LongerPump BT100-1F User manual

BT100-1F Longer Peristaltic Pump
Baoding Longer Precision Pump Co.,Ltd.
BaodingLongerPrecisionPumpCo.,Ltd
ADD: Building A,Chuangye Center Baoding
National High - Tech Industrial Development Zone
Baoding,Hebei,China 071051
TEL: 86 - 312 - 3110087 3138553
FAX: 86 - 312 - 3168553
E - mail: longer@longerpump.com
Www.longerpump.com
BT100-1F PERISTALTIC PUMP
OPERATING MANUAL

Table of Contents
Instruction
Operation Panel
Basic Operation
Flow Rate Display
Suitable Pump Head and Tubing
Pump Head Installation
System Setting
Pump Head and Tubing Setting
External Control
Address
Footswitch Setting
Back Suction Time Setting
Control Mode
Dispensing Setting
Dispensing Volume
Copy No
Calibration
Flow Rate Calibration
Dispensing Calibration
Operation
Flow Rate Mode Operation
Dispensing Mode Operation
External Control Input
Footswitch Function
Communication Function
Running Interface
Dispensing Display
Menu Diagram
Flow Rate
Pause Time
Volume
Procedure
Procedure
Procedure
External Control Output
Maintenance
Warranty
Technical Specification
1
2
2
3
3
3
4
5
5
6
6
6
6
7
7
8
8
8
8
9
9
9
9
10
11
11
12
14
14
15
15
15
16
11
IMPORTANT INFORMATION:
Please read operation manual carefully before operation.
WARNINGS:
Tubing breakage may result in fluid being sprayed from pump. Use
appropriate measures to protect operator and equipment. Please check the
tubing frequently and change the tubing in time.
If the power line or the plug is worn or damaged please pull out the plug (Hold
the plug not the power line when pulling out).
Please shut down the power supply and pull out the plug when meet below
circumstances (Hold the plug not the power line when pulling out).
The fluids splash on the body of the pump.
You think the pump needs to be maintained or repaired.
Please shut down the power supply before install the external control
equipments.
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

Table of Contents
Instruction
Operation Panel
Basic Operation
Flow Rate Display
Suitable Pump Head and Tubing
Pump Head Installation
System Setting
Pump Head and Tubing Setting
External Control
Address
Footswitch Setting
Back Suction Time Setting
Control Mode
Dispensing Setting
Dispensing Volume
Copy No
Calibration
Flow Rate Calibration
Dispensing Calibration
Operation
Flow Rate Mode Operation
Dispensing Mode Operation
External Control Input
Footswitch Function
Communication Function
Running Interface
Dispensing Display
Menu Diagram
Flow Rate
Pause Time
Volume
Procedure
Procedure
Procedure
External Control Output
Maintenance
Warranty
Technical Specification
1
2
2
3
3
3
4
5
5
6
6
6
6
7
7
8
8
8
8
9
9
9
9
10
11
11
12
14
14
15
15
15
16
11
IMPORTANT INFORMATION:
Please read operation manual carefully before operation.
WARNINGS:
Tubing breakage may result in fluid being sprayed from pump. Use
appropriate measures to protect operator and equipment. Please check the
tubing frequently and change the tubing in time.
If the power line or the plug is worn or damaged please pull out the plug (Hold
the plug not the power line when pulling out).
Please shut down the power supply and pull out the plug when meet below
circumstances (Hold the plug not the power line when pulling out).
The fluids splash on the body of the pump.
You think the pump needs to be maintained or repaired.
Please shut down the power supply before install the external control
equipments.
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

1
Operation Panel
2
Start/Stop Key
Exit Key
Direction Key
Prime Key
Knob
S1
S2
P3
P2
P4
P1
LCD Display
Power Switch
Introduction
Applicable pump heads for BT100-1F are YZ1515x, YZ2515x, YZII15, YZII25 and
DG-1, DG-2, DG-4. The pump delivers flow rates from 0.16 L to 500 mL / min
and dispensing volume is from 0.01 mL to 9.99 L. Combining 128 32 LCD
display with membrane keypad and rotary coded switch makes the operation
easy and prompt. The speed can be adjusted manually or automatically through
external control interface.
μ
×
Note:
Ref. page 11 to know detailed external control function.
The pump must be grounded properly before operating in order to protect the
operator in moist environment.
S Standard Configuration
S1 Pump Head
S2 Drive
PPart
P1 Signal Output Port
P2 External Control Input Port
P3 Power Socket
P4 Infuse Holder
Basic Operation
Start/Stop Key
Direction Key
Knob
Knob
Knob
Knob
Prime Key
Prime Key
Prime Key
Exit Key
Exit Key
Power Switch
Press to start or stop the pump
Press to change the rotation direction of the pump
Function 1: Turn the to adjust flow rate under none-prime working status.
Function 2: Menu selection function, turn the to select menu and set the
parameters, press to confirm the selection.
In common state, press the to enter prime state when the pump runs
at full speed for empting, filling and rinsing operation; Press the
again to return to common state. In prime state, other keys are invalid.
Function 1: Cancel current operation and return to previous menu.
Function 2: In dispensing working mode, press to check current speed
and the needed time for dispensing the set volume.
Turn on or off the pump.
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

