Long-run LRF-6000H Series User manual


LRF-6000H Ultrasonic flowmeter
Page 2of 29
Notice!
Thank you for choosing model LRF-6000H Portable Ultrasonic Flowmeter.
This instruction manual contains the important using and operation information of the flow
meter. Please read carefully the reference manual before operations to make your portable
ultrasonic flowmeter exert the best performance.
If the operation is wrong, it will affect the normal operation of the flow meter, reduce the
service life of it or cause some malfunctions.
This manual will present how to use the flow meter step by step. Beginning with the product
component, installation, wiring, quick setting and so on which will let you very easily operate
the flow meter.
By knowing more menu settings, the powerful function options and output functions of the
flow meter can meet your higher requirements.
Note: Some content in this manual may be different from the flow meter you purchased,
depending on the optional configuration requirements at the time of purchase. On the
other hand, due to the design changes and upgrade needs of the product, there is no
indication in the manual. Please pay attention to the version number and the added
Attached page description.

LRF-6000H Ultrasonic flowmeter
Page 3of 29
Product Component
Inspection should be made before installing. Check if the spare parts are in accordance with the
packing list. Make sure that there is no potential damage to the enclosure due to a loose screw
or loose wire during transportation. Any questions, please contact your representative as soon
as possible.
Transmitter
Transducers with Rack
Accessories:
Case
Coupling compound
Battery changer
4~20mA Wire
Transducer cable
Flexible rule
SD card
SD card reader
Documents:
Instruction Manual
Certified 3-point factory calibration

LRF-6000H Ultrasonic flowmeter
Page 4of 29
Content
Transmitter Installation................................................................. 6
Powering On......................................................................................................................... 7
Keypad Operation................................................................................................................7
Main screen..................................................................................... 8
Setup Menu....................................................................................10
Setup Menu – Measurement Parameter..........................................................................10
Setup Menu – System Setting............................................................................................11
Setup Menu – Calibration................................................................................................. 13
Setup Menu – Output Setting........................................................................................... 15
Setup Menu – Energy Setting ......................................................................................... 17
Setup Menu – Historical Data...........................................................................................17
Setup Menu – System Information...................................................................................18
Setup Menu – Manual Totalizer....................................................................................... 18
Setup Menu – Flow Batch Controller (For wall mount model).................................... 19
Setup Menu – Language Setting.......................................................................................19
Measurement Site Selection.........................................................20
Transducer Installation................................................................21
Transducer spacing............................................................................................................ 21
Transducer Mounting Methods........................................................................................ 21
V Method...................................................................................................................... 22
Z Method.......................................................................................................................22
N Method (not commonly used)...................................................................................23
W Method (rarely used)................................................................................................23
Transducer Mounting Inspection..................................................................................... 24
Signal Strength..............................................................................................................24
Signal Quality (SQ value).............................................................................................24
Total Time and Delta Time........................................................................................... 24
Transit Time Ratio........................................................................................................ 25
Operating Instructions.................................................................26
Zero Cut.............................................................................................................................. 26

LRF-6000H Ultrasonic flowmeter
Page 5of 29
Zero Setting.........................................................................................................................26
Scale Factor.........................................................................................................................26
System Lock (Unlock)........................................................................................................26
4~20mA Current Loop calibration...................................................................................27
Analog Output Calibration............................................................................................... 27
ESN...................................................................................................................................... 27
Appendix Error Codes and Solutions............................................28

LRF-6000H Ultrasonic flowmeter
Page 6of 29
Transmitter Installation
Take out the transmitter. Shown from left to right on the top of model 6000H are the upstream
transducer connector, downstream transducer connector, and 4~20mA output connector.
On the left side of the transmitter, are the power recharging and the power switch.
One rechargeable 11.1V Lithium battery and matching battery charger are coming with.
Please refer to the below diagram for specific connection information:
CAUTION
Caution
Wire when it is power-off. The flow meter must be grounded reliably before
installation and use.

LRF-6000H Ultrasonic flowmeter
Page 7of 29
Powering On
When power on, the self-diagnosis program will start to run. If any error is detected, an error
code will be displayed on the screen (see Error Codes and Solutions). After that, the system
will run automatically using the programmed parameters.
If it is the first time to use or install on a new site, new installation site parameters are needed
to enter. Any parameters which are set by the user will be saved permanently until next change.
When the user reedits the parameters or removes the transducers, the flow meter will
recalculate automatically, and operate normally with the newly set parameters.
Keypad Operation
Follow these guidelines when using the Flowmeter keypad:
for setting and display mode, when in setting mode, press it to turn to previous menu,
and for scrolling up and down to select the menu, when press to move to
next digit, press and the numbers scroll from 0 to 9. Select them. Press to
confirm, press to delete.
For displaying flow rate and totalize; For displaying flow rate curve;
For displaying signal diagnosis information; For SD card data collection
operation.

