Lenntech TESTOMAT 2000 THCl User manual

GB
TESTOMAT®2000 THCl
Combined Process-Photometer for total chlorine 0 - 2.5ppm
and Process-Titrator for water hardness 0.25 - 2.50°dH
Operating Instructions
Lenntech
Lenntech
info@lenntech.com Tel. +31-152-610-900
www.lenntech.com Fax. +31-152-616-289
info@lenntech.com Tel. +31-152-610-900
www.lenntech.com Fax. +31-152-616-289
Lenntech

Contents
Introduction ..................................... 1
Short description ................................ 1
Notes for the user ................................ 2
Notes about the operational reliability ............... 2
Safety notes .................................... 2
Installation and commissioning .................... 3
Installation ..................................... 3
Electrical connection ............................. 3
Mains water supply .............................. 3
Water to drain ................................ 3
Commissioning ................................. 4
Installation diagram (example) ................... 4
General description .............................. 5
Internal construction ............................. 5
Description of the electrical connections .............. 6
Terminal block identification ..................... 6
Description of displays and operating features ......... 7
Display functions ................................ 7
Status of limit value displays ..................... 7
Measured value display ......................... 7
Limit value display ............................. 7
Alarm and maintenance message ................. 7
Status display .................................. 8
Description of the relay outputs ..................... 8
(Flush) External flush valve ...................... 8
LV1 and LV2 Limit value outputs ................. 8
1 or 2 Measuring points switch-over ............... 8
AUX Programmable function output ............... 9
Alarm fault message output ..................... 9
Maintenance message ......................... 9
Description of the signal inputs and outputs .......... 10
Start external analysis start ..................... 10
Stop external analysis stop ..................... 10
IN1 and IN2 universal inputs .................... 10
OUT Interface ouput (optional) .................. 10
Function characteristics ......................... 11
Mode of operation .............................. 11
Analysis cycle ................................. 11
Displayed unit ................................. 11
Measuring points switch-over ..................... 11
Setting the timer ................................ 12
Monitoring of limit value .......................... 12
Hysteresis .................................... 12
Logic functions ................................. 13
Function IN1 .................................. 13
Alarm / Message ............................... 13
Function AUX .................................. 14
Service II ..................................... 14
Reset operating time .......................... 14
Maintenance interval .......................... 14
Interfaces (optional) ............................. 14
Structure of the menu ........................... 15
Selection and input ............................. 15
Start menu .................................. 15
Selection ................................... 15
Input ...................................... 15
End menu .................................. 15
Information menu "i" ............................ 16
Structure of the "i" menu ......................... 16
Call .......................................... 16
Customer service ............................... 16
Operating values ............................... 16
Program values ................................ 16
Error registration ............................... 16
Maintenance .................................. 16
Program menu "M" .............................. 17
Service I ...................................... 17
Input reagent ................................ 17
Manual operation ............................. 17
Confirm maintenance ......................... 18
Diagnosis .................................. 18
Date/time of day ................................ 18
Basic program ................................. 18
Program values ................................ 18
Structure of the basic program .................... 19
Error messages / trouble shooting ................. 20
Further information ............................. 21
Maintenance ................................... 21
Description of maintenance work ................... 21
Cleaning the measuring chamber & sight-glass windows 21
Cleaning the filter housing ....................... 21
Notes on unit care .............................. 21
Spare parts list Testomat®2000 ................... 22
Technical appendix ............................. 23
Block diagram "Testomat®2000" ................... 23
Technical data ................................. 23

1 / 23
Introduction
These operating instructions describe the installation, operation and programming of the combined process photometer
and titrator Testomat®2000 TH/Cl.
We recommend that, whilst familiarising yourself with the operation of the unit aided by this manual, you have
immediate access to the unit in order to perform the functions and combinations as described. As certain functions are
interrelated, it is advisable to follow the instructions in the given order.
Should problems or questions arise which are not described in this manual and/or cannot be solved, our customer
service is always at your disposal.
Try to identify the problem as accurately as possible and record the conditions under which it has occurred. This will
enable us to offer you swift effective assistance.
Symbols and abbreviations used in these instructions:
LNote for the user "STANDBY" = STANDBY lamp is “ON”
Important to observe / warning note "M" = Press menu key M TTip: Helpful hint
ºSERVICE ºMANUAL OPERATION ºFLUSH = Sequence in a menu selection
Short description
The Testomat®2000 THCl is an online monitor for monitoring the concentration of total chlorine in the measuring range
0 - 2.5 mg/l (ppm) using the colorimetric analysis principle and determines the water hardness through titration.
The analysis of total chlorine is implemented by adding three reagents, after a reaction time of approx. 1 minute
(measuring time without flush time) the analysis result is shown.
Areas of use are, e.g. the protection of reverse osmosis membranes from blocking due to hardness and from
destruction due to a high content of chlorine. Two independent programmable limit value contacts are available for
monitoring. The analysis result can be recorded using the optional printer board (interface SK910, item no. 270305) and
a dotted or continuous line printer (0/4-20mA).
Parameter / Measuring range
water hardness
TH 2025 total chlorine
Unit
°dH
(Resolution) 0.25 - 2.50
(0.05) -
°f
(Resolution) 0.45 - 4.48
(0.1) -
ppm CaCO3
(Resolution) 4.47 - 44.7
(0.9) -
mmol/l
(Resolution) 0.04 - 0.45
(0.01) -
mg/l (ppm) -0.0 - 2.5
(0.1)
LAconsistently reliable result can be guaranteed by the Testomat®2000 units only with the use of
original HEYL Testomat®2000 indicators and reagents! Failure or problems which are caused by the use
of different indicators are not covered by the warranty or refund services.
Larger quantities of heavy-metal ions in the softened water can affect the colour reaction, especially iron above
0.5 mg/l, copper above 0.1 mg/l and aluminium above 0.1 mg/l (brownish-red colour indication).The measuring
is possible in a range of pH 4 - 10.5.
At a concentration of more than 100 mg/l of CO2(carbon dioxide) in the water the analysis can be affected.
Fault influences:
Concentration values of these ingredients can be determined safely and simply with our TESTOVAL®colorimetric test
comparators.

