Lenntech Testomat 2000 Fe User manual

1
Testomat 2000®Fe
Process Photometer for
Iron (Dissolved) 0 –1.0 mg/l
Operating Instructions
Lenntech
Lenntech
info@lenntech.com Tel. +31-152-610-900
www.lenntech.com Fax. +31-152-616-289
info@lenntech.com Tel. +31-152-610-900
www.lenntech.com Fax. +31-152-616-289
Lenntech

Contents
2
Contents
Contents...............................................................................................2
Important safety information.............................................................4
Intended use .........................................................................................4
Qualification of the staff ........................................................................4
Warning notices in these instructions ...................................................5
Further documents................................................................................5
Pay particular attention to .....................................................................5
General instructions..............................................................................5
Installation.............................................................................................6
Operation ..............................................................................................6
After switch-off and longer downtime....................................................6
Cleaning................................................................................................6
De-installation .......................................................................................6
Disposal ................................................................................................6
Scope of delivery ................................................................................7
Performance specifications...............................................................7
Application instructions.....................................................................8
Installation...........................................................................................9
Operating Testomat 2000®Fe in the pressure range 0.3 to 1 bar........9
Installing Testomat 2000®Fe................................................................9
Connecting the water inlet and outlet .................................................10
Water inlet...........................................................................................10
Water outlet.........................................................................................10
Connecting the power supply and devices .........................................11
Block diagram Testomat 2000®Fe .....................................................11
Internal design Testomat 2000®Fe ....................................................12
Connecting the mains voltage.............................................................13
Connecting the plant components ......................................................14
Connecting the inputs and outputs .....................................................15
Commissioning.................................................................................16
Inserting reagent bottles .....................................................................16
Filling the reagent ...............................................................................16
Opening the water inlet.......................................................................16
Instrument settings and data input......................................................17
Functions of the operating and display elements.........................17
Switching Testomat 2000® Fe on/off...................................................17
Display functions.................................................................................18
Operating elements and function keys ...............................................19
Operating system................................................................................20
Password protection and basic program.......................................21
Entering basic program data...............................................................21
Selecting the operating mode.............................................................21
Selecting the displayed unit................................................................24
Entering further basic program data ..............................................25
Internal flushing...................................................................................25
External flushing .................................................................................25
Interval pause .....................................................................................26
Limit value monitoring.........................................................................26
Hysteresis ...........................................................................................27

Contents
3
Switch functions of the limit value outputs LV1 and LV2....................28
Switch function 0, duration..................................................................28
Switch function 1, impulse ..................................................................28
Switch function 2, interval...................................................................28
Switch function 3, two-point................................................................28
Function IN1........................................................................................29
Water meter ........................................................................................29
Plant monitoring..................................................................................29
Alarm / Message.................................................................................30
Function AUX......................................................................................30
Service II.............................................................................................31
Calibration...........................................................................................31
Resetting the operating time...............................................................33
Maintenance interval...........................................................................33
Water quantity RESET.......................................................................33
Plant control RESET..........................................................................33
Description of the signal inputs/outputs........................................34
Interfaces (optional) ............................................................................35
Installing the interface cards SK910/RS910.......................................35
Monitoring of one measuring point .....................................................36
Monitoring of two measuring points....................................................36
Calculating the output currents...........................................................37
Serial interface RS232........................................................................37
SD card data logger............................................................................38
Description of the relay outputs......................................................38
Flushing (external flush valve)............................................................38
LV1 and LV2 limit value outputs .........................................................38
Measuring point 1 or 2 (measuring point switch-over)........................39
AUX (programmable function output) .................................................39
Alarm (fault message output)..............................................................40
Maintenance (output for maintenance message) ...............................40
Information menu "i" ........................................................................41
Program menu "M" ...........................................................................42
Structure of the basic program .......................................................44
Error messages / Troubleshooting .................................................45
Further information..............................................................................46
Maintenance ......................................................................................47
Description of maintenance work........................................................47
Service instructions.............................................................................48
Testomat 2000®Fe spare parts and accessories ..........................49
Accessories.........................................................................................50
Reagents.............................................................................................50
Technical data...................................................................................51
Conformity Declaration .......................................................................52
Product overview Testomat 2000®instruments ............................52
Check List Testomat 2000®................................................................54

