Klein Tools ET200 User manual

Rev. 12/13 B139575T Rev. 12/13 B
ENGLISH
• MODERN
SOLID-STATE
DESIGN
• INDICATES VOLTAGE
WITHOUT BATTERY
• LOW IMPEDENCE
• CONTINUITY FUNCTION WITH
LIGHT, VIBRATION AND SOUND
INDICATIORS
600
480
277
240
208
120
48
24
12
ENGLISH
CLEANING
Clean the instrument by using a damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of
time. Do not expose to high temperatures or humidity. After a period of storage
in extreme conditions exceeding the limits mentioned in the Specifications
section, allow the instrument to return to normal operating conditions before
using it.
DISPOSAL /RECYCLE
Caution: This symbol indicates that equipment and its accessories
shall be subject to a separate collection and correct disposal.
480
277
240
208
120
48
24
600
12
480
600
GFCI Testing
Check the GFCI Receptacle user manual for more information.
Always contact a qualified electrician to resolve wiring problems.
Operate the Push Button on the GFCI Receptacle. If the GFCI circuit
does not trip, the receptacle is not operating properly.
To test the functionality of a Ground Fault Circuit Interrupter receptacle,
apply leads to the hot and ground terminals for 7 seconds.
• If the GFCI receptacle is wired properly, power to the circuit
will go out and the ET200 will stop indicating voltage.
• If the GFCI receptacle is wired improperly, power to the circuit
will remain and the ET200 will continue to indicate voltage.
To resolve an issue, contact a qualified electrician.
Battery Replacement
Continuity Testing
Continuity is indicated at less than 270kΩof resistance.
Batteries are not required for visual voltage indication.
1. Remove test leads from voltage source.
2. Unscrew battery cover and replace 3xAAA batteries.
600V
Instruction Manual
ET200
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
ENGLISH
600
480
277
240
208
120
48
24
12
120
48
24
12
600
480
277
240
208
120
48
24
12
120
48
24
12
FEATURE DETAILS
ET200
Instruction Manual
GENERAL SPECIFICATIONS
The Klein Tools ET200 is a solid-state electrical tester. It measures AC/DC
voltage, tests continuity, and features a bright LED worklight.
• Voltage Level Indicators: 12, 24, 48, 120, 208, 240, 277, 480, 600
• Voltage Type Indicators: AC, DC Positive, DC Negative
• AC Frequency: 45Hz to 66Hz
• Maximum Voltage To Earth Ground: 600V RMS
• Maximum Measurable Voltage: 600V RMS (displayed by LEDs)
• Input Impedance: 15kΩon connection, up to 150kΩsteady-state.
• Continuity: < 270kΩ
• Usage Duty Cycle (Above 277V): 1 minute on, 5 minutes off
• Accuracy: 70% to 100% of indicated voltage
• Battery Type: 3 x AAA
• Operating Altitude: 2000 m
• Relative Humidity: 32°F to 86°F (0°C to 30°C) 90%
86°F to 104°F (30°C to 40°C ) 75%
104°F to 131°F (40°C to 55°C) 45%
• Operating Temperature: 32°F to 131°F (0°C to 55°C)
• Storage Temperature: 32°F to 131°F (0°C to 55°C)
• Dimensions: 7.4 x 2.4 x 1.25 in. (187 x 60 x 32 mm)
• Weight: 8.1oz. (230 g)
• Safety: CSA C22.2-1010.1, IEC 61010-1, Pollution Degree II
• Safety Rating: CAT IV 600V, CAT III 1000V
• Drop Rating: 10 ft. (3 m)
WARNINGS
To ensure safe operation and service of the tester, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
• Before each use, verify meter operation by measuring
a known voltage or current.
• Never use the meter on a circuit with voltages that exceed the
category based rating of this meter.
• Do not use the meter during electrical storms, or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Ensure meter leads are fully seated, and keep fingers away from
the metal probe contacts when making measurements.
• Do not open the meter to replace batteries while the probes
are connected to a voltage source.
• Use caution when working with voltages above 60V DC,
or 25V AC RMS. Such voltages pose a shock hazard.
• Shut off and lock out power before testing continuity.
• Always adhere to local and national safety codes. Use individual
protective equipment to prevent shock and arc blast injury where
hazardous live conductors are exposed.
Indicators
Lights
• LEDs light up numerical voltage level indicators,
as well as AC and DC +or DC –indicators.
• Continuity LED will light up on detection.
• Voltage detection will function without batteries installed.
Vibration
• Instrument will vibrate in the presence of voltage
measurements greater than 48V.
• Vibration will function only with batteries installed.
Sound
• Instrument will emit a tone in the presence of voltage
measurements greater than 48V.
• Instrument will emit a tone on detection of continuity.
• Sound will function only with batteries installed.
Accuracy
Accuracy of indicator lights is 70% to 100% of the voltage displayed
to account for normal variations in supply voltage.
Work Light
Press the light bulb button above the Klein Tools logo to toggle the
worklight on and off.
Ghost Voltages
The approximate input resistance of 15kΩwill reduce the possibility of
false readings due to ghost voltages on de-energized lines.
Lead Spacing
Molded lead holders on the back of the ET200 are spaced correctly
to test tamper-resistant outlets.
OPERATING INSTRUCTIONS
SYMBOLS
+
Positive DC Voltage Warning or Caution
–
Negative DC Voltage Dangerous levels
~
AC Voltage Double Insulated Class II
Earth Ground ETL Certication
This product has been tested to the requirements of
CAN/CSA-C22.2 No. 61010-1, second edition, including
Amendment 1, or a later version of the same standard
incorporating the same level of testing requirements
CAT III CAT IV
Designed for transient protection
on indoor measurements at
installations connected directly
to the primary source. Examples
include distribution panels and
lighting systems in large buildings.
Designed for transient protection
on outdoor measurements taken
at the primary supply voltage
source. Examples include outdoor
overhead or underground utility
service and electricity meter.
Measuring Voltages
Time spent operational at voltages above 277V must be less than
60 seconds, with a cool-off time of at least 5 minutes.
Tester will turn on automatically when voltage is applied.
AC Voltage Measurement
DC Voltage Measurement

