LSI LASTEM ALIEM EL03305 User manual

Environmental monitoring solutions
ALIEM
Alpha-Log Input Extension Module
User manual

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 2 / 32

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 3 / 32
Document ALIEM –User manual
Pages 32
Revision list
Issue
Date
Description of changes
Origin
06/05/2020
1 (a)
21/06/2021
General revision based on a new board called “HW V3”
2
21/09/2022
Added the Declaration of conformity. Made minor changes
Notes on this manual
The information included in this manual may be modified without notice. No part of this manual may be
reproduced in any form or by any electronic or mechanical means, for any use, without LSI LASTEM’s written
permission.
LSI LASTEM reserves the right to intervene on the product, without the obligation to promptly update this
document.
Copyright 2019-2022 LSI LASTEM. All rights reserved.

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 4 / 32
Table of contents
1 Introduction............................................................................................................................................... 6
2 Instrument description.............................................................................................................................. 6
2.1 Power supply ..................................................................................................................................... 7
2.2 Inputs and actuators.......................................................................................................................... 7
2.2.1 Differential mode inputs............................................................................................................ 8
2.2.2 Single-ended inputs................................................................................................................... 9
2.3 Serial ports....................................................................................................................................... 10
3 Product installation ................................................................................................................................. 13
3.1 General safety standards................................................................................................................. 13
3.2 Mechanical installation and placing ................................................................................................ 13
4 3DOM software ....................................................................................................................................... 14
4.1 Connection to PC............................................................................................................................. 14
4.2 Entering the instrument in 3DOM and receiving the configuration ............................................... 14
5 Configuration........................................................................................................................................... 15
6 Use of ALIEM ........................................................................................................................................... 16
6.1 Start-up/Shutdown.......................................................................................................................... 16
6.2 Use of the keyboard ........................................................................................................................ 16
6.3 Operating status LEDs...................................................................................................................... 16
7 Insight ALIEM........................................................................................................................................... 17
7.1 Acquisition and calculation of measures......................................................................................... 17
7.2 Acquisition from thermocouples..................................................................................................... 17
7.3 Details about the measure acquisition process .............................................................................. 17
7.4 Acquisition from status signal ......................................................................................................... 18
7.5 Details about calculated measures.................................................................................................. 19
7.6 Actuation logics ............................................................................................................................... 19
7.6.1 Eolic alarm ............................................................................................................................... 20
7.6.2 Evaporimeter filling ................................................................................................................. 20
7.6.3 Start precipitation alarm ......................................................................................................... 21
7.6.4 Flood alarm.............................................................................................................................. 21
7.6.5 Threshold value compare........................................................................................................ 21
7.6.6 Timer........................................................................................................................................ 24
7.6.7 Snow level alarm...................................................................................................................... 24
7.6.8 System errors........................................................................................................................... 24

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 5 / 32
7.7 Function library for derived calculations......................................................................................... 25
8 Maintenance............................................................................................................................................ 26
9 Configuration templates.......................................................................................................................... 27
9.1 ALIEM configuration for Type 1: Alpha-Log + ALIEM....................................................................... 27
10 Technical specifications........................................................................................................................... 28
11 Declaration of conformity ....................................................................................................................... 31
12 Disposal.................................................................................................................................................... 32
13 Contacting LSI LASTEM ............................................................................................................................ 32

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 6 / 32
1Introduction
ALIEM (Alpha-Log Input Extension Module) is the input extension module of Alpha-Log.
The module acquires the punctual values of the measurements relating to the sensors connected to it and
makes them available to Alpha-Log via the Modbus RTU® protocol.
2Instrument description
ALIEM is presented as in Fig. 1.
Fig. 1 –Instrument description.
On the ALIEM frontal side are located the 4-buttons keyboard (5), reserved for the use of LSI LASTEM staff,
the LEDs (4) that indicate the operating status of the instrument, and the terminal board (6) with removable
screw connectors for the connection of sensors, power supply and external devices. For details, refer to Fig.
2.
On the left side panel there is the power switch (1) of the instrument, the RS-232 Com1 serial port (2) for PC
connection and the Com2 serial port (3), RS-232 or RS-485 depending on the model, for connection to Alpha-
Log.
(1) - On/Off switch
(2) - Com1 serial port (PC)
(3) - Com2 serial port (Alpha-Log)
(4) - Operating status LEDs
(5) - Keyboard
(6) - Terminal board

