Masibust VT7S12E User manual

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 1 -
USER MANUAL
VT7S12E
VIBRATION TRANSMITTER
Masibus Automation & Instrumentation Pvt. Ltd.
B/30, GIDC Electronics Estate,
Sector-25, Gandhinagar-382044, Gujarat, India
+91 79 23287275-79 +91 79 23287281-82
Email: support@masibus.com
Web: www.masibus.com

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 2 -
CONTENTS
CONTENTS ............................................................................................................................................................. 2
LIST OF TABLE ........................................................................................................................................................ 3
LIST OF FIGURE ...................................................................................................................................................... 3
1. INTRODUCTION.................................................................................................................................................. 4
2. INSTALLATION.................................................................................................................................................... 5
3. HARDWARE SPECIFICATION DETAIL ................................................................................................................... 6
3.1 Input specification............................................................................................................................6
3.2 Power supply specification ..............................................................................................................6
3.3 Display specification ........................................................................................................................6
3.4 Mechanical specification..................................................................................................................6
3.5 Environmental specification.............................................................................................................7
3.6 Output relay specification ................................................................................................................7
3.7 Analog output specification..............................................................................................................7
3.8 Rs-485 communication specification (optional)...............................................................................7
3.9 Buffer output specification ...............................................................................................................7
3.10 Signal isolations and insulation specification.................................................................................7
4. FRONT PANEL DESCRIPTION............................................................................................................................... 8
5. KEY FUNCTION DESCRIPTION ............................................................................................................................. 8
6. CONNETION DIAGRAM ...................................................................................................................................... 9
7. MENU LAYOUT................................................................................................................................................. 10
7.1 Flow Diagram.................................................................................................................................10
7.2 Menu Parameters .........................................................................................................................12
7.2.1 Relay Configuration.................................................................................................................13
8. RELAY OUTPUTS............................................................................................................................................... 17
9. CALIBRATION PROCEDURE ............................................................................................................................... 21
10. MODBUS COMMUNICATION DETAIL .............................................................................................................. 22
11. MISCELLANEOUS ............................................................................................................................................ 27
12. TROUBLESHOOTING ....................................................................................................................................... 29

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 3 -
LIST OF TABLE
Table 1 Product Ordering Code.......................................................................................................4
Table 2 Input Type Selection Table .................................................................................................6
Table 3 Signal Isolation Specification ..............................................................................................7
Table 4 Front Panel Description ......................................................................................................8
Table 5 Level –1 Parameter description .......................................................................................12
Table 6 Level –2 Parameter Description......................................................................................13
Table 7 Relay Group –1 Configuration Detail...............................................................................13
Table 8 Relay Group –2 Configuration Detail...............................................................................14
Table 9 Relay Group –4 Configuration Detail...............................................................................14
Table 10 Relay as Control Output Configuration Detail.................................................................14
Table 11 Level –3 Parameter Description....................................................................................15
Table 12 AO (Analog Output)(Retransmission Output) description...............................................16
Table 13 Calibration Mode Parameter Description........................................................................16
Table 14 Factory Reset Mode Parameter Description...................................................................16
Table 15 Modbus Function code description.................................................................................22
Table 16 Exception codes..............................................................................................................22
Table 17 Modbus Parameter Details for Process Value................................................................22
Table 18 Details of abnormal Conditions of Process Value...........................................................22
Table 19 Modbus Parameter Details for Input Register.................................................................24
Table 20 Modbus Parameter Details for Calibration Register .......................................................24
Table 21 Modbus Parameter Details for Read Output Status Register.........................................25
Table 22 Input Type Selection Table .............................................................................................25
Table 23 Relay Logic Selection .....................................................................................................25
Table 24 Relay Operation Selection ..............................................................................................25
Table 25 Relay Function Selection ................................................................................................25
Table 26 Relay Open Sensor Selection.........................................................................................25
Table 27 Relay Group Selection....................................................................................................26
Table 28 Relay Group-1 Selection.................................................................................................26
Table 29 Relay Group-2 Selection.................................................................................................26
Table 30 Relay Group-4 Selection.................................................................................................26
Table 31 Relay Type For Group-2 ................................................................................................26
Table 32 Relay Type For Group-4 & 1...........................................................................................26
Table 33 Relay Latch Selection .....................................................................................................26
Table 34 Baud-Rate For Communication Selection ......................................................................26
Table 35 Parity/Stop bit Selection..................................................................................................26
Table 36 Retransmission Open Sensor Selection.........................................................................26
Table 37 Retransmissoin Type Selection ......................................................................................26
Table 38 Retransmission Direction Selection................................................................................26
Table 