Important safety information
4
Important safety information
Please read these operating instructions carefully and completely
prior to working with the instrument.
Ensure that these operating instructions are always available for
all users.
These operating instructions must always be passed on to the
new owner should Testomat 2000®Fe change hands.
Intended use
The process photometer Testomat 2000®Fe is a robust wet-chemical
online monitor for monitoring the dissolved iron content (II)+(III) from
0 to 1.0 mg/l using a colourimetric analysis principle.
Analysis is carried out by adding two reagents. The analysis result is
displayed after a reaction time of approx. 7 minutes (absolute meas-
uring time without flushing times).
Applications include, e.g., the monitoring of de-ironing plants or the
monitoring of process or drinking-water flows.
Always adhere to the performance limits stated in the section entitled
“Technical data”.
Always observe the requirements of the medium being measured.
To ensure correct and intended usage, always read and understand
these instructions, especially the section entitled “Important safety
information”, prior to use.
The instrument is not used as intended if
it is used in areas not specified in these instructions,
it is used in areas which do not correspond to the ones described
in these instructions.
Qualification of the staff
Assembly and commissioning require fundamental electrical and
process engineering knowledge as well as knowledge of the respec-
tive technical terms. Assembly and commissioning should therefore
only be carried out by a specialist or by an authorised individual su-
pervised by a specialist.
A specialist is someone who due to his/her technical training, know-
how and experience as well as knowledge of relevant regulations can
assess assigned tasks, recognise potential hazards and ensure ap-
propriate safety measures. A specialist should always adhere to the
relevant technical regulations.
Always adhere to hazard warnings and safety tips when using
reagents, chemicals and cleaning agents. Please adhere to the
respective safety data sheet! Download the safety data sheets for
the supplied reagents at www.lenntech.com .

Important safety information
5
Warning notices in these instructions
The warning notices in these instructions warn the user about poten-
tial dangers to individuals and property resulting from incorrect han-
dling of the instrument. The warning notices are structured as follows:
Description of the type or source of danger
Description of the consequences resulting from non-observance
Preventive measures. Always adhere to these preventive
measures.
“DANGER” indicates an immediate hazardous situation which, if not
avoided, will result in death or serious injury.
“WARNING” indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
“CAUTION” indicates a potentially hazardous situation which, if not
avoided, could result in minor or moderate injuries or property dam-
age.
“NOTE” indicates important information. If this information is not ob-
served, it may result in an undesirable result or state.
Further documents
Testomat 2000®Fe is a plant component. Therefore, always observe
the maintenance manual of Testomat 2000® /Testomat ECO®and the
documentation of the plant manufacturer.
Pay particular attention to
General instructions
Please adhere to health and safety regulations, electrical equip-
ment safety regulations, and environmental protection regulations
valid in the country of use and at the installation site.
Adhere to national and local regulations during installation and
commissioning.
Always protect the instrument against moisture and humidity. It
should never come into contact with condensation or splash water.
Do not carry out any changes or actions at the instrument which
are not described in these instructions; failure to adhere to these
instructions will negatively affect any warranty claims that you
make thereafter.
SIGNAL WORD
DANGER
!
CAUTION
!
WARNING
!
WARNING
!
NOTE

Important safety information
6
Installation
Always completely disconnect the relevant plant part before in-
stalling the instrument or connecting/disconnecting it to/from the
power supply. Secure the plant against reconnection.
Only connect the instrument to the mains voltage specified on the
rating plate.
Always observe technical data and ambient parameters.
Testomat 2000®Fe requires an interference free and stable power
supply. If necessary, use a mains filter to protect Testomat 2000®
Fe against interference voltages caused, e.g., by solenoid valves
or large motors. Never lay connecting cables parallel to power ca-
bles.
Operation
Ensure that the maximum electrical load capacity of the relay out-
puts is never exceeded.
Immediately switch off Testomat 2000®Fe and contact service
staff if malfunctioning occurs. The warranty will be void if you tam-
per with or attempt to repair Testomat 2000®Fe. Repairs must be
carried out by authorised service staff.
After switch-off and longer downtime
Ventilate the indicator leads as described in the start-up, because
longer downtimes (more than 6 hours) can cause the indicator in
the leads to retract.
Do not switch off the appliance for longer periods (e.g. over the
weekend) via the start/stop output. The indicator can retract from
the leads. This results in measurement errors after the appliance is
switched on.
Cleaning
Only use a dry, lint-free cloth for cleaning.
De-installation
Prior to de-installing a defective instrument, always write down a
description of the error (failure effect). It is only possible to repair a
defective instrument (irrespective of the warranty period) if it has
been de-installed and returned to us with a description of the error.
Please also inform us of the measured medium. Before you return
the instrument for repair work, remove the reagent bottles and en-
sure that the measuring chamber has been flushed out and is
empty.
Disposal
Dispose of the instrument in accordance with national regulations.
WARNING
!
NOTE

