Konica Minolta AeroDR System User manual

EN 30


3
Contents
Introduction � � � � � � � � � � � � � � � � 5
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
USA only. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Summary of usability specications (for IEC/
EN 60601-1-6, IEC/EN 62366). . . . . . . . . . . . . . 7
Disclaimer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Trademark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Name correspondence table . . . . . . . . . . . . . . . 9
Term description. . . . . . . . . . . . . . . . . . . . . . . . 10
Chapter 1
Chapter1 Safety Precautions &
Warnings� � � � � � � � � � � � � � � � 11
1.1 Symbols relating to safety . . . . . . . . . . . . 12
1.1.1 Safety alert symbol . . . . . . . . . . . . . . 12
1.1.2 Warning notice (signal words) . . . . . . 12
1.1.3 Description of graphic symbols . . . . . 12
1.2 Warning labels. . . . . . . . . . . . . . . . . . . . . 13
1.2.1 DR Detector . . . . . . . . . . . . . . . . . . . . 13
1.2.2 AeroDR Generator Interface Unit2. . . 14
1.2.3 AeroDR Battery Charger . . . . . . . . . . 14
1.2.4 AeroDR Battery Charger2 . . . . . . . . . 14
1.3 Safety precautions. . . . . . . . . . . . . . . . . . 15
1.3.1 Precautions before usage . . . . . . . . . 15
1.3.2 Precautions for usage . . . . . . . . . . . . 15
1.3.3 Precautions regarding
electromagnetic waves . . . . . . . . . . . 20
1.3.4 Precautions regarding wireless
communication. . . . . . . . . . . . . . . . . . 25
1.3.5 Precautions for installing, moving,
and storing . . . . . . . . . . . . . . . . . . . . . 27
1.3.6
Precautions regarding maintenance. . . .
28
1.3.7 Precautions on service life. . . . . . . . . 28
Chapter 2
Chapter2 Product Overview � � � � � � � � 29
2.1 Overview of this device . . . . . . . . . . . . . . 30
2.1.1 Functions . . . . . . . . . . . . . . . . . . . . . . 30
2.1.2 System conguration and
connection examples . . . . . . . . . . . . 31
2.2 Component names and functions . . . . . . 39
2.2.1 AeroDR Detector . . . . . . . . . . . . . . . . 39
2.2.2 AeroDR 2 Detector. . . . . . . . . . . . . . . 40
2.2.3 AeroDR 3 Detector. . . . . . . . . . . . . . . 41
2.2.4 AeroDR Interface Unit . . . . . . . . . . . . 42
2.2.5 AeroDR Interface Unit2 . . . . . . . . . . . 43
2.2.6 Detector Interface Unit. . . . . . . . . . . . 44
2.2.7 Detector Interface Unit 2 . . . . . . . . . . 45
2.2.8 Power Supply Unit . . . . . . . . . . . . . . . 46
2.2.9 AeroDR Generator Interface Unit. . . . 47
2.2.10 AeroDR Generator Interface Unit2. . . 48
2.2.11 Generator Interface Unit 3 . . . . . . . . . 49
2.2.12 AeroDR Battery Charger . . . . . . . . . . 50
2.2.13 AeroDR Battery Charger2 . . . . . . . . . 51
2.2.14 AeroDR Access Point. . . . . . . . . . . . . 52
2.2.15 I/F Cable . . . . . . . . . . . . . . . . . . . . . . 53
2.2.16 AeroDR UF Cable . . . . . . . . . . . . . . . 53
Chapter 3
Chapter3 General Operations � � � � � � � 55
3.1 Startup and shutdown . . . . . . . . . . . . . . . 56
3.1.1 Startup of this device . . . . . . . . . . . . . 56
3.1.2 Shutdown of this device. . . . . . . . . . . 59
3.2 Operation of DR Detector . . . . . . . . . . . . 63
3.2.1 Exposure with basic connection or
Aero Sync connection . . . . . . . . . . . . 63
3.2.2 Exposure under the S-SRM
connection example . . . . . . . . . . . . . 63
3.2.3 Exposure using Aero Storage . . . . . . 64
3.2.4 Serial exposure . . . . . . . . . . . . . . . . . 67
3.2.5 DR Detector orientation . . . . . . . . . . . 67
3.2.6 Precautions for exposure. . . . . . . . . . 68
3.3 Charging of DR Detector . . . . . . . . . . . . . 74
3.3.1
Charging with AeroDR Battery Charger . .
74
3.3.2 Charging with AeroDR Battery
Charger2 . . . . . . . . . . . . . . . . . . . . . . 75
3.3.3 Charging with the wired cable . . . . . . 76
3.3.4 Charging time guide. . . . . . . . . . . . . . 77
3.3.5
DR Detector battery level display . . . . . .
78
3.4 Registration and selection of the DR
Detector. . . . . . . . . . . . . . . . . . . . . . . . . . 79
3.4.1 Registration with AeroDR Battery
Charger . . . . . . . . . . . . . . . . . . . . . . . 79
3.4.2 Registration with AeroDR Battery
Charger2 . . . . . . . . . . . . . . . . . . . . . . 80
3.4.3
Registration with the wired cable . . . . . .
80
3.4.4 Selection of the DR Detector . . . . . . . 81
3.4.5
Manual selection of the DR Detector . . . .
81
3.5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . 82
3.6
Position to ax DR Detector identication
label and AeroDR Grip sheet . . . . . . . . . . .
83
3.6.1 Position to ax DR Detector
identication label . . . . . . . . . . . . . . . 83
3.6.2 AeroDR Grip sheet. . . . . . . . . . . . . . . 83

