LAUMAS W100RIP User manual

ENGLISH ENGLISH ENGLISH ENGLISH
W100RIP − W200RIP
WDESKLIGHTRIP − WDESK-L/R-RIP
WINOX-L/R/2L-RIP − WTAB-L/R-RIP
WEIGHT REPEATER
version 1.10

SYMBOLS
Here are the symbols used in the manual to draw the reader's attention:
Caution! Risk of electric shock.
Caution! This operation must be performed by skilled personnel.
Pay particular attention to the following instructions.
Further information.
WARRANTY
24 months from the date of the delivery note. Warranty covers only failures of defective components (due to construction
defects or defects in materials) and includes replacement or repair of the components and related labor costs.
Warranty is automatically forfeited in the event of:
- tampering, deletion, removal of the identification label and/or serial number of the product
- misuse, transformation, alteration, repair of products not carried out by Laumas personnel
Laumas provides a 1-year warranty from the date of the delivery note on defects in material or manufacture of the battery.
GUIDELINES FOR PROPER DISPOSAL
This symbol on the product or packaging indicates that:
- This is electrical/electronic equipment and cannot be disposed of as municipal solid waste, but must be delivered
to a recycling center
- Improper use or disposal can pollute the environment or damage human health
- Non-compliance with these guidelines will be penalized in accordance with the regulations in force in the country
of destination
- It is recommended to dispose of the packing and packaging as required by local regulations

TABLE OF CONTENTS
USER WARNINGS...................................................................................................................1
RECOMMENDATIONS FOR CORRECT INSTALLATION OF WEIGHING INSTRUMENTS.1
MAIN CHARACTERISTICS OF THE INSTRUMENTS............................................................2
BUFFER BATTERY....................................................................................................................2
TECHNICAL SPECIFICATIONS..............................................................................................3
W100, W200...............................................................................................................................3
WDESK-LIGHT...........................................................................................................................3
WDESK-L/R, WINOX-L/R/2L, WTAB-L/R .................................................................................3
ELECTRICAL CONNECTIONS................................................................................................4
BASIC INFORMATION...............................................................................................................4
WDESK-LIGHT...........................................................................................................................4
TERMINALS LEGEND........................................................................................................................4
W100 – W200.............................................................................................................................5
CHANGING VOLTAGE 115 VAC/230 VAC (W200)...........................................................................6
TERMINALS LEGEND........................................................................................................................6
WDESK-L/R – WINOX-L/R/2L – WTAB-L/R .............................................................................7
CHANGING VOLTAGE 115 VAC/230 VAC (WDESK-L/R) ...............................................................7
INSTRUMENTS BACK TYPES...........................................................................................................8
KEY TO P, Q, X TYPE CONNECTORS..............................................................................................8
KEY TO D TYPE CONNECTORS.......................................................................................................9
RS485 SERIAL COMMUNICATION ........................................................................................10
RS232 SERIAL COMMUNICATION ........................................................................................10
LED, SYMBOLS AND KEY FUNCTIONS..............................................................................11
MENU MAP ............................................................................................................................12
SETPOINT................................................................................................................................12
SYSTEM PARAMETERS.........................................................................................................12
PROGRAMMING OF SYSTEM PARAMETERS....................................................................13
SERIAL COMMUNICATION SETTINGS .................................................................................13
SETTING THE NUMBER OF DECIMALS................................................................................13
SETTING UNIT OF MEASURE................................................................................................13
SELECTION OF THE WEIGHT TYPE TO DISPLAY...............................................................14
OUTPUTS AND INPUTS CONFIGURATION ..........................................................................14
TEST.........................................................................................................................................14
DATE AND TIME SETTING.....................................................................................................14
SETPOINT PROGRAMMING.................................................................................................15
ALARMS.................................................................................................................................15
RECEIVING PROTOCOL.......................................................................................................16
RESERVED FOR THE INSTALLER......................................................................................17
MENU LOCKING......................................................................................................................17
MENU UNLOCKING.................................................................................................................17