Scope of delivery
7
Scope of delivery
1x Testomat 2000®Fe
1x plastic bag with two screw caps with a hole and inserts for the
screw caps of the reagent bottles
1x operating instructions
Performance specifications
The process photometer Testomat 2000®Fe is used for automatically
monitoring the dissolved iron content (II)+(III) from 0 to 1.0 mg/l using
a colourimetric analysis principle. Analysis is carried out by adding
two reagents.
Simple, menu-driven operating and programming via a plain text
display
High measuring accuracy provided by a precise piston-dosing
pump
Range I: 0 - 0.65 mg/l, resolution: 0.01 mg/l
Range II: 0.7 –1.0 mg/l, resolution 0.1 mg/l
Analysis initiation:
- Automatic interval operation
(interval pause can be set from 0-99 minutes)
- External control
- Dynamic (exhaustion-dependent interval operation)
- Quantity dependent via impulse water meter
Two independent limit values with hysteresis (1, 2 or 3 bad anal-
yses) and settable switch functions
Monitoring of two measuring points (change-over via external
solenoid valves)
Internal error documentation
Programmable service address
Programmable maintenance interval for a maintenance request
Extended operating periods due to 500 ml reagent storage bottle
Optional
Interface card (0/4-20 mA or 0/2-10 V) or
Interface card RS232 (for a protocol printer)

Application instructions
8
Application instructions
Wait at least 5 seconds before switching the instrument on and
then off again at the main switch.
Trouble-free operation of Testomat 2000®Fe is only possible
when using Heyl Testomat 2000® reagents FE2005A and
FE2005B
!
Faults or problems resulting from the use of other rea-
gents will not be accepted as a warranty or fair dealing claim.
Larger quantities of heavy metal ions in the water influence the
colour reaction, especially copper above 0.01 mg/I and alumini-
um above 0.1 mg/l. Measurement is possible in the pH range 4 –
10.5. The concentration of influencing contents can be deter-
mined by using our colourimetric TESTOVAL®test kits.
Only use Testomat 2000®Fefor its intended purpose.
Careful handling of the instrument increases both its operational
reliability and service life! Therefore, carry out a visual inspection
at regular intervals as described below:
- Are the hose connections of the dosing pump free of leaks?
- Is there air inside the dosing hoses?
- Are all the water connections free of leaks?
- Are the doors of the instrument closed properly?
- Is the instrument heavily soiled?
- Are the measuring chamber and the drain duct/drain hose
clean?
Trouble-free operation is only possible when maintenance is car-
ried out on a regular basis! For maintenance and service instruc-
tions, please refer to the section entitled “Maintenance” and the
"Maintenance manual of Testomat 2000®/Testomat ECO®".
If problems occur, please refer to the section entitled "Error mes-
sages/Troubleshooting”.

Installation
9
Installation
Risks resulting from incorrect installation!
Install Testomat 2000® Fe at a location where it is protected
against dripping or splash water, dust and aggressive substances
–e. g. in a switch cabinet or on a suitable wall.
Only authorised technicians should carry out installation and
commissioning!
Information for trouble-free operation
Install the instrumentvertically and without mechanical stress.
Install the instrumentat a vibration-free site.
Operating Testomat 2000®Fe in the pressure
range 0.3 to 1 bar
Prior to installation, please check whether lower operating pressure is
required. The instrument is factory set for the operating range 1 to 8
bar. Remove the flow controller valve body to operate the instru-
ment in the operating range 0.3 to 1 bar (e.g. when using a type R
aerator). This involves removing the retaining pin from the control-
ler / filter receiver . Subsequently use the metal bracket to remove
the controller plug from the borehole. Subsequently remove the
flow controller valve body and reinsert the controller plug and the
retaining pin.
Installing Testomat 2000®Fe
Select an installation site where the water inlet hose can be kept as
short as possible (max. 5 m).
Please leave sufficient space on the left-hand side of the instru-
ment to open the door.
Drill the mounting holes as shown in the drawing on the left.
Use three screws to attach the instrument at a suitable position in
the switch cabinet or on a wall.
Remove for
pressure
range 0.3 to 1
bar
4
3
2
1
WARNING
!
NOTE

