Lloydspharmacy BM75 User manual

User Manual
BM75
Advanced
Blood Pressure
Monitor
+ HealthManager App


Contents
Introduction
Important information
Parts
Information on the display
USB Interface
Setting up your blood pressure monitor
Using your blood pressure monitor
Evaluating your Results
Displaying and deleting measurements
Transferring measurements
Cleaning your blood pressure monitor
Rectifying faults
Specifications
Electromagnetic compatibility information
1
2-5
6
7
8
9-10
11-14
15 -17
18-19
20
21
21
22-23
24-27

Thank you for purchasing the LloydsPharmacy Advanced Blood Pressure Monitor.
Please read these instructions for use carefully and keep them for later use, be sure to make them accessible
to other users.
The upper arm blood pressure monitor is used for non-invasive measurement and monitoring of adults arterial blood
pressure. You can use it to measure your blood pressure quickly and easily, storing the results and displaying the
progression of readings together with the average reading.
• This symbol is displayed for anyone suffering from cardiac arrhythmia.
•
The values determined are classified and graphically evaluated according to the World Health Organisation (WHO)
guidelines.
This blood pressure monitor also has a haemodynamic stability display, which is referred to as a resting
indicator throughout these instructions for use. This shows whether you, and consequently your circulatory system,
are sufficiently at rest when the blood pressure measurement is being taken and is therefore a more precise indicator
of your resting blood pressure. Read more about this on page 16-17.
1
Introduction

Important information
Caution
Note
Note on important information
Follow instructions for use
This symbol indicates that the Blood Pressure
Monitor is a type BF device. The cuff is type BF
applied part.
Direct current
Disposal in accordance with EC Directive
WEEE (Waste Electrical and Electronic
Equipment).
Batch code
Date of manufacture
The name and address of the manufacturer
Storage
RH 10-90
%
-10°C
60°C
Permissible transport and storage
temperature. Permissible transport and
storage humidity.
Operating
10°C
40°C
RH 15%-85%
Permissible operating temperature and humidity
IP 21
Protected against foreign objects equal to or
greater than 12.5 mm in diameter and against
vertically falling drops of water
SN Serial number
0483
The CE labelling certifies that the product com-
plies with the essential requirements of Directive
93/42/EEC on medical products.
Model number
2
SN
SN
SN
Signs and symbols
The following symbols are used in these instructions

Advice on use
•
In order to ensure comparable values, always measure your blood pressure at the same time of day.
•
Before every measurement, relax for about five minutes.
•
If you want to perform several measurements on the same person, wait five minutes between each measurement.
•
Do not take a measurement within 30 minutes after eating, drinking, smoking or exercising.
•
Repeat the measurement if you are unsure of the measured value.
•
The measurements taken by you are for your information only – they are not a substitute for a medical examination.
Discuss the measurements with your doctor, and never base any medical decisions on them (e.g. medicines and
their administration).
•
Do not use the blood pressure monitor on newborns or patients with pre-eclampsia. We recommend consulting a doctor
before using the blood pressure monitor during pregnancy.
•
Cardiovascular diseases may lead to incorrect measurements or have a detrimental effect on measurement accuracy.
The same also applies to very low blood pressure, diabetes, circulatory disorders and arrhythmias as well as chills
or shaking.
•
The blood pressure monitor must not be used in connection with a high-frequency surgical unit.
•
Only use the device on people who have the specified upper arm measurement for the device.
•
Please note that when inflating, the functions of the limb in question may be impaired.
•
During the blood pressure measurement, blood circulation must not be stopped for an unnecessarily long time. If the
device malfunctions, remove the cuff from the arm.
•
Avoid any mechanical restriction, compression or bending of the cuff tube.
Important information
3