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 7 / 32
2.1 Power supply
Please refer to the table below for the inbound power terminal connections to the instrument and outbound
to the sensors or devices that need to be powered.
Line
Connection
Terminal
Inbound
0 Vdc battery
64
+ 8÷30 Vdc battery
65
GND
66
Outbound
+ Vdc fixed to power sensors / external devices
31
0 Vdc
32
+ Vdc actuated to power sensors / external devices
33
Whenever available, connect the ground wire (GND) to terminal 66. If the round wire (GND) isn’t available,
you make sure to connect terminals 60 and 61 (short circuit). This measure will improve the immunity from
electromagnetic noises and the protection from induced and conduced electrical discharges.
WARNING! In case the 31 and 32 terminals feed outside equipments, they must be equipped with power fail
circuit against short circuits or absorbed currents above 1 A.
2.2 Inputs and actuators
The instrument is fitted with 7 actuators used to power the sensors connected to the terminal board, (4
actuators for 8 analogue inputs in differential mode and 16 in single-ended mode, 2 actuators for 4 digital
inputs, 1 actuator for other functions); the actuators can also be used by the actuation programmable logics,
than can produce alarms according to the values acquired by the sensors. The voltage available on these
terminals depends on the kind of power supply received by the instrument.
Use the program 3DOM (see SWUM_00339 available on www.lsi-lastem.com website) to configure the op-
eration to switch on the sensors by means of the switching powers. In order to choose the actuation time is
advisable to consider both the energetic saving and the time the sensor needs to initialize.
The association between input and switching power is fixed, as shown in the table below. The number of the
terminal is indicated in italics; for example, that inputs 1 and 2 both make use of the first actuator; therefore,
it cannot be used for the other inputs. In case of sensors that generate two signals (like the thermo-hygro-
metric sensor), it’s suitable to select both inputs that use the same actuator.

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 8 / 32
Analogue
input
Terminal board
GND
Actuator
A
B
C
D
Number
+V
0 V
An1
1
2
3
4
7
Act1
5
6
An2
8
9
10
11
An3
12
13
14
15
18
Act2
16
17
An4
19
20
21
22
An5
34
35
36
37
40
Act3
38
39
An6
41
42
43
44
An7
45
46
47
48
51
Act4
49
50
An8
52
53
54
55
Digital
input
Terminal board
GND
Actuator
E
F
G
Number
+V
0V
Dig9
23
24
25
28
Act5
26
27
Dig10
56
57
58
Dig11
-
29
30
61
Act6
59
60
Dig12
-
62
63
28
Act7
33
32
Fig. 2 –Terminal board.
2.2.1 Differential mode inputs
The differential mode is the one commonly used for the connection of sensors, in which a sensor occupies a
single input.
It is preferable to the single-ended mode as it is less subject to electrical noise.
The following pictures explain in detail the connections of all types of sensors, both analogue in differential
mode and digital.

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 9 / 32
Fig. 3 –Connection diagram for sensors with analog signal (differential mode).
The drop resistance, indicated by Rx, is used to return a voltage signal from the current generated by the
sensor. Program 3DOM supplies a library for the setting of the LSI LASTEM sensors, including some powered
outputs; for such models the settings have been arranged to use the energized scale -300÷1200 mV, thus
being able to use 50 Ω drop resistances.
Fig. 4 –Connection diagram for sensors with digital signal.
2.2.2 Single-ended inputs
Single-ended is a connection mode that allows you to double the number of analog inputs of ALIEM. In this
mode two sensors that have voltageor current signals can be connected to the same input. It is not applicable
to sensors with resistive signals.

