Lambda TA018 User manual

When the unitisconnected toaDMM whichaccuracy is
supposed tobe 0.5%in200mVrange,the reading displayed
on theDMM will be 10.25mVmax.(10.20mVx(1+0.5%)=10.
25mV)and 9.75mVmin.(9.80mVx(1-0.5%)=9.75mV)
CurrentClampAccuracy
DCA range: 1mV/10mA
± (2.0%±5mA) 10mA~20A
DCA range: 1mV/100mA
± (2.0%±20mA) 100mA~40A
± (4.0%±0.3A) 40A~60A
ACA range: 1mV/10mA
±(2.0%±5mA) 10mA~10A (40Hz~2kHz)
±(4.0%±30mA) 10mA~10A (2kHz~10kHz)
±(6.0%±30mA) 10mA~10A (10kHz~20kHz)
±(8.0%±30mA) 10A~15A (40Hz~20kHz)
ACA range: 1mV/100mA
±(2.0%±30mA) 100mA~40A (40Hz~1kHz)
±(4.0%±30mA) 100mA~40A (1kHz~2kHz)
±(6.0%±30mA) 100mA~40A (3kHz~5kHz)
±(8.0%±0.3A) 40A~60A (40Hz~5kHz)
Load Resistance: 10kΩ typical
Temperature Coefficient: 0.1x(specified accuracy) per
degree C. (0°C to 18°C, 28°C to 40°C)
SAFETY INFORMATION
Cleaning
Wipe the case with a damp cloth and mild detergent. Do not
use abrosives or solvents. Dirt or moisture in the terminals
can affect readings.
Safety:Conforms to EN 61010-1, EN 61010-2-032,
CAT II 300V, Class II, Pollution degree 2 Indoor use.
CAT II: Is for measurements performed to the low-
voltage on circuits directly connected.
EMC: Conforms to EN 61326-1.
The symbols used on this instrument are:
Dangerous voltage.
Caution, refer to accompanying documents
Equipment protected throughout by Double
insulation (Class II)
Alternating current Direct current
AC/DC
mACURRENTCLAMP
OPERATOR’SMANUAL
CLAMP 300V
60A
CAT.II
7000-1698A

INTRODUCTION
The AC/DC mACurrentClampisatransducerwhichwill
allowyourmultimetertomeasurelowelectricalor/and elec-
troniccurrentup to60 amperesAC/DC,withafrequency
responseup to20kHz.When measuring currentwiththis
clamp,thereisno need tobreakacircuitortoaffectthe
isolation.
The extended measurementjawsallowperforming mea-
surementsinamarrowspace.When measuring DC current,
asimpleoperating pushbutton isdesigned forzero
adjustment. The clampadapterisapplicabletoleakge detec-
tion ormonitering.
APPLICATIONPROCEDURES
1. Insert the black banana plug into the COM jack and the red
banana plug into the V-Ωjack of any multimeter with
a minimum input impedance of 10k ohms.
2. Set the power switch from “OFF” to the desired range,
1mV/10mA or 1mV/100mA position. The green LED will
light to indicate that the clamp is switched on.
3. For current measurement below 2A, set the unit to 1mV/
10mA range and set the multimeter to 200mV AC range for
AC current measurements, or 200mV DC range for DC
current measurements. If the measured current exceeds
2A, set the unit to 1mV/100mA range.
4. When perform DC current measurement, always push the
zero adjustment button on the clamp until the multimeter
reads zero.
5. Clamp the jaws around the current-carring conductor and
interpret the reading according to Step 3 above.
6. When 1mV/10mA range of clamp unit is selected,
multiple the reading displayed on the multimeter by “10”
for interpreting the measured current value in mA. For
example, if the multimeter reads 10mV, the measured
current is 10x10=100mA.
When 1mV/100mA range is selected, multiple the reading
displayed on the multimeter by “100” for interpreting the
measured current value in mA. For example, if the multim
eter reads 5mV, the measured current is 5x100=500mA.
APPLICATION NOTES
1.In the case of DC current, the output is positive when the
current flows from the upside to the underside of the
clamp. The red banana plug is positive.
2.In the case of DC current measurement, a hysteresis ef-
fect can occur so that it is impossible to zero the clamp
properly. To eliminate this effect, open and close the jaws
several times and push zero adjustment button.
OPERATOR SAFETY
1.Do not clamp around conductors with voltages equal to or
exceeding 300VDC or 240V rms AC.
2.To avoid physical injury, measurements on bare conduc-
tors or conductors with cracked or frayed insulator are
forbidden.
SPECIFICATIONS
GENERAL
Captured ConductorSize:9mm maximum.
LowBatteryIndicator:Red LEDlighting.
Operating Temperature:0°Cto40°C,70%R.H.
Storage Temperature:-20°C+70°C,80%R.H.
BatteryType:9VDC,NEDA1604,6F22,006P.
BatteryLife:100 hourstypicalwithalkaline.
Weight: 250 gmtypical
Dimensions:195mm (H)x70mm (W)x33mm (D).
Output: Coil cablewithstraightbanana plug.
ELECTRICAL (At 23± 5°C, 70% R.H. maximum )
Effective Measurement Range
1mV/10mA: 10mA to 20A DC or rms AC for 200mV range
of the multimeter.
1mV/100mA: 10mA to 20A DC or rms AC for 200mV range
of the multimeter.
20A to 60A DC or rms AC for 2V range of the
multimeter.
Accuracy
System accuracy: Current clamp accuracy + DMM
accuracy.
For example, if the measured conductor carrys a
100mA current, set the current clamp to 1mV/10mA
range to get an output signal of 10mV. Suppose the
accuracy of the units is 2.0%, the tolerance limit
should be between 10.20mV maximun and 9.80mV
minimum.