•
Do not allow sustained pressure in the cuff or frequent measurements. The resulting restriction of the blood flow may
cause injury.
•
Ensure that the cuff is not placed on an arm in which the arteries or veins are undergoing medical treatment, e.g.
intravascular access or therapy, or an arteriovenous (AV) shunt.
•
Do not use the cuff on people who have undergone a mastectomy.
•
Do not place the cuff over wounds as this may cause further injury.
•
You can either use the blood pressure monitor with batteries or with a mains charger. Please note that data transfer
and data storage is only possible when your blood pressure monitor is supplied with power. As soon as the batteries are
empty or the mains disconnected from the power supply, the blood pressure monitor loses the date and time.
•
To conserve the batteries, the monitor switches off automatically if no buttons are pressed for 3 minutes.
•
The device is only intended for the purpose described in these instructions for use. The manufacturer is not liable for
damage resulting from improper or careless use.
Storage and Care
•
The blood pressure monitor is made up of precision electronic components. Accuracy of readings and the instrument’s
service life depend on careful handling.
– You should protect the device from impact, moisture, dirt, major temperature fluctuations and direct exposure to the
sun’s rays. – Never drop the device. – Do not use near strong electromagnetic fields, i.e. keep it away from any
radiosystems and mobile phones. – Only ever use the cuffs provided with the monitor or original replacement cuffs
Important information
otherwise erroneous results will be recorded.
•
Do not press any buttons until the cuff is in position.
•
If the instrument is not to be used for any length of time, we recommend removing the batteries.
4

.
Advice on batteries
•
Please use 4 x AAA batteries supplied
•
Remove the batteries if the device is not in use for long periods of time
•
Do not mix old and new batteries or different types of batteries
•
Warning: If batteries leak and come into contact with skin or eyes, wash immediately with plenty of water
•
Batteries must be handled by an adult. Keep batteries out of reach of children
•
Only batteries of the same or equivalent type are recommended
•
Do not use rechargeable batteries
•
Remove exhausted batteries from the unit
•
Supply terminals are not to be short circuited
•
Dispose of batteries safely according to battery manufacturer’s instructions
Repair and disposal
Important information
5
Batteries do not belong in domesitc refuse. Used batteries should be disposed at collections points provided.
• Never open the instrument. If these instructions are not followed, the warranty will be null and void.
•
Never attempt to repair the instrument or adjust it yourself. We can no longer guarantee perfect functioning if you do.
Please note there are no servicable parts included in this monitor
•
Always check the batteries and replace them if necessary prior to making any complaint.
•
The appliance should be disposed of according to EC directive 2012/19/EC WEEE (Waste Electrical and Electronic
Equipment). In case of queries, please contact the municipal authorities responsible for waste disposal in your area.

Parts
6
Cuff
Cuff tube
Cuff connector
Cuff holder (insert cuff holder into position shown)
Connection for mains charger
(optional) and USB interface
NFC detection zone
Display
Memory buttons M1/M2
(Touch screen)
Start/stop button
WHO scale
(World Health Organization)
Connection for cuff connector
(left-hand side)
Resting indicator
display
Cuff holder
NOTE: USB cable included

Information on the Display
7
Date / time
Systolic pressure
Diastolic pressure
Pulse value
Pulse symbol
Release air (arrow)
Number of memory space/
memory display average
value ( ), morning ( ),
evening ( )
Battery replacement symbol
Multi-user memory
WHO classification
Cardiac arrhythmia symbol

USB Interface
The blood pressure monitor also allows you to transfer your measured values to a PC.
To do this, you need a regular USB cable (included) and the HealthManager PC software.
The software can be downloaded free of charge from the download area under Service at www.beurer.com
System requirements for the Beurer HealthManager PC software
1. Supported operating systems:
• Windows Vista SP1 or later
• Windows 7
• Windows 7 SP1
• Windows 8
2. Supported architectures:
• x86 (32 bit)
• x64 (64 bit)
3. Hardware requirements:
• Recommended: At least Pentium 1 GHz or faster with at least 1 GB RAM
• Free memory on the primary partition of at least:
– x86 – 600 MB
– x64 – 1.5 GB
• Graphic resolution from: 1024 x 768 pixels
• USB port 1.0 or later
8