LRF-6000H Ultrasonic flowmeter
Page 8of 29
Main screen
When power on, the self-diagnosis program will automatically start to run.
First comes the welcome screen:
Signal Quality (SQ value), Net Totalizer (NET), Positive Totalizer (POS), Negative Totalizer
(NEG), Flow Velocity (Vel.).
Note: SQ value is short for Signal Quality. It indicates the level of the signal detected. SQ
value is indicated by numbers from 0~99. 00 represents the minimum signal detected
while 99 represent the maximum.

LRF-6000H Ultrasonic flowmeter
Page 9of 29
Display the velocity and flow curve. The data will be updated only stay in this screen.
Display the signal strength, Date and time, Signal quality, Speed ratio (TOM/TOS), Sound
Velocity, Total transit time (TOF), Delta time (DT).
About TOM/TOS: Display the ratio of actual measured transmission time to calculated
transmission time according to customer requirements. Normally the ratio should be 100±3%.
If the difference is too large, the user should check that if the parameters are entered correctly,
especially the sound velocity of the fluid and the installation of the transducers.
Display the signal strength, flow rate, flow totalizer and collection time. Press ENT to choose
stop or continue, Double-click ENT to stop data collection, Flow data will be stored on the SD
card. Cannot any operation when the meter run collection, once it exits, the collection is stop.

LRF-6000H Ultrasonic flowmeter
Page 10 of 29
Setup Menu
Press will display Setup menu.
The following options are available (by pressing or buttons), select the option
Setup Menu – Measurement parameter
For example, to measure a DN1000 PVC pipe with medium as water and no liner. Input
settingss are as above.
When finish entering parameters and selections, the flow meter will display transducer spacing,
then we can install the transducers.
0. Pipe Outer diameter range: 25~6000mm
1. Pipe Wall thickness range: 0~600mm
2. Pipe material:
0. Carbon Steel
1. Stainless Steel
2. Cast Iron
3. Ductile Iron
4. Copper
5. PVC
6. Aluminum
7. Asbestos
8. Fiber Glass Epoxy
9. Other

LRF-6000H Ultrasonic flowmeter
Page 11 of 29
3. Fluid type:
0. Water
1. Sea Water
2. Kerosene
3. Gasoline
4. Fuel Oil
5. Crude Oil
6. Propane
7. Butane
8. Other
4. Transducer Type:
0. Standard
1. Wetted45
2. Wetted23
5. Transducer installing:
0. V Method
1. Z Method
2. N Method
3. W Method
6. Liner Material:
0. None, Liner
1. Tar Epoxy
2. Rubber
3. Mortar
4. Polypropylene
5. Polystyrol
6. Polystyrene
7. Polyester
8. Polyethylene
9. Ebonite
10. Teflon
11. Other
7. Transducer Spacing: calculated automatically by the flow meter instead of manual selection.
Setup Menu – System setting
0. System unit: select the measurement unit as follows: (Factory default is metric)
0. Metric
1. English
1. Flow rate: the following flow rate units are available: (day, hour, min, sec)
0. m3
1. L
2. GAL
3. ig
4. mg
5. cf
6. bal
7. ib
8. ob

LRF-6000H Ultrasonic flowmeter
Page 12 of 29
2. The following units are available:
0. m3
1. L
2. GAL
3. IG
4. mg
5. cf
6. UB
7. IB
8. OB
3. Total RESET:
ENT TO RESET: reset all totalizer
System RESET: factory reset
Note: Except during the initial installation, it is generally not necessary to activate
this feature.
4. Time set:
Generally, it is unnecessary to modify date time as the system is provided with a highly
reliable perpetual calendar chip.
The format for time setting is 24-hour.
5. System lock:
System lock is readable but uneditable to prevent operation error due to unauthorized
tampering.
6. Damping:
The damping function will stabilize the flow displaying. Essentially, it is a part of the
signal filter. Enter the coefficient and increase it to improve stability. However, due to
over-damping, the displayed measurement may be slightly delayed. The measurement
time is too long, especially when the flow fluctuates greatly, and there is no response to
real-time fluctuations. Therefore, the damping should be kept to a minimum and increased
just enough to reduce the fluctuation to an acceptable level for 3 to 10 seconds.
The damping coefficient is between 0~999 seconds. 00 means no damping; 999 means
maximum damping. In applications, a damping factor of 3~10 is usually recommended.
7. Back light:
Brightness is adjustable between 5% to 100%

