
Section 1. General information
Sensor recalibration
Factory recalibration of LI-COR radiation sensors is recommended every two years.
Sensors can be returned to LI-COR for this service – please contact our technical sup-
port staff for more information.
Getting started
All LI-COR radiation sensors produce a current proportional to the radiation intens-
ity. During factory calibration, sensor output (in microamps) is measured while the
sensor is exposed to a standard lamp of known intensity. The sensor output at this
intensity has general units of microamps per radiation unit and is called the Cal-
ibration Constant (Calconstant). Each sensor has a slightly different output at a
given radiation intensity and will therefore have a unique Calconstant.
Relationship between the Calibration Constant and Calibration
Multiplier
LI-COR Light Meters and dataloggers measure the current output of the sensor in
units of microamps, and convert the measured current to units of radiation. To
make this conversion, LI-COR instruments use the sensor Calibration Multiplier. The
Calibration Multiplier is the negative reciprocal of the Calconstant.
1-1
The LI-192 and LI-193 calibration certificates contain calibration multipliers for both
in air and in water operation. The in water multiplier includes an immersion effect
correction.
To use a type SA sensor with the LI-250, the calibration multiplier is entered in con-
figuration mode, using the up and down arrow keys to scroll the multiplier value(s).
When using the LI-1400 DataLogger, the calibration multiplier must be entered into
the software as described in the Instruction Manual.
When a LI-COR light meter or data logger is not used, your sensor can be used with
other millivolt recorders or data loggers by connecting a millivolt adapter. The LI-
192SA and LI-193SA underwater sensors require the 2291 Millivolt Adapter, which
has a resistance of 1210 Ohms.
1-2 LI-192 and LI-193 Quantum Sensors