1. Inserting the batteries
•
Remove the battery compartment lid on the rear of the device.
•
Insert four 1.5V AAA (alkaline type LR03) batteries. Make sure that the batteries are
inserted the correct way round observing the correct polarity.
Do not use rechargeable batteries.
•
Close the battery compartment lid again carefully.
All display elements are briefly displayed, flashes in the display. Set the date and time as
described below.
If the battery replacement symbol is permanently displayed, you can no longer perform
any measurements and must replace all batteries. Once the batteries have been removed
from the device, the date and time must be set again. Any saved measurements are retained.
Used batteries should not be disposed of in normal household waste. Dispose of them via
your electronics retailer or your local recycling point. You are legally obligated to do so.
The codes below are printed on batteries containing harmful substances:
Pb: Battery contains lead, Cd: Battery contains cadmium, Hg: Battery contains mercury.
Please note that the M1 and M2 memory buttons are sensor touch pads and not actual buttons.
4 x 1,5V AAA (LR03)
Setting up your Blood Pressure Monitor
9

2. Setting the date and time
Setting the hour format, date and time. This menu allows
you to set the following functions, one after another.
Hour format Date Time
It is essential to set the date and time. Otherwise, you will
not be able to save your measured values correctly with a
date and time and access them again later.
Hour format
•
Press and hold the Start/stop button for 5 seconds.
•
Choose the desired hour format with the M1/
M2 memory buttons and confirm with the
Start/stop button .
Date
The year flashes on the display.
•
Choose the desired year with the M1/M2
memory buttons and confirm with the Start/
stop button .
The month flashes on the display.
•
Choose the desired month with the M1/M2
memory buttons and confirm with the Start/
stop button .
Setting up your Blood Pressure Monitor
10
If you press and hold the M1 or M2 memory
button, you can set the values more quickly.
Date
The day flashes on the display.
•
Choose the desired day with the M1/M2
memory buttons and confirm with the Start/
stop button .
If the hour format is set as 12h, the day/month display
sequence is reversed.
Time
The hour flashes on the display.
•
Choose the desired hour with the M1/M2
memory buttons and confirm with the Start/
stop button .
The minute flashes on the display.
•
Choose the desired minute with the M1/M2
memory buttons and confirm with the Start/
stop button .

Measuring blood pressure
Your first test
We recommend that for your first test you test both arms. This will determine which arm should be used for future
measurements. Take note of your systolic (higher number) reading on both arms. The arm that gives you the higher
systolic reading should be used for future testing.
Please ensure the device is at room temperature before measuring.
The measurement can be performed on the left or right arm.
Reading example: left arm: 132 (systolic) Right arm: 128 (systolic)
86 (diastolic) 84 (diastolic)
This example shows that you should use your left arm. (Note: Please ensure you do not have
any medical complications which will prevent you from having your blood pressure measured in your arms.
You can also operate this device with a mains charger (not included).
When doing so, there must not be any batteries in the battery compartment. The mains charger can be obtained from
specialist retailers, please contact Beurer for details.
•
To prevent possible damage to the device, the blood pressure monitor must only be used with the mains part described
here.
•
Insert the mains charger into the connection provided for this purpose on the right-hand side of the blood pressure
monitor. The mains charger must only be connected to the mains voltage that is specified on the type plate.
How to use your Blood Pressure Monitor
11

How to use your Blood Pressure Monitor
If the measurement is performed on the right upper arm, the line should be located on the inside of your elbow.
Ensure that your arm is not pressing on the line.
12
Fit the cuff on your arm.
Blood circulation in the arm
should not be restricted by tight
clothing or other objects.
The cuff should be placed on the upper
arm so that the lower edge is about 2 to 3
cm from the bend of the elbow and above
the artery. The tube should be in line with
the centre of the palm.
Tighten the free end of the cuff,
but make sure that it is not too
tight around the arm and close to
the hook-and-loop fastener. The
cuff should be fastened so that
two fingers fit under the cuff.
•
Then insert the mains plug into the mains socket.
•
After using the blood pressure monitor, unplug from the mains socket first and then disconnect it from the blood
pressure monitor. As soon as you unplug the mains, the blood pressure monitor loses the date and time setting but
the saved measurements are retained.
Insert the cuff line into the
connection for the cuff connector.
Positioning the cuff
1.
3.
2.
4.