INTRODUCTION
The AC/DC mACurrentClampisatransducerwhichwill
allowyourmultimetertomeasurelowelectricalor/and elec-
troniccurrentup to60 amperesAC/DC,withafrequency
responseup to20kHz.When measuring currentwiththis
clamp,thereisno need tobreakacircuitortoaffectthe
isolation.
The extended measurementjawsallowperforming mea-
surementsinamarrowspace.When measuring DC current,
asimpleoperating pushbutton isdesigned forzero
adjustment. The clampadapterisapplicabletoleakge detec-
tion ormonitering.
APPLICATIONPROCEDURES
1. Insert the black banana plug into the COM jack and the red
banana plug into the V-Ωjack of any multimeter with
a minimum input impedance of 10k ohms.
2. Set the power switch from “OFF” to the desired range,
1mV/10mA or 1mV/100mA position. The green LED will
light to indicate that the clamp is switched on.
3. For current measurement below 2A, set the unit to 1mV/
10mA range and set the multimeter to 200mV AC range for
AC current measurements, or 200mV DC range for DC
current measurements. If the measured current exceeds
2A, set the unit to 1mV/100mA range.
4. When perform DC current measurement, always push the
zero adjustment button on the clamp until the multimeter
reads zero.
5. Clamp the jaws around the current-carring conductor and
interpret the reading according to Step 3 above.
6. When 1mV/10mA range of clamp unit is selected,
multiple the reading displayed on the multimeter by “10”
for interpreting the measured current value in mA. For
example, if the multimeter reads 10mV, the measured
current is 10x10=100mA.
When 1mV/100mA range is selected, multiple the reading
displayed on the multimeter by “100” for interpreting the
measured current value in mA. For example, if the multim
eter reads 5mV, the measured current is 5x100=500mA.
APPLICATION NOTES
1.In the case of DC current, the output is positive when the
current flows from the upside to the underside of the
clamp. The red banana plug is positive.
2.In the case of DC current measurement, a hysteresis ef-
fect can occur so that it is impossible to zero the clamp
properly. To eliminate this effect, open and close the jaws
several times and push zero adjustment button.
OPERATOR SAFETY
1.Do not clamp around conductors with voltages equal to or
exceeding 300VDC or 240V rms AC.
2.To avoid physical injury, measurements on bare conduc-
tors or conductors with cracked or frayed insulator are
forbidden.
SPECIFICATIONS
GENERAL
Captured ConductorSize:9mm maximum.
LowBatteryIndicator:Red LEDlighting.
Operating Temperature:0°Cto40°C,70%R.H.
Storage Temperature:-20°C+70°C,80%R.H.
BatteryType:9VDC,NEDA1604,6F22,006P.
BatteryLife:100 hourstypicalwithalkaline.
Weight: 250 gmtypical
Dimensions:195mm (H)x70mm (W)x33mm (D).
Output: Coil cablewithstraightbanana plug.
ELECTRICAL (At 23± 5°C, 70% R.H. maximum )
Effective Measurement Range
1mV/10mA: 10mA to 20A DC or rms AC for 200mV range
of the multimeter.
1mV/100mA: 10mA to 20A DC or rms AC for 200mV range
of the multimeter.
20A to 60A DC or rms AC for 2V range of the
multimeter.
Accuracy
System accuracy: Current clamp accuracy + DMM
accuracy.
For example, if the measured conductor carrys a
100mA current, set the current clamp to 1mV/10mA
range to get an output signal of 10mV. Suppose the
accuracy of the units is 2.0%, the tolerance limit
should be between 10.20mV maximun and 9.80mV
minimum.