Other Klein Tools Measuring Instrument manuals

Klein Tools CL900 User manual

Klein Tools

Klein Tools CL900 User manual

Klein Tools CL120 User manual

Klein Tools

Klein Tools CL120 User manual

Klein Tools DTL304 User manual

Klein Tools

Klein Tools DTL304 User manual

Klein Tools RT390 User manual

Klein Tools

Klein Tools RT390 User manual

Klein Tools CL150 User manual

Klein Tools

Klein Tools CL150 User manual

Klein Tools ET140 User manual

Klein Tools

Klein Tools ET140 User manual

Klein Tools ET920 User manual

Klein Tools

Klein Tools ET920 User manual

Klein Tools CL220 User manual

Klein Tools

Klein Tools CL220 User manual

Klein Tools CL120 User manual

Klein Tools

Klein Tools CL120 User manual

Klein Tools ET910 User manual

Klein Tools

Klein Tools ET910 User manual

Klein Tools Probe-Pro User manual

Klein Tools

Klein Tools Probe-Pro User manual

Klein Tools CL320 User manual

Klein Tools

Klein Tools CL320 User manual

Klein Tools ET130 User manual

Klein Tools

Klein Tools ET130 User manual

Klein Tools CL390 User manual

Klein Tools

Klein Tools CL390 User manual

Klein Tools CL210E User manual

Klein Tools

Klein Tools CL210E User manual

Klein Tools CL800 User manual

Klein Tools

Klein Tools CL800 User manual

Klein Tools ET110 User manual

Klein Tools

Klein Tools ET110 User manual

Klein Tools MM1300 User manual

Klein Tools

Klein Tools MM1300 User manual

Klein Tools 93LDM65 User manual

Klein Tools

Klein Tools 93LDM65 User manual

Klein Tools MM5000 User manual

Klein Tools

Klein Tools MM5000 User manual

Klein Tools CL3200 User manual

Klein Tools

Klein Tools CL3200 User manual

Klein Tools CL2200 User manual

Klein Tools

Klein Tools CL2200 User manual

Klein Tools CL330 User manual

Klein Tools

Klein Tools CL330 User manual

Klein Tools MM2000 User manual

Klein Tools

Klein Tools MM2000 User manual

Popular Measuring Instrument manuals by other brands

FISCHER ECO-LINE FT80 operating manual

FISCHER

FISCHER ECO-LINE FT80 operating manual

TIF TIF3710 Operator's manual

TIF

TIF TIF3710 Operator's manual

YOKOGAWA 960 34 user manual

YOKOGAWA

YOKOGAWA 960 34 user manual

Hioki 3126-01 user manual

Hioki

Hioki 3126-01 user manual

Bince LS533H instruction manual

Bince

Bince LS533H instruction manual

Hanna Instruments HI 93751 instruction manual

Hanna Instruments

Hanna Instruments HI 93751 instruction manual

vinmetrica SC-100A user manual

vinmetrica

vinmetrica SC-100A user manual

Tenmars TM-740 user manual

Tenmars

Tenmars TM-740 user manual

L3Harris Narda SignalShark 3330 Quick Start and Safety Instructions

L3Harris

L3Harris Narda SignalShark 3330 Quick Start and Safety Instructions

truflo TK Series manual

truflo

truflo TK Series manual

Textron GREENLEE CMGRT-100A-C instruction manual

Textron

Textron GREENLEE CMGRT-100A-C instruction manual

Thermo Scientific iS5N Series user guide

Thermo Scientific

Thermo Scientific iS5N Series user guide

DeWalt DW080LGS instruction manual

DeWalt

DeWalt DW080LGS instruction manual

Apera Instruments LabSen 241-3SP user manual

Apera Instruments

Apera Instruments LabSen 241-3SP user manual

VOH VISIOTEST user guide

VOH

VOH VISIOTEST user guide

Shinkawa VM-7 Series System manual

Shinkawa

Shinkawa VM-7 Series System manual

ARRI Sound Devices PIX 240i Interconnection and Setup

ARRI

ARRI Sound Devices PIX 240i Interconnection and Setup

Hanna Instruments HI 9126 instruction manual

Hanna Instruments

Hanna Instruments HI 9126 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.