Using your Blood Pressure Monitor
13
If the values between the two arms are significantly different, please consult your pharmacist or doctor to determine which arm
should be used for the measurement.
Important: The unit may only be operated with the original cuff. The cuff is suitable for an arm circumference of 24 to 36 cm.
A larger cuff for upper-arm circumferences of 35 to 44 cm can be obtained from www.lloydspharmacy.com
Correct posture
•
Rest for approx. 5 minutes before each measurement.
Otherwise there may be divergences.
•
You can perform the measurements either sitting or lying down.
Always make sure that the cuff is on a level with your heart.
•
To carry out a blood pressure measurement, make sure you are sitting comfortably with your arms and back leaning on
something. Do not cross your legs. Place your feet flat on the ground.
•
In order not to distort the result, it is important to keep still during the measurement and do not talk.

Using your Blood Pressure Monitor
14
Measurement
•
appears if the measurement could not
be performed properly. Observe page 21 on
error messages/troubleshooting in
these instructions for use and repeat
the measurement.
•
Now select the desired user memory by pressing the
M1 or M2 memory buttons. If you do not select a
user memory, the measurement is stored in the most
recently used user memory. The relevant symbol
or appears on the display.
•
Using the Start/stop button , switch off the blood
pressure monitor. The measurement is then stored in
the selected user memory.
If you forget to turn off the device, it will switch off
automatically after approx. 3 minutes. In this case too,
the value is stored in the selected or most recently used
user memory.
•
Wait at least 5 minutes before taking
another measurement.
Measurement
As described on page 13, attach the cuff and adopt the
posture in which you want to perform the measurement.
•
To start the blood pressure monitor, press
the Start/stop button .
All display elements are briefly displayed.
The blood pressure monitor will begin the
measurement automatically after 3 seconds.
The cuff automatically inflates.
Measuring can be cancelled at any time by
pressing the start/stop button .
The cuff’s air pressure will be slowly released. If you already
recognise a tendency for high blood pressure, you should
reinflate the cuff and increase the cuff’s pressure again. As
soon as a pulse is found, the pulse symbol is displayed.
•
Systolic pressure, diastolic pressure and
pulse measurements are displayed. The
resting indicator display (see page 16-17)
illuminates in accordance with the
positive or negative classification.

Evaluating Your Results
15
Cardiac arrhythmia:
This instrument can identify possible cardiac arrhythmia disorders during measurement and if necessary indicates the
measurement with the flashing icon .
This may be an indicator for arrhythmia. Arrhythmia is a condition where the heart rhythm is abnormal as a result of defects in the
bioelectrical system controlling the heart beat. The symptoms (omitted or premature heart beats, slow or excessively fast heart
rate) may be caused, among other things, by heart disease, age, physical predisposition, excessive use of stimulants, stress or lack of
sleep. Arrhythmia can only be ascertained through examination by your doctor.
Repeat the measurement if the flashing icon is displayed after the measurement. Please note that you should rest
for 5 minutes between measurements and not talk or move during the measurement. If the icon appears often,
please contact your doctor. Any self-diagnosis and treatment based on the test results may be dangerous. It is vital to
follow your doctor‘s instructions.
WHO Classification:
In accordance with the guidelines/definitions of the World Health Organisation and the latest findings,
the measurements can be classified and assessed according to the table on page 16.
However, these standard values serve only as a general guideline, as the individual blood pressure varies
in different people and different age groups etc.
It is important to consult your doctor regularly for advice. Your doctor will tell you your individual values for
normal blood pressure as well as the value above which your blood pressure is classified as dangerous.