2 / 23
Notes for the user
•Repeated switching 'on' and 'off':
Wait at least for 5 seconds before you switch the unit repeatedly 'on' or 'off' at the main switch.
• Observation of the environmental conditions:
In order to guarantee a reliable operation, the unit must only be used under the environmental conditions described
in the technical data. Protect the controller against excessive humidity, condensation and water-splash.
• Safety seal:
The original seals attached during manufacture (e.g. EPROM labels) must not be broken, otherwise all warranty
rights are lost.
• Malfunction / repair of defective units:
The repair of a defective unit is only possible when the unit is dismantled and returned to us with a description of the
fault. In addition, please inform your supplier of the reagent type used.
Before you return the unit for repair, remove the reagent bottle and ensure that the measuring chamber is flushed out
and empty.
• Electrical load capacity
The maximum electrical load capacity of the relay outputs and the total power rating must not be exceeded.
• Operate the Testomat® 2000 THCl strictly in accordance with the manufacturer's instructions.
• Environmental protection regulations
Please observe the environmental protection regulations, collect large amounts of any unused reagent which you can
no longer use and send this to us for correct disposal.
Operational checks
Careful handling of the unit increases both the operational reliability and the service life!
Therefore, the following visual inspection should be carried out at regular intervals.
•Are the hose connections with the dosing pump free of leaks?
•Is there any air inside the dosing hoses?
•Are all the water connections free of leaks?
•Are the doors of the unit properly closed? (Check door seals!)
•Is the unit unduly contaminated with dirt?
Maintenance and servicing notes
(See page 21)
Safety notes
• The unit must be installed and operated in compliance with the relevant standards (e.g. DIN, VDE, UVV), or in
accordance with the regulations laid down by the individual country.
• Some functions (e.g. the manual analysis) allow the direct manipulation of the installation without locking or
monitoring. These functions may only be used by trained staff and for this reason are only accessible after a
password has been entered.
• If you observe the unit malfunctioning, switch it off immediately. Then shut off the water supply and contact your
supplier.
•Do not attempt to repair the unit yourself (loss of warranty rights); instead always get in touch with authorised service
staff. This is the only way a reliable and safe operation of the unit can be ensured.
•After a protective circuit (fuse) has tripped, first of all, attempt to correct the cause of the malfunctioning (e.g. replace
the solenoid valve) before reactivating the protective circuit. Frequent tripping is always due to a fault which, in
certain circumstances, may also cause damage to the unit.
• Observe the safety notes about working with reagents, chemicals and cleaning agents.
Ignoring these notes can result in damage to the unit as well as to the installation and may result in a loss of
warranty rights.

3 / 23
360
266
Dimensioned sketch
Installation and commissioning
Installation and commissioning must be undertaken only
by authorised technicians!
Installation
The unit should be fixed vertically!
Avoid tension of the housing.
The unit doors swing to the left when they are opened. Please
ensure that there is sufficient space for opening them. In this
way, you facilitate the electrical installation and later
maintenance and service work.
Electrical connection
Check supply voltage on the rating plate!
Basic requirement
The external cables (e.g. water meter, interface) should be kept as short as possible and clear of power cables.
Connection
Loosen both fastening screws and open the upper door. Pierce the required rubber cable glands with a screwdriver and
insert the cable (1). Then pull back the cable until the bush (2) has been turned over. Ensure that the leads are held
securely in the terminals, then close the upper door after the installation is completed using the two fastening screws.
Mains water supply
The temperature of the sample water must be between 10 °C and 40 °C. Higher water temperatures can lead to
damage to all parts coming into contact with water (e.g. filter housing, measuring chamber)!
Lower water temperatures can cause misting on the sight-glass windows.
LIn the event of higher temperatures, the KCN type cooler must be installed in the branch line of the Testomat®
2000 THCl.
Hot water can cause scalding!
The sampling tube to the Testomat®2000 THCl must be equipped with a hand-operated shut-off valve and be kept as
short as possible (not longer than a maximum of 5 metres). It is important that the branch line connection is taken
vertically from the top of the main soft water line in order to prevent dirt particles from entering into the measuring
chamber. Remove the valve body out of the regulator/ filter housing when operating the Testomat®2000 THCl with a
pressure range of 0.1 to 1 bar.
Plug connector
The unit is equipped as a standard with a plug connector for opaque plastic hoses 6/4 x 1 (external diameter 6 mm/
internal diameter 4 mm).
Quick-acting coupling (accessory: optional adapter for water inlet, item no. 40123)
When fabric reinforced pressure hoses (e.g. for existing installations) are used, please replace the plug connector at the
controller and filter housing with a plug for the quick-acting coupling (not included with the delivery).
Water to drain
The feed water flows through the measuring chamber then, via the outlet pipe, to the drain (hose connection internal
diameter 14 mm). Make sure, by using a funnel for example, that the water can run freely to drain and cannot back-up
into the measuring chamber. A hose impervious to light should also be used for the drain pipe (to discourage algae
formation).