1
Operation Panel
2
Start/Stop Key
Exit Key
Direction Key
Prime Key
Knob
S1
S2
P3
P2
P4
P1
LCD Display
Power Switch
Introduction
Applicable pump heads for BT100-1F are YZ1515x, YZ2515x, YZII15, YZII25 and
DG-1, DG-2, DG-4. The pump delivers flow rates from 0.16 L to 500 mL / min
and dispensing volume is from 0.01 mL to 9.99 L. Combining 128 32 LCD
display with membrane keypad and rotary coded switch makes the operation
easy and prompt. The speed can be adjusted manually or automatically through
external control interface.
μ
×
Note:
Ref. page 11 to know detailed external control function.
The pump must be grounded properly before operating in order to protect the
operator in moist environment.
S Standard Configuration
S1 Pump Head
S2 Drive
PPart
P1 Signal Output Port
P2 External Control Input Port
P3 Power Socket
P4 Infuse Holder
Basic Operation
Start/Stop Key
Direction Key
Knob
Knob
Knob
Knob
Prime Key
Prime Key
Prime Key
Exit Key
Exit Key
Power Switch
Press to start or stop the pump
Press to change the rotation direction of the pump
Function 1: Turn the to adjust flow rate under none-prime working status.
Function 2: Menu selection function, turn the to select menu and set the
parameters, press to confirm the selection.
In common state, press the to enter prime state when the pump runs
at full speed for empting, filling and rinsing operation; Press the
again to return to common state. In prime state, other keys are invalid.
Function 1: Cancel current operation and return to previous menu.
Function 2: In dispensing working mode, press to check current speed
and the needed time for dispensing the set volume.
Turn on or off the pump.
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

Flow Rate Display
b.
c. d.
e.
Running Interface
Turn on the pump, the pump initializes itself first then display the running interface.
a.
3 4
7.000 mL/min
50.0
INT
g.
f. Vol 13.0 mL:
Copy 9999:
b.
c.
e.
h.
c.
e.
Disp T. 79.0":
100 RPM
Dispensing Display: Dispensing status is displayed by two interfaces below,
these interface can be switched by pressing .Exit Key
Suitable Pump Heads and Tubing
Suitable Pump Heads Tubing (mm) Ref. Flow Rate(mL/min)
15#
#
24
0.17-170
0.29-290
13
14
19
16
25
17
18
#
#
#
#
#
#
#
YZ2515
YZ 25
x
Ⅱ
==
DG-1/2/4(10 )Rollers
DG-1/2/4(6 )Rollers
7 L-7mL/min
0.027-27
0.051-51
0.082-82
0.17-170
0.29-290
0.38-380
μ
15
24
5
#
#
#
#
3
36
0.17-170
0.29-290
0.38-380
0.50-500
0.13 0.91
0.25 0.91
0.51 0.91
1.02 0.86
1.65 0.86
2.4 0.8
2.79 0.86
3.17 0.86
×
×
×
×
×
2×1
×
×
×
0.13 0.91
0.25 0.91
0.51 0.91
1.02 0.86
1.65 0.86
2.4 0.8
2.79 0.86
3.17 0.86
×
×
×
×
×
2×1
×
×
×
0.25 L/min-160 L/min
0.52 L/min-520 L/min
2.0 L/min-2.0mL/min
4.6 L/min-4.6mL/min
0.014 - 14
0.016 - 16
0.022 - 22
0.028 - 28
0.035 - 32
μμ
μμ
μ
μ
0.16 L/min-160 L/min
0.56 L/min-560 L/min
2.2 L/min-2.2mL/min
5.8 L/min-5.8mL/min
0.018-18
0.02-21
0.026-26
0.035-35
0.044-44
μμ
μμ
μ
μ
a. Current flow rate: Display current flow rate. Turning the can increase or
decrease the flow rate.
b. Current speed: Display current speed. the can increase or
decrease the speed.
c. Running direction: Indicate the direction of the pump.
d. Running status: stands for pump running; stands for pump stopping;
Stands for pump pausing. Press to control the working state of
the pump.
e. Control Mode: stands for internal control mode; stands for closing the
external control by setting menu; stands for inserting analog voltage input
control module into ; stands for inserting analog
current control module into ; stands for inserting 0-10
KHz frequency input module into .
f. Dispensing Volume: Display dispensing volume under dispensing mode.
g. Copy number: Display copy number under dispensing mode.
h. Dispensing time: Display the dispensing time. When the dispensing procedure
starts, the dispensing time will count down. Adjust the Flow rate to change
Dispensing time in dispensing parameters setting interface.
Knob
Knob
Start/Stop Key
INT Off
V
External Control Port mA
External Control Port Hz
External Control Port
Turning
YZ1515 YZ 15xⅡ
Pump heads and tubing can be selected referring to above table. It is necessary to
select suitable pump heads and tubing to get higher flow rate and higher dispensing
accuracy. We suggest your needed flow rate range is in the range of 20% to 80% of
ref. flow rate range.
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