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 10 / 32
Single-ended mode is more subject to electrical noise, therefore the differential mode, in which a sensor
occupies a single input, is preferable when you want to obtain the best measurement performance from the
instrument.
Fig. 5 –Connection diagram for sensors with analog signal (single-ended mode).
The drop resistance, indicated by Rx, is used to return a voltage signal from the current generated by the
sensor. Program 3DOM supplies a library for the setting of the LSI LASTEM sensors, including some powered
outputs; for such models the settings have been arranged to use the energized scale -300÷1200 mV, thus
being able to use 50 Ω drop resistances.
When a probe is added using the 3DOM probes library, those sensors are always added in differential mode;
for this reason is not possible add a single-ended probe to a configuration where there is only one single-
ended sensor free; when this situation occurs set the measure parameters manually matching the single-
ended probe needs; otherwise change, if possible, different measures from differential to single-ended mode
until obtaining a complete differential input free (two single-ended sequential inputs).
2.3 Serial ports
MDMMB1110 is fitted with two RS-232 serial ports through 9-pole female standard connectors (DB9F).
Serial port 1 (2) (located on the left-hand side panel, at the bottom near the terminals) is used to program
the instrument's operative modes and data download, by means of the LSI CISS communication protocol.

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 11 / 32
Serial port 2 (3), in addition to the proprietary protocol, supports the TTY, LSI Probe and Modbus RTU® pro-
tocols.
The factory configuration sets the RS-232 serial ports as follows:
Serial port
1
2
Protocol
Native CISS
Modbus RTU
Speed
9600 bps
115200 bps
Data bit
8
8
Stop bit
1
1
Parity
None
None
Network address
1
1
Flow control
RTS only
RTS only
The instrument can be programmed; however only the Speed, the Network address and the Flow control can
be changed; all other parameters cannot. The speed can be programmed from 1200 to 115200 bps.
Electrically speaking, both RS-232 ports are configured as a DCE device. The following table shows the mean-
ing of each serial connectors' pin:
Signal
Pin
TD
2
RD
3
GND
5
RTS
8
CTS
7
The MDMMB1110.1 model has RS-485 serial line 2 instead of RS-232. The DB9 adapter/terminal block, shown
in the figure, is supplied for connecting the RS-485 device.
It must be connected as per the following table:
Signal
Pin
D+ / Data+
2
D-/Data-
3
GND
5

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 12 / 32
The factory configuration of the RS-485 serial port is as follows:
Serial port
2
Protocol
Modbus RTU
Speed
115200 bps
Data bit
8
Stop bit
1
Parity
None
Network address
1
Flow control
RTS only
The configurable parameters are the same as the RS-232 port, namely the Speed, the Network Address and
the Flow Control.
For more information on changing the parameters, refer to the user manual SWUM_00339.

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 13 / 32
3Product installation
3.1 General safety standards
Read the following safety standards to avoid personal injuries and to prevent damages to this product or to
the devices connected to it. Use this product strictly in the indicated way to avoid damages.
Only the support staff is authorized to perform the setup and managing procedures.
Install the instrument in a clean, dry and safe place. Humidity, dust and extreme temperatures tend to ruin
or damage the instrument. In these particular environments, it’s recommended to install the instrument in
appropriate protective boxes.
Power the instrument properly. Observe the power voltage indicated for the instrument model owned.
Connect the instrument properly. Follow meticulously the wiring diagram provided with the equipment.
Do not use the product if a malfunction presence is suspected. If the existence of a malfunction is suspected,
do not power the instrument and ask for assistance to the qualified support staff.
Before any operation on electrical connections, power, sensors and communication devices:
-turn off the power
-discharge the accumulated electrostatic charges touching a conducting material or a grounded de-
vice
Do not start up the product if water or condensing humidity is present.
Do not start up the product in an explosive atmosphere.
Lithium ion battery inside. Replacing the battery with an incorrect type can cause an explosion hazard.
3.2 Mechanical installation and placing
ALIEM is usually used outdoor inside the appropriate protective boxes. However, indoor use is possible by
placing the instrument on shelves or attaching it to a wall.
For its functioning it requires the dedicated power supply or a photovoltaic module with the proper battery.
For the mechanical installation see the documentation provided with the equipment.

