LabRepCo Torbal AG100 User manual

101 Witmer Road, Suite 700 • Horsham, PA 19044 • Phone 1-800-521-0754 • Fax 215-442-9202
www.labrepco.com • sales@labrepco.com
Instruction Manual for
Torbal AG and AGN Series Balances

- 1 -
Table of Contents
1. Cautionary Notes and Precautions…………………………………………………….
…….
3
1.1 Important Handling Cautions and Warnings……………………………..…………
…...
4
2.
Specifications………………………………………………...
……………………….……….
5
3.
Keys, Display Indicators and Abbreviations……………………………
…………………….
7
4.
Parts Descriptions…………………………………………………..
…………………………
8
5.
Unpacking Balance and Getting Started…………………………………………...
………….
10
6.
External Keyboard Connection (optional)………………………………………...
.………….
12
7.
Functions, Descriptions and Definitions…………………………………….
………………..
13
8.
Main Menu…………..……………………..
…….………………………………….………..
14
8.1.Mode Setup……………………………………
………….………………………………..
14
8.1.1.
Weighing………………………….
……………………….………………….………
15
8.1.2.
Parts Counting…………………………………
…………………………….……….
16
8.1.3.
Check Counting………..………………………
………………………….………….
18
8.1.4.
Percent Weighing……...………………….
……………………………….………….
19
8.1.5.
Totalizing………………………….
………………………………………………….
21
8.1.6.
Compounding………………………
………………………………………………...
22
8.1.7.
Check Weighing…………………………………………
…………………………...
23
8.1.8.
Animal and Dynamic Weighing………………….…………………..
………………
25
8.1.9.
Display Hold…………………………………….
……………………………………
27
8.1.10.
SQC Process Control………………………………….……………………………...
28
8.1.11.Tare Setup…………………………………….
………………………………………
30
8.2.
Units of Measure……………...…………………...
…………………………………..…..
31
8.3.
Configurations…………………………...
………………………………………………...
32
8.3.1.Auto-Zero…………………………………………………………………………..
32
8.3.2.Print……………………………………
……………………………………………...
33
8.3.2.1GLP……..…………………..
………………………………………………….
33
8.3.2.2Print Details……….…………..
…………………………………………………
34
8.3.3.Filter……………………….
……………………………………………….………….
34
8.3.4.Power Save and LCD Contrast…………………………...
…………………………..
35
8.3.5.RS232………………………..
…………………………………………….…………
36
8.3.6.USB Configuration….………………...
……......………………………….…………
36
8.3.6.1PC Connection via USB…………………..……...
…………………..………….
37
8.3.7.Time and Date Adjustment…………………….
……………………………..………
38
8.3.8.Reset and Default Restore………………...
………………………………………….
39
8.3.9.Erasing the Database……………………
…………………………………………….
39
8.4.
Calibration……………………………………………………………………….………
40
8.4.1.Auto Calibration Setting and Adjustment…………………
………………………….
41
9.
Using Functions and Modes………………………………..
…………………………………
42
9.1.
Weighing…………………………………………………………………………………
42
9.1.1.Taring……………………………………
……………………………………………
42

