece [4].
▸ Use the supplied adjustment tool [11]to turn the adjustment screw [3]behind
the prism surface [2](under the rubber cap), and adjust the scale so that the
measuring device is optimally set up by moving the boundary line upwards or downwards.
Calibration value ORA 4FA/4FB: 0 °C EG / PG / CW
▸ Repeat step 4 (cleaning).
Important!
The ambient/room temperature and the sample temperature influ-ence
the refractometer measuring result. The scales of the refractometer
models, which have "FB" in its name, are designed for an ambient
7.2 Additional advice
It is important that the samples being measured are representative samples. For
samples that evaporate easily, measurements should be carried out quickly. The
samples should be at the same temperature as the measuring instrument in order
to achieve an accurate result.
9
11 ORA_1SB_2SB_3SB-BA-e-1510
13 Brix to refractive index (nD) conversion table
Data from "ICUMSA" International Commission for Uniform Methods
of Sugar Analysis, at 20 °C and 589 nm wavelength.
8. Cleaning and maintenance
Clean the refractometer using a so, lint-free cloth moistened with either water
or, if necessary, alcohol. Do not use any aggressive or abrasive cleaning agents.
Never immerse the device in water or hold it under running water. Never handle the
device with wet or damp hands.
Never touch the measuring prism [2]with hard tools made from plastic, wood,
rubber, metal, glass etc. Hard objects can quickly damage the relatively so prism
glass, resulting in measurement errors.
The refractometer is maintenance-free.
Cleaning should be carried out immediately before and aer each use of the refrac-
tometer to maximise its life and optimise measurement results.
9. Storage
Store the refractometer in a dry, non-corrosive environment, preferably
between 10 °C and 30 °C.
10. Service
Aer reading this operating manual, if you have any questions about setting up
or using the refractometer, or if any unexpected problem occurs, please contact
your dealer. The device housing may only be opened by trained service technicians
authorised by KERN.
11. Disposal
The packaging consists of environmentally friendly materials which can be disposed
of via local recycling facilities.
The device and storage box should be disposed of by the operator in accordance
with applicable national or regional regulations at the place of use.
12. Additional information
The product may differ slightly from the illustrations. Avoid exposing the refracto-
meter to direct sunlight. Never bring the refractometer into contact with solvents.
!
ORA_1SB_2SB_3SB-BA-e-1510
Table 1: International Temperature Correction Table for°Brix (% sugargradient)
temperature of +20 °C! If the measurements are carried out at a temperature other than +20 °C,
the results must be correspondingly corrected.
Refractometer models, which have "FA" in its name, are equipped with an automatic
temperature compensation (ATC). Between +10°C and +30°C measurement differences due to
temperature changes are automatically compensated.
7.3 Measuring procedure for antifreeze testing
▸ Open the prism cover [1]and use the supplied pipette [8]to apply a few drops of
the sample coolant [8]onto the prism surface [2], then close the prism cover [1].
Spread the liquid evenly by pressing down on the prism cover [1]and eliminate any
air bubbles present.
▸ Close the prism cover [1]and view the measurement scale through the eyepiece
[4]. Point the prism surface [2]of the refractometer at a bright light source while
doing this.
▸ Rotate the adjustment ring [5]on the eyepiece [4]to adjust the focus.
▸ The boundary line will move on the measurement scale depending on the anti-
freeze concentration. This light/dark boundary on the antifreeze scale [B]directly
indicates the freezing point of the antifreeze mixture in °C. The figures show the
measured value for propylene* (G13) and the measured value for ethylene**
(G11/G12). If the antifreeze used in not known or it is a mixture of both, the
propylene scale must always be used. Only then can sufficient coolant antifreeze
protection be reliably guaranteed.
▸ The measurement result is not influenced by the temperature of the coolant.
▸ Carefully clean the supplied pipette [8]and the refractometer aer carrying out the
measurement.
7.4 Measuring procedure for windscreen wiper water testing
▸ The measuring procedure is identical to that for antifreeze testing.
▸ The measured value can be read off the windscreen wiper water testing scale [C]in
°C and provides a guide value, because the composition of the individual antifreeze
agent varies.
▸ Carefully clean the supplied pipette [8]and the refractometer aer carrying out the
measurement.
!
7.5 Measuring procedure for determining battery acid density
▸ Instead of antifreeze solution, place some battery acid on the prism surface [2]
using the pipette [10].
▸ Using the light/dark boundary on the measuring scale for determining battery
acid density [A]„- / +“, read off the acid density in g/cm³ (kg/l).
▸ The scale [A]is divided into three regions for determining the charge condition.
RECHARGE = replace/recharge (<1.20 g/cm³);
Charge ~50%
FAIR = acceptable (1.20 to 1.25 g/cm³);
Charge ~75%
GOOD = good (>1.25 g/cm³);
Charge ~100%
* Propylenes are preferentially used as antifreezes in the European region.
** Ethylenes are preferentially used as antifreezes outside of the European region.
Important!
Aer every measurement, use a lint-free, absorbent cloth to remo-
ve the fluids from the prism surface [2]and prism cover [1]. Then
carefully clean the prism and prism cover using a cloth moistened
with water or if necessary alcohol, and dry both parts using a so,
dry and lint-free cloth. Avoid rubbing the prism [2].
11