Flow Rate Display
b.
c. d.
e.
Running Interface
Turn on the pump, the pump initializes itself first then display the running interface.
a.
3 4
7.000 mL/min
50.0
INT
g.
f. Vol 13.0 mL:
Copy 9999:
b.
c.
e.
h.
c.
e.
Disp T. 79.0":
100 RPM
Dispensing Display: Dispensing status is displayed by two interfaces below,
these interface can be switched by pressing .Exit Key
Suitable Pump Heads and Tubing
Suitable Pump Heads Tubing (mm) Ref. Flow Rate(mL/min)
15#
#
24
0.17-170
0.29-290
13
14
19
16
25
17
18
#
#
#
#
#
#
#
YZ2515
YZ 25
x
Ⅱ
==
DG-1/2/4(10 )Rollers
DG-1/2/4(6 )Rollers
7 L-7mL/min
0.027-27
0.051-51
0.082-82
0.17-170
0.29-290
0.38-380
μ
15
24
5
#
#
#
#
3
36
0.17-170
0.29-290
0.38-380
0.50-500
0.13 0.91
0.25 0.91
0.51 0.91
1.02 0.86
1.65 0.86
2.4 0.8
2.79 0.86
3.17 0.86
×
×
×
×
×
2×1
×
×
×
0.13 0.91
0.25 0.91
0.51 0.91
1.02 0.86
1.65 0.86
2.4 0.8
2.79 0.86
3.17 0.86
×
×
×
×
×
2×1
×
×
×
0.25 L/min-160 L/min
0.52 L/min-520 L/min
2.0 L/min-2.0mL/min
4.6 L/min-4.6mL/min
0.014 - 14
0.016 - 16
0.022 - 22
0.028 - 28
0.035 - 32
μμ
μμ
μ
μ
0.16 L/min-160 L/min
0.56 L/min-560 L/min
2.2 L/min-2.2mL/min
5.8 L/min-5.8mL/min
0.018-18
0.02-21
0.026-26
0.035-35
0.044-44
μμ
μμ
μ
μ
a. Current flow rate: Display current flow rate. Turning the can increase or
decrease the flow rate.
b. Current speed: Display current speed. the can increase or
decrease the speed.
c. Running direction: Indicate the direction of the pump.
d. Running status: stands for pump running; stands for pump stopping;
Stands for pump pausing. Press to control the working state of
the pump.
e. Control Mode: stands for internal control mode; stands for closing the
external control by setting menu; stands for inserting analog voltage input
control module into ; stands for inserting analog
current control module into ; stands for inserting 0-10
KHz frequency input module into .
f. Dispensing Volume: Display dispensing volume under dispensing mode.
g. Copy number: Display copy number under dispensing mode.
h. Dispensing time: Display the dispensing time. When the dispensing procedure
starts, the dispensing time will count down. Adjust the Flow rate to change
Dispensing time in dispensing parameters setting interface.
Knob
Knob
Start/Stop Key
INT Off
V
External Control Port mA
External Control Port Hz
External Control Port
Turning
YZ1515 YZ 15xⅡ
Pump heads and tubing can be selected referring to above table. It is necessary to
select suitable pump heads and tubing to get higher flow rate and higher dispensing
accuracy. We suggest your needed flow rate range is in the range of 20% to 80% of
ref. flow rate range.
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

6
5
VOL 0.25mL:
Copy 991:
YZ1515 ID 0.80
YZ1515
YZ2515
Select Tubing ID:
0.80 1.60 2.40
Ext Control :
ON OFF
Pump Address:
1
Menu Diagram
Running Interface
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1. Pump head Tube&
YZ1515 ID 0.80
2.Ext Cont. ON:
3.Address 1:
4.Footswitch G:
5.Back T. 1 2:."
Flowrate
Dispensing
Disp Vol 2.02mL:
Copy No. 10:
Flow 2.043mL/min:
Pause T. 1.0:"
Flowrate
Dispensing
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1.System Setting
2.Control Mode
1. Pump head Tube&
1.System Setting
2.Control Mode
2.Ext Cont. ON:
3.Address 1:
1.System Setting
2.Control Mode
2.Ext Cont. ON:
3.Address 1:
Pump Head Installation
The pump head is mounted on drive before the pump leaves the factory. Follow
below procedure for changing another pump head.
Loose the two M4 screws which connect the pump head and the drive.
Dismantle the pump head slightly.
Insert the flat end of the pump head's shaft to the slot of drive's coupling.
Make the location hole of the pump head match the location pin of the drive.
Insert the two mounting screws into the mounting holes of the pump head.
Then tighten the mounting screws to connecting hole of drive. (For more
information, please refer to pump head's operation manual)
Turn off the power before changing the pump head.
Note:
System Setting:
All the setting can not be done when pump is running.
Pump head and tubing setting
Press the to enter interface, press again to enter
submenu of , then press again the pump will show all
the suitable pump head, turn to select needed pump head, press the
again to confirm the selection of pump head and the pump will show all the
suitable tubing, turn to select needed tubing and press to confirm or
press to return to previous menu.
Knob System Setting Knob
Pump Head & Tube Knob
Knob Knob
Knob Knob
Exit Key
External Control
Set the pump to enable the or not.
On: Enable
Off: Disable
Set the
Press and turn to highlight the external control line. Press to enter
next interface, turn to select or then press to confirm or press
to cancel the selection and return to previous menu.
In external control mode, the status of external control can be changed when the
pump stops.
External Control
External Control
External Control
External Control
Knob Knob
Knob On Off Knob
Exit Key
Press and turn Knob
Press and turn Knob
Address:
When control computer controls many pumps through RS485 interface, it must
identify each pump's I. D. This pump I.D. should be unique. It's the identification
of this pump. Maximum 30 pieces BT100-1F can be controlled through RS485 at
the same time.
Setting of each pump
Press and turn the to highlight the line. Press again to enter
next interface, turn Knob to select , and then press to confirm or
press Exit Key to cancel the selection and return to previous menu.
Address
Knob Address Knob
Pump ID Knob
1.
2.
3.
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