4
Contents
Chapter 4
Chapter4 Status (LED) Display � � � � � � 85
4.1 LED display of respective devices. . . . . . 86
4.1.1 AeroDR Detector . . . . . . . . . . . . . . . . 87
4.1.2 AeroDR 2 Detector. . . . . . . . . . . . . . . 88
4.1.3 AeroDR 3 Detector. . . . . . . . . . . . . . . 89
4.1.4 AeroDR Interface Unit . . . . . . . . . . . . 92
4.1.5 AeroDR Interface Unit2 . . . . . . . . . . . 93
4.1.6 Detector Interface Unit/Detector
Interface Unit 2. . . . . . . . . . . . . . . . . . 94
4.1.7 Power Supply Unit . . . . . . . . . . . . . . . 94
4.1.8 AeroDR Generator Interface Unit. . . . 95
4.1.9 AeroDR Generator Interface Unit2. . . 95
4.1.10 Generator Interface Unit 3 . . . . . . . . . 96
4.1.11 AeroDR Battery Charger . . . . . . . . . . 96
4.1.12 AeroDR Battery Charger2 . . . . . . . . . 97
4.1.13 AeroDR Access Point. . . . . . . . . . . . . 98
Chapter 5
Chapter5 Troubleshooting� � � � � � � � � � 99
5.1 Support ow during trouble . . . . . . . . . . 100
5.2
Various problems and countermeasures . . . .
102
5.2.1 DR Detector . . . . . . . . . . . . . . . . . . . 102
5.2.2 AeroDR Interface Unit . . . . . . . . . . . 106
5.2.3 AeroDR Interface Unit2 . . . . . . . . . . 107
5.2.4 Detector Interface Unit. . . . . . . . . . . 107
5.2.5 Detector Interface Unit 2 . . . . . . . . . 108
5.2.6 Power Supply Unit . . . . . . . . . . . . . . 108
5.2.7 AeroDR Generator Interface Unit. . . 108
5.2.8 AeroDR Generator Interface Unit2. . 109
5.2.9 Generator Interface Unit 3 . . . . . . . . 109
5.2.10 AeroDR Battery Charger . . . . . . . . . 109
5.2.11 AeroDR Battery Charger2 . . . . . . . . 110
5.2.12 AeroDR Access Point. . . . . . . . . . . . 110
5.2.13
Image processing controller/Images . .
110
Chapter 6
Chapter6 Maintenance � � � � � � � � � � � � 111
6.1 Maintenance and inspection items . . . . 112
6.1.1 Maintenance schedule. . . . . . . . . . . 112
6.1.2 Cleaning. . . . . . . . . . . . . . . . . . . . . . 112
6.1.3 Disinfection of the DR Detector . . . . 115
6.1.4 Consumables. . . . . . . . . . . . . . . . . . 115
6.1.5 AeroDR Grip sheet axing and
replacement . . . . . . . . . . . . . . . . . . . 115
Chapter 7
Chapter7 Specications � � � � � � � � � � 117
7.1 Specications . . . . . . . . . . . . . . . . . . . . 118
7.1.1 DR Detector . . . . . . . . . . . . . . . . . . . 118
7.1.2 AeroDR Interface Unit . . . . . . . . . . . 122
7.1.3 AeroDR Interface Unit2 . . . . . . . . . . 122
7.1.4 Detector Interface Unit. . . . . . . . . . . 123
7.1.5 Detector Interface Unit 2 . . . . . . . . . 123
7.1.6 Power Supply Unit . . . . . . . . . . . . . . 124
7.1.7 AeroDR Generator Interface Unit. . . 124
7.1.8 AeroDR Generator Interface Unit2. . 125
7.1.9 Generator Interface Unit 3 . . . . . . . . 125
7.1.10 AeroDR Battery Charger . . . . . . . . . 126
7.1.11 AeroDR Battery Charger2 . . . . . . . . 126
7.1.12 AeroDR Access Point. . . . . . . . . . . . 127
7.1.13 I/F Cable . . . . . . . . . . . . . . . . . . . . . 127
7.1.14 AeroDR UF Cable . . . . . . . . . . . . . . 128
7.1.15 Cables and minor components . . . . 129
7.1.16 AeroDR SYSTEM . . . . . . . . . . . . . . 130
7.1.17 AeroDR SYSTEM 2 . . . . . . . . . . . . . 130
7.1.18 SKR 3000. . . . . . . . . . . . . . . . . . . . . 131
7.1.19 Label . . . . . . . . . . . . . . . . . . . . . . . . 132

