Mangar International Leglifter Installation guide

Mangar International
Presteigne
Powys
Wales, UK
LD8 2UF
tel: +44 (0)1544 267674
fax: +44 (0)1544 260287
www.mangarinternational.co.uk
Patented
freedom through lightweight solutions
MI0003 issue 5

user instructions and warranty
leglifter
freedom
through lightweight solutions

leglifter
2
MI0003 issue 5
The Leglifter by Mangar International enables people to raise their legs onto the bed independently or
with minimal assistance. Power is provided by the Mangar International Airo compressor.
PLEASE READ the separate Airo compressor instructions before using your Leglifter. When properly
maintained, the Leglifter will give many years of trouble free service.
WARNINGS
Ensure that you have read and fully understand
these and the separate Airo compressor user
instructions before using the equipment.
Only use a Mangar International air supply to
inate the Leglifter.
Ensure that the Leglifter does not come into
contact with sharp objects.
Ensure that the Airo compressor, air hose and
hand control are positioned so as not to present
a tripping hazard.
•
•
•
•
A
C
B
Standard Contents
1 x Leglifter (A)
1 x Fabric Cover (B)
2 x Fitting Straps (C)
1 x User Instructions

leglifter
Standard tting instructions
Fit the looped end of each strap around the
legs, castors or bed frame on the opposite side
of the bed (A g. 1) and pass the straps under
the mattress.
Fit the hook of each strap onto the Leglifter
ap (B g.1) and position the Leglifter loosely
alongside the bed, clear of any obstacles such
as a bedside table.
Tighten the straps to pull the edge of the ap
(‘Vee’ of deated Leglifter) under the mattress
by approximately 1 to 2 ngers width (C g.2).
Connect the Leglifter air hose and hand control
(D g.3) to the Airo as indicated in the separate
Airo compressor user instructions.
Test that the Leglifter inates to fully vertical
against the mattress and further tighten the
straps if necessary.
NOTE: Upon nal adjustment, the edge of the
ap settles to sit on the frame/divan (under the
mattress) by about 1 to 2 ngers width. This will
vary according to bed type e.g. divan, sprung
divan, narrow framed bed.
Place the Airo compressor by the side or
under the bed in an out of the way position.
•
•
•
•
•
•
3
freedom through lightweight solutions
MI0003 issue 5
2
3
1
A
B
D
1 to 2 ngers
width
‘Vee’
C

leglifter
4
Fitting instructions (Proling Beds)
The exible construction of the deated Leglifter
makes it ideal for use with most types of proling
bed. However, please carefully follow these
guidelines to avoid any conict with moving
sections of the bed.
Remove the mattress from the bed.
Raise and lower the backrest and legrest
sections of the bed to identify strap positions
which best avoid chang/fouling of the straps.
On the opposite side of the bed to where the
Leglifter is to be tted, secure the loop end
of each strap to the inner frame as indicated
(A g.4). NOTE: Wrap the strap around the
frame several times so as to shorten the strap
length, allowing the adjuster buckle to lie near
the centre line of the bed.
WARNING: When tting to a bed with a Low/High
function, ensure that the straps are attached to the
frame which moves up and down.
•
•
•
MI0003 issue 5
Hook the two straps onto the ap of the Leglifter as indicated in (B g.4) and pull on the strap at the
adjuster buckle to draw the ‘Vee’ of the deated Leglifter inwards over the bed frame and under the
mattress by about 1 to 2 ngers width (C g.5).
Raise and lower the backrest and legrest sections of the bed, checking to ensure that the straps do
not interfere with the bed’s functions.
Replace the mattress.
Connect the Leglifter to the Airo compressor and with the backrest and legrest sections of the bed
in the horizontal position, test that the Leglifter inates to fully vertical against the mattress. Tighten
the straps further if necessary. Upon nal adjustment, the edge of the ap settles to sit on the frame
(under the mattress) by about 1 to 2 ngers width.
NOTE: Normally, the Leglifter should be operated with all bed sections in the horizontal position. If
necessary, the backrest can be raised up to 45° to assist upper body stability while the legs are raised.
WARNINGS:
Do not operate the Leglifter when the bed’s legrest is raised.
Do not raise the bed’s legrest when the Leglifter is inated.
Always deate the Leglifter after the legs have been safely positioned on the bed.
NOTE: Where tted, any integral cotside on the transfer side of the bed must be removed to permit
effective tting and operation of the leglifter.
•
•
•
•
•
•
•
4
A
B
5
CNOTE: If the bed sections are of
metal grid construction, ensure the
strap hook has its open side facing
upwards, to avoid fouling the grid.