6
5
VOL 0.25mL:
Copy 991:
YZ1515 ID 0.80
YZ1515
YZ2515
Select Tubing ID:
0.80 1.60 2.40
Ext Control :
ON OFF
Pump Address:
1
Menu Diagram
Running Interface
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1. Pump head Tube&
YZ1515 ID 0.80
2.Ext Cont. ON:
3.Address 1:
4.Footswitch G:
5.Back T. 1 2:."
Flowrate
Dispensing
Disp Vol 2.02mL:
Copy No. 10:
Flow 2.043mL/min:
Pause T. 1.0:"
Flowrate
Dispensing
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1.System Setting
2.Control Mode
1. Pump head Tube&
1.System Setting
2.Control Mode
2.Ext Cont. ON:
3.Address 1:
1.System Setting
2.Control Mode
2.Ext Cont. ON:
3.Address 1:
Pump Head Installation
The pump head is mounted on drive before the pump leaves the factory. Follow
below procedure for changing another pump head.
Loose the two M4 screws which connect the pump head and the drive.
Dismantle the pump head slightly.
Insert the flat end of the pump head's shaft to the slot of drive's coupling.
Make the location hole of the pump head match the location pin of the drive.
Insert the two mounting screws into the mounting holes of the pump head.
Then tighten the mounting screws to connecting hole of drive. (For more
information, please refer to pump head's operation manual)
Turn off the power before changing the pump head.
Note:
System Setting:
All the setting can not be done when pump is running.
Pump head and tubing setting
Press the to enter interface, press again to enter
submenu of , then press again the pump will show all
the suitable pump head, turn to select needed pump head, press the
again to confirm the selection of pump head and the pump will show all the
suitable tubing, turn to select needed tubing and press to confirm or
press to return to previous menu.
Knob System Setting Knob
Pump Head & Tube Knob
Knob Knob
Knob Knob
Exit Key
External Control
Set the pump to enable the or not.
On: Enable
Off: Disable
Set the
Press and turn to highlight the external control line. Press to enter
next interface, turn to select or then press to confirm or press
to cancel the selection and return to previous menu.
In external control mode, the status of external control can be changed when the
pump stops.
External Control
External Control
External Control
External Control
Knob Knob
Knob On Off Knob
Exit Key
Press and turn Knob
Press and turn Knob
Address:
When control computer controls many pumps through RS485 interface, it must
identify each pump's I. D. This pump I.D. should be unique. It's the identification
of this pump. Maximum 30 pieces BT100-1F can be controlled through RS485 at
the same time.
Setting of each pump
Press and turn the to highlight the line. Press again to enter
next interface, turn Knob to select , and then press to confirm or
press Exit Key to cancel the selection and return to previous menu.
Address
Knob Address Knob
Pump ID Knob
1.
2.
3.
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

78
Flowrate
Dispensing
Disp Vol 2.02mL:
Copy No. 10:
Disp Vol:
2.02 mL
Copy No.:
10
4.Footswitch G:
5.Back T. 1 2:."
Footswitch:
Gated Trigger
Back Time :
"1.2
1.System Setting
2.Control Mode
1.System Setting
2.Control Mode
4.Footswitch G:
5.Back T. 1 2:."
1.System Setting
2.Control Mode
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Disp Vol 2.02mL:
Copy No. 10:
Footswitch Setting
Footswitch has two kinds of working status.
Press footswitch, the pump starts running; Press footswitch again,
the pump stops.
The pump runs as long as the footswitch is pressed.
Set Footswitch
Press and turn to highlight the Footswitch line. Press to enter next
interface. Turn to select or , then press to confirm or
press to cancel the selection and return to previous menu.
The pump identifies the Longer Footswitch automatically. When the pump
connects the footswitch, the Key on operating panel is invalid no
matter the remote control is on or off.
Trigger:
Gated:
Knob Knob
Trigger Gated Knob
Exit Key
Note:
Start/Stop
Knob
Press and turn Knob
Press and turn Knob
Back Suction Time Setting
In dispensing and filling state, to prevent the liquids from dropping to cause
error when the filling stops, the pump will back-turn a suction angle to suck
back the liquids. The back suction is the same for each filling operations, it
doesn't influence the filling accuracy.
Set :
Press and turn to highlight the back suction time line. Press the to
enter next interface, turn to adjust , press to
confirm or press to cancel the adjustment and return to previous
menu.
Back Suction Time
Knob Knob
Knob Back Suction Time Knob
Exit Key
Control Mode
The pump has two kinds of working mode: and .
Set
Press and turn the to highlight control mode line. Press to enter next
interface, turn Knob to select control mode, press to confirm or press
to cancel the selection and return to previous menu.
Flowrate Dispensing
Control Mode:
knob Knob
Knob Exit
Key
Dispensing Setting
Before dispensing process, the parameters including Pump Head, Tubing,
Dispensing Volume, Copy number, Flow Rate and Pause time must be set.
Dispensing Volume: the volume that the pump dispenses between every time
interval.
Press and turn to highlight dispense setting line. Press to highlight
dispensing volume line. Press again to highlight the value of the volume.
Turn to select Dispensing Volume. Press to confirm or press
to cancel the selection and return to previous menu.
Knob Knob
Knob
Knob Knob Exit Key
Copy No.: the total filling number. Its range is 0 to 9999.
Press to highlight copy number line. Press Knob to enter next interface
Turn to select copy number. Press to confirm or press to
cancel the selection and return to previous menu.
If the is “0”, the dispensing process of the pump will continue until
press key or turn off the pump to stop dispense process.
form
Knob .
Knob Knob Exit Key
Copy No.
Start/Stop
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

78
Flowrate
Dispensing
Disp Vol 2.02mL:
Copy No. 10:
Disp Vol:
2.02 mL
Copy No.:
10
4.Footswitch G:
5.Back T. 1 2:."
Footswitch:
Gated Trigger
Back Time :
"1.2
1.System Setting
2.Control Mode
1.System Setting
2.Control Mode
4.Footswitch G:
5.Back T. 1 2:."
1.System Setting
2.Control Mode
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Disp Vol 2.02mL:
Copy No. 10:
Footswitch Setting
Footswitch has two kinds of working status.
Press footswitch, the pump starts running; Press footswitch again,
the pump stops.
The pump runs as long as the footswitch is pressed.
Set Footswitch
Press and turn to highlight the Footswitch line. Press to enter next
interface. Turn to select or , then press to confirm or
press to cancel the selection and return to previous menu.
The pump identifies the Longer Footswitch automatically. When the pump
connects the footswitch, the Key on operating panel is invalid no
matter the remote control is on or off.
Trigger:
Gated:
Knob Knob
Trigger Gated Knob
Exit Key
Note:
Start/Stop
Knob
Press and turn Knob
Press and turn Knob
Back Suction Time Setting
In dispensing and filling state, to prevent the liquids from dropping to cause
error when the filling stops, the pump will back-turn a suction angle to suck
back the liquids. The back suction is the same for each filling operations, it
doesn't influence the filling accuracy.
Set :
Press and turn to highlight the back suction time line. Press the to
enter next interface, turn to adjust , press to
confirm or press to cancel the adjustment and return to previous
menu.
Back Suction Time
Knob Knob
Knob Back Suction Time Knob
Exit Key
Control Mode
The pump has two kinds of working mode: and .
Set
Press and turn the to highlight control mode line. Press to enter next
interface, turn Knob to select control mode, press to confirm or press
to cancel the selection and return to previous menu.
Flowrate Dispensing
Control Mode:
knob Knob
Knob Exit
Key
Dispensing Setting
Before dispensing process, the parameters including Pump Head, Tubing,
Dispensing Volume, Copy number, Flow Rate and Pause time must be set.
Dispensing Volume: the volume that the pump dispenses between every time
interval.
Press and turn to highlight dispense setting line. Press to highlight
dispensing volume line. Press again to highlight the value of the volume.
Turn to select Dispensing Volume. Press to confirm or press
to cancel the selection and return to previous menu.
Knob Knob
Knob
Knob Knob Exit Key
Copy No.: the total filling number. Its range is 0 to 9999.
Press to highlight copy number line. Press Knob to enter next interface
Turn to select copy number. Press to confirm or press to
cancel the selection and return to previous menu.
If the is “0”, the dispensing process of the pump will continue until
press key or turn off the pump to stop dispense process.
form
Knob .
Knob Knob Exit Key
Copy No.
Start/Stop
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