TEMPORARY MENU UNLOCKING.........................................................................................17
DATA DELETION AND PROGRAM SELECTION...................................................................17
KEYPAD OR DISPLAY LOCKING ..........................................................................................19
DECLARATION OF CONFORMITY - EU...............................................................................20
DECLARATION OF CONFORMITY - UKCA.........................................................................21

- 1 -
USER WARNINGS
RECOMMENDATIONS FOR THE PROPER USE OF WEIGHING INSTRUMENT
- Keep away from heat sources and direct sunlight
- Repair the instrument from rain (except special IP versions)
- Do not wash with water jets (except special IP versions)
- Do not dip in water
- Do not spill liquid on the instrument
- Do not use solvents to clean the instrument
- Do not install in areas subject to explosion hazard (except special Atex versions)
RECOMMENDATIONS FOR CORRECT INSTALLATION OF WEIGHING INSTRUMENTS
The terminals indicated on the instrument’s wiring diagram to be connected to earth must
have the same potential as the weighed structure (same earthing pit or earthing system). If
you are unable to ensure this condition, connect with an earthing wire the terminals of the
instrument (including the terminal – SUPPLY) to the weighed structure.
The cell cable must be individually led to its panel input and not share a conduit with other cables;
connect it directly to the instrument terminal strip without breaking its route with support terminal
strips.
Use “RC” filters on the instrument-driven solenoid valve and remote control switch coils.
Avoid inverters in the instrument panel; if inevitable, use special filters for the inverters and separate
them with sheet metal partitions.
The panel installer must provide electric protections for the instruments (fuses, door lock switch
etc.).
It is advisable to leave the equipment always switched on to prevent the formation of condensation.
MAXIMUM CABLE LENGTHS
- RS485: 1000 metres with AWG24, shielded and twisted cables
- RS232: 15 metres for baud rates up to 19200
- Analog current output: up to 500 metres with 0.5 mm2 cable
- Analog voltage output: up to 300 metres with 0.5 mm2cable

- 2 -
MAIN CHARACTERISTICS OF THE INSTRUMENTS
- Weight remote display with RS-232 or RS-485 serial connection;
- DIN box for panel front mounting:
W100 48x96x130 mm −drilling template 45x91 mm
W200 96x96x130 mm −drilling template 91x91 mm
WDESK-LIGHT −WDESK-L/R 122x 226x164 mm
WINOX-L/R/2L 206x286x108 mm
WTAB-L/R 315x315x180 mm
- Front panel protection rating:
W100 −W200 IP54 (IP67 front optional)
WDESK-LIGHT −WDESK-L/R IP67
WINOX-L/R/2L IP68
WTAB-L/R IP40
- Display:
W100 −W200 semi-alphanumeric, red LED, six 14 mm digits,
7 segments, 8 indicator LEDs
WDESK-LIGHT −WDESK-L
WINOX-L – WTAB-L
semi-alphanumeric backlit LCD, six 20 mm digits,
7 segments; 46 signalling symbols
WDESK-R −WINOX-R
WTAB-R
semi-alphanumeric, red LED, six 20 mm digits,
7 segments; 16 indicator LEDs
WINOX-2L semi-alphanumeric backlit LCD, six 40 mm digits,
7 segments; 46 signalling symbols
- Membrane keypad with buzzer:
W100 4 keys
W200 5 keys
WDESK-LIGHT −WDESK-L/R −WINOX-L/R/2L 6 keys
WTAB-L/R 8 keys
- Real-time clock/calendar with buffer battery.
BUFFER BATTERY
The instrument is equipped with an internal battery that allows to keep active the internal clock even
in the event of power failure.
At the first start and after long periods of inactivity, leave the instrument on for at least 12
hours to fully charge the battery.