4 / 23
Commissioning
1. Full reagent bottles must be connected before commissioning and switching on. Attach the vacuum connection onto
the reagent bottle using the union nut. Please observe the correct allocation of the reagents A, B and C to the dosing
pumps: TH2025 = DOSIClip 1 (left), Cl2025A = DOSIClip 2, Cl2025B = DOSIClip 3, Cl2025C = DOSIClip
4 (right).
2. Switch the unit 'on' and press the "STANDBY" key. In this mode, an analysis is prevented from being carried out
before a correct program has been entered which could possibly result in an error or alarm message.
3. Then bleed the dosing pump and the capillary by repeatedly pressing the "Manual" key on the dosing pump. Ensure
all air is removed from the tubing! (If necessary tighten up the connections.)
4. Program the unit as applicable to your requirements, e.g.: - Mode of operation - Displayed units
-Limit values -Flush time / interval
For a description of programming see from page 17 onward.
5. Enter the correct fill status of the reagent bottles:
"M" ºSERVICE IºINPUT REAGENTS ºREAGENT TH FILLING (100%)
"M" ºSERVICE IºINPUT REAGENTS ºREAGENT A FILLING (100%)
"M" ºSERVICE IºINPUT REAGENTS ºREAGENT B FILLING (100%)
"M" ºSERVICE IºINPUT REAGENTS ºREAGENT C FILLING (100%)
6. Then remove all air from the water supply side of the unit by manual flushing.
"M" ºSERVICE ºMANUAL OPERATION ºFLUSH ("ENTER" press repeatedly).
Continue flushing until no bubbles can be seen in either the measuring chamber or the filter housing.
7. Check all connections for tightness.
8. Carry out the first analysis by pressing the "Manual" key.
Installation (example):

5 / 23
Terminal block for inputs
Start, Stop, IN1, IN2 and output OUT
F7 T0,08A F5 T0,2A F6 T0,315A
F8 T1A
F1 T0,1A/ 240V
F4 T0,16A
230 V
F1 T0,1A
230 V
F9 T4A
H 250 V
J6
J4 DOSI 3
J3 DOSI 2
J2 DOSI 1
J5 DOSI 4
J11
J13J8J7
E-Ventil A-Ventil 24 V
Steckplatz Flexkabel
Messkammer
17
21
18
*
2
14 13 *
Mains water supply:
Inlet with primary filter and pressure regulator,
outlet
Valve
block
Dosing pump 1
TH2025
Plug-socket holder J4 = Dosing pump 3
Plug-socket holder J5 = Dosing pump 4
Dosing pump 2
Cl2250A
TH2025
TESTOMAT
2000
Dosing pump 4
Cl2250C
Plug-socket holder J2 = Dosing pump 1
Plug-socket holder J3 = Dosing pump 2 Dosing pump 3
Cl2250B
CL2250C
TESTOMAT
2000
General description
Internal construction

6 / 23
Description of the electrical connections
Terminal block identification
IN = input, OUT = output
No. Terminal Type Function Note
-PE IN Mains-Protective earth (5 x)
-
-N
LIN Mains, N = Neutral
Mains, L = Live Mains-input 230 - 240 V AC
(or 24V AC)
-
-n
lOUT Neutral switched (8 x)
Mains switched (8 x) Mains voltage, max. 4 A
1
2Flush OUT External flush valve Volt-free relay output
max. load 240 V AC, 4 A
3
4
5LV1 OUT Limit value output 1 - Normally closed
Limit value output 1 - Normally open
Limit value output 1 - Common
Residual hardness
Volt-free relay output
max. load 240 V AC, 4 A
6
7
8LV2 OUT Limit value output 1 - Normally closed
Limit value output 1 - Normally open
Limit value output 1 - Common
Chlorine
Volt-free relay output
max. load 240 V AC, 4 A
9
10
11
Measuring point
1/2 OUT Measuring point 1 - Normally closed
Measuring point 2 - Normally open
Measuring point switch-over -Common
Volt-free relay output
max. load 240 V AC, 4 A
12
13 AUX OUT Universal output Volt-free relay output
max. load 240 V AC, 4 A
14
15
16 Alarm OUT Fault message - Normally closed
Fault message - Normally open
Fault message - Common
Volt-free relay output
max. load 240 V AC, 4 A
17
18
19 Maintenance OUT Maintenance message - Normally closed
Maintenance message - Normally open
Maintenance message - Common
Volt-free relay output
max. load 240 V AC, 4 A
20
21 Start
2
IN External analysis start
Common earth for inputs Only for volt-free normally open
contact!
22
23 Stop
2
IN External analysis stop
Common earth for inputs Only for volt-free normally open/
normally closed contact!
24
25 IN1
2
IN External measuring point switch
Common earth for inputs Only for volt-free normally open/
normally closed contact!
26
27 IN2
2
IN Universal input 2 (Water meter)
Common earth for inputs Only for volt-free normally open
contact!
28
29 OUT OUT 0/4 - 20 mA or serial interface Potential output !
28 = 0/4 - 20 mA, 29 =
2