910
If the actual volume is known in advance, enter the calibration interface and
input the actual volume directly.
Disp Vol 2.02mL
Disp T. 60.5"
:
:
Disp T. 60.5
Act Vol 2.020 mL
:
:
"
Act Vol :
Flow 2.043mL/min:
Pause T. :"1.0
Pause Time :
"1.1
Flowrate
Dispensing
Test T. 0.5 ’
Cal Vol 132.3 L
:
:μ
Test T.
Cal Vol 132.3 L
:
:μ
30"
2.02 mL
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Flow 2.043mL/min:
Pause T. :"1.0
Flow 2.043mL/min:
Pause T. :"1.0
Test T. 0.5 ’
Cal Vol 132.3 L
:
:μ
Test T. 0.5 ’
Cal Vol 132.3 L
:
:μ
Test T.
Cal Vol 133.0 L
:
:μ
30"
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Flowrate
Dispensing
Test T. 0.5 ’
Cal Vol 132.3 L
:
:μ
Test T. 0.5 ’
Cal Vol 132.3 L
:
:μ
:Act Vol
133.0 Lμ
Flowrate
Dispensing
Disp T. 60.5
Act Vol 2.040 mL
:
:
"
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Flowrate
Dispensing
Disp Vol 2.02mL
Disp T. 60.5"
:
:
Flow Rate: Adjusting the dispensing flow rate can change the dispensing time.
Press and turn to highlight dispensing flow rate line. Press to enter
next interface, turn to select suitable flow rate. Press to confirm or
press to cancel the selection and return to previous menu.
Knob Knob
Knob Knob
Exit Key
Press
Knob
Pause Time: Time Interval between every dispense process.
Press and turn to highlight pause time line. Press to enter next
interface. Turn to select pause time. Press to confirm or press
to cancel the selection and return to previous menu.
Knob Knob
Knob Knob Exit
Key
Calibration
Pump should be calibrated when flow rate or dispense volume is beyond the
tolerance. Flow rate and dispensing volume can be calibrated separately in
different application.
Flow Rate Calibration
Press and turn to highlight calibration line. Press to enter next
interface. Turn to select Flowrate. Press to highlight test time line.
Press again to highlight the value of time. Turn to select test time
(its range is from 0.5 to 30 minutes). Press to confirm or press
to cancel the selection and return to previous menu.
Press Start/Stop Key, the LCD displays running time and actual volume.
Running time counts down.
Turn to input actual tested volume. Press to confirm. The
calibration operation can be repeated.
Knob Knob
Knob Knob
Knob Knob
Knob Exit Key
Knob Knob
Turn Knob to input
Actual Volume
If the actual volume is known in advance, enter the calibration interface and
input the actual volume directly.
Turn to input
Actual Volume
Knob
Dispensing Volume Calibration
The parameters can not be changed under dispensing volume calibration
operation. The defaulted parameters are the parameters set under Dispensing
Setting.
Pump stops under dispense mode, press and turn to highlight calibration
line. Press to enter next interface. Turn to select Dispensing. Press
to enter next interface.
Press Key, the LCD displays running time and actual volume.
Running time counts down.
Turn to input actual tested volume. Press to confirm.
Knob
Knob
Start/Stop
Knob Knob
Knob
Knob
Turn to input
actual volume
Knob
Turn to input
actual volume
Knob
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