5
Introduction
Introduction

6
Introduction
The digital radiography AeroDR SYSTEM, AeroDR SYSTEM 2 and SKR 3000 (hereinafter, both are referred to as
this device) perform X-ray imaging of the human body using an X-ray planar detector that outputs a digital signal,
which is then input into an image processing device, and the acquired image is then transmitted to a ling system,
printer, and image display device as diagnostic image data.
• Diagnostic X-ray image data of this device does not provide mammographic images.
• This device is also used for carrying out exposures on children.
The CS-7 or ImagePilot (hereafter referred to as the image processing controller), which controls the receiving, pro-
cessing, and output of image data, is required for operation. For the operation of the image processing controller,
refer to the "Operation Manual" of the image processing controller. Also, regarding the connectable devices, contact
Konica Minolta technical representative.
This operation manual provides instructions on the basic functions for operation of this device. Those operating this
device for the rst time should read this manual beforehand.
Also, store this manual close to this device after reading it through, so it can be used as a guide to allow optimum
operating conditions.
IMPORTANT
• This operation manual (English version) is eective only outside Japan.
• This Operation Manual is common to three types of the digital radiography AeroDR SYSTEM, digital radiography AeroDR SYS-
TEM 2 and digital radiography SKR 3000.
• If the pages of the operation manual are smudged and illegible, replace it with a new one. (a fee is required.)
• The illustrations in this manual use the AeroDR 3 1417HD and AeroDR Battery Charger2 as the example.

7
Introduction
USA only
Indications for Use
This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace
radiographic lm/screen system in general-purpose diagnostic procedures.
This device is not indicated for use in mammography, uoroscopy, tomography and angiography applications.
CAUTION
• Federal law restricts this device to sale by or on the order of a physician.
Summaryofusabilityspecications(forIEC/EN60601-1-6,IEC/EN62366)
(1) Medical purposes
• Provision and reading of disease and injury diagnostic images.
(2) Patient groups
• No patient population exists who uses the device.
• Patient population for the X-ray images read is not specied.
(3) Parts of body or organizations to which the device is mounted or that interact with the device.
• DR Detector contacts the body surfaces of a patient and an operator.
• Devices other than DR Detector contact the body surface of an operator.
(4) Operating principle
• DR Detector forms the still images according to the X-ray energy passing through the human and animal
body; after digitizing the exposed image, it is transmitted to the console (the image processing controller) with
wired communication or wireless communication.
• In the serial radiography mode, it allows to capture multiple frames of radiography image serially taken by the
pulse irradiation from an X-ray generator device during one time. This device continues not to be intended to
connect with the x-ray device for mammography, uoroscopy, angiography and cone-beam CT, and not to be
used for such procedures.
• AeroDR Interface Unit and AeroDR Interface Unit2 supply the power to DR Detector, AeroDR Generator In-
terface Unit, AeroDR Generator Interface Unit2, and access point (radio communication device). It also relays
wired communication.
• Detector Interface Unit and Detector Interface Unit 2 supply the power to DR Detector. It also relays wired
communication.
• Power Supply Unit supply the power to Detector Interface Unit, Detector Interface Unit 2, AeroDR Generator
Interface Unit, AeroDR Generator Interface Unit2 and Generator Interface Unit 3. It also relays wired commu-
nication.
• AeroDR Interface Unit2, AeroDR Generator Interface Unit, AeroDR Generator Interface Unit2 and Generator
Interface Unit 3 interfaces with an X-ray device.
• Access point (radio communication device) performs a wireless communication with DR Detector.
• The console (the image processing controller) processes the image data into the diagnostic image, and then
stores and outputs the images added with the patient information.
• The AeroDR Battery Charger and AeroDR Battery Charger2 charge the DR Detector. It registers the using DR
Detector with the exposure room.
(5) Signicant physical characteristics
• Refer to "7.1 Specications".
(6) Signicant performance characteristics
• Refer to "2.1 Overview of this device".
(7) User of this device
• No special training is required to use this device. The intended users of this device are as follows.
A professional in good health with specialist knowledge/qualications who has fully understood the content of
this document. (Such as a doctor or radiological technologist)