Installation
10
Connecting the water inlet and outlet
Information for trouble-free operation
The water pressure must be between 0.3 bar and 8 bar.
Avoid strong pressure fluctuations.
The measuring water temperature must be between 10°C and
40°C. Higher water temperatures can damage the parts which
come into contact with the water (e.g. filter housing, measuring
chamber)!
For temperatures above 40°C, install the KCN type cooler in the
branch line of Testomat 2000®Fe.
Water inlet
Equip the sampling line to Testomat 2000®Fe with a manually oper-
ated shut-off valve and keep it as short as possible (the max. length
of 5 metres should not be exceeded). The instrument is equipped
with a plug connector for opaque plastic hoses 6/4 x 1 (external di-
ameter 6 mm/internal diameter 4 mm, wall thickness 1 mm) as stand-
ard.
Install the connection for the branch line of Testomat 2000®Fe
directly at the main water line .
It is important that the branch line connection is laid vertically up-
wards in order to prevent dirt particles from entering the instru-
ment from the main water line.
Install a manually operated shut-off valve in the branch line to
Testomat 2000®Fe.
Use an opaque plastic hose 6/4 x 1 (max. length 5 m) for the wa-
ter inlet .
Flush the inlet to remove any dirt particles.
When operating within a pressure range of 0.3 to 1 bar or with a sup-
ply via a booster pump, please remove the valve body from the con-
troller and the filter housing. The pump should have a feeding capaci-
ty of between 25 and 35 litres/hour and be resistant to the medium
being measured.
When using a cooler
The hot water can cause burns and damage wetted parts of
Testomat 2000®Fe.
Water outlet
The feed water flows through the measuring chamber to the drain via
the outlet hose.
Connect the outlet connection of Testomat 2000®Fe to an outlet
hose (internal diameter 12 mm).
Lay this hose without backwater development and any syphon-
ing effect, e.g. via an open funnel, to the drain .
NO !!
"Sagging“
causes backwa-
ter!
CAUTION
!
NOTE

Installation
11
Connecting the power supply and devices
Risk of electric shocks during installation!
If the power supply is not disconnected prior to installation, it may
result in personal injuries, destruction of the product or damage to
plant parts.
Always disconnect the relevant plant parts before installing Tes-
tomat 2000®Fe.
Only use tested cables with sufficient cross-sections for the con-
nections.
Risk of damages caused by electromagnetic fields!
If Testomat 2000® Fe or the connecting cables are installed paral-
lel to power cables or in close proximity to electromagnetic fields,
the instrument may be damaged or measurements may be incor-
rect.
Ensure that connecting cables are as short as possible.
Always install connecting cables and power cables separately.
Connect the instrument to the protective earth conductor (for
230/115 VAC).
Protect Testomat 2000® Fe against interference voltages –e.g. via
a mains filter.
Shield the instrument against strong electromagnetic fields.
Block diagram Testomat 2000®Fe
Drawn relay positions: Instrument de-energised
WARNING
!
NOTE

Installation
12
Internal design Testomat 2000®Fe
1
Terminal block for inputs Start, Stop, IN1, IN2, and output OUT
2
Mains switch
3
Terminal block for mains inputs and mains outputs
4
Terminal block relay outputs
5
Dosing pumps
6
Water connections, inlet and outlet
7
Flow controller
8
Measuring chamber

Installation
13
Connecting the mains voltage
Only connect the instrument to the specified mains voltage. Refer to
the rating plate for the appropriate mains voltage. Connect the cables
as follows:
Loosen both fastening screws and open the upper door
.
The
terminal box is now accessible.
Pierce the required rubber cable glands with a screwdriver and
insert the cable through the bush into the terminal box (1).
Subsequently pull back the cable until the bush has been turned
over (2).
Connect the power supply to terminals PE,N,L or for 24 V instru-
ments to terminals U, V
Connect the conductor to the terminal block as shown on the left
.
Ensure that the leads are held securely in the terminals.
Proceed as shown in figure to loosen the connection.
Terminals
Terminal
description
Type
Function
Comment
PE
IN
Protective earth conductor
(5x)
Only for mains
115/230 V !
N (U)
L (V)
IN
Mains, N=neutral conduc-
tor (U=24 V)
Mains, L=phase (V=24
V)
Mains input
24 V / 115 V / 230 V
n
l
OUT
Neutral, switched (8x)
Phase, switched (8x)
Mains for consumers,
max. 4 A
Insert the conductor with
ferrule or the solid conductor
into the round cable input.
1
2
1. Insert a screwdriver into the
square opening without force in
order to open the terminal.
2. Once the terminal has been
opened, remove the conductor.