910
If the actual volume is known in advance, enter the calibration interface and
input the actual volume directly.
Disp Vol 2.02mL
Disp T. 60.5"
:
:
Disp T. 60.5
Act Vol 2.020 mL
:
:
"
Act Vol :
Flow 2.043mL/min:
Pause T. :"1.0
Pause Time :
"1.1
Flowrate
Dispensing
Test T. 0.5 ’
Cal Vol 132.3 L
:
:μ
Test T.
Cal Vol 132.3 L
:
:μ
30"
2.02 mL
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Flow 2.043mL/min:
Pause T. :"1.0
Flow 2.043mL/min:
Pause T. :"1.0
Test T. 0.5 ’
Cal Vol 132.3 L
:
:μ
Test T. 0.5 ’
Cal Vol 132.3 L
:
:μ
Test T.
Cal Vol 133.0 L
:
:μ
30"
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Flowrate
Dispensing
Test T. 0.5 ’
Cal Vol 132.3 L
:
:μ
Test T. 0.5 ’
Cal Vol 132.3 L
:
:μ
:Act Vol
133.0 Lμ
Flowrate
Dispensing
Disp T. 60.5
Act Vol 2.040 mL
:
:
"
1.System Setting
2.Control Mode
3.Disp Setting
4.Calibration
Flowrate
Dispensing
Disp Vol 2.02mL
Disp T. 60.5"
:
:
Flow Rate: Adjusting the dispensing flow rate can change the dispensing time.
Press and turn to highlight dispensing flow rate line. Press to enter
next interface, turn to select suitable flow rate. Press to confirm or
press to cancel the selection and return to previous menu.
Knob Knob
Knob Knob
Exit Key
Press
Knob
Pause Time: Time Interval between every dispense process.
Press and turn to highlight pause time line. Press to enter next
interface. Turn to select pause time. Press to confirm or press
to cancel the selection and return to previous menu.
Knob Knob
Knob Knob Exit
Key
Calibration
Pump should be calibrated when flow rate or dispense volume is beyond the
tolerance. Flow rate and dispensing volume can be calibrated separately in
different application.
Flow Rate Calibration
Press and turn to highlight calibration line. Press to enter next
interface. Turn to select Flowrate. Press to highlight test time line.
Press again to highlight the value of time. Turn to select test time
(its range is from 0.5 to 30 minutes). Press to confirm or press
to cancel the selection and return to previous menu.
Press Start/Stop Key, the LCD displays running time and actual volume.
Running time counts down.
Turn to input actual tested volume. Press to confirm. The
calibration operation can be repeated.
Knob Knob
Knob Knob
Knob Knob
Knob Exit Key
Knob Knob
Turn Knob to input
Actual Volume
If the actual volume is known in advance, enter the calibration interface and
input the actual volume directly.
Turn to input
Actual Volume
Knob
Dispensing Volume Calibration
The parameters can not be changed under dispensing volume calibration
operation. The defaulted parameters are the parameters set under Dispensing
Setting.
Pump stops under dispense mode, press and turn to highlight calibration
line. Press to enter next interface. Turn to select Dispensing. Press
to enter next interface.
Press Key, the LCD displays running time and actual volume.
Running time counts down.
Turn to input actual tested volume. Press to confirm.
Knob
Knob
Start/Stop
Knob Knob
Knob
Knob
Turn to input
actual volume
Knob
Turn to input
actual volume
Knob
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

Select pump head
and tubing refer to
thetableofpage4
External Control Input
Analog Signal Input function
Communication Control
The pump can connect to control computer (computer, PLC, SCM) through
RS485 serial communication module (shown as below). Please contact Longer
Company for communication protocol.
COM
B
A
11 12
Set the “External Control” in “on” state. External Control Module is shown as below.
Note:
Five kinds of standard external control module need to be ordered separately
according to special requirements.
5
4
3
2
1
External Control Module
Analog Control
4 - 20 mA Speed Control
0-5VSpeed Control
0 - 10 V Speed Control
0 10 KHz Speed Control-
External Control
DB15
Dispensing Mode operation procedure
Confirm the
volume to be
dispensed
Press to
return to running
interface
Exit Key
Operation Procedure
Select pump head and tubing refer to “Suitable Pump head, Tubing, Ref.
Flow Rate” table before working. Thick wall tubing is preferred for longer
service life.
Flow rate mode operation procedure
Confirm the flow
rate of the fluid
which need to be
pumped
Press to
enter
interface
to set pump head
and tubing
Knob
System
Setting
Enter
interface to
select
Control
Mode
Flow Mode
Press to
return to running
interface, turn
to select
needed flow rate
Exit Key
Knob
Press to
enter calibration
selection interface
and select
when
calibration is
needed
Knob
Flowrate
Set ,
press
to start the pump.
Input the actual
volume
Test Time
Start/Stop
Press to
return to running
interface
Exit Key
Select pump
head and tubing
refer to the table
of page 4
Press to
enter
interface
to set pump head
and tubing
Knob
System
Setting Set
time
Back
Suction
Enter control
mode interface
to select
Dispensing
Enter
interface
to set all the
dispensing
parameters.
Dispense
Setting
Press to
enter calibration
selection
interface and
select
when calibration
is needed
Knob
Dispensing
Press
to start the pump.
Input the actual
volume
Start/Stop
Communication Control RS485
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

Select pump head
and tubing refer to
thetableofpage4
External Control Input
Analog Signal Input function
Communication Control
The pump can connect to control computer (computer, PLC, SCM) through
RS485 serial communication module (shown as below). Please contact Longer
Company for communication protocol.
COM
B
A
11 12
Set the “External Control” in “on” state. External Control Module is shown as below.
Note:
Five kinds of standard external control module need to be ordered separately
according to special requirements.
5
4
3
2
1
External Control Module
Analog Control
4 - 20 mA Speed Control
0-5VSpeed Control
0 - 10 V Speed Control
0 10 KHz Speed Control-
External Control
DB15
Dispensing Mode operation procedure
Confirm the
volume to be
dispensed
Press to
return to running
interface
Exit Key
Operation Procedure
Select pump head and tubing refer to “Suitable Pump head, Tubing, Ref.
Flow Rate” table before working. Thick wall tubing is preferred for longer
service life.
Flow rate mode operation procedure
Confirm the flow
rate of the fluid
which need to be
pumped
Press to
enter
interface
to set pump head
and tubing
Knob
System
Setting
Enter
interface to
select
Control
Mode
Flow Mode
Press to
return to running
interface, turn
to select
needed flow rate
Exit Key
Knob
Press to
enter calibration
selection interface
and select
when
calibration is
needed
Knob
Flowrate
Set ,
press
to start the pump.
Input the actual
volume
Test Time
Start/Stop
Press to
return to running
interface
Exit Key
Select pump
head and tubing
refer to the table
of page 4
Press to
enter
interface
to set pump head
and tubing
Knob
System
Setting Set
time
Back
Suction
Enter control
mode interface
to select
Dispensing
Enter
interface
to set all the
dispensing
parameters.
Dispense
Setting
Press to
enter calibration
selection
interface and
select
when calibration
is needed
Knob
Dispensing
Press
to start the pump.
Input the actual
volume
Start/Stop
Communication Control RS485
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