- 2 -
9.1.2.Clearing a Tare………………………...
……………………………………………..
43
9.2.
Parts Counting……………………
……………………………………..…….………….
44
9.2.1.Automatic………
………………………………………………………...…………..
44
9.2.1.1
Counting Parts for the First time…………..………………………………
.……
44
9.2.1.2
Counting Parts which Have Been Previously Stored in the Scale’s Database……
46
9.2.2.
Manual………………………………………………………………….…………..
48
9.3.
Check Counting……………………………………………………………………………
49
9.3.1.Storing a Limit Set into the Scale’s Database……………...
……………………..….
52
9.4.Percent Weighing…...……………….
…………………………………………………….
52
9.4.1.Automatic…………….
………………………………………………………………
52
9.4.1.1
Percent Weighing for the First Time…………………………………………….
52
9.4.1.2
Percent Weighing with Stored Reference Weights…………
……….…………..
55
9.4.2.Manual…………………..
…………………………………………….……………
56
9.5.Totalizing…………………………………………………………………….
…………
57
9.6.Compounding……………..
………………………………..……………….………….
59
9.7.Check Weighing……………………
……………………………………….………….
61
9.7.1.Storing a Limit Set into the Scale’s Database…….…………………….………….
62
9.8.Animal and Dynamic Weighing…...…………….…………………………………….
63
9.9.Display Hold…………………..
………………………………………………………..
64
9.10.SQC Process Control……………………………………..
………………….…………..
65
9.10.1.Storing a Limit Set into the Scale’s Database…………………………..…….……
66
10.
Using the Scale’s Alphabet for Data Entry…………………………………………………
67
11.
Connecting and Using a Barcode Scanner with the Scale’s Database……………………...
68
11.1.Connecting the Scanner……………………………………………………………….
..
68
11.2.Assigning Barcodes to Samples or Limit Sets………………………..
………………….
68
11.3.Recalling a Previously Stored Sample or a Limit Set from the Database………
………
68
12.
Technical Information………………………………………………………………………
69
13.
Common Error and Troubleshoots…………………….……...……………………………. 70
14.
Maintenance…………………………..
…………………………………..…………………
71
15.
Accessories…………………
………………………………….……………………………
71
16.
Replacement Parts……………………………………………….………………………….
72
17.
Limited Warranty…………….….…………………………………….……………………
72

- 3 -
The TORBAL Precision scale is a sensitive and a delicate instrument. Always handle your scale with care.
The correct location and proper environment makes an important contribution to the accuracy of the weighing
results of TORBAL precision scales.
The optimum location for your scale:
•Stable, vibration-free base as horizontal as possible
•Away from direct sunlight
•Not exposed to high temperature variations
•Away from direct drafts
•The best location is on a stable bench away from direct drafts, doors, windows, radiators and air
conditioner vents.
CAUTION:
•The scale is designed for indoor use only..
•Do not operate the scale in hazardous areas or conditions.
•Do not use the scale in locations subject to high humidity or dust levels.
•Do not connect cables in ways other than those mentioned in this manual.
•Be sure to set the scale on a firm, stable, horizontal surface.
•Never stand on or lean on this product. Equipment may fall or collapse,
causing breakage and possible injury.
•Before moving the product, unplug it and unplug all cables connected to it.
•When storing, transporting or returning the scale for service, always use the
original packaging.
WARNING:
•Never attempt to repair, disassemble or modify the scale yourself. Tampering
with the scale may result in injury and cause more damage to the equipment.
•Be sure to use the specified power source.
•Do not allow foreign matter to fall into the scale.
•If water or other liquid spills into the scale, do not continue to use it. Unplug
the power cord immediately and contact technical support.
Chapter 1: Cautionary Notes

- 4 -
1.1. Important handling Cautions and Warnings
Always handle your scale with care. Damage caused by improper handling is
not covered under the scale’s warranty.
Never drop or throw any articles
onto the scale’s pan or onto any
other parts of the unit!
DO NOT let the scale fall or drop
from its tabletop surface!
When moving the scale do not
press or apply force onto the
scales pan!
DO NOT pass or handle liquids
directly over the scale to avoid
spillage and liquid damage!