7 / 23
STANDBY
Description of displays and operating features
(1) Power switch
The 'ON'/'OFF' switch is located on the right-
hand side panel. Use it to switch the unit on
and off.
(2) Unit fuse (inside the unit)
Protects outputs against overload and short
circuit.
3Status of limit value displays
Displays the status of the limit values of
residual hardness LV1 (1) and chlorine LV2 (2).
4Text display
Displays the current analysis, all important
status results and programming data in a 4-line
LC-Display.
5Alarm
Displays a function fault.
6Maintenance message
Displays a maintenance request.
7Status display
6LEDs signal the current status (analysis and unit
status) of the Testomat®2000 THCl.
8Programming keys (Cursor block with ENTER)
These keys are used to enter all values and
programming data.
Function keys:
9Manual" = manual start of analysis
10 STANDBY = manual analysis stop/standby
11 "Horn" = cancels alarm message
12 i-key
Call-up all unit information (see i-Menu).
13 M-Taste
Call-up the programming menu for user and specific unit
settings (see M-Menu).
Display functions
Status of limit value displays 1 and 2
The display signals the status of the limit values.
1: The red display lamp lights up if the limit value 1 (residual hardness) is reached or exceeded.
Lights up green if the measured value falls below the set limit value.
2: The red display lamp lights up if the limit value 2 (chlorine) is reached or exceeded.
Lights up green if the measured value falls below the set limit value.
Measured value display
The current measured value for the residual hardness (TH:) and for the total content of chlorine (CI:) are shown in line
2 and 3, e.g. TH: 0.37 °dH or Cl: 0.73ppm
When the measured value exceeds the measuring range ">" is displayed, e.g. TH:>2.50 °dH or Cl: > 2.5 mg/l
Limit value displays
The set limit values are shown in the bottom display lines 3 and 4.
1: residual hardness, 2: total chlorine
Alarm and maintenance message
Display of present error messages (red) and of present maintenance request (yellow).
LError messages are displayed alternately with the normal display text and can only be deleted by cancellation and
correction of the fault.

8 / 23
Status display
The display indicates active
unit components.
Description of relays outputs
Flush external flush valve
We recommend the fitting of a programmed flush valve upstream of the unit if, when installing the unit, the use of a long
sampling tube is unavoidable
Also, in cases where the unit is used for monitoring two measuring points, an external flush valve should be installed in
order to prevent faulty measurements caused by mixing of the samples.
Immediately before each analysis the external flush valve is opened for a programmed period, thereby allowing the tube
up to the Testomat®2000 THCl to fill with the water to be measured. Please ensure that the programmed flush time is
sufficiently long.
The flush time is set with menu item
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºFLUSH TIMES/INTERVAL ºEXTERNAL FLUSH TIME
LV1 and LV2 Limit value outputs
Two volt-free relay contacts are available for reporting that a limit value has been exceeded. For both contacts the limit
values, the hysteresis and the function can be independently programmed:
Function Type of contact Action
LV1
Sactive at
limit value in excess of
limit value 1 (residual hardness)
(Measuring point 1)
volt-free
change-over contact
programmable:
SContinuous contact
SImpulse (1 - 99 seconds/minutes)
SInterval (1 - 99 seconds/minutes)
SHysteresis (1, 2 or 3 limit value in excess)
LV2
Sactive at
limit value in excess of
limit value 2 (total chlorine)
(Measuring point 2)
volt-free
change-over contact
programmable:
SContinuous contact
SImpulse (1 - 99 seconds/minutes)
SInterval (1 - 99 seconds/minutes)
SHysteresis (1, 2 or 3 limit value in excess)
Menu values:
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºLIMIT VALUES
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºFUNCTION LV1
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºFUNCTION LV2
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºHYSTERISIS LV1
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºHYSTERISIS LV2
(Measuring points 1 or 2) Measuring points switch-over
If the unit is used for monitoring two measuring points, the solenoid valves (individual valves or one 3/2-way distributing
valve) must be connected to the corresponding sampling tubes. The switch-over can occur automatically: The analyses
are carried out alternately from measuring point 1 or 2.
Or via an external request: Inlet IN1 active = next analysis from measuring point 1 (hardness)
It is important that the terminals are strictly allocated to the appropriate measuring points and to the parameter to be
measured:
Terminal 9 = Measuring point 1= Residual hardness
Terminal 10 = Measuring point 2 = total chlorine
IN1 active = measurements only from measuring point 1
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºMEASURING POINTS 1/2

9 / 23
AUX Programmable function output
The function of this volt-free relay output is programmable:
1. Message of a current analysis
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºFUNCTION AUX ºCONTACT DURING ANALYSIS
and/or
2. Contact before an analysis, e. g. control of a cooler
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºFUNCTION AUX ºCONTACT BEFORE ANALYSIS
or
3. Contact after an analysis
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºFUNCTION AUX ºCONTACT AFTER ANALYSIS
Alarm Fault message
The "Alarm" output is a volt-free change-over relay contact. During trouble-free operation the contact between the
terminals 14 - 16 is closed and the one between terminals 15 - 16 is open. When the voltage fails, the contact between
terminals 15 - 16 is closed and the one between 14 - 16 is open.
The unit is equipped with a range of monitoring functions. You can define certain situations as a fault and program a
corresponding message either as continuous contact (A) or as a message impulse (M).
-With a continuous contact, the "Alarm" output remains activated (with terminals 15 and 16 closed) as long as the fault
persists.
-For a message impulse, the output is alternately switched 'on' for 2 seconds and then switched 'off' for 5 seconds.
-If several faults with different programmed messages are reported simultaneously, the output is switched to
continuous contact.
-The red LED "Alarm (5)" and the text on the display indicate a fault.
-The error signal at the "Alarm" output is deleted by cancelling the fault with the "Horn" key.
-An error message can only be deleted after the fault has been corrected.
-Exception:The maintenance date has been exceeded. This message is confirmed in the M-menu, see below
(maintenance).
-Each new fault is entered into the fault history (also see i-menu, page 16).
The following faults activate the "Alarm" output and are displayed:
The following conditions always trigger an error
message:
Power failure
Low-water pressure
Function fault optics
Function fault dosing pump
Function fault outlet to drain
Function fault failure 24V
Conditions which can be programmed as faults:
Indicator low level
Reagent low level (A, B, C)
Measuring fault dirtiness
Measuring fault turbid
Transfer error
Measuring range exceeded
Maintenance interval exceeded
Error messages are described on page 20.
Maintenance, maintenance message
The
"Maintenance" output is a volt-free change-over relay contact. During trouble-free operation and without a
programmed maintenance interval the contact between terminals 17 and 19 is closed and the one between 18 and 19
is open.
The unit is equipped with a range of monitoring functions and a programmable maintenance interval. The maintenance
message is always a continuous contact.
A maintenance request is displayed by the yellow "Maintenance" LED.
The maintenance display can only be deleted after the condition has been corrected, or after the maintenance request
has been confirmed.
"M" ºSERVICE ºCONFIRM MAINTENANCE
The following conditions activate the "Maintenance" output:
Reagent low level
Measuring chamber dirty (Measuring fault dirtiness)
Maintenance date reached
For a more detailed description of the programming refer to page 17 onward. For general maintenance refer to page 21.

