Series TID Temperature/Process Indicator
Specications - Installation and Operating Instructions
Bulletin E-90-TID
The affordable Series TID allows user to monitor temperature or a process value.
Temperature ranges are available from -58 to 302°F using one of our PTC or NTC
thermistors. Process values can be displayed from -999 to 999 counts using a 4-20 mA
signal from one of our various transmitters. The process indicator has an adjustable
span and zero on the 4-20 mA models.
Installation
Note: Unit must be mounted away form vibration, impacts, water, and corrosive gases.
• Cut hole in panel 2.80 x 1.14 in (71 x 29 mm).
• Apply silicone (or rubber gasket) around the perimeter of the hole to prevent leakage.
• Insert unit into hole of panel.
• Slideremovablettingclipsontounitfromthebackuntilsecuretopanel.
• Remove back cover to wire unit.
• Wiring diagram is displayed on the top of the unit.
SPECIFICATIONS
Inputs: PTC/NTC thermistor or 4-20 mA.
Display: 3-digits; Red, green or blue display.
Range: -58 to 302°F (thermistor); -999 to 999 counts (4-20 mA).
Power Requirements: 110 VAC, 230 VAC, 12 VAC/DC, 24 VAC/DC.
Accuracy: 1% FS.
Resolution: 1° or 0.01 count.
Front Panel Rating: IP64.
Agency Approvals: CE.
INPUT
GND
1234
57
8
IN 12 VDC
POWER
SUPPLY
CURRENT CONNECTION DIAGRAM
MODEL CODE
TID-
INPUT
1 =
THERMISTOR PTC
2 =
THERMISTOR NTC
3 = 4-20 mA
1 = 110 VAC
2 = 230 VAC
3 = 12 VAC/DC
4 = 24 VAC/DC
SUPPLY POWER UNITS
0 = RED
1 = GREEN
2 = BLUE
0 = NONE
1 = °F
2 = °C
DISPLAY
3
[76.2]
1-11/32
[34.13]
2-3/8
[60.38]
1-7/64
[28.18]
THERMISTOR CONNECTION DIAGRAM
PROBE
POWER
12 3 4
t°
LOVE CONTROLS
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
A DIVISION OF
DWYER INSTRUMENTS, INC. www.lovecontrols.com
e-mail: [email protected]
Printed in U.S.A. 8/19 FR# 443527-00 Rev. 3©Copyright 2019 Dwyer Instruments, Inc.
Parameter Descriptions
Lc = Value for 4 mA input
If P1 = Yes the value is displayed with decimal point
Hc = Value for 20 mA input
If P1 = Yes the value is displayed with decimal point
P1 = Decimal Point
If P1= Yes LC, Hc and probe values are displayed with decimal point
H5 = Access code to parameters
Parameter Programming
• Press SET for 8 seconds. The Access code value 00 is shown on the display (unit
comes with code set at 00 from factory).
• With the UP and DOWN arrows, code can be set to user needs.
• PressSETtoenterthecode.Ifthecodeiscorrect,therstparameterlabelisshown
on the display (Lc).
• Move to the desired parameter with the UP and DOWN arrow keys.
• Press SET to view the value on the display.
• ThevaluecanbemodiedwiththeUPandDOWNarrows.
• Press ENTER to enter the value and exit.
• Repeatuntilallnecessaryparametersaremodied.
• Press SET and DOWN at the same time to quit programming or wait one minute and
the display will automatically exit programming mode.
*The keyboard code can be reset to ZERO by turning off the controller and turning it
on again while keeping the SET key depressed.
Display Message
The display normally shows the temperature of the probe. In case of error the following
messages are shown:
• Er = Memory error
• 00 = Open probe
• -- = Short circuit probe
Maintenance, Cleaning and Repair
Afternalinstallationoftheunit,noroutinemaintenanceisrequired.Cleanthesurface
of the display controller with a soft and damp cloth. Never use abrasive detergents,
petrol, alcohol, or solvents. All repairs must be made by authorized personnel.
LIST OF PARAMETERS (CURRENT MODELS ONLY)
Description Units Range
Lc Value for 4 mA input Range -999 to 999
Hc Value for 20 mA input Range -999 to 999
P1 Decimal point Selection no yes
H5 Access code Numeric 0 to 99
LOVE CONTROLS
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
www.lovecontrols.com
e-mail: [email protected]

Other Love Measuring Instrument manuals

Love HM35 User manual

Love

Love HM35 User manual

Love CR10 User manual

Love

Love CR10 User manual

Love LDPM-01 User manual

Love

Love LDPM-01 User manual

Popular Measuring Instrument manuals by other brands

Diatest DIATRON 2200 instruction manual

Diatest

Diatest DIATRON 2200 instruction manual

Remocon HCD900 Operation manuals

Remocon

Remocon HCD900 Operation manuals

TESTO 622 instruction manual

TESTO

TESTO 622 instruction manual

Rohde & Schwarz ZVL series operating manual

Rohde & Schwarz

Rohde & Schwarz ZVL series operating manual

Hameg HZ350 manual

Hameg

Hameg HZ350 manual

Sonel MIC-10s1 operating manual

Sonel

Sonel MIC-10s1 operating manual

BW SENSING LEC315 Series Technical manual

BW SENSING

BW SENSING LEC315 Series Technical manual

Bitronics QTWIE2 user manual

Bitronics

Bitronics QTWIE2 user manual

Spectrum Technologies FIELDSCOUT 3413F user manual

Spectrum Technologies

Spectrum Technologies FIELDSCOUT 3413F user manual

Tektronix KEITHLEY 2450 quick start guide

Tektronix

Tektronix KEITHLEY 2450 quick start guide

Circutor DM45 AD manual

Circutor

Circutor DM45 AD manual

Tektronix CDC250 user manual

Tektronix

Tektronix CDC250 user manual

Veris Industries E50B1 installation guide

Veris Industries

Veris Industries E50B1 installation guide

Megger MPQ1000 user manual

Megger

Megger MPQ1000 user manual

Amprobe AT-6000-R quick start guide

Amprobe

Amprobe AT-6000-R quick start guide

UEi C20 instruction manual

UEi

UEi C20 instruction manual

Agatec CL250 user guide

Agatec

Agatec CL250 user guide

Stahl 8404/6 Series operating instructions

Stahl

Stahl 8404/6 Series operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.