- 5 -
Model
AG100 AG200 AG300 AG500 AG1000 AG2000 AG3000 AG4000
Capacity 100g 200g 300g 500g 1000g 2000g 3000g 4000g
Readability (d) 0.001g 0.01g
Repeatability (Standard
Divination) 0.001g 0.01g
Linearity +/-0.002g +/-0.02g
Stabilization time (typical)
Approx. 3 sec Approx. 3 sec
Tare Range -100g -200g -300 -500 -1000 -2000 -3000 -4000
Accuracy Class I
Automatic Internal
Calibration YES
Time Configuration 1h, 2,h, 3h, 4h, 5h, 6h
Temperature
Configuration 0.5°C, 1.0°C, 1.5°C, 2.0°C
Automatic Calibration
Ad ustment Masses (g) 20,50, 100 50, 100,
200
100, 200,
300
100, 200,
500
200, 500,
1000
500, 100,
2000
1000,
2000,
3000
1000, 2000,
4000
Pan Dimension 120 mm 165 mm x 165 mm
Draft Shield Optional Not Required
Scale Dimension 216 x 345 x 90 mm
Operating Temperature +15°C to +30°C
PS2 Port Standard
RS232 Port Bidirectional
USB Standard
Power Supply Input: 120VAC 60Hz 15W Output: 12VDC 500mA
Display Type Graphical LCD
Display Size 30mm x 100mm
Application Modes Weighing, Parts Counting, Animal Weighing, Check Weighing, Display Hold, Percent Weighing, Totalizing, Check
Counting, Compounding, Statistics, SQC Quality Control, 512 KB Library Bank
Weighing Units Grams, Milligrams, Kilograms, Carats, Pounds, Newton, Grains, Ounces Troy, Pennyweight
Scale’s Net Weight (lb/kg)
5kg/11lbs
Sample Size 1 through 100 or any
Parts
Counting Min Piece
Weigh 30mg 300mg
Sample Size 1 through 100 or any
Check
Counting Min Piece
Weigh 30mg 300mg
Min
Reference
Weight
100mg 1 g
% Resolution
Displayed
Percent
Weighing
(Reference
Weight)
0.1% (100mg to 7g), 0.01% (7g to 70g)
0.001% (>70g)
0.1% (1g to 70g), 0.01% (70g to 700g)
0.001% (>700g)
Tiny Intervals
3sec., 5sec., 10sec., 15sec.
Animal
Weighing Modes (Min
Weight) Manual, Auto (10mg), Semi-Auto (10mg) Manual, Auto (100mg), Semi-Auto (100mg)
Totalizing Modes (Min
Weight) Manual, Auto (10mg), Semi-Auto (10mg) Manual, Auto (100mg), Semi-Auto (100mg)
Display
Hold
Modes (Min
Weight) Manual, Auto (10mg), Semi-Auto (10mg) Manual, Auto (100mg), Semi-Auto (100mg)
Chapter 2: Specifications

- 6 -
Model
AGN100 AGN200
Capacity 100g 200g
Readability (d) 0.0001g
Repeatability (Standard
Divination) 0.0001g
Linearity +/-0.0002g
Stabilization time (typical) Approx. 3 sec
Tare Range -100g -200g
Accuracy Class I
Automatic Internal Calibration
YES
Time Configuration 1h, 2,h, 3h, 4h, 5h, 6h
Temperature Configuration 0.5°C, 1.0°C, 1.5°C, 2.0°C
Automatic Calibration
Ad ustment Masses (g) 20,50, 100 50, 100, 200
Pan Dimension 120 mm
Draft Shield Optional
Scale Dimension 216 x 345 x 90 mm
Operating Temperature +15°C to +30°C
PS2 Port Standard
RS232 Port Bidirectional
USB Standard
Power Supply Input: 120VAC 60Hz 15W Output: 12VDC 500mA
Display Type Seven Segment LCD
Display Size 30mm x 100mm
Application Modes Weighing, Parts Counting, Animal Weighing, Check Weighing, Display Hold, Percent Weighing,
Totalizing, Check Counting, Compounding, Statistics, SQC Quality Control, 512 KB Library Bank
Weighing Units Grams, Milligrams, Kilograms, Carats, Pounds, Newton, Grains, Ounces Troy, Pennyweight
Scale’s Net Weight (lb/kg) 6kg/14lbs
Sample Size 1 through 100 or any
Parts
Counting Min Piece Weigh 3e (30mg)
Sample Size 1 through 100 or any
Check
Counting Min Piece Weigh 3e (30mg)
Min Reference
Weight 1e (1mg)
Percent
Weighing
% Resolution
Displayed
(Reference
Weight)
0.1% (100mg to 7g), 0.01% (7g to 70g), 0.001% (>70g)
Tiny Intervals 3sec., 5sec., 10sec., 15sec. or any
Animal
Weighing Modes (Min
Weight) Manual, Auto 1e (1mg), Semi-Auto 1e (1mg)
Totalizing Modes (Min
Weight) Manual, Auto 1e (1mg), Semi-Auto 1e (1mg)
Display
Hold
Modes (Min
Weight) Manual, Auto 1e (1mg), Semi-Auto 1e (1mg)