10 / 23
Description of the signal inputs and output
LConnect the signal inputs "Start", "Stop", "IN1" and "IN2" only with volt-free contacts!
Start external analysis start
Stop external analysis stop
Function Type of contact Test time Action
Start
SExternal analysis start
(e.g. from the process controller)
normally open
volt-free! none
SMode of operation EXTERNAL starts
an analysis by triggering a contact at
the input
Stop
SExternal analysis stop
(e.g. via flow controller or from the
process controller)
programmable
normally closed/
normally open
volt-free!
none
SAs long as the contact at the input is
'open' or 'closed', no analyses are
carried out.
M" ºBASIC PROGRAM ºPROGRAM VALUES ºINPUT STOP
IN1 and IN2 Universal inputs
Function Type of contact Test time Action
IN1
SExternal measuring point switch-over
(suppression of measuring point 2)
programmable
normally closed/
normally open
volt-free!
none
SAs long as the contact at the input is
'open' or 'closed', the analyses are
carried out by measuring point 1
IN2
SWater meter input
normally open
volt-free! none SQuantity recording to start an analysis
and for plant monitoring
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºINPUT IN1
OUT Interface output (optional)
Function Terminal Test time Action
Programmable interface
S0 - 20 mA
S4 - 20 mA
max. load
500 Ohm -SMeasuring value residual hardness
(measuring range 0-10 / 4-12 mA)
SMeasuring value total chlorine
(measuring range 10-20 / 12-20 mA)
Serial interface RS 232 serial bus
(2-cable-line) -SSee description of the plug-in circuit
card RS910
Change the function of the output by changing the plug-in circuit board.
"M" ºBASIC PROGRAM ºPROGRAM VALUES ºINTERFACES

11 / 23
MODE OF OPERATION
sstt
M E
External (START)
Volume interval
TIME-CONTROLLED *
DISPLAY OF UNIT Cl
sstt
M E
Display in ppm
DISPLAY IN mg/l *
4, 5, 6
7
8
3
2
2
1
Function characteristics
Mode of operation (analysis controller)
1. Time control: Internal triggering by a timer. The shortest interval = 0
minutes between analyses, longest interval = 99 minutes.
(see page 12, interval periods)
LThe analysis interval is determined by the duration of the supplementary
program AUX, the set flush times (internal and external), the programmed
pause (interval) and the duration of the analysis. The analysis duration is a
direct function of the value to be measured.
2. Quantity control: Triggered by the water meter.
Minimum interval = 1 litre, maximum interval = 9999 litres.
After the programmed water quantity is measured, the analysis is carried
out. Prior to the analysis, the capillary and the measuring chamber are
flushed (observe the programmed flush times).
3. External analysis start via contact “Start” input
LThe current analysis interval can be interrupted by making contact at the
"Stop" input.
Analysis cycle (example with schematic cycle diagram)
1Supplementary program AUX before analysis
2Flush branch line and measuring chamber (note flush time of the sampling
line), TSE and TSI
3Fill measuring chamber
4, 5 Check the sample for dirtiness (stirring mechanism is "ON"),
Chlorine: Measure out reagents: 12 x reagent A, 5 x reagent B
and 2 x reagent C, then a reaction time of 2 minutes
Residual hardness: Measure out indicator (titration)
6Evaluate and display reaction (approx. 1 minute)
7Drain measuring chamber
8Pause period up to the next analysis (time or quantity analysis interval), Tp.
TI= total analysis interval, ch = measuring chamber
Displayed unit
It is possible to program the units of the displayed value. You have the choice of
°dH, °f, ppm CaCO3and mmol/l for the residual hardness and mg/l und ppm for
the chlorine. All the following inputs and displays will then be displayed in the
programmed unit.
Measuring points switch-over
The Testomat 2000 THCl enables the monitoring of 2 measuring points. The measuring points switch-over occurs
automatically, but can be suppressed by a contact at input IN1.
The active status of IN1 should be programmed correspondingly.
The parameters are assigned to the measuring points:
Measuring point 1 = Residual hardness
Measuring point 2 = Total chlorine
IN1 active = Measurements only from measuring point 1