- 3 -
TECHNICAL SPECIFICATIONS
W100, W200
POWER SUPPLY and CONSUMPTION (VDC) 12/24 VDC; ±10%; 5 W (standard)
POWER SUPPLY and CONSUMPTION (VAC) 115/230 VAC; 50-60 Hz; 6 VA
(optional for W200)
DISPLAY RANGE ±999999
NO. OF DECIMALS 0÷4
RELAY LOGIC OUTPUTS N.5 - max 115 VAC; 150 mA
SERIAL PORTS RS485, RS232
BAUD RATE 2400, 4800, 9600, 19200, 38400, 115200
HUMIDITY (non condensing) 85%
STORAGE TEMPERATURE -30°C +80°C
WORKING TEMPERATURE -20°C +60°C
WDESK-LIGHT
POWER SUPPLY and CONSUMPTION (VDC) 12/24 VDC; ±10%; 6 W (standard)
DISPLAY RANGE ±999999
NO. OF DECIMALS 0÷4
RELAY LOGIC OUTPUTS N.2 - max 115 VAC; 150 mA
SERIAL PORTS RS485, RS232
BAUD RATE 2400, 4800, 9600, 19200, 38400, 115200
HUMIDITY (non condensing) 85%
STORAGE TEMPERATURE -30°C +80°C
WORKING TEMPERATURE -20°C +60°C
WDESK-L/R, WINOX-L/R/2L, WTAB-L/R
POWER SUPPLY and CONSUMPTION (VDC) 12/24 VDC; ±10%; 6 W (standard)
POWER SUPPLY and CONSUMPTION (VAC) 115/230 VAC; 50-60 Hz; 6 VA
(optional for WDESK-L/R only)
DISPLAY RANGE ±999999
NO. OF DECIMALS 0÷4
RELAY LOGIC OUTPUTS N.5 - max 115 VAC; 150 mA
SERIAL PORTS RS485, RS232
BAUD RATE 2400, 4800, 9600, 19200, 38400, 115200
HUMIDITY (non condensing) 85%
STORAGE TEMPERATURE -30°C +80°C
WORKING TEMPERATURE -20°C +60°C

- 4 -
ELECTRICAL CONNECTIONS
BASIC INFORMATION
- It is recommended that the power supply negative pole be grounded (WDESK-D, WINOX,
WTAB: connect the earthing system to the dedicated external terminal ).
- Connect terminal “– SUPPLY” to the RS485 common of the connected instruments in the event
that these receive alternating current input or that they have an optically isolated RS485.
- In case of an RS485 network with several devices it is recommended to activate the 120 ohm
termination resistance on the two devices located at the ends of the network, as described in the
paragraph RS485 SERIAL CONNECTION.
- E/EC option (if available): selects 12 groups of 5 setpoint.
WDESK-LIGHT
2 outputs: settable setpoint.
TERMINALS LEGEND
1 12
2 13
3 14
4 15
5 16
6 17 OUTPUT No. 1
7 RS232: TXD 18 OUTPUT No. 2
8 RS232: RXD 19 OUTPUT COMMON
9 RS485: - 20 + SUPPLY (12/24 VDC)
10 RS485: + 21 - SUPPLY (12/24 VDC)
RS232, RS485: SHIELD, GND
11
To access the terminal strip you need to remove the front panel of the instrument
unscrewing the eight screws.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
12-24
VDC
supply
RS485
-
+
SUPPLY +
SUPPLY -
TXD
RXD
OUT 1
OUT 2
OUT COMMON
OUTPUTS
max 115 VAC
150 mA
RS232

- 5 -
W100 – W200
5 outputs: settable setpoint.
WARNING: connect power supply specified on the plate found on the back of the instrument.
In 115 V and 230 V versions, terminals “+ SUPPLY” and “– SUPPLY” generate continuous voltage
at 24 VDC, only to be used as power supply for instrument inputs.
RXD
19
1
13
23456
16
14 15 17 18
TXD
12
24
7811
109
21
20 22 23
12
EC OPTION
3
2
1
7
68
5
4
10
11
9
8
E OPTION
1 324567 1411910 1312 1615
162
162
to instrument
OUT 1
OUT 2
OUT 3
OUT 4
OUT COMMON
RS485
OUT 5
to OPT. E - EC
RS232
- SUPPLY
+ SUPPLY
Buttons not included in the supply
N
L
115/230
VAC
123
- SUPPLY OUT
+ SUPPLY OUT
L N
WARNING! W200 - WDOS
115 V / 230 V OPTIONS
INSTRUMENT
24 VDC
OUTPUT
OUTPUTS
max 115 VAC
150 mA
12-24 VDC
supply