- 7 -
Key Primary Function Secondary
Function
Power On and Power Off -
Tare – used to tare the weighing pan N/A
0Zero – used to zero the scale (Legal for Trade models onl ) N/A
Enter – used to accept and confirm commands N/A
MENU Menu – used to access the main menu Number “9” ke
CLR Clear – used to clear and cancel operation N/A
▲▼◄► Navigation Ke s – used to navigate through the menus and functions N/A
Function Ke – used access database and secondar function Number “0” ke
Data Transfer – used to print data or transfer data to a PC via the RS232
communication port Number “7” ke
▼Calibration Ke - used to initialize internal calibration Number “8” ke
F1 Function Ke Number “1” ke
F2 Function Ke Number “2” ke
F3 Function Ke Number “3” ke
F4 Function Ke Number “4” ke
F5 Function Ke Number “5” ke
Display Indicator Description
Stabilit Indicator - The weighing result has stabilized and an accurate reading ma be taken.
Zero - The scale is maintaining a “center of zero” condition. (Legal for Trade models onl )
AUT AZSM (Automatic Zero Setting Mechanism) is enabled. Alwa s enabled in Legal for Trade models.
OFF Power Off - The scale is turned OFF and in standb mode.
Diff. Difference
SPL / SAPL Sample or Samples
LOCK Displa Locked
PCS Parts Counting
CHKC Check Counting
PERC Percent Weighing
TOT Totalizing
COMP Compounding
CHKW Check Weighing
ANL Animal Weighing
HLD Displa Hold
SQC SQC Process control
Chapter 3: Keys, Display Indicators and Abbreviations

- 8 -
AG Models:
Front View Rear View Side View
Pan Base Pan AC Adapter
RS232 Serial Port Service Switch Level Indicator AC Adapter Socket
Chapter 4: Parts Description

- 9 -
AGN Models:
1 - Pan, 2 - Pan Base, 3 - Draft Ring, 4 - Weighing Chamber Base, 5 – Keypad, 6 - Adjustable Legs,
7 - Level Indicator, 8 - LCD Display, 9 - Weighing Chamber, 10 - AC Adapter Socket,
11 – Service Switch, 12 - RS232 Serial Port
1 - Pan, 2 - Pan Base, 3 - Draft Ring,
4 - Chamber Base, 5 - Level Indicator,
6 - Adjustable Legs

- 10 -
1. Carefully remove the scale, pan, and all of its components out of the
packaging and place them on a stable base where the scale will not
be affected by any mechanical vibrations or air movements.
2. AG Modes: After removing the pan base and the pan from their packaging, carefully install the pan
base onto the scale by seating it on the pan support located in the middle of the scale. Once the pan
base has been installed, carefully place the pan on the base as shown below.
AGN Models:
After removing the chamber base (4), draft
ring (3), pan base (2), and the pan (1) from
their packaging, carefully install the chamber
base (4) on the case of the balance as shown
below. Once the chamber base is in place,
install the draft ring (3). Once components 4
and 3 are in place install the pan base onto the
scale by placing it on the pan support located
in the middle of the scale. Once the pan base
has been installed, carefully place the pan on
the base as shown.
Chapter 5: Unpacking the Balance and Getting Started

- 11 -
3. Once the pan has been installed, level the scale by adjusting the rear feet until the level indicator
shows the “air bubble” is in the center position of the sight glass. The level indicator is located in the
rear left side of the scale.
4. After leveling the scale, you may plug the AC adaptor to the AC adaptor socket located in the rear of
the scale.
5. When the AC adapter is plugged into the wall outlet, the scale will turn “ON” automatically, initialize
itself, perform automatic internal calibration and enter weighing mode.
6. To put the scale into standby mode, leave the AC adaptor plugged into both the scale and the wall
outlet and press the Power “OFF” button ( ). The “OFF” indicator will light up in the upper left
corner of the display signaling the scale is in standby mode.