12 / 23
FLUSH TIMES/INTERVAL
sstt
M E
INTERNAL FLUSH TIME 00s
Interval pause 1m
External flush time 00s
LIMIT VALUES
sstt
M E
Limit val. Cl: 0.50 mg/l
LIMIT VAL. TH: 0.30°dH
HYSTERESIS LV1
sstt
M E
Analysis (1,2,3) 1
Setting the timer
Internal flush
In order to ensure the analysed sample represents the current value, the sampling
tube must be sufficiently well flushed taking its length into consideration. After the
installation has been out of operation for a longer period, or in the case of long
analysis intervals, it is sensible to select a flushing time in excess of 60 seconds. Flush is initiated by simultaneously
opening the inlet and the outlet valve of the Testomat®2000 THCl.
LThe analysis interval directly depends on the programmed flush time. If a flush time of, for example, 60 seconds is
set, the analysis interval itself cannot be shorter than 60 seconds.
External flush
An external flush valve should be installed upstream of the Testomat®2000 THCl if very short analysis intervals are
required, the sampling tube is very long (several metres), or when using a tube with a large diameter. This is connected
to the "Flush" outlet. The external flush time for the valve depends, like the flush time for flush by the unit, on length and
diameter of the connection to the Testomat®2000 THCl.
Example: For connections longer than 3 m and an internal tube diameter of 6 mm, a minimal internal flushing time
of 10 seconds is required to ensure a current sample is taken from the sampling tube. The quantity of
flush water for an internal flush of 1minute is 0.5 litres.
Interval pause
In the case of timed triggering of the analysis, the interval between two analyses (plus flush time) is determined by the
interval pause. The shortest interval can be 0 minutes. In this case, the analyses are carried out continuously. The
longest interval is 99 minutes.
Monitoring of limit value
The limit values can be programmed on a continuous scale. The range for the limit
value depends on the measuring range of the respective parameter.
Two outputs for the limit value are available for monitoring.
Limit value Parameter Measuring
point Function Hysteresis
LV1 Residual
hardness 1duration, impulse,
interval 1st, 2nd or 3rd exceeds
LV2 Total chlorine 2duration, impulse,
interval 1st, 2nd or 3rd exceeds
The functions of these outputs can be programmed independently from each other:
If the limit value LV1 is exceeded, the limit value control display lights up LV1 RED and the relay output LV1 reacts as
programmed in the switch function. If this limit value is not exceeded, the display lights up GREEN.
The unit operates in the same way for the limit value LV2.
Hysteresis
Each limit value output reacts only after the 1
st,2
nd or 3
rd limit value excess
(suppression of the first or the second measured value).
This increases the reliability during the evaluation of the analysis, e.g. after the
measuring point has been switched over or if the sampling line has possibly not been
flushed sufficiently. The hystereses of the two outputs LV1 and LV2 can be set independently from each other.
Operation: With a hysteresis of 2 a further analysis is carried out immediately after the limit value has been exceeded.
Only if the limit value of this analysis is exceeded again, is the corresponding output energised. If you have set a
hysteresis of 3, the corresponding output reacts only after the limit value has been exceeded for the 3rd time in
succession.
(Basic setting is 1 for LV1 and LV2)

13 / 23
Schematic representation of logic function
TYPE OF WATER METER
sstt
M E
1 LITRE/IMPULSE
2.5 Litres/Impulse
5 Litres/Impulse
10 Litres/Impulse
100 Litres/Impulse *
500 Litres/Impulse
1000 Litres/Impulse
ALARM/MESSAGE
sstt
M E
REAGENT Cl LOW LEVEL A
Low water pressure A
Indicator TH low level A
Ff. optics A
Ff. dosing pump A
Ff. outlet to drain A
Mf. dirtiness M
power failure 24 V A
Mf. turbid M
Transfer error M
A = Alarm / M = Message / - = no action
Ff. = Function fault
Mf. = Measuring fault
A/M/-
A/M
A/M
A/M
A/M
A/M/-
A/M/-
A/M
A/M/-
Meas. range exceeded M
A/M/-
Maint. int. exceeded M
A/M/-
A/M/-
FUNCTION AUX
sstt
M E
Contact during analysis
CONTACT BEFORE ANALYSIS *
Contact after analysis
Time: 00m 10s
Logic functions of the limit value outputs LV1 and LV2
Logic function 0, Duration
Output relay LV1 or LV2 pulls up when the measured value rises
above limit value LV1 or LV2. If the measured value falls below
the limit value LV1 or LV2 without lock out, the relevant relay
drops out again.
Logic function 1, Impulse
If the measured value rises above limit value LV1 or LV2, the
relevant output pulls up for a set time t.
Independent of the time taken for the limit value to rise above the
set limit, the relevant output always remains in the "ON" position
for the set time.
Logic function 2, Interval
If the measured value rises above one limit value, the relevant output pulls up at intervals with the time set
(Impulse/Pause).
Function IN1
When measurements are carried out by two measurement points, the measuring points switch-over can be suppressed
by activating IN1. If IN1 is active, measurements are only carried out by measuring point 1. A corresponding controller
has to be connected to the IN1 (e.g. timer) (a volt free contact is required!). The energised status of IN1 must be
programmed correspondingly.
Water meter
For quantity dependent analysis triggering it is necessary to connect a water
meter to the IN2 input.Program the corresponding water meter rating under
menu item "WATER METER".
Alarm / Message
The unit is equipped with an alarm relay output for reporting errors. The events
which could mean a fault at the unit, or are intended to trigger a message, can
either trigger an alarm (continuous contact) or a message (2 second impulse).
For certain unit faults it is already pre-determined whether they trigger an alarm or
a message
The faults are entered into the fault history and stored if the event is programmed
as an alarm or message. For example, if reagent low is not programmed as
ALARM or MESSAGE, then it will not be registered in the fault history. Up to 20
error messages can be stored. This list can be called up via the information menu.
The information stored is the time (day, month, year and hour) and the type of the
fault.
LAfter a power failure, the error messages are lost.
Function AUX
The AUX relay output can be programmed for the following control functions:
SIt can be programmed as a functional output for providing a contact of a
programmable duration before and/or during, or after an analysis.
For example, the cooling water flow in an upstream cooler can be controlled by means of a solenoid valve. Cooling
water will then only flow when necessary, i.e. while an analysis is being carried out.

