Meter Type
At the METERTYP prompt, press ENTER. The current meter type displays. If the current selection is correct, press ENTER
to advance to the next parameter. To change the type, press UP or RIGHT. Press ENTER to save and advance to the next
parameter. If GAS is selected, the operating pressure (psig) and operating temperature (° F) must be entered. If LIQUID is
selected, the monitor advances to the DAMPING parameter.
OTE:N The 4…20 mA zero and span values are not factory set.
Operating Pressure
At the OP PRES prompt, press ENTER. The first digit of the current setting flashes. If the current setting is correct, press
ENTER to advance to the next paramter. To change the current setting, press RIGHT to move to the correct digit. Press UP to
increment the digit until it matches the new operating pressure. Repeat for all digits. Press ENTER to save and advance to the
OP TEMP parameter.
Operating Temperature
At the OP TEMP prompt, press ENTER. The first digit of the current setting flashes. If the current setting is correct, press
ENTER to advance to the next paramter. To change the current setting, press RIGHT to move to the correct digit. Press UP to
increment the digit until it reaches the new operating temperature. Repeat for all digits. Press ENTER to save and advance to
the DAMPING parameter.
Damping Factor
The damping factor is increased to enhance the stability of the flow readings, and decreased to allow the flow meter to react
faster to changing values of flow. This parameter can be any value 0…99 with 0 being the default.
At the DAMPING prompt, press ENTER. The first digit of the current damping setting flashes. If the current selection is correct,
press ENTER to advance to the next parameter. To change the damping factor, press UP to increment the digit. Press RIGHT
to move to the next digit. After all digits are entered, press ENTER to save and advance to the PULS OUT parameter.
Totalizer Pulse Output
The pulse output parameter can be either enabled or disabled. When enabled, this output generates a 20 mS duration pulse
for every time the least significant digit of the totalizer increments. The amplitude of the pulse is dependent on the voltage
level of the supply connected to the pulse output and is limited to a maximum 30V DC.
Linearization
Linearization enhances the accuracy of the display. The linearization accepts a maximum of ten points and requires additional
calibration data from the meter being used with the monitor. Typically, calibration information can be obtained in three,
five and ten points from the flow meter’s manufacturer. If linearization is not needed, press RIGHT to advance to the next
parameter. To complete linearization, press ENTER at the LINEAR prompt. The meter advances to the NUM PTS parameter.
Number of Points
If the number of points is set to 0, linearization is disabled. Press ENTER. The most significant digit of the number of points
entry begins to flash. The first number can be a 1or a 0only. Press UP to increment the first digit. Press RIGHT to move to the
least significant digit. When both digits have been entered, press ENTER to save and advance to the FREQ 1 parameter.
Frequency
At the FREQ 1 prompt, press ENTER. The first digit of the first linear point’s frequency input flashes. If the first frequency is
correct, press ENTER to advance to the next parameter. To change the frequency, press UP to increment the digit. Press
RIGHT to move to the next digit. After all digits are entered, press ENTER to save and advance to the COEFF 1 parameter.
Coefficient
The coefficient is the value applied to the nominal K-factor to correct it to the exact K-factor for that point. The coefficient is
calculated by dividing the actual K-factor for that point by the average K-factor for the flow meter.
Coecient = Average K-Factor
Actual K-Factor
At the COEFF prompt, press ENTER. The first digit of the coefficient flashes. If the first coefficient is correct, press ENTER
to advance to the next parameter. To change the frequency, press UP to increment the digit. Press RIGHT to move to the
next digit. Repeat for all digits. After all digits are entered, press ENTER to save and advance to the next set of linear points.
Continue entering pairs of frequency and coefficient points until all data has been entered. Press MENU twice at the NUM PTS
prompt to exit to the LINEAR prompt. Press RIGHT to advance to the PASSWORD parameter.
Programming