When the unitisconnected toaDMM whichaccuracy is
supposed tobe 0.5%in200mVrange,the reading displayed
on theDMM will be 10.25mVmax.(10.20mVx(1+0.5%)=10.
25mV)and 9.75mVmin.(9.80mVx(1-0.5%)=9.75mV)
CurrentClampAccuracy
DCA range: 1mV/10mA
± (2.0%±5mA) 10mA~20A
DCA range: 1mV/100mA
± (2.0%±20mA) 100mA~40A
± (4.0%±0.3A) 40A~60A
ACA range: 1mV/10mA
±(2.0%±5mA) 10mA~10A (40Hz~2kHz)
±(4.0%±30mA) 10mA~10A (2kHz~10kHz)
±(6.0%±30mA) 10mA~10A (10kHz~20kHz)
±(8.0%±30mA) 10A~15A (40Hz~20kHz)
ACA range: 1mV/100mA
±(2.0%±30mA) 100mA~40A (40Hz~1kHz)
±(4.0%±30mA) 100mA~40A (1kHz~2kHz)
±(6.0%±30mA) 100mA~40A (3kHz~5kHz)
±(8.0%±0.3A) 40A~60A (40Hz~5kHz)
Load Resistance: 10kΩ typical
Temperature Coefficient: 0.1x(specified accuracy) per
degree C. (0°C to 18°C, 28°C to 40°C)
SAFETY INFORMATION
Cleaning
Wipe the case with a damp cloth and mild detergent. Do not
use abrosives or solvents. Dirt or moisture in the terminals
can affect readings.
Safety:Conforms to EN 61010-1, EN 61010-2-032,
CAT II 300V, Class II, Pollution degree 2 Indoor use.
CAT II: Is for measurements performed to the low-
voltage on circuits directly connected.
EMC: Conforms to EN 61326-1.
The symbols used on this instrument are:
Dangerous voltage.
Caution, refer to accompanying documents
Equipment protected throughout by Double
insulation (Class II)
Alternating current Direct current
AC/DC
mACURRENTCLAMP
OPERATOR’SMANUAL
CLAMP 300V
60A
CAT.II
7000-1698A
Table of contents
Other Lambda Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Cadex
Cadex C7000 user manual

Hach
Hach CL17sc user manual

Siemens
Siemens Class 87 Additional Installation, Operation and Maintenance Instructions

Sper scientific
Sper scientific 840006 instruction manual

Rhopoint Instruments
Rhopoint Instruments IQ Flex 20 operating instructions

MEGABRAS
MEGABRAS MAR83s user manual