8
Introduction
Disclaimer
(1) This manual may not be reproduced in whole or in part without the permission of Konica Minolta, Inc.
(2) The contents of this manual may be subject to change without prior notice.
(3) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by installation,
relocation, modication, maintenance, and repair made by anyone except Konica Minolta and contractors desig-
nated by Konica Minolta.
(4) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage to Konica Minolta prod-
ucts, caused by third-party products not installed by Konica Minolta.
(5) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by maintenance
and repair using maintenance parts other than those specied by Konica Minolta.
(6) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by not observing
the precautions and operation methods described in the operation manual.
(7) Konica Minolta, Inc. is not responsible for any claims made for malfunction or damage caused by the environ-
ment that is not suitable for the installation requirements such as power source and installation environment de-
scribed in the Installation Requirement or operation manual.
(8) Konica Minolta, Inc. is not responsible for any claims for malfunction or damage caused by acts of nature such
as res, earthquakes, oods, or lightning strikes.
(9) Konica Minolta, Inc. is not responsible for any claims for malfunction or damage caused by using this device for
any purpose other than that specied for this device.
(10) Diagnostic and treatment action is performed under the responsibility of the physician(s). Konica Minolta, Inc. is
not responsible for any diagnostic/treatment conditions or diagnostic/treatment results.
(11) Konica Minolta, Inc. is unable to assume any liability relating to the following events arising due to systems in-
fected by computer viruses, worms and other malicious software potentially harmful to computers (hereunder
referred to as "malwares"):
— The erasure, corruption, or unwanted disclosure of data stored within this device, including personal and
sensitive patient information.
— Malfunction of the device or accidents occurring as a result of the malfunction of the device.
— Infection of other products upon connection with the device, or damages arising therefrom.
(12) Whenever you use this service, you are required to take some security or safety measures for the following
items related to the service:
Please understand that Konica Minolta, Inc. is unable to assume liability for certain events, such as malfunctions
or incidents of information leakage resulting from such events.
— Client and network equipment managed by our facility
— The network and ISP in our facility used for connection with this service
— Your equipment to be connected to the same network as that of this service
(13) When this device is used, various user and patient data stored within the device and external storage devices
must be adequately managed under the full responsibility of the user. Konica Minolta, Inc. is unable to assume
any liability relating to damages arising due to leakage of various data at the time of disposal.
Trademark
Company names and product names in this manual are trademarks or registered trademarks of their respective own-
ers.
Please note that ©, ® and ™ marks are omitted hereafter.
Copyright © 2011 - 2020 Konica Minolta, Inc. All Rights Reserved.

9
Introduction
Name correspondence table
Operation Manual name Model name (nameplate name)
DR Detector
AeroDR Detector
AeroDR 1417HQ AeroDR P-11
AeroDR 1417S AeroDR P-12
AeroDR 1717HQ AeroDR P-21
AeroDR 1012HQ AeroDR P-31
AeroDR 2 Detector AeroDR 2 1417HQ AeroDR P-51
AeroDR 2 1417S AeroDR P-52
AeroDR 3 Detector
AeroDR 3 1417HD P-61
AeroDR 3 1717HD P-71
AeroDR 3 1012HQ P-81
AeroDR 3 1417HD2 P-65
AeroDR 3 1717HD2 P-75
AeroDR Interface Unit AeroDR B-1
AeroDR Interface Unit2 AeroDR Interface Unit2
Detector Interface Unit G-21
Detector Interface Unit 2 G-52
Power Supply Unit G-11
AeroDR Generator Interface Unit AeroDR X-1
AeroDR Generator Interface Unit2 AeroDR Generator Interface Unit2
Generator Interface Unit 3 G-51
AeroDR Battery Charger AeroDR D-1
AeroDR Battery Charger2 AeroDR Battery Charger2
AeroDR Access Point AeroDR C-1

10
Introduction
Term description
Terms Explanation
DR Detector
Collective term indicating AeroDR 1417HQ, AeroDR 1417S, AeroDR 1717HQ,
AeroDR 1012HQ, AeroDR 2 1417HQ, AeroDR 2 1417S, AeroDR 3 1417HD,
AeroDR 3 1717HD, AeroDR 3 1012HQ, AeroDR 3 1417HD2, and AeroDR 3 1717HD2.
AeroDR Detector Collective term indicating AeroDR 1417HQ, AeroDR 1417S, AeroDR 1717HQ,
AeroDR 1012HQ.
AeroDR 2 Detector Collective term indicating AeroDR 2 1417HQ, AeroDR 2 1417S.
AeroDR 3 Detector Collective term indicating AeroDR 3 1417HD, AeroDR 3 1717HD, AeroDR 3 1012HQ,
AeroDR 3 1417HD2, and AeroDR 3 1717HD2.
Image processing controller The CS-7 or ImagePilot is referred to as the image processing controller.
Calibration Processing for calibrating the characteristics of the DR Detector for each pixel.
PoE An abbreviation for Power over Ethernet. Provides power using an Ethernet cable.
Aero Sync This is a mode in which exposure is performed without being synchronized with the X-ray
device.
Aero Storage This function allows you to expose without using the image processing controller.
Aero Indicator When the Aero Storage mode is used, this function allows you to view information in the DR
Detector on a mobile terminal.
Mobile terminal This is an external device to use for Aero Indicator.
Access Point Collective term indicating the AeroDR Access Point and general-purpose access points.
The AeroDR Access Point 2 is included.
I/F Cable Collective term indicating the AeroDR I/F Cable, AeroDR I/F Cable2, I/F Cable3(0.67mU, 1mU,
8mD, 8mU ), and I/F Cable4(1.5mD, 8mD, 8mU).
Wired cable Collective term indicating I/F Cable and AeroDR UF Cable.
Serial exposure An exposure method to capture multiple frame images serially taken by the pulse irradiation
from an X-ray generator device.