- 12 -
Any PC external keyboard can be connected to the scale via the PS2. An external keyboard allows you to
input data quickly and easily. To connect a PC keyboard to the scale, follow the steps below.
1. Clear the pan and turn the scale off.
2. Locate the keyboard connection port at the rear of the scale, marked “keyboard.”
3. Insert the keyboard connector into the scale’s port.
4. Turn the scale on.
Note: Some keyboard keys remain disabled and do not work with the scale (i.e. function keys, Shift, Control
etc…). Do not use the number keypad located to the right of your keys. To key in numbers always use the
number keys in the upper row of the keyboard.
Chapter 6: External Keyboard Connection (Optional)

- 13 -
oParts Counting is used to count parts or pieces of articles based on their average piece weight.
oCheck Counting is used to verify whether a parts counting result falls within specified threshold
limits.
oPercent weighing is used to establish the relative weight of an unknown, compared to that of a stored
reference weight and expressed as a percent.
oTotalizing is used for summation of sequential weighing results of items or articles.
oCompounding is used for weighing and summing individual ingredients such as powers or liquids
during recipe making or mixing compounds.
oCheck weighing is used to check whether a weight of an object falls within specified threshold limits.
oAnimal and dynamic weighing is used to weigh animals or objects which are in motion while on the
pan.
oDisplay hold is used to lock and retain a weighing result on the display until cleared.
oStatistical quality control (SQC) is used to establish a trend analysis and statistical data of samples in
relation specified threshold limits.
Chapter 7: Functions, Descriptions and Definitions

- 14 -
Main Menu is used to configure the scale and it’s weighing modes. Before entering the menu complete
and clear active functions by pressing the CLR key.
1. To enter the Main Menu press the MENU key. Once the MENU key is pressed the menu options will
be displayed.
8.1. Mode Setup
Torbal AG and AGN Scales can operate in ten application modes: Weighing, Parts Counting, Check
Counting, Percent Weighing, Totalizing, Compounding, Check Weighing, Animal Weighing, Display
Hold, and SQC Process Control. Before an application mode can be used it must be enabled and
configured. To use an application mode a quick key (Function keys F1 through F5) should be assigned
(recommended). Quick keys allow quick toggling between functions. By factory default the scale is in the
weighing mode. To enable other application modes follow the steps below.
1. From the Main Menu select “Mode Setup” and press the ENTER key. A list of available
applications and modes will be displayed.
Chapter 8: Main Menu

- 15 -
8.1.1. Weighing
1. To configure the Weighing Function use the navigation keys to select “Weighing” in “Mode Setup”
and press ENTER. The status and available weighing options will be displayed. If the function is
currently in use the status will be set to “ON”.
2. To assign a “Quick Key” use the navigation keys to select “Quick Key” and press the ENTER key.
Use the navigation keys to select a desired quick key (F1 through F5). Once the desired key is selected
press the ENTER key to accept the selection.
If the key is not available (used by a different function) a warning message will be displayed
“Warning Quick Key Already In Use!” To override the key press ENTER. To abort and select a
different function key press the CLR key.
3. To exit the Weighing set up select “Exit” and press the ENTER key.

- 16 -
8.1.2. Parts Counting
1. To configure the Parts Counting Function use the navigation keys to select “Parts Counting” in
“Mode Setup” and press ENTER. The status and available parts counting options will be displayed. If
the function is currently in use the status will be set to “ON”.
2. To assign a default “Sample Size” that will be used in calculating the average piece weight use the
navigation keys to select “Sample Size” and press ENTER. To select a preset “Sample Size” use the
navigation keys to toggle until a desired sample size is displayed (1, 5, 10, 20, 50, 100, 200, 500). To
assign a custom “Sample Size” use the number keys on the scale’s keypad to key in a desired value.
Once the desired “Sample Size” is selected or entered, press the ENTER key to set and confirm the
selection.
3. To select an operating mode use the navigation keys to select “Mode” and press ENTER. By default
parts counting is set to “Auto” (Automatic). To operate parts counting in “Manual” use the navigation
keys to select “Manual” and press ENTER.
Parts Counting Operating Modes
Automatic
When set to automatic the balance allows you to create a database of samples and
average piece weights. Each sample can be labeled with a unique name or an alpha
numeric code and stored in the scale’s internal memory along with its average piece
weight. This eliminates the need for establishing a new average piece weight every time
a part is counted.
Manual
When using parts counting in manual, the database feature is disabled and samples size,
as well as, average piece weight is not stored in the scale’s memory. The sample size
must be manually counted and an average piece weight must be established every time
part counting is performed.