leglifter
5
freedom through lightweight solutions
MI0003 issue 5
Operating instructions
Sit on the bed at your normal entry point and
as far across the bed as possible. Angle your
legs towards the foot of the bed as far as is
comfortable (g. 6). If possible, use one arm
for support as illustrated, until you are ready to
lie down. Some clients prefer to use a bed lever
for this purpose.
To lift your legs onto the bed, press ‘up’ on the
hand control. Release the ‘up’ button to stop at
the half-way stage and/or any convenient point
and adjust your position to improve the angle
of your legs (g. 7).
Fully raise the Leglifter, allowing your legs to
slide along the cushion and gently drop onto
the bed (gs. 8 & 9).
Deate the Leglifter by holding the ‘down’
button for a few seconds; release the button
and the Leglifter will carry on deating.
•
•
•
•
6
7
8
9

leglifter
To deate the Leglifter without the use of a hand
control, insert the Vent Plug (A g. 10). Air will
continue to be released until the Leglifter is
fully deated.
•
6
MI0003 issue 5
10
Cleaning
routine cleaning
Remove the cover. Fully inate the Leglifter, sponge clean with a detergent solution and dry thoroughly.
Do not dry using hot air.
The cover can be machine washed up to 70°C tumble dried and hot ironed. The cover is manufactured
from a material which has been PROBAN™ treated for re retardancy. To retain the re retardancy
properties, do not use soap or soap powder and do not use hypochlorite bleach.
NOTE: The following detergents/washing powders may be used to launder the cover:- Ariel, Daz, Tide,
Fairy Non Bio, Surf automatic, Dreft, Surf, Persil System Automatic. Some brands now state ‘suitable for
FR fabrics’.
The following washing powders, which are soap based must not be used:- Acdo, Lux soap akes. If a
washing product is soap-based, it will be stated on the pack.
disinfection
Remove the cover. Fully inate the Leglifter and pre-clean by washing with detergent and water and
then disinfect using either buffered hypochlorite (such as ‘Milton’ solution) or a hard surface, alcohol
impregnated wipe. Do not dry using hot air.
The cover can be thermal disinfected at 70°C for 3 minutes, followed by tumble drying.
Storage
Ensure that the Leglifter is thoroughly dry before storing. Store deated. Insert the Vent Plug (g. 10) in
the air hose tting to assist in releasing any trapped air.
Maintenance
Periodically check the air hose for signs of damage or wear.
Periodically check the tting straps for signs of damage or wear.
Regularly launder the cover and allow to thoroughly dry before retting.
We recommend that your Leglifter is inspected annually by either your supplier or other suitably qualied
person. If required, Mangar International can provide full maintenance within an annual service contract.
Please contact our Customer Care department for further details (See page 7 for further information).
Repairs or servicing must only be carried out by your supplier or other authorised person.
•
•
•
•
•
•
•
A

leglifter
7
freedom through lightweight solutions
MI0003 issue 5
57cm
124cm
90cm (min)
236cm (max)
10cm
Specications
(Inated) 124cm x 57cm x 10cm
Max. Pressure: 0,35 bar (5psi)
Max. Temperature: 50°C
Warranty
Mangar International guarantees this product for a period of two years.
The guarantee is given against defects in material and workmanship and comprises
piece, parts and labour. The guarantee is not valid in the case of fair wear and
tear, wilful damage, common neglect, misuse or unauthorised alteration or repair by
unqualied persons.
The following statements do not affect the consumer’s statutory rights. This product is sold on the
understanding that in the event of any defect of manufacture or material appearing within two years of the
date of receipt of this product, the defect will be rectied free of charge providing that:
(a) Reasonable evidence is provided (e.g. purchase invoice, delivery note) that the product was not
purchased more than two years prior to the date of the claim.
(b) The defect is not attributable to accidental damage (either in transit or otherwise), misuse or
unauthorised repair.
Leglifter serial number: ................................
Date of purchase: ................................
Invoice/Delivery note number: ................................
Mangar International reserves the right to alter product specications and/or any of the information contained within this document without notice.
We strive to provide our customers with reliable and durable products of the highest quality.
In addition to a two year guarantee, and for added peace of mind, our servicing division,
Primeserve, also offers a variety of service options. Choose from an Extended Parts Warranty,
Call-Out Protection and an Annual Service and Clean.
To nd out how we can provide complete cover for your Mangar International product call our
Customer Care Department on Freephone 0800 280 0485.
Table of contents