14 / 23
SERVICE II
sstt
M E
Reset operating time
Maintenance interval
MAINTENANCE INTERV.
stst
M E
001d
INTERFACES
stst
M E
Type 0-20 mA
Type RS232
Type 4-20 mA *
Service II
The service II menu contains a number of functions for monitoring the operation of the
unit.
LThe functions in the service II menu have a direct influence on the operation, and
monitoring functions of the unit!
Reset operating time
After replacing the metering pump or the holder for the measuring chamber, the current operating time can be reset to
0 hours:
"M" ºBASIC PROGRAM ºSERVICE II ºRESET OPERATING TIME
The current operating time can be called up in the information menu:
"i" ºINFORMATION ºOPERATING VALUES
Maintenance interval
The observance of the maintenance interval is monitored and displayed by the
Testomat®2000 THCl. Program the desired maintenance interval in days.
( 0 days = no maintenance interval. )
Interfaces (optional)
Interface 0/4 - 20 mA
Another possibility for monitoring the analysis is the connection of
arecorder. For this purpose the unit is equipped with a
programmable current output.
Standard values of 0 - 20 mA or 4 - 20 mA can be selected.
LA maximum working resistance of 500 Ùmust not be
exceeded!
In situations where interferences might occur, and where very
long cables are used (approx. 20 m), a screened cable should
be used if possible.
Serial interface RS232
The connection of the Testomat®2000 to a protocol printer via the serial interface RS232 enables the print out of
measuring results and error messages. This means the analyses can be continuously recorded. This option is only
possible in conjunction with the plug in circuit card RS910 (Item No. 270310).

15 / 23
Displayed function
"i" (Key) "M"
Information Menu Program Menu
Structure of the menu
Selection and input
Start menu
Select one of the two menus by pressing either the "M" or the "i" key.
Selection
The active line position is displayed in CAPITALS. Use the "ENTER" key to activate the line, i.e. to enter a submenu.
By using the arrow key "w", the next parameter appears below the lowest display line:
The menu can be scrolled.
Input (only possible in the "M" menu)
•Select a programming step by means of the arrow keys "w" and "v" and activate the entry function by pressing the
"ENTER" key.
•In the case of digits to be entered, the first digit to be modified flashes.
•The value can be changed using the arrow keys "w" and "v".
•By pressing the arrow keys ">" and "<", it is possible to confirm the entry and simultaneously change to the next or the
preceding digit (which then starts flashing).
•Terminate the entry function by pressing "ENTER".
•The following line is activated.
•Change to the menu one level up by pressing the "M" key.
End menu
Return to the menu one level up by pressing either the "M" or the "i" key. After returning from the highest menu level the
unit returns to the display mode.

16 / 23
i button
CUSTOMER SERVICE i
GEBRUEDER HEYL
i
i
E
E
i
ERROR REGISTRATION st i E
Error display, if it is programmed as
alarm
MODE OF OPERATION
Displayed unit
PROGRAM VALUES st i E
Measuring points 1 or 2
Flush times/interval
Limit values
INDICATOR TH LEVEL 030%
Residual analyses 1386
Reagent B level 030%
Reagent A level 030%
OPERATING VALUES st i
Res. operat. time 00126h
D-31135 HILDESHEIM
TEL. 05121-7609-0
Maintenance
Program values
Error registration
CUSTOMER SERVICE
Operating values
INFORMATION st i E
Maintenance interval 000d
MAINTENANCE i
Maintenance in 000d
Up to 20
error messages
Low water pressure
Ff. optics
Ff. outlet to drain
Ff. dosing pump
Mf. dirtiness
Transfer error
power failure 24 V
REAGENT Cl LOW LEVEL
Indicator TH low level
Power failure
TH: 0.20 °dH
Cl: 0.24 mg/l
^1: 0.20 °dH
^2: 0.40 mg/l
Ei
1
2
3
4
5
6
i
E
POWER FAILURE i
E
Reagent C level 030%
4 lines are displayed
(black scale)
Water quantity 0063m3
Software version 48EXXX
Operating time 00629h
i
E
to
From
Information Menu "i"
Structure of the "i" - menu
The information menu can be used
to call up the active settings and
status of the unit, the error
registration, the date for the next
maintenance and the customer
service address.
Call (1)
Call up the information menu "i" by
pressing the "i" key.
Customer Service (2)
Display of the customer service
address or, e.g. of a service
telephone number.
These three lines can be
independently programmed in the
basic program (protected by a
password):
"M" ºBASIC PROGRAM º
CUSTOMER SERVICE
Operating values (3)
Display of the current values.
Program values (4)
Call up the menu item "program values" by pressing the arrow keys. By pressing "ENTER" the list with the set values
can be opened. The active setting of a parameter can be queried by pressing "ENTER".
The selected function is marked by a star. (In this context there are no active lines.)
Error registration (5)
Open the error registration by pressing the "i" key and then "ENTER". The error registration is a list of the errors or
status, that have been programmed as faults. This list is lost during a power failure and the recording is started afresh
Provided no faults have accumulated since the start-up, the display will show the last time the unit was switched on, e.g.:
POWER FAILURE
from 16.06.99 06:56
to 16.06.99 07:09
Maintenance (6)
Display of the date of the next maintenance and of the programmed maintenance interval.
The maintenance interval can be set in the basic program (protected by a password):
"M" ºBASIC PROGRAM ºSERVICE II
For further maintenance information refer to page 21 onward.