13 14
Pump is equipped with output port in order to monitor the status of pump. It
adopts DB9 male socket as output port (see as below). External control output
start/stop output, CW/CCCW output and pulse output ( 0 - 100 rpm
corresponding to 0 - 10 kHz). Output signal adopts optoelectronic isolation
circuit. A pull-up resistor and power supply must be added when using.
Terminals Definition
External Control Output
External Control Output Port Terminal Drawing: DB9
Footswitch Function
Connect to to control Start/Stop of
pump.
Footswitch External Control Input Port
Terminals Definition
0-5V,0-10V 4-20mA
Analog input to control the speed
,CC/CCW Signal Start/Stop Signal
1
4
3
4
2
0 5V, 0 10V, 4 20mA
External Control Input Module Connection Diagram
-- -
Pulse Signal Input Function:
0 - 10 kHz pulse input
+
-
4
5
CC/CCW Signal Start/Stop Signal
+
-
3
1
+
-
2
COM1 COM
0 - 10 kHz Pulse Signal Input Connection Diagram
Analog input in 4 20 mA,0-5V,0 10Vexternalcontrolmodule
to control the speed of pump.
External control start/stop input. Open, the pump runs; Close, the
pump stops.
External control direction input. Open, the pump runs clockwise;
Close, the pumps runs counter clockwise.
COM of external control input.
--
Terminal 1
Terminal 2
Terminal 3
Terminal 4
COM of external control input
External control start/stop input. Open, the pump runs; Close, the
pump stops.
External control direction input. Open, the pump runs clockwise;
Close, the pumps runs counter clockwise.
Pulse input in 0 10 kHz pulse external control module to control
the speed of the pump.
COM of pulse input
-
Terminal 1
Terminal 2
Terminal 3
Terminal 4
Terminal 5
Terminal 1
Terminal 2
Terminal 8
Terminal 4, 6, 7
Start/Stop output. Pump outputs low level when it runs; pump
output high level when it stops.
CW/CCW output. Pump output low level when it runs clock
wise; pump output high level when it runs count clock wise.
Pulse output, 0 100 rpm corresponding to 0 10 kHz.
COM
--
Terminal 13
Terminal 5, 11
Note:
Footswitch is optional accessories. It can only control the start and stop of the
pump. The speed and direction of the pump are controlled by membrane keypad
and knob on the operation penal.
Start/Stop control. Close or input
pulse, the pump stops; Open or
input pulse again, the pump runs
COM when connecting terminal 5
and11together.
Connection Diagram for Footswitch
External Control Port
Relay
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

13 14
Pump is equipped with output port in order to monitor the status of pump. It
adopts DB9 male socket as output port (see as below). External control output
start/stop output, CW/CCCW output and pulse output ( 0 - 100 rpm
corresponding to 0 - 10 kHz). Output signal adopts optoelectronic isolation
circuit. A pull-up resistor and power supply must be added when using.
Terminals Definition
External Control Output
External Control Output Port Terminal Drawing: DB9
Footswitch Function
Connect to to control Start/Stop of
pump.
Footswitch External Control Input Port
Terminals Definition
0-5V,0-10V 4-20mA
Analog input to control the speed
,CC/CCW Signal Start/Stop Signal
1
4
3
4
2
0 5V, 0 10V, 4 20mA
External Control Input Module Connection Diagram
-- -
Pulse Signal Input Function:
0 - 10 kHz pulse input
+
-
4
5
CC/CCW Signal Start/Stop Signal
+
-
3
1
+
-
2
COM1 COM
0 - 10 kHz Pulse Signal Input Connection Diagram
Analog input in 4 20 mA,0-5V,0 10Vexternalcontrolmodule
to control the speed of pump.
External control start/stop input. Open, the pump runs; Close, the
pump stops.
External control direction input. Open, the pump runs clockwise;
Close, the pumps runs counter clockwise.
COM of external control input.
--
Terminal 1
Terminal 2
Terminal 3
Terminal 4
COM of external control input
External control start/stop input. Open, the pump runs; Close, the
pump stops.
External control direction input. Open, the pump runs clockwise;
Close, the pumps runs counter clockwise.
Pulse input in 0 10 kHz pulse external control module to control
the speed of the pump.
COM of pulse input
-
Terminal 1
Terminal 2
Terminal 3
Terminal 4
Terminal 5
Terminal 1
Terminal 2
Terminal 8
Terminal 4, 6, 7
Start/Stop output. Pump outputs low level when it runs; pump
output high level when it stops.
CW/CCW output. Pump output low level when it runs clock
wise; pump output high level when it runs count clock wise.
Pulse output, 0 100 rpm corresponding to 0 10 kHz.
COM
--
Terminal 13
Terminal 5, 11
Note:
Footswitch is optional accessories. It can only control the start and stop of the
pump. The speed and direction of the pump are controlled by membrane keypad
and knob on the operation penal.
Start/Stop control. Close or input
pulse, the pump stops; Open or
input pulse again, the pump runs
COM when connecting terminal 5
and11together.
Connection Diagram for Footswitch
External Control Port
Relay
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