LRF-6000H Ultrasonic flowmeter
Page 13 of 29
Setup Menu – Calibration
0. Scale factor:
Scale factor refers to the ratio between “actual value” and “reading value”. For example,
when the measurement is 2.00, and it is indicated at 1.98 on the instrument, the scale
factor reading is 2/1.98. This means that the best scale factor constant is 1.
During operation, there still exists possible difference in pipe parameters, etc. The “scale
factor” may be necessary when use on different pipes. Thus, scale factor calibration is
specially designed for calibrating the differences that result from applications on different
pipes. The scale factor entered must be one that results from actual calibration.
1. 4-20mA Calibration:
Check if the current loop has been calibrated before leaving factory.
Select 4mA, and at the same time check with an ammeter to verify that output terminals
agree with the displayed values. It is necessary to re-calibrate the current loop if it display
over the permitted tolerance.
20mA calibration is with the same method.
Note: When verifiering the current loop, make sure the flow meter has run for more
than 20 minutes.
2. Set zero:
When the fluid is in the static state, the displayed value is called “Zero Point”. When
“Zero Point’ is not at zero in the flow meter, the difference is going to be added into the
actual flow values and measurement differences will occur in the flow meter.
Set zero must be carried out after the transducers are installed and the flow inside the pipe
is in the absolute static state (no liquid movement in the pipe). Thus, the “Zero Point”
resulting from different pipe mounting locations and parameters can be eliminated. The
measuring accuracy at low flow is enhanced by doing this and flow offset is eliminated.
3. Low flow cut:
If the flow rate falls below the low flow cutoff value, the flow indication is driven to zero.
This function can prevent the flow meter from reading flow after a pump as shut down but
there is still liquid movement in the pipe, which will result in tantalization error.

LRF-6000H Ultrasonic flowmeter
Page 14 of 29
Generally, 0.03m/s is recommended to enter as the low flow cutoff point. The low flow
cutoff value has no relation to the measurement results once the velocity increases over
the low flow cutoff value.
4. Manual zero point:
This method is not commonly used. It is only suitable for experienced operators to set
zero point under conditions when it is not preferable to use other methods. Enter the value
manually to add to the measured value to obtain the actual value.
For example:
Actual measured value =250 m3/h
Value Deviation =-10 m3/h
Flow meter Display =240 m3/h
Normally, set the value as “0”.
5. AI1 Calibration: (For wall mount models)
Calibration analog input channel 1.
6. AI2 Calibration: (For wall mount models)
Calibration analog input channel 2.
7. Automatic Correction:
With the function of automatic flow correction, the flow lost in an offline session can be
estimated and automatically adjusted. The estimated value is based on the average value
multiplied by the time period during which the meter is offline. The average value is
derived from the flow before the offline and the flow measured after the offline. Select
"No" to cancel this function.
8. Linear correction:
Sometimes there is a small deviation between the flow model and the actual measured
value. In order to make the measured value more closed to reality, the flow data can be
corrected according to the calibration value.
Up to 16 segments can be corrected according to the flow rate.

LRF-6000H Ultrasonic flowmeter
Page 15 of 29
Setup Menu – Output setting
0. RS485 Setup:
The window used to set serial ports. It’s for communications with other equipment. It
connects with the equipment through serial ports. The parameters must be set to match
with each other. Select the baud rata 9600, 19200, 38400, 56000, 57600, 115200. The
second option is that in check, None.
Data length is fixed to 8
Fixed length stop bit
Default serial port parameters: 9600, 8, None, 1
1. 4-20mA Setup:
0. 4mA value: Set the CL output value according to the flow value
1. 20mA value: Set the CL output value according to the flow value
2. CL mode: Select the current loop mode
0. 4-20mA Set up the output range from 4-20mA
1. 20-4-20mA Set up the output range from 20-4-20mA
2. 4-20mA vs Vel. Set up the CL output range from 4-20mA
3. 4-20mA vs Energy Set up the CL output range from 4-20mA
2. FO flow range: (For wall mount models)
Set up FO flow rate, which is the corresponding flow value when output signal frequency
is at the lowest or Highest FO frequency.
3. FO Frequency: (For wall mount models)
Set up low FO frequency and high FO frequency range.
4. OCT output: (For wall mount models)
The OCT output in the flow meter is a kind of isolated collector open circuit output with
programmable open and close qualifications. The user can program the open and close
functions under the following conditions: the system alarm signals are being activated or
the totalizer pulse is being transmitted.
The frequency output signal is also transmitted from the OCT. When it functions as the