- 6 -
CHANGING VOLTAGE 115 VAC/230 VAC (W200)
Remove the instrument’s power board and connect together the welding points marked in the
diagram using a stiff wire (on welding side).
TERMINALS LEGEND
1 + SUPPLY (12/24 VDC) 13
2 - SUPPLY (12/24 VDC)
RS232, RS485: SHIELD, GND
E/EC OPTION: GND
14
3 RS232: TXD 15
OUTPUT No. 5
4 RS232: RXD 16
E/EC OPTION
5 OUTPUT No. 1 17
RS485: -
6 OUTPUT No. 2 18
RS485: +
7 OUTPUT No. 3 19
8 OUTPUT No. 4 20
9 OUTPUT COMMON 21
10 22
11 23
12 24
230 VAC
115
115
230
115
115
230
115 VAC

- 7 -
WDESK-L/R – WINOX-L/R/2L – WTAB-L/R
5 outputs: settable setpoint.
WARNING: connect power supply specified on the plate found on the back of the instrument.
In 115 V and 230 V versions, terminals “+ SUPPLY” and “– SUPPLY” generate continuous voltage
at 24 VDC, only to be used as power supply for instrument inputs.
CHANGING VOLTAGE 115 VAC/230 VAC (WDESK-L/R)
Access instrument board by removing the six bottom screws and work on the welding side: join the
red points using a stiff wire.
12
EC OPTION
3
2
1
7
68
5
4
10
11
9
8
E OPTION
1 32 45671411910 1312 1615
4
to instrument
Buttons not included in the supply
1 2 3 4 5 6 7 8 9 101112131415161718192021222324
12-24 VDC
supply
RS485
OUT 5
to E-EC OPT.
-
+
SUPPLY +
SUPPLY -
TXD
RXD
OUT 1
OUT 2
OUT 3
OUT 4
OUT COMMON
RS232
OUTPUTS
max 115 VAC
150 mA
8
84
115/230
VAC
WARNING! WDESK
115 V / 230 V OPTIONS
24 VDC
OUTPUT
78
SUPPLY OUT +
SUPPLY OUT -
1LN
230 VAC
230
115
115
115 VAC
230
115
115

- 8 -
INSTRUMENTS BACK TYPES
Instrument Back type
WDESK – WINOX
P– M16x1.5 cable gland (*)
Q– Removable terminal strip
X– Atex cable gland
D– D-Sub tray (*)
WTAB D– D-Sub tray (*)
(*): including switching power supply plug 24 V 450 mA, input 100÷240 VAC, 3 meters long cable.
KEY TO P, Q, X TYPE CONNECTORS
Terminal Signal Terminal
Signal
1 14 OUTPUT No. 4
2 15 OUTPUT COMMON
3 OUTPUT No. 5 16
4 E/EC OPTION 17
5 RS485: - 18
6 RS485: + 19
7
+ SUPPLY (12/24 VDC)
115/230 VAC optional version:
+ OUTPUT (24 VDC)
20
8
- SUPPLY (12/24 VDC)
RS232, RS485: SHIELD, GND
E/EC OPTION: GND
115/230 VAC optional version:
- OUTPUT (24 VDC)
RS232, RS485: SHIELD, GND
E/EC OPTION: GND
21
9 RS232: TXD 22
10 RS232: RXD 23
11 OUTPUT No. 1 24
12 OUTPUT No. 2 L PHASE (115/230 VAC optional ver.)
13 OUTPUT No. 3 N NEUTRAL (115/230VACoptional ver.)
GROUND (115/230 VAC optional ver.)
To access the terminal strip on the WDESK instruments with cable glands, you need to
remove the bottom of the instrument unscrewing the six screws.

