
Programming
Page 14 March 2016
Select Rate Interval*
At the RateInt prompt, press ENTER. The monitor flashes the current time interval. If the current selection is correct, press
ENTER to advance to the next parameter. To change to an alternate time interval, press UP or RIGHT to scroll to the required
time interval. Press ENTER to save and advance to the RateUnt parameter.
Select Flow Rate Units*
At the RateUnt prompt, press ENTER. The monitor flashes the current rate unit. If the current selection is correct, press ENTER
to advance to the next parameter. To change to an alternate unit, press UP or RIGHT to scroll to the required rate unit and
press ENTER to save and advance to the TotlUnt parameter.
Select Total Units of Measure*
At the TotlUnt prompt, press ENTER. The monitor flashes the current total units. If the current selection is correct, press ENTER
to advance to the next parameter. To change to an alternate unit, press UP or RIGHT to scroll to the required totalization unit.
Press ENTER to save and advance to the TotlMul parameter.
Select Total Multiplier*
This parameter displays the accumulated flow total in multiples of 10. For example, if the optimum totalization unit is 1000
gallons, the unit total display increments by one digit for every 1000 gallons monitored. In Run mode, at 1000 gallons the total
monitor reads 1, at 3000 gallons, the total display reads 3. This feature eliminates having to look at a total, count the digits,
and mentally insert commas for each 1000 multiple.
At the TotlMul prompt, press ENTER . The monitor shows the current total multiplier. If the selection is correct, press ENTER to
advance to the next parameter. To change to an alternate multiplier, press UP or RIGHT to scroll to the required multiplier unit
and press ENTER to and advance to the next parameter.
OTE:N If the RateUnt or TotlUnt parameter has been set to pounds or kilograms, the monitor advances to the Spec Gr
parameter. At any other setting, the monitor advances to PulsOut in Programming mode. See Totalizer Pulse Output*
on page 16.
Enter Specific Gravity Value*
Mass readings in the BK2900 monitor are not temperature or pressure compensated so it is best to enter the specific gravity
of the fluid as close to the system running temperature as possible. As liquids are essentially incompressible, pressure
compensation is not necessary for liquids.
At the Spec Gr prompt, press ENTER. The most significant digit of the current specific gravity flashes. If the current specific
gravity is correct, press ENTER to advance to the next parameter. To change to an alternate specific gravity, press UP to
increment the flashing digit until you reach the first digit of the new specific gravity. Press RIGHT to move to the next digit.
When all digits have been entered, press ENTER to save and advance to the next parameter.
OTE:N If Gas was chosen as the fluid, see Gas on page 21 and follow the directions for the gas parameters.
In Programming mode, the monitor advances to the PulsOut parameter, see Totalizer Pulse Output* on page 16.
Enter a Scale Factor
The scale factor is used to force a global span change. For example, in Run mode the display is reading a consistent three
percent below the expected values at all flow rates. Rather than changing the K-factor and linearization parameters
individually, the scale factor can be set to 1.03 to correct the readings. The range of scale factors is from 0.10…5.00. The
default scale factor is 1.00.
At the Scale F prompt, press ENTER. The first digit of the existing scale factor flashes. If the current selection is correct, press
ENTER to advance to the next parameter. To change to an alternate scale factor, press UP to increment the display digit until
it matches the first digit of the new scale factor. Press RIGHT to advance to the next digit. Repeat for all digits. Press ENTER to
save and advance to the SetTotl parameter.
OTE:N If the number you enter is out of range, the display flashes Limit and refuses the entry.