11
Chapter 1
Chapter 1 Safety Precautions & Warnings
Chapter 1
Safety Precautions & Warnings

12
1�1 Symbols relating to safety
1�1�1 Safety alert symbol
This is a "safety alert symbol". This symbol alerts you
to matters and/or operation potentially hazardous to
yourself and other people. Read these messages and
follow the instructions carefully.
1�1�2 Warning notice (signal words)
Signal words indicate the degree of potential hazards
in the use of the product.
Signal words include the following three types, which
are used according to risk of damage caused by dan-
ger and the severity of damage.
DANGER
Indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if
not avoided, could result in minor or moderate injury.
It may also be used to indicate hazardous situation
where only physical damage is likely to occur.
1�1�3 Description of graphic symbols
Graphic
symbols
Explanation
Indicates the on or standby position for the
power switch of this device.
Indicates the o position for the main power
switch of this device.
Indicates the on position for the main power
switch of this device.
Indicates the position where the ground is at-
tached.
Indicates a B-type Applied Parts medical device.
Indicates that full caution is required when han-
dling this device.
Indicates that it is necessary to read the User's
Manual before use or operation of this device.
Indicates devices including radio frequency
transmitters.
Indicates that the level of water ingress protec-
tion is equivalent to IPX6.
0197
This CE mark on this product indicates that
this product is in conformity with the applicable
requirements set out in the Directive 93/42/
EEC (Medical Device Directive) and in Directive
2011/65/EU (RoHS Directive).
0197 indicates the identication number of the
notied body responsible only for implementa-
tion of the Directive 93/42/EEC (Medical Device
Directive).
EC Directive 93/42/EEC does not cover animal
use. So, the notied body whose identication
number is 0197 is not responsible for animal
use.
It means conformity with the Radio Equipment
Directive 2014/53/EU.

1
2
3
4
5
6
7
8
9
10
11
12
13
1�2 Warning labels
Various warning labels are attached to this device on the locations shown below. Do not remove these labels from
this device. Warning labels are there to make sure that the user recognizes potential hazards when operating this
device.
IMPORTANT
• If a warning label is too dirty or damaged to read, contact Konica Minolta technical representatives to have a new warning label
attached, and redisplay by parts replacement. (a fee is required.)
1�2�1 DR Detector
IPX6 IPX6

14
Warning labels
1�2�2 AeroDR Generator Interface Unit2
1�2�3 AeroDR Battery Charger
1�2�4 AeroDR Battery Charger2

15
1
2
3
4
5
6
7
8
9
10
11
12
1�3 Safety precautions
Read all safety precautions thoroughly before using
this device.
Be sure to observe the safety precautions described in
this section.
1�3�1 Precautions before usage
CAUTION
• The operators (hospitals and clinics) hold responsibility
for the usage and maintenance of this device. Do not
use this device unless you are a physician or certied
person under law.
• This device excluding the image processing controller is
suitable for use in the patient environment. (PC used for
the image processing controller is not suitable for use in
the patient environment.)
• Confirm that this device is operating normally before
using.
• When a problem occurs with this device, turn the power
o, attach an appropriate sign, such as "malfunction",
on this device, and contact Konica Minolta technical
representatives.
• This device is not explosion-proof, so do not use any
ammable or explosive gas near this device.
• For the basic operation of the computer, display moni-
tor, and optional parts for this device, refer to their op-
eration manuals.
Please follow the rules and regulations of your relevant
authorities in the disposal of this product, accessories,
options, consumables, media and their packing materi-
als.
This symbol means: Do not dispose of this
product together with your household waste!
Please refer to the information of your local
community or contact our dealers regarding
the proper handling of end-of-life electric and
electronic equipments.
Recycling of this product will help to conserve
natural resources and prevent potential negative
consequences for the environment and human
health caused by inappropriate waste handling.
1�3�2 Precautions for usage
WARNING
• The DR Detector is a precision instrument. If it is
dropped or hit against any object, a failure may occur
due to strong impact or excessive load.
• Take note of the following when using this device:
— Do not subject the DR Detector to strong shocks or
excessive loads by dropping it, etc.
— Do not exceed the specied overall load range when
applying a load to an DR Detector.
— Do not disassemble or modify this device.
— Do not connect any devices that were not purchased
from Konica Minolta.
—
Do not turn the power switch o or pull out the power cable,
Ethernet cable while the system is operating.
—
Be careful not to drop the DR Detector on any part of a per-
son's body by catching the wired cable on your feet.
— Do not lean on or put pressure on the AeroDR Bat-
tery Charger installed on a wall.
•
The wired cable is connected to the DR Detector using mag-
netic force. When moving the DR Detector, do not hold onto the
wired cable, and always hold on the DR Detector. Also, do not
grasp and pull the DR Detector forcefully.