- 9 -
KEY TO D TYPE CONNECTORS
Connector Pin Signal Internal
terminal Internal
colour
P1
Power supply
+ SUPPLY (12/24 VDC) 7 red
- SUPPLY (12/24 VDC) 8 black
D3
Male
OUT 1 ÷ 5
E/EC option
1 OUTPUT No. 1 (max 24 V) 11 yellow
2 OUTPUT No. 2 (max 24 V) 12 blue
3 OUTPUT No. 3 (max 24 V) 13 white
4 OUTPUT No. 4 (max 24 V) 14 green
5 OUTPUT No. 5 (max 24 V) 3 orange
6 OUTPUT COMMON (max 24 V) 15 purple
7
8
9
10
11 E/EC OPTION 4 red
12 E/EC OPTION: GND 8 black
13
14
15
D4
Male
RS232 serial port
1
2 RS232: RXD 10 yellow
3 RS232: TXD 9 blue
4
5 RS232: SHIELD, GND 8 black
6
7
8
9
D5
Male
RS485 serial port
1
2
3
4 RS485: + 6 yellow
5 RS485: SHIELD, GND 8 black
6 RS485: - 5 blue
7 RS485: - 5 blue
8
9 RS485: + 6 yellow

- 10 -
RS485 SERIAL COMMUNICATION
Instrument RS485 + RS485 – SUPPLY –
W100 – W200 18 17 2
WDESK-LIGHT 10 9 21
WDESK L/R – WINOX L/R/2L – WTAB L/R 6 5 8
If the RS485 network exceeds 100 metres in length or baud-rate over 9600 are used, the network
must be terminated at its ends:
- connect a 120 ohm resistor between the ”+” and “–“ terminals of the line, on the terminal strip of
the instrument;
- close the RS485 termination jumpers on the furthest instrument (if present), otherwise connect a
120 ohm resistor on the terminal strip of the instrument.
RS232 SERIAL COMMUNICATION
Instrument RS232 TX RS232 RX SUPPLY –
W100 – W200 3 4 2
WDESK-LIGHT 7 8 21
WDESK L/R – WINOX L/R/2L – WTAB L/R 9 10 8
RS485
+-
0 VDC
-+
SUPPLY
12-24 V
+-
RS485
RS232 0 VDC
TX RX
SUPPLY
12-24 V
+-
RS232
TX RX
Transmitting instrument
(
connected to the load cell
)
Receiving instrument
Transmitting instrument
(
connected to the load cell
)
Receiving instrument

- 11 -
LED, SYMBOLS AND KEY FUNCTIONS
LED / SYMBOL Function
NET net weight (semi-automatic tare or preset tare)
zero (deviation from zero not more than ±0.25 divisions)
stability
g – kg – t – lb - pcs unit of measure in use
KEYS: W100 – W200 – WDESK-LIGHT – WDESK-L/R – WINOX-L/R/2L – WTAB L/R
KEY Short press
Long press
(3 s) Into menus
Power-on Power-off
Cancel or return to
previous menu
Select figure to be
modified or return to
previous menu item
Modify selected figure or
go to next menu item
Setting setpoint and
hysteresis Confirm or enter in
submenu
+ +
Setting general parameters
(press MENU immediately
followed by ESC)
The LEDs light up in sequence (or, in instruments with LCD display, symbol “*” is on) to
indicate that a setting and not a weight is being viewed
0

- 12 -
MENU MAP
Into menus changes are applied right after pressing the ENTER key (no further confirmation is
required).
SETPOINT
SYSTEM PARAMETERS
WDESK-LIGHT does not support E/EC option: just class 1 and just two setpoint
are available.
…
MENU
…
…
MENU + ESC

- 13 -
PROGRAMMING OF SYSTEM PARAMETERS
From the weight display, press MENU and ESC simultaneously to access the parameter setting.
SERIAL COMMUNICATION SETTINGS
MENU + ESC →:
- / : selects the communication port to configure.
- : disables any type of communication (default);
- : enables weight receiving on selected port;
- : transmission rate (2400, 4800, 9600, 19200, 38400, 115200; default: 9600);
- :
- :no parity (default);
- :even parity;
- : odd parity;
- :stop bit (1 – 2; default: 1).
SETTING THE NUMBER OF DECIMALS
MENU + ESC →:
(from 0 to 4): sets the number of decimals; it MUST be equal to the number of decimals
used by the transmitting instrument.
SETTING UNIT OF MEASURE
MENU + ESC →:
(default: ): sets the unit of measure to be used.
- : kilograms;
- : grams;
- : tons.

