LaserPoint srl - Fit IPL-R Instruction Manual - page 5
capacity. If the bar graph is below full scale, one or more measurements can be
carried out before the sensor head needs to be cooled.
A new measurement can be done after the LED automatic shuts off (approx. after
20 seconds). To perform a different measurement return to Paragraph 3.
ENERGY MEASUREMENTS:
8-Resetting
Beforestartingthefirstenergymeasurement(seePar3)or anew
one, the instrument has to be reset; to do so, just push and soon
release the ON/Zero button. The LCD displays “rdY” and the
LED on the right side of the screen, turns steady green. At this
point, any previously recorded value is zeroed and the instrument is ready to measure.
No light must strike on the sensor during the resetting period.
An automatic Zero is performed by Fit after every measurement.
9 –Measurement Cycle
Check that the light beam is not available; align the probe to
the beam axis and start the measurement cycle by removing
the beam stop. Avoid taking measurements during the
machine warm up or in case of beam instabilities.
Always position and keep the beam in the centre of detector.
The measurement starts automatically (during the measurement the display shows
tr’d) During the acquisition time the LED blinks green . The bar graph indicates the
head temperature on an arbitrary scale.
10 –Measurement Display
As soon as the measurement is over the LED turns into steady
red. Remove the probe from the light beam: the display shows
the actual power, and the remaining thermal capacity. If the bar
graph is below full scale, one or more measurements can be
carried out before the sensor head needs to be cooled.
AnewmeasurementcanbedoneaftertheLEDautomaticshutsoff (approximately after
20 seconds). To performa new measurement go back to Paragraph 3
11-Customer’s calibration
This instrument can be re-calibrated by the user to match a
custom reference. Remove the plastic cover of the adjustment
switch located at the right side of the instrument body. Zero the
instrument as indicated in Paragraph 2 of this Operation Section
and afterwards carry out a measurement of a known light source.