16
Safety precautions
WARNING
• If there is any smoke, odor, or abnormal sound, it may
cause a re if use is continued, so immediately turn the
power switch o, unplug the power plug from the wall
outlet, and contact Konica Minolta technical representa-
tives.
• Take note of the following to reduce the risk of re, elec-
tric shock, or electrical leakage:
— Use specied cables for the power cable, etc and AC
adapter.
— Use a wall outlet with the correct rating as a power
source.
— Conrm that the power plug is properly connected to
the wall outlet without any slack.
— Use a grounded power source.
— If you do not plan to use this device for an extended
period of time, unplug the power plug.
— The supplied power cable and AC adapter are dedi-
cated for this device, so do not use it elsewhere.
— Avoid exposure to liquid such as water.
— Make sure that foreign material, such as pieces of
metal or wire, does not get inside.
— Do not allow any metal or conductive objects to come
into contact with the spring connector of the AeroDR
Battery Charger2 or wired cable.
— Do not handle the power plug with wet hands.
— Do not let soil or dust accumulate on the power plug,
AeroDR Battery Charger2 or wired cable.
— Do not use extension cords.
— Do not connect many plugs to a single electrical out-
let.
— Do not damage the power cable, AC adapter or wired
cable. Also, do not use damaged cables.
— Do not block the ventilation openings.
— Do not apply a load on the power cable.
• If there is any abnormality in appearance such as defor-
mation of the housing or a crack, stop using the device
immediately and contact Konica Minolta technical rep-
resentatives.
• Images exposed using Aero Storage do not have pa-
tient information. It is recommended, therefore, to keep
record so that you can associate the patient exposed
with the number of saved images displayed on DR De-
tector.
• When the images exposed using Aero Storage are
imported, the history of exposure sequence, exposure
failure, and divider in the DR Detector is updated. Be-
fore importing images, check the information associated
with the image information recorded on paper.
• Exposure failure or division display functions of Aero
Indicator is not associated to the exposed images.
Record the associated information on a piece of paper
with the patient information.
CAUTION
• Take note of the following when using this device:
— Do not use devices that emit electromagnetic waves
such as high-frequency therapy equipment, mobile
phones, or pocket pagers, close to this device.
— Use under the specified environmental conditions.
Failure to do so may result in degradation of perfor-
mance or malfunction.
— Limit the duration of continuous use of the AeroDR
1012HQ and AeroDR 3 1012HQ in a hot and humid
environment (35 °C to 37 °C/95 % or lower) in an in-
cubator to 25 minutes or less.
— Do not insert the AeroDR 1012HQ and AeroDR 3
1012HQ into the AeroDR Battery Charger because
the charger cannot be used for registration or charg-
ing the battery.
— Note that the AeroDR Battery Charger cannot charge
the AeroDR 2 1417S and AeroDR 3 Detector.

1
2
3
4
5
6
7
8
9
10
11
12
Safety precautions
17
CAUTION
• Take the following notes when using the DR Detector:
— Use the DR Detector on a at base. If the DR Detec-
tor is placed on a sloping surface and subjected to a
load, its internal sensor may be damaged.
— When laying the patient onto the DR Detector during
positioning, do so very slowly.
— When placing the DR Detector under the patient as
part of the exposure process, insert and pull out the
detector slowly.
— Use the recommended adapter when you need to
perform exposure on a stretcher or a place where
load is to be applied locally.
— Exposing or immersing the DR Detector to patient
body uids, chemicals, water, etc. may allow such liq-
uids to enter through exterior gaps and cause a fail-
ure. Make sure to cover the DR Detector with a vinyl
sheet, etc.
— Be careful of the ingress of patient body uids, chem-
icals, water, etc., because the AeroDR Detector is not
designed to be waterproof.
— The AeroDR 2 Detector and AeroDR 3 Detector are
designed to have waterproof capabilities, but be care-
ful when handling them because they are not guar-
anteed to be completely waterproof. Also, be careful
because subjecting them to strong impacts or exces-
sive loads by dropping or hitting them against other
objects may diminish their waterproof capabilities.
— Never leave DR Detectors in hot and humid environ-
ments for long periods of time.
— The DR Detector has wireless antennae positioned
at 2 places. Do not block them with the body or metal
because that will interfere with, disconnect or slow
down wireless communication.
— Pressing the power switch on the AeroDR Detector
and AeroDR 2 Detector causes each LED (green,
orange, blue) to light temporarily or ash. After this,
only green begins to flash in a slow cycle. Please
check the lighting or ashing of each color once.
— Pressing the Power SW of the AeroDR 3 Detector
causes the Information LED (white) to ash or light.
After this, the Battery LED (blue) lights. Please check
the ashing or lighting of each color.