- 14 -
SELECTION OF THE WEIGHT TYPE TO DISPLAY
MENU + ESC →:
: sets the weight type to display:
- : gross weight;
- : net weight (preceded by letter in 7 segments display).
OUTPUTS AND INPUTS CONFIGURATION
MENU + ESC →:
The outputs are set as follows by default: /
Possible operation modes:
- (normally open): the relay is de-energised and the contact is open when the weight is
lower than the programmed setpoint value; it closes when the weight is higher than or equal to
the programmed setpoint value.
- (normally closed): the relay is energised and the contact is closed when the weight is
lower than the programmed setpoint value; it opens when the weight is higher than or equal to
the programmed setpoint value.
- : the contact will switch on the basis of gross weight.
- : the contact will switch on the basis of net weight (If the net function is not active, the
contact will switch on the basis of gross weight).
TEST
MENU + ESC →:
- Outpu Test:
: setting ensure that the corresponding output opens. Setting ensure that the
corresponding output closes.
- E/EC Option Test:
: it shows the group number of setpoint selected by the E/EC option, if the option is not
present or is not active, the message is displayed
DATE AND TIME SETTING
MENU + ESC →:
Pressing ENTER several times scrolls through days - months – years and hours – minutes.

- 15 -
SETPOINT PROGRAMMING
From the weight display, press MENU to access the setpoint setting.
MENU → → →:
- : connecting the E/EC option you can program 12 groups (classes) of different setpoint
values; otherwise you can only set the first class. The E/EC position selects the valid relay
release values.
- (from 0 to max full scale; default: 0): setpoint, weight value above which contact switching
occurs. The type of switching can be set (see section OUTPUTS AND INPUTS
CONFIGURATION).
MENU →:
- (from 0 to max full scale; default: 0): hysteresis, value to be subtracted from the setpoint
to obtain contact switching threshold for the decreasing weight. For example, with SET = 100
and hysteresis = 10, switching occurs at 90 for the decreasing weight.
ALARMS
- : no serial port set as .
- : communication problems between transmitter and receiver; check instrument electrical
connections and settings.
- : maximum displayable value exceeded (value higher than 999999 or lower than -999999).
- : the weight display exceeds 110% of the full scale of transmitting instrument.
- : in transmitting instrument the protocol has been set; change it to protocol.
- : in transmitting instrument the protocol has been set; change it to protocol.

- 16 -
RECEIVING PROTOCOL
The following string is received from the transmitting instrument:
&NxxxxxxLyyyyyy\ckckCR
where: &= 1 character of string start (38 ASCII).
N= 1 character of net weight identification (78 ASCII).
xxxxxx = 6 characters of net weight or PEAK if present (4857 ASCII).
L= 1 character of gross weight identification (76 ASCII).
yyyyyy = 6 characters of gross weight (4857 ASCII).
\=1 character of separation (92 ASCII).
ckck = 2 ASCII checksum characters calculated considering the characters between
“&”and “\”excluded. The checksum value is obtained from the calculation of XOR
(exclusive or) of the 8-bit ASCII codes of the characters considered. This obtains a
character expressed in hexadecimals with two digits that can have the values from “0” to “9”
and from ”A” to “F”. “ckck” is the ASCII code of the two hexadecimal digits.
CR = 1 character of string end (13 ASCII).
In case of negative weight, the first character on the left acquires the value “–” (minus sign - ASCII 45).
In case of weight value is under -99999, the minus sign (“-”) is sent alternated with the most
significant figure.
In case of error or alarm, the 6 characters of the gross weight are substituted by an error
message (see section ALARMS on transmitting instrument manual).
This manual suits for next models
5
Table of contents