Installation
14
Connecting the plant components
Connect the plant components to the output terminals of relays 1
to 19 (e.g. valves)
If the plant components require mains voltage, connect the
switched mains voltage (l) to the common contact of the re-
spective relay (see the connection example for 230 VAC on the
left)
Connect the neutral conductor of the plant component to one of
the terminals (n)
For components with a protective earth conductor connection,
connect it to the PE connection
Ensure that the leads are held securely in the terminals
No.
Terminal de-
scription
Typ
e
Function
Comment
1
2
Flushing
OUT
External flush valve
Volt-free relay output,
max. 240 VAC, 4 A
3
4
5
LV1
OUT
Limit value output 1 –Normally closed
Limit value output 1 –Normally open
Limit value output 1 –Common
Volt-free relay output,
max. 240 VAC, 4 A
6
7
8
LV2
OUT
Limit value output 2 –Normally closed
Limit value output 2 –Normally open
Limit value output 2 –Common
Volt-free relay output,
max. 240 VAC, 4 A
9
10
11
Measuring
point 1/2
OUT
Measuring point 1 –Normally closed
Measuring point 1 –Normally open
Measuring point switch-over - Common
Volt-free relay output,
max. 240 VAC, 4 A
12
13
AUX
OUT
Universal output
Volt-free relay output,
max. 240 VAC, 4 A
14
15
16
Alarm
OUT
Fault message output –Normally closed
Fault message output –Normally open
Fault message output –Common
Volt-free relay output,
max. 240 VAC, 4 A
17
18
19
Mainte-
nance
OUT
Maintenance message –Normally closed
Maintenance message –Normally open
Maintenance message –Common
Volt-free relay output,
max. 240 VAC, 4 A
Connection example
Limit value contact LV 1
switches mains voltage
n
l
3
4
5
e.g.
solenoid
valve
PE

Installation
15
Connecting the inputs and outputs
Testomat 2000® Fe has the following connections for control and
monitoring functions.
Do not connect external voltage to these connections!
Ensure that the leads are held securely in the terminals.
Use the two fastening screws to close the upper door once instal-
lation has been completed.
For more information, please refer to the section entitled "Description
of the signal inputs/outputs".
No.
Terminal
description
Typ
e
Function
Comment
20
21
Start
IN
External analysis start
Common earth for inputs
Only for volt-free normally
open
22
23
Stop
IN
External analysis stop
Common earth for inputs
Only for volt-free normally
closed/normally open!
24
25
IN1
IN
Universal input 1
Common earth for inputs
Only for volt-free normally
closed/normally open!
26
27
IN2
IN
Universal input 2 (water meter)
Common earth for inputs
Only for volt-free normally
open
┴
28
29
OUT
OUT
Earth
0/4 - 20 mA galvanically separated
or serial interface RS232
Earth = ┴
28 = (+) or (TxD)
29 = (-) or (RxD)

Commissioning
16
Commissioning
Handling reagents
Please adhere to the respective safety data sheet!
Trouble-free operation of Testomat 2000®Fe is only possible
when using Heyl Testomat 2000® reagents FE2005A and
FE2005B!
Caution: Observe correct allocation of the reagents A and B to
the dosing pumps: A = left, B = right!
Inserting reagent bottles
Open the lower housing door by pulling on the right-hand side.
Remove the cap from the reagent bottle.
Remove the plastic bag from inside the lower housing door. The
plastic bag contains the screw cap with hole and the insert
for the screw cap.
Connect the parts as shown on the left.
Screw the hose connector of the intake hose hand-tight into
the insert .
Place the insert with the screwed-in intake hose into the reagent
bottle.
Now screw the screw cap with hole hand-tight onto the rea-
gent bottle .
Filling the reagent
Switch the instrument on and press the "STANDBY" key
.
During operation, the pumps (DOSIClip) automatically extract
the reagents.
To ensure that reagent is available for the initial analyses, the
intake hose and the transport hose must be filled with rea-
gent from the pump up to the measuring chamber.
Press the "manual" key several times until the intake hose
and the transport hose are filled with reagent up to the measur-
ing chamber.
If necessary, manually tighten the hose connectors of the intake
and transport hose slightly in case of bubble formation.
Opening the water inlet
Open the lower housing cover.
Slowly open the manually operated shut-off valve to prevent the
measuring chamber overflowing. The flow regulator requires a few
seconds to function correctly.
Make sure that the water conducting parts are not leaky.
CAUTION
!
NOTE