17 / 23
Prozeß
M - key
SERVICE I
sstt
M E
INPUT REAGENT
Manual operation
Confirm maintenance
DATE / TIME OF DAY
sstt
M E
DATE 11.01.00
Time of day 23.59
PROGRAM MENU
Testomat 2000 TH/Cl
BASIC PROGRAM
INPUT REAGENTS
sstt
M E
INDICATOR TH LEVEL: 100%
Reagent A level: 100%
Reagent B level: 100%
E
Day of week Monday
MANUAL OPERATION
sstt
M E
FLUSH *
Drain chamber
Fill chamber
Flush chamber
TH: 0.20 °dH
Cl: 0.24 mg/l
^1: 0.20 °dH
^2: 0.40 mg/l
NOTE ! M E
Manual function without
monitoring!
go on with enter key
DIAGNOSIS
sstt
M E
OUTPUT LV1
Output LV2
Output MP 1/2
Output Flush
Output AUX
Output Alarm *
Output Maintenance *
Output OUT 255
Input Start
Input Stop
Diagnosis
E
1
2
3
4
4
5
6
7
8
10
11
12
E
MAINTENANCE M E
Confirm the course with enter!
E
PROGRAM
sstt
M E
SERVICE I
Basic program
Date/Time of day
E
E
Input IN1 *
Input IN2 *
E
E
Description on the next
page
Reagent C level: 100%
Program Menu "M"
Call (1)
The program menu "M" is selected by
pressing the "M" key
Except for the basic programming, all
functions can be called up without password
protection.
Service I (2)
Input reagent (3)
The new filling level has to be entered after
every refill or replacement of the reagent
bottle. As soon as the menu item for the
filling level "Reagent X filling (0 - 100%)" has
been selected by pressing "ENTER", the
value is preset to 100%. Confirm the
connection of a full bottle pressing "ENTER".
If only a partially filled bottle is connected,
enter the corresponding % value.
Manual operation (4)
After the information message (4) has been
confirmed by pressing ENTER, the desired
function can be selected using the arrow
keys and then activated by pressing
"ENTER".
These functions are used for monitoring the
operation and for commissioning.
LAll manual functions can only be selected during an interval between two analyses. No analyses are carried out
during manual operation. All signal inputs and outputs are locked.
Flush (5)
Start the flushing of the sampling tube through the internal valve by pressing "ENTER". This function is terminated by
pressing the "ENTER" key again.
Flush chamber (6)
By pressing "ENTER", the measuring chamber is flushed once.
Drain chamber (7)
The outlet valve to drain the water from the measuring chamber is opened by pressing "ENTER".
This function is terminated by pressing the "ENTER" key again.
Fill chamber (8)
The measuring chamber is filled by pressing "ENTER".

18 / 23
Confirm maintenance (10)
After maintenance has been carried out confirm this by pressing "ENTER", and leave this item by pressing "M". The
maintenance interval is started afresh.
LAmaintenance request at the end of the maintenance interval must be confirmed in the M-menu. The message on
the display is deleted and the maintenance output is reset.
For information on which maintenance has to be carried out and in which time intervals refer to the section
maintenance on page 22.
Diagnosis (11)
The current status of the signal inputs and outputs can be called up from a list.
An active status is marked with a * (see menu structure).
Date/time of day (12)
Set the time of day and date by selecting the desired function using the arrow keys and activating them by pressing
"ENTER". Then press the "M" key again to store the setting and to return to the display mode.
Basic program
This menu item is only reached after entering the password!
Example for a password entry: >BASIC PROGRAM 07:25
21.04.00 07:25
Password: ____ _ _ _ _
(5270)
After entering the password and confirming it by pressing "ENTER", it is possible to carry out the basic programming of
the unit and to select a number of service functions (e.g. calibration).
Program values
To call up the factory-set basic default setting, briefly hold down the "M" and "i" keys while switching the unit "ON".
Values and settings are described in the structure of the basic program on page 19.
The following program values can be entered and stored in the basic program:
Abbreviations: s = seconds
m = minutes
h = hours
d= days
l = litres
Table of contents
Other Lenntech Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Keysight
Keysight AE6941A User Guide and Method of Implementation

Shodex
Shodex IC SI-35 4D Operation manual

EmerCom Technologies
EmerCom Technologies EMC-ALRM user guide

GrainSense
GrainSense A-2 user manual

GEX Corporation
GEX Corporation Thermo Scientific GENESYS 30 Product Specifications and Usage Information

RayTek
RayTek Thermalert ET Series Operator's manual

turck
turck B N QR20-2UPN Series Instructions for use

Aaronia
Aaronia SPECTRAN HF-2025E manual

Kamstrup
Kamstrup MULTICAL 602 Installation and user guide

PCB Piezotronics
PCB Piezotronics IMI SENSORS 628F01 Installation and operating manual

GREISINGER
GREISINGER GLF 100 RW operating manual

OWL
OWL FO-4V Operation guide