Evaluating Your Results
16
The classification on the display and the scale on the unit show which category the recorded blood pressure values fall into.
If the values of systole and diastole fall into two different WHO categories (e.g. systole in the ‘High normal’ category and
diastole in the ‘Normal’ category), the graphical WHO classification on the unit always shows the higher category.
Resting indicator measurement (using HSD diagnostics)
The most frequent error made when measuring blood pressure is taking the measurement when not at rest (haemodynamic stability),
which means that both the systolic and the diastolic blood pressures are incorrect in this case.
While measuring the blood pressure, the device automatically determines whether you are at rest or not.
If there is no indication that the circulatory system is not sufficiently at rest, the symbol (haemodynamic stability)
lights up green and the measurement can be recorded as a reliable resting blood pressure value. If the symbol does not light up
green, then it could not be determined whether the circulatory system was sufficiently at rest or not.
Source: WHO, 1999
Range of blood pressure values Systolic
(in mmHg)
Diastolic
(in mmHg)
Advice on results
Optimal < 120 < 80 Practice a healthy lifestyle. Consult your doctor only if suffering
symptoms of low BP.
Normal 120 – 129 80 – 84 Practice a healthy lifestyle.
High-normal 130 – 139 85 – 89 Practice a healthy lifestyle.
Grade 1: Mild hypertension 140 – 159 90 – 99 Practice a healthy lifestyle. Re-measure BP monthly over the next
3 weeks. If high levels ( ≥140/90) persist (e.g. 2 high readings on 2
separate occasions) consult your pharmacist or doctor.
Grade 2: Moderate hypertension 160 – 179 100 – 109 Practice a healthy lifestyle. Re-measure BP monthly over the next
4 weeks. If high levels ( ≥140/90) persist (e.g. 2 high readings on 2
separate occasions) consult your pharmacist or doctor.
Grade 3: Severe hypertension ≥ 180 ≥ 110 We would recommend that you see your doctor urgently.

Evaluating your Results
17
GREEN: Haemodynamic stability
Measurement of the systolic and diastolic pressure is increased when the circulatory system is sufficiently at rest and
is a very reliable indicator of resting blood pressure.
However, if there is an indication that the circulatory system is not sufficiently at rest (haemodynamic instability),
the symbol lights up red.
In this case, the measurement should be repeated after a period of physical and mental rest. The blood pressure measurement must be
taken when the patient is physically and mentally rested, as it will be the basis for diagnosing the blood pressure level and regulating the
patient’s medical treatment.
RED: Lack of haemodynamic stability
It is very probable that the systolic and diastolic blood pressures have not been measured whilst the patient is at rest
and the resting blood pressure measurement has therefore been distorted.
Repeat the measurement after a rest and relaxation period of at least five minutes. Go to a sufficiently quiet
and comfortable area and remain calm, close your eyes and breathe deeply and evenly and try to relax.
If the next measurement also shows insufficient stability, you can repeat the measurement after another resting period. If the measure-
ments continue to show some instability, identify these blood pressure measurements as having been taken when the circulatory system had
not been sufficiently rested. In this case, nervousness or inner anxiety may be the cause and this cannot be cured by brief periods of rest. Ex-
isting cardiac arrhythmias may also prevent a stable blood pressure measurement. A lack of resting blood pressure can have various causes,
such as physical or mental strain or distraction, speaking or experiencing cardiac arrhythmias during the measurement. In an overwhelming
number of cases, the HSD diagnosis will give a very good guide as to whether the circulatory system is rested when taking the measurement.
Certain patients suffering from cardiac arrhythmia or chronic mental conditions can remain haemodynamically unstable in the long-term,
something which persists even after repeated periods of rest.
If the resting indicator symbol does not light up green or red, then it could not be determined whether the circulatory system was sufficiently
at rest or not.
Table of contents
Other Lloydspharmacy Blood Pressure Monitor manuals