Functions of the operating and display elements
17
If water sprays from the vent hole of the measuring chamber
, reduce the amount of inlet water via the manually operated
shut-off valve. It should take 2 to 6 seconds to fill the measuring
chamber!
Instrument settings and data input
Please read the following information before carrying out settings
and entering data for operating the instrument.
Functions of the operating and dis-
play elements
The Testomat 2000® Fe display shows operating statuses and meas-
ured values. The input keys for programming (cursor block) and the
function keys are located underneath the display.
Switching Testomat 2000® Fe on/off
(1) Mains switch
Use this switch to switch the instrument on or off.
(2) Instrument fuse (inside the instrument)
This fuse protects the instrument and the outputs against
overloads and short circuits.
1
(2)

Functions of the operating and display elements
18
Display functions
1 Status of limit value displays (red/green)
The display 1 illuminates red if limit value 1 has been reached or
exceeded. The display 1 illuminates green if the value falls below the
limit value. The same principle applies to limit value 2 and display 2.
2 Text display (4 lines)
Displays the current analysis result as well as all important statuses
and programming data
2a = The current measured value for measuring point 1 (M1:) and 2
(M2:) is displayed in line 2 and 3.
Value exceeds the measuring range = ">" e.g. M1: > 1.0 mg/l
2b = The set limit values LV1 and LV2 are displayed in line 4.
3Alarm (red)
Indicates malfunctioning/error message or warning message.
4 Maintenance message (yellow)
Indicates current maintenance requests
5 Status display of the active instrument components (line)
Six displays indicate the current instrument and analysis status
Cancelling error messages/warning messages
Press to acknowledge the message and, if necessary, elimi-
nate the cause of the fault.
Inlet
valve
Analysis
running
Outlet
valve
Dosing
message
Analysis
stop valve
No
function
All error and
warning messages
are alternately
shown in line 1 of
the standard
display!
NOTE

Functions of the operating and display elements
19
Operating elements and function keys
If you wish to carry out settings or enter data, or if alterations are
necessary, press the "M" key to open the programming mode. Use
this key when in the menu to go to the higher order menu items or to
exit the programming mode. Please note: The basic program is
password protected!
Programming keys (cursor block)
Use the programming keys (cursor block) to navigate in the menu, to
select the desired functions and to enter necessary instrument and
plant specific data. Press the "ENTER" key to select the submenu
item and to confirm and accept the selection or data input.
Function keys
Manually start an analysis via the "Manual" key (1).
Set the instrument to standby mode via the "STANDBY"
key (2) (automatic analyses are not carried out: analysis
stop)
Acknowledge error and warning messages via the
"Horn" key (3).
Open the program menu for user-specific and instrument-
specific settings via the "M" key (4).
Retrieve all instrument information and settings via the
"i" key (5).
(M)enu key
1
2
3
4
5
Programming
keys
(cursor block)
Cursor block

Functions of the operating and display elements
20
Operating system
Instrument settings and data input
Date, time and weekday input
Press the "M" key
The basic menu " " appears
Use the cursor block to select the desired menu item
" "
The selection appears in CAPITAL LETTERS.
Press "ENTER" to confirm your selection
The selected submenu " " appears.
The menu item " has already been selected (capital letters).
Press "ENTER" to confirm the menu item " ".
The cursor flashes in the date field:
Use the cursor keys to select the desired number.
Use the keys to move the cursor to the next input field.
Repeat this input process until the year has been entered.
Press "ENTER" to confirm the entry.
The date has now been entered.
Exit the menu item " " in order to set the time.
Use the cursor block to select the desired menu item
" ".
Press "ENTER" to confirm your selection.
The cursor flashes at the first position of the time:
Use the cursor keys to select the desired number.
Use the keys to move the cursor to the next input field.
Repeat this input process until the seconds have been entered.
Press "ENTER" to confirm the entry.
The time has now been entered.
Exit the menu item " " to set the weekday.
Use the cursor block to select the desired menu item
"".
Press "ENTER" to confirm your selection.
Use the cursor block to select the current weekday.
Press "ENTER" to confirm the entry.
Press the "M" key twice to end programming.
The standard measured value display appears on the display
Testomat instrument
(display) in display mode
Table of contents
Other Lenntech Measuring Instrument manuals