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 14 / 32
43DOM software
3DOM is the LSI LASTEM’s configuration software for the E/R/M/S-Log, Pluvi-ONE, Alpha-Log and ALIEM in-
struments. In Alpha-Log, it allows the user to:
•Create and modify the configurations.
•Send and receive the configurations through different kinds of connection.
•Import and export the configurations between different instruments.
•Check the constantly refreshed measurement values in the instrument.
3DOM can be downloaded from www.lsi-lastem.com website. To install the software, follow the instructions
displayed on the Setup program.
See the online help guide for more information on 3DOM’s usage.
4.1 Connection to PC
To connect ALIEM to PC, proceed as follows:
1. Insert the ELA105 serial cable to the ALIEM serial
port Com1 (2) and to that of the computer. If the
latter does not have an RS-232 serial port, use the
USB RS-232 adapter supplied. In this case, allow the
installation of the device driver.
2. Connect the power supply to the terminal (64-, 65+)
of the terminal block (6).
3. Turn on ALIEM via the On/Off switch (1).
4.2 Entering the instrument in 3DOM and receiving the configuration
On 3DOM’s first start, after the installation, you need to register the instrument you are using. Proceed as
follows:
1. Start 3DOM.
2. Select New… from the Instrument menu
3. Select E-Log –R/M-Log, S-Log, ALIEM… and press [Con-
tinue] and then [Next].
4. In the Communication parameters Window set:
•Communication type: Serial.
•Serial port: the port number of the PC to which it is
connected (e.g. COM1).
•Bit rate (bps): 9600.
5. Press [Save],then [Next] to connect to the instrument.
Continue by pressing [Next] and then [Finish].
In Microsoft Windows, to check the PC serial port
where the USB adapter is connected, select System
in the Control Panel and select Hardware setup. Ex-
pand the Ports list (COM and LPT). Identify the port
number called USB Serial Port. In the case more than
one USB port is listed, disconnect and reconnect the
adapter and
check with
port will disap-
pear and re-
appear. Set
the same
COM number
in the 3DOM

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 15 / 32
6. To download the configuration choose Yes,
then [Continue]. At the end, press [Close], as-
sign a name to the configuration (e.g. “Factory”)
and press [Ok].
3DOM will upload Instruments Browser and Confi-
gurations with the serial number of the instrument
and the newly downloaded configuration.
5Configuration
ALIEM comes with a factory standard configuration: set to acquire the internal temperature measurement
(the related sensor is built-in) and the battery level.
If the factory configuration does not suit your usage wish, use the 3DOM software to customize ALIEM. This
operation requires the instrument to be connected to a PC with 3DOM software installed (see §4).
For more information about the configuration of ALIEM, refer to the 3DOM manual.

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 16 / 32
6Use of ALIEM
6.1 Start-up/Shutdown
Start-up and shutdown are controlled by the On/Off switch (1).
6.2 Use of the keyboard
The keyboard consists in four buttons: two of them are directional and the other two are functional. Its use
is reserved for LSI LASTEM personnel.
6.3 Operating status LEDs
On ALIEM’s front panel are some LEDs that indicates the instrument operative status: Err/Ok e Tx/Rx.
Err/Ok
These LEDs determine the operating status of the instrument. In particular:
-Err: it is red; if on it indicates the presence of an error in the instrument
-Ok: it is green; it has the following meanings:
ooff: instrument switched off, or low feeding voltage (min. 7 V)
osingle quick blinkings: instrument switched on; regular operation (acquisition mode)
othree quick blinkings: instrument switched on; it operates in reconfiguration mode
ofive quick blinkings: the instrument is on; it operates in quick acquisition mode (reserved for
the use of LSI LASTEM staff)
Tx/Rx
These LEDs light up to indicate that communication is in progress on any of the serial ports. In particular:
-Blinking Tx: it is red; it indicates the instrument being transmitted
-Blinking Rx: it is green; it indicates the instrument being received
-Tx and Rx off: no communication activity in progress
Interval between blinking cycles: 5 s.
Quick blinking period: 125 ms
Slow blinking period: 375 ms.