LRF-6000H Ultrasonic flowmeter
Page 16 of 29
frequency output, other functions are unavailable.
The following signal options are available:
0. No Signal
1. Alarm #1
2. Alarm #2
3. Batch Control
4. POS Int Pulse
5. NEG Int Pulse
6. NET Int Pulse
7. Energy Pulse
8. FO
9. Not Using
Note: 3 and 7 is optional.
5. Relay output: (For wall mount models)
The relay output in the flow meter is programmable. The user can program the open and
close functions under the following conditions, the system alarm signals are activated or
the Totalizer pulse is transmitting. The relay is single-pole and constant-on for external
flow metercontrols.
The following signal options are available:
0. No Signal
1. Alarm #1
2. Alarm #2
3. Batch Control
4. POS Int Pulse
5. NEG Int Pulse
6. NET Int Pulse
7. Energy Pulse
8. Not Using
Note: 3 and 7 is optional.
6. Beep output: (For wall mount models)
The beep output in the flow meter is programmable. The user can program the open and
close functions under the following conditions.
The following signal options are available:
0. No Signal
1. Alarm #1
2. Alarm #2
3. Batch Control
4. POS Int Pulse
5. NEG Int Pulse
6. NET Int Pulse
7. Energy Pulse
8. KEY input
9. Not Using
Note: 3 and 7 are optional.
7. Alarm value: (For wall mount models)
Enter the low and high alarm value. Both relevant alarms are on OCT output; any of the
measured flow, which is lower or higher than the alarm value will activate the alarm in the
OCT hardware or relay output signal.
8. AI range: (For wall mount models)
Analog input range: 0-9999
Enter range value which 4mA and 20mA analog input represented.
9. Output Multiplier:
Select pulse output multiplier.

LRF-6000H Ultrasonic flowmeter
Page 17 of 29
Setup Menu – Energy setting
0. Energy unit:
0. GJ
1. MBtu
2. KWh
3. MWh
4. RT
1. Temperature unit:
0. ℃
1. ℉
2. Flow site:
0. Inlet
1. Outlet
Setup Menu – Historical data
Review the historical flow data totalizer for any day in the last 64 days, any month in last 64
months and any year in the last 10 years.

LRF-6000H Ultrasonic flowmeter
Page 18 of 29
Setup Menu – System information
Display electronic serial number (ESN) of the flow meter, power on/off time, run time and flow
rate.
This ESN is the only one assigned to each flow meter when ready to leave factory.
Setup Menu – Manual totalizer
The manual totalizer is a separate totalizer. Press to start, and press again to stop it. It is
used for flow measurement and calculation.

LRF-6000H Ultrasonic flowmeter
Page 19 of 29
Setup Menu – Flow Batch controller (For wall mount model)
Batch control can be carried out, and the total amount can be controlled and operated by button
setting.
Setup Menu – Language setting
Language setting: 0. English; 1. Chinese Simplified; 2. Chinese Traditional

LRF-6000H Ultrasonic flowmeter
Page 20 of 29
Measurement Site Selection
When selecting a measurement site, it is important to select an area where the fluid flow profile
is fully developed to guarantee a highly accurate measurement. Please follow these guidelines
for selecting a proper measurement installation site:
Choose a section of pipe, which is always full of liquid, such as a vertical pipe with flow in the
upward direction or a full horizontal pipe.
Make sure that the pipe surface temperature at the measuring point is within the transducer
temperature limits.
Consider the inside condition of the pipe carefully. If possible, select a section of pipe where its
inside is free excessive corrosion or scaling.
Examples acceptable measurement site selection are shown in the figure below.
Site
Installation point front straight section
Straight pipe section after installation point
Elbow
Tee
Expanded
pipes
Reducing
pipe
Valve
Pump
Table of contents
Other Long-run Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

WPI
WPI EVOM EVA-MT-02-01 install guide

Voltminder
Voltminder Voltminder owner's manual

Victor
Victor 852A manual

Pentax
Pentax Honeywell 1/21 Degree Spotmeter operating manual

Swan Analytical Instruments
Swan Analytical Instruments AMI Deltacon DG Operator's manual

GANN
GANN HYDROMETTE HB 30 operating instructions