Technical Specifications
15
Functions
Acceptable Pump Head: YZ1515X, YZ2515X, YZII15, YZII25, DG-1, DG-2, DG-4
Operation Mode: Membrane keypad and Knob
Direction Control: CW and CCW, reversible
Prime: Full speed for fast filling and emptying
Back Suction: Preventing drippage of the liquid
Display: 128*32 graphic LCD displays all the information
External Control Input: Control Start/Stop, direction and flow rate of the pump under
Flowrate mode
Footswitch: Control Start/Stop of the pump
External Control Output: Output the signals of Start/Stop, direction and speed
Communication: Communicate with control computer
Flowrate Function: Deliver fluid at set flow rate
Dispensing Function: Set dispensing volume, copy number and pause time
Memory Function: Store all the running information automatically
Calibration: Acquire higher accuracy
Cooling Mode: Heat-emitting fan
16
Communication Function
BT100-1F can connect to the control computer (computer, PLC, SCM) through
RS485 serial communication interface. Max. 30 pumps can be controlled by a
control computer.
Note:
1. The pump ID which connects to the control computer through RS485 serial
communication interface must be unique to prevent the communication
error.
2. Please contact Longer Company for communication protocol.
Maintenance
When the pump is idle, we recommend you to release the tubing from pressure.
This helps to protect the tubing from unnecessary strain and prolongs its
service life
Keep rollers clean and dry. This will prolong the service lives of tubing and
pump head.
The surface of drive and the pump head are not organic solvent and aggressive
liquids resistant. Please pay attention when using.
If a trouble happens, please contact us or our dealers.
Note :
Warranty
The warranty period for this product is one year. If repair or adjustment is
necessary within the warranty period, the problem will be corrected at no
charge if it is not due to misuse or abuse on your part, as determined by the
manufacturer. Repair costs outside the warranty period, or those resulting
from product misuse or abuse, may be invoiced to you.
Specification
Flow Rates: 0.16 L /min to 500 mL/min
Dispensing Volume: 0.01mL to 9.99L
Copy Number: 0 to 9999
Pause Time: 0.1s to 99.9min
Back Suction Time: 0 to 99.9s
External Control Input: Start/Stop control, direction control, flow rates control (0 - 5V ,
0-10V,4-20mA,0-10KHz optional )
External Control Output: Start/Stop, Direction,0-10kHzspeed signal, OC Gate output
Communication Interface: RS485
Applicable Power: AC 90 - 264V 50Hz/60Hz
Power Consumption: <40W
Operating Condition: Temperature 0 to 40 C
Relative humidity < 80%
Dimension (L W H): 292.3 185 180.5 (mm)
Weight: 3.8 kg
IP Rating: IP 31
μ
°
×× × ×
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

Technical Specifications
15
Functions
Acceptable Pump Head: YZ1515X, YZ2515X, YZII15, YZII25, DG-1, DG-2, DG-4
Operation Mode: Membrane keypad and Knob
Direction Control: CW and CCW, reversible
Prime: Full speed for fast filling and emptying
Back Suction: Preventing drippage of the liquid
Display: 128*32 graphic LCD displays all the information
External Control Input: Control Start/Stop, direction and flow rate of the pump under
Flowrate mode
Footswitch: Control Start/Stop of the pump
External Control Output: Output the signals of Start/Stop, direction and speed
Communication: Communicate with control computer
Flowrate Function: Deliver fluid at set flow rate
Dispensing Function: Set dispensing volume, copy number and pause time
Memory Function: Store all the running information automatically
Calibration: Acquire higher accuracy
Cooling Mode: Heat-emitting fan
16
Communication Function
BT100-1F can connect to the control computer (computer, PLC, SCM) through
RS485 serial communication interface. Max. 30 pumps can be controlled by a
control computer.
Note:
1. The pump ID which connects to the control computer through RS485 serial
communication interface must be unique to prevent the communication
error.
2. Please contact Longer Company for communication protocol.
Maintenance
When the pump is idle, we recommend you to release the tubing from pressure.
This helps to protect the tubing from unnecessary strain and prolongs its
service life
Keep rollers clean and dry. This will prolong the service lives of tubing and
pump head.
The surface of drive and the pump head are not organic solvent and aggressive
liquids resistant. Please pay attention when using.
If a trouble happens, please contact us or our dealers.
Note :
Warranty
The warranty period for this product is one year. If repair or adjustment is
necessary within the warranty period, the problem will be corrected at no
charge if it is not due to misuse or abuse on your part, as determined by the
manufacturer. Repair costs outside the warranty period, or those resulting
from product misuse or abuse, may be invoiced to you.
Specification
Flow Rates: 0.16 L /min to 500 mL/min
Dispensing Volume: 0.01mL to 9.99L
Copy Number: 0 to 9999
Pause Time: 0.1s to 99.9min
Back Suction Time: 0 to 99.9s
External Control Input: Start/Stop control, direction control, flow rates control (0 - 5V ,
0-10V,4-20mA,0-10KHz optional )
External Control Output: Start/Stop, Direction,0-10kHzspeed signal, OC Gate output
Communication Interface: RS485
Applicable Power: AC 90 - 264V 50Hz/60Hz
Power Consumption: <40W
Operating Condition: Temperature 0 to 40 C
Relative humidity < 80%
Dimension (L W H): 292.3 185 180.5 (mm)
Weight: 3.8 kg
IP Rating: IP 31
μ
°
×× × ×
BT100-1F OPERATING MANUAL BT100-1F OPERATING MANUAL

BT100-1F Longer Peristaltic Pump
Baoding Longer Precision Pump Co.,Ltd.
BaodingLongerPrecisionPumpCo.,Ltd
ADD: Building A,Chuangye Center Baoding
National High - Tech Industrial Development Zone
Baoding,Hebei,China 071051
TEL: 86 - 312 - 3110087 3138553
FAX: 86 - 312 - 3168553
Www.longerpump.com
BT100-1F PERISTALTIC PUMP
OPERATING MANUAL
Table of contents
Other LongerPump Water Pump manuals
Popular Water Pump manuals by other brands

STI
STI SC/Vo instruction manual

Einhell Royal
Einhell Royal HW 811 operating instructions

GORMAN-RUPP PUMPS
GORMAN-RUPP PUMPS PA6H60-4045H FT4-ESP Installation, operation and maintenance manual

Becker
Becker SVw 380 Series operating instructions

BUSCH
BUSCH COBRA NC 1000 B instruction manual

SAER Elettropompe
SAER Elettropompe CB Series Use and maintenance manual