- 17 -
4. To enable sample labeling use the navigation keys to select “Sample Labeling” and press the ENTER
key. Use the navigation keys once again to enable or disable sample labeling. Once a desired selection
is made press the ENTER key.
“Sample Labeling” applies to the “Automatic Mode” only. It allows the user to assign a unique name
to a sample which is then stored in the scales database. When set to “OFF” the scale will not assign or
display names of the samples.
5. To assign a “Quick Key” use the navigation keys to select “Quick Key” and press the ENTER key.
Use the navigation keys to select a desired quick key (F1 through F5). Once the desired key is selected
press the ENTER key to accept the selection.
If the key is not available (used by a different function) a warning message will be displayed
“Warning Quick Key Already In Use!” To override the key press ENTER. To abort and select a
different function key press the CLR key.
6. To configure the scale’s display during parts counting use the navigation keys to select “Display
Setup” and press the ENTER key. To disable information from displaying during the function use the
navigation keys to highlight the desired selection and press the ENTER key to remove the check mark.
Once the check mark has been removed the information will not display while performing parts
counting.
7. To exit Parts Counting set up use the navigation keys to select “Exit” and press ENTER.

- 18 -
8.1.3. Check Counting
1. To configure the Check Counting Function, use the navigation keys to select “Check Counting” in
“Mode Setup” and press ENTER. The status and available Check Counting options will be displayed.
If the function is currently in use the status will be set to “ON”.
2. To assign a default “Sample Size” that will be used in calculating the average piece weight use the
navigation keys to select “Sample Size” and press ENTER. To select a preset “Sample Size” use the
navigation keys to toggle until a desired sample size is displayed (1, 5, 10, 20, 50, 100, 200, 500). To
assign a custom “Sample Size” use the number keys on the scale’s keypad to key in a desired value.
Once the desired “Sample Size” is selected or entered press, the ENTER key to set and confirm the
selection.
3. To enable Limit Set Labeling use the navigation keys to select “Limit Set Labeling” and press the
ENTER key. Use the navigation keys once again to enable or disable Limit Set labeling. Once a
desired selection is made press the ENTER key.
“Limit Set Labeling” allows the user to assign a unique name to a Limit Set which is then stored in the
scales database. When set to “OFF” the scale will not assign or display names of the Limit Set.
4. To assign a “Quick Key” use the navigation keys to select “Quick Key” and press the ENTER key.
Use the navigation keys to select a desired quick key (F1 through F5). Once the desired key is selected
press the ENTER key to accept the selection.
If the key is not available (used by a different function) a warning message will be displayed
“Warning Quick Key Already In Use!” To override the key press ENTER. To abort and select a
different function key press the CLR key.

- 19 -
5. To configure the scale’s display during check counting use the navigation keys to select “Display
Setup” and press the ENTER key. To disable information from displaying during the function use the
navigation keys to highlight the desired selection and press the ENTER key to remove the check mark.
Once the check mark has been removed the information will not display while performing check
counting.
6. To exit Check Counting set up use the navigation keys to select “Exit” and press ENTER.
8.1.4. Percent Weighing
1. To configure the Percent Weighing Function, use the navigation keys to select “Percent Weighing” in
“Mode Setup” and press ENTER. The status and available Percent Weighing options will be
displayed. If the function is currently in use the status will be set to “ON”.
2. To select an operating mode use the navigation keys to select “Mode” and press ENTER. By default
Percent Weighing is set to “Auto” (Automatic). To operate Percent Weighing in manual use the
navigation keys to select “Manual” and press ENTER.
This manual suits for next models
9
Table of contents
Popular Scale manuals by other brands

SR Instruments
SR Instruments SRScales SRV949 Operating and service manual

IVDT
IVDT SkidWeigh ED2-AT Series Quick Reference Calibration & Operation Guide

AWS
AWS NutriBalance-5000 user manual

Salter
Salter SA 9096 GN3R Instructions and guarantee

Cardinal
Cardinal Detecto 6437 operating instructions

OHAUS
OHAUS AV53 instruction manual