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 17 / 32
7Insight ALIEM
7.1 Acquisition and calculation of measures
It is possible to acquire measures of sensors with a rate from 1 second to 12 hours; this allows a better
representation of both fast-changing quantities (wind speed), as well as slow-changing ones (air tempera-
ture). It is possible to acquire as many as 10 measures per second.
For sensors which have to be powered, ALIEM is fitted with own outputs with switched power supply (see
“actuators”§2.2).
The instrument is able to calculate quantities deriving from measures that sample the signs from the inputs:
ALIEM is fitted with a calculation library dedicated to environmental applications, also able to carry out useful
mathematical functions (see §0). ALIEM is able to acquire and calculate up to a maximum of 99 total
measures. Furthermore, it’s possible to program some calculated measures according to the data generated
by other calculated measures.
Program 3DOM can be used to program the measure sampling sequence: the program interface allows
choosing the sequence in which the measures are displayed; hence which is the sampling sequence taking
place during the survey. With 3DOM you can also set automatically the measure sequence according to their
acquisition rate (from the shortest to the slowest). This is important to gather measures which, during the
acquisition process, must be sampled within the shortest intervals possible among each other (if they are
allocated to only one actuator).
If configured with the same acquisition rate, the analogue channels following the first one will be sampled
about 80 ms after the previous channel. This means that as many as eight analogue channels are sampled
within a total time of 700 ms. Digital channels are instead sampled almost instantly.
7.2 Acquisition from thermocouples
ALIEM is able to acquire the signals from many types of thermocouples. The instrument uses the internal
temperature value as reference of value of the cold junction.In these cases it is necessary to program, by
means of the 3DOM program, the measure of the internal temperature. The measure of the internal tem-
perature must precede, in the measure sequence, all measures of quantities which use such reference.
7.3 Details about the measure acquisition process
The sampling of the signals produced by the sensors connected to the instrument's terminal board takes
place according to the following logical procedure:
1) Measurement of the electrical signal based on its type (voltage, resistance, frequency, etc.) and its
digital conversion into a 16-bit numeric value; the physical type of the sensor is programmed through
parameter Electrical measure type;
2) Data validation: during this operation the value is limited within the scale values allowed by the phys-
ical type of measure;
3) Possible thermocouple value correction through the measure of the cold junction temperature (in-
ternal temperature of the instrument);

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 18 / 32
4) Linearization of non-linear signals based on the setting of parameter Linearization type; the lineari-
zation may also take place through the setting of a polynominal function whose factors are specified
until 10° degree (section Linear parameter of 3DOM);
5) Recalculation of the value according to numeric parameters defined in section Parameters:
•Computation of the measured quantity through the defined initial and final scale values;
•Application of the calibration factor of the specific used sensor (radiometers, rain gauges, etc.);
•Selection of the logical state with reference to the analogue signal thresholds;
•Measure validation after processing accompanied by error indication if greater (by 0.5%) than
the limits set in output; wind direction and relative humidity are excluded;
•Linearized quantity control: the instrument produces null output when receiving null input.
All above parameters are indicated in section Measure properties in the measure modification window of
program 3DOM.
7.4 Acquisition from status signal
ALIEM is able to acquire different types of digital status; they have to be configured for their connection to
9, 10, 11 and 12 inputs.
Ther are 3 different signal types: frequency signals, digital status and counters.
The default configuration of acquirer has been made for:
give logical status = 1
In case of short circuit or 0 V
give logical status = 0
In case of opened contact or 3 V
Furthermore, if one configuration for low power consumption is selected:
•for signals with frequencies over 1000 Hz it’s better the input 9;
•for signals with frequencies under 1000 Hz, for counters and logical status they’re better the inputs
11 and 12;
•do not configure the input 10 because it doesn’t reduce the instrument’s low power consumption.
If it’s been selected one configuration with no attention to power consumption:
•for signals with frequencies over 1000 Hz use the inputs 9 and 10;
•for signals with frequencies under 1000 Hz and for logical status use any inputs among the available
ones.
The sensors with status output that produce voltage (e.g. they aren’t pure contacts “open/closed”) but with
variable voltage according to measured status, can be connected to ALIEM through one diode; in this way
every connection is always the right one, apart from the output voltage (no divider is required). The anode
of diode must be placed on clamp F of terminal board’s entrance and the cathode towards the sensor.

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 19 / 32
LSI LASTEM recommended default configuration:
•input 9 for wind speed (frequency signal);
•input 10 for rain gauge (counter);
•input 11 for logical status.
7.5 Details about calculated measures
If the instrument has been programmed to process one or more calculated measures, the logical process will
be the following:
1) Acquisition of all primary measures that allow the estimate of calculated measures; a calculated
measure can be a primary measure for new calculated measure;
2) Collection of the value of primary measures; if at least one of these values is found to be in error,
its calculated measure will also be indicated in error;
3) Collection of the value of standard parameters, whenever used in the calculation; the value of these
parameters is decided during the configuration process and cannot therefore be modified during
the survey;
4) Execution of the estimate;
5) Allocation of the calculated value to the measure's instant datum.
The acquisition rate of a specific calculated measure is setted by 3DOM so that it may correspond to the
acquisition rate smaller than the calculated measures it depends on.
7.6 Actuation logics
ALIEM has got an actuation logics library. It’s useful to switch-on all type devices (alarms, solenoid valves,
motors) according to the surveyed parameters in surrounding environment. The actuation logics are based
on the instantaneous value of the acquired and calculated measures. They can be programmed up to 20
calculation algorithms, that use same or different logics. One or more algorithms can be combined in two
differed modes, in order to switch-on the selected actuator:
1) All algorithms must be in alarm at the same time (AND logic);
2) One algorithm can be in alarm at least (OR logic).
The actuator’s switch-on logic can operate according to low power consumption mode (the actuator is usually
deactivated, and starts in case of alarm), or according to safety mode (the actuator is usually activated, and
it shuts-down in case of alarm). Summarizing explanation in the table below.
Operating logic
State
Actuator
Low consumption mode
No alarm
Actuation output switched-off
Alarm
Actuation output switched-on
Safety mode
No alarm
Actuation output switched-on
Alarm
Actuation output switched-off

ALIEM –User manual
LSI LASTEM SRL
INSTUM_03766
Page 20 / 32
In case of one or several measures’ error (for example owing to sensor breaking, no-scale acquisition, or
disconnected cable) it doesn’t modify the present state of the actuator piloted by the logic that uses the
same measures.
The actuation logic deactivates the actuator only if it has previously activated from the logic itself.
The actuation logic begins operation after the first activation of the actuator, even if this occurs temporally
after the deactivation.
The programming of the actuation logics happen by means of the 3DOM program according to two different
phases:
1) Selection of the logics and their calculation parameters (section Logics);
2) Selection of the actuation outputs and their AND or OR modes correlation with the preset logics
(section Actuators); note that the same logic can be combined with other different logics several
times, in order to switch over different effectors.
7.6.1 Eolic alarm
The logic uses a wind direction measure to establish the condition of a wind position in a specified sector for
a predefined time. Can be set:
•The measure that samples wind direction (degrees);
•The starting angle (extreme included) of the direction sector;
•The ending angle (extreme included) of the direction sector;
•The continuous permanence time of the wind direction inside the defined sector in order to detect
the alarm condition;
•The continuous permanence time of the wind direction outside the defined sector in order to detect
the end of the alarm condition.
Both times can be set from 0 seconds to 12 hours; if both times are set to zero, the wind direction in or out
conditions inside the sector are immediately detected.
It is possible to join this logic with an another logic of threshold bypass type (see §7.6.5) applied to a wind
speed measure: in this way it’s possible to refine further the alarm activation (e.g. to activate the alarm when
the wind is more than 5 m/s for at least 3 minutes and inside the Est sector of 45 degree for at least 1 minute).
7.6.2 Evaporimeter filling
The logic uses a water level probe inside the evaporimeter to establish the need for filling up it.
It can be set:
•The measure that samples the water level;
•The filling-up start time (it’s recommended the automatic filling-up programmed in the morning,
before the sunrise, in order to avoid temperature changing that could alter the evaporation
measure);
This manual suits for next models
5
Table of contents
Popular Control Unit manuals by other brands

HMS
HMS CM CANopen user manual

Wieland
Wieland gesis FLEX-0/2W DC F installation instructions

Idex
Idex WARREN RUPP SANDPIPER Signature Series Service & operating manual

Sun Microsystems
Sun Microsystems Sun Storage J4200 Replacement guide

Allen-Bradley
Allen-Bradley POINT I/O PROFIBUS 1734-APB user manual

Quectel
Quectel RM510Q-GL Hardware design

TECH
TECH ST-65 user manual

Microchip Technology
Microchip Technology MM7150 user guide

Agilent Technologies
Agilent Technologies 81618A user guide

Broan
Broan VAUTOW Installation and user guide

Viconics
Viconics VT8000 Series installation guide

TechnipFMC
TechnipFMC Smith Meter microLoad.net installation manual