18
Safety precautions
CAUTION
— The battery capacity is designed to allow the life pe-
riod calculated based on the standard usage of the
DR Detector, but replace with a new battery as nec-
essary.
— Check that the buzzer sounds when inserted in the
AeroDR Battery Charger or AeroDR Battery Char-
ger2.
— Check that the LED indicates the fully charged state
when inserted in the AeroDR Battery Charger.
— Check that the LED indicates recharging status when
inserted in the AeroDR Battery Charger2.
— Be careful not to get your hand caught when inserting
the DR Detector into the AeroDR Battery Charger or
AeroDR Battery Charger2.
— If the DR Detector is inserted into the AeroDR Battery
Charger or AeroDR Battery Charger2, make sure that
the AeroDR Battery Charger or AeroDR Battery Char-
ger2 will not tip over due to any shock applied to the
DR Detector.
— The DR Detector is precision equipment, and there-
fore impact or vibration during radiography or image
transfer may aect the image quality. Be careful when
handling the DR Detector during and just after radi-
ography.
— When using the remote table for exposure, take cau-
tion to prevent shocks and vibrations during expo-
sure and/or image transmission, because aect from
shocks and vibrations may be large.
— Never leave DR Detectors around heat generators
such as electric carpet.
— Do not damage or deform the DR Detector; doing so
may have an effect on exposed images, or lead to
injury.
— Do not use a sharp object to operate the switch;
doing so may damage the AeroDR 2 Detector and
AeroDR 3 Detector.
— When using a tool to ax the AeroDR Grip sheet or
perform replacement, be careful not to get injured by
a tip or edge of the tool.
• When using the wired cable, observe the following:
— Remove the cable by holding the connector housing.
— Do not let the cable get pinched by doors and do not
place heavy objects on it.
— Do not bend or pull the cable excessively.
— Make sure that the cable is properly connected to the
DR Detector without wobbling.
— Do not connect the connector housing backwards.
• Do not sit on or put your feet on the AeroDR Battery
Charger or AeroDR Battery Charger2.
• Take note of the following if the DR Detector is dam-
aged and uorescent medium or lead is exposed:
— Immediately stop using the device, and contact Koni-
ca Minolta technical representatives.
— If the uorescent medium gets in your eyes, do not
rub and instead wash with running water immediately.
— If you have swallowed any of the uorescent medium
or if the fluorescent medium has gotten into your
eyes, administer rst-aid treatment immediately, and
consult a doctor.
— If the medium comes into contact with your skin di-
rectly, wash the aected area thoroughly with water.
— Use and store in a location inaccessible to children.
• When the DR Detector is used for exposure, pay atten-
tion to the following items.
— Start exposure after conrming on the display of the
image processing controller that the machine is ready
for exposure.
— Perform exposure under the X-ray imaging conditions
that has been conrmed by us.
— Use the specied grid to perform exposure.
— Apply the specified operation methods to use the
grid.
— Do not let the DR Detector vibrate or receive shock
until the preview image is displayed.
— If the irradiation eld is larger than DR Detector when
you take an image, an artifact may occur due to the
X-ray backscatter eect. Adjust the irradiation led to
t within the imaging range.

1
2
3
4
5
6
7
8
9
10
11
12
Safety precautions
19
CAUTION
• Precautions when performing exposure in Aero Sync
mode.
— Conrm that the image processing controller is ready
for exposure through its display before performing
exposure.
— Conrm that a conrmation is sounded from the im-
age processing controller after the start of exposure.
— Do not let the DR Detector vibrate or receive shock
until the preview image is displayed.
— Conrm that the DR Detector's battery level is full be-
fore performing exposure.
— Use the system under the exposure condition con-
rmed prior to exposure.
• Please pay attention on the followings when using Aero
Storage and Aero Indicator:
— Ensure that the DR Detector LED display or Aero In-
dicator display indicates that it is ready for exposure,
before performing an exposure operation.
— If Aero Indicator experiences a connection failure,
check the DR Detector LED display before perform-
ing an exposure operation.
— After an exposure operation, do not give impact or
vibration to the DR Detector until the DR Detector
LED display or Aero Indicator display indicates that
the device is ready for exposure. In addition, if the
display cannot be veried, do not give shocks and/or
vibrations to the DR Detector for six (6) seconds after
an exposure operation.
— Ensure that the battery level of the DR Detector is 10
% or more before performing an exposure operation.
— Note that the exposed image is not displayed in Aero
Indicator when the connection to the DR Detector is
disconnected.
— When exposing images using Aero Indicator, check
the number of saved images on the DR Detector
information LED display, or ensure that the exposed
image is displayed in Aero Indicator.
— Do not give medical diagnoses using the images dis-
played in Aero Indicator.
• If there are any radioactive substances remaining in the
patient's body after nuclear medicine testing or some
other reason, the AeroDR 2 Detector and AeroDR 3 De-
tector may recognize that X-rays were emitted and start
image transfer. When the image is transferred, change
the X-ray sensitive setting of the image processing con-
troller to "Large" before performing exposure.
• If you change the X-ray sensitivity to "Large" in the
AeroDR 2 Detector, make sure that all of the following
conditions are met before performing exposure. If you
perform exposure under other conditions, you may not
be able to get an image.
— On the DR Detector, set up an area (2 cm x 2 cm or
larger) on which the X-rays are emitted directly.
— Emit the X-rays onto the entire DR Detector.
— Emit the X-rays by using a tube current of 140 mA or
higher or a tube voltage of 90 kV or higher.
• If you change the X-ray sensitivity to "Large" or "Med"
in the AeroDR 3 Detector, make sure that all of the fol-
lowing conditions are met before performing exposure.
If you perform exposure under other conditions, you
may not be able to get an image.
— Emit the X-rays onto the entire DR Detector.
— Emit the X-rays by using a tube current of 140 mA or
higher or a tube voltage of 90 kV or higher.
• Set the maximum exposure time at the suitable expo-
sure time with the image processing controller when
performing exposure.
• When setting the maximum exposure time to 4.0, 6.7,
10.3, be sure to contact Konica Minolta technical repre-
sentatives.
• When the maximum exposure time of the AeroDR 3
Detector is set at 10.3, the interval between exposures
should be 2 minutes or longer. The following exposures
also require 2 minutes intervals.
• If you use Ethernet cables, please pay attention to the
following.
— If the power of this device is on, do not connect nor
remove the Ethernet cables.
— Install the cables so that you do not catch them on
your feet and so on.
— Do not apply a load.
— Use shielded Ethernet cables. However, for connec-
tions between AeroDR Interface Unit and AeroDR
Battery Charger or AeroDR Battery Charger2 and be-
tween AeroDR Interface Unit and access points, you
can use unshielded Ethernet cables.
— Communication speed is low or communication is
down, it is necessary to check the Ethernet cable
replacement. Contact Konica Minolta technical repre-
sentatives.

20
Safety precautions
1�3�3 Precautions regarding
electromagnetic waves
EMC Statement
This device has been tested and found to comply with the
limits for medical devices in IEC 60601-1-2: 2007 or IEC
60601-1-2: 2014.
These limits are designed to provide reasonable protec-
tion against harmful interference in a typical medical
installation. The device generates, uses and can radiate
radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful in-
terference to other devices in its vicinity. However, there is
no guarantee that interference will not occur in a particular
installation.
Whether this device does cause harmful interference to
other devices can be determined by turning this device o
and on. If it causes harmful interference, the user is en-
couraged to try to correct the interference by 1 or more of
the following measures:
• Reorient or relocate the receiving device.
• Increase the separation between the devices.
• Connect this device into a wall outlet on a circuit dier-
ent from that to which the other devices are connected.
• Contact Konica Minolta technical representatives.
WARNING
• Do not use mobile RF communication devices such as
mobile phones or pocket pagers in the vicinity of this
device. Use of such RF devices near this device can
cause errors in operation due to electromagnetic wave
interference. When using a mobile RF device, ensure
that the distance between the RF device in use and this
device is at least 30 cm (12 inches).
• The use of accessories, transducers and cables other
than those sold by Konica Minolta, Inc. as internal
components, may result in increased emissions or de-
creased electromagnetic immunity of this device, pos-
sibly causing malfunction.
• Do not use this device adjacent to or stacked with other
devices. If adjacent or stacked use is necessary, con-
rm normal operation in the conguration in which this
device will be used.
Supplementary information regarding IEC
60601-1-2:2007andIEC60601-1-2:2014
(1) Take precautions against this device especially re-
garding EMC. Install and put into service according
to the electromagnetic compatibility (EMC) informa-
tion provided in the manual (Table 1 - Table 5).
(2) This device is suitable for use in medical institu-
tions such as hospitals and clinics. However, this
device is not suitable for use in environments with
high degree of electromagnetic wave interference
(such as near an electric operation device in use
for output, or near a high frequency treatment de-
vice).
(3) Cable list
• Power cable (3.048 m/3-Wire/No Shielding)
• Ethernet cable (max 30 m/Shielding)
• Non-shielded Ethernet cable can be used for con-
necting to the AeroDR Battery Charger and the
AeroDR Battery Charger2
(4) Specications regarding RF transmitters frequency:
• Frequency
— AeroDR SYSTEM:
5150 MHz to 5350 MHz, 5470 MHz to
5850 MHz
— AeroDR SYSTEM 2:
2412 MHz to 2472 MHz
5180 MHz to 5320 MHz, 5500 MHz to
5825 MHz
— SKR 3000:
2412 MHz to 2472 MHz
5180 MHz to 5320 MHz, 5500 MHz to
5825 MHz
• Modulation
— AeroDR SYSTEM:
OFDM
— AeroDR SYSTEM 2:
2412 MHz to 2472 MHz: DSSS/CCK/OFDM
5180 MHz to 5320 MHz, 5500 MHz to 5825
MHz:OFDM
— SKR 3000:
2412 MHz to 2472 MHz: DSSS/CCK/OFDM
5180 MHz to 5320 MHz, 5500 MHz to 5825
MHz:OFDM
• Maximum eective radiation power
— AeroDR SYSTEM:
+15 dBm
— AeroDR SYSTEM 2:
+10 dBm
— SKR 3000:
+10 dBm
• This device may be interfered with by other devices
that conform to CISPR emission requirements.
Other manuals for AeroDR System
3
This manual suits for next models
2
Table of contents
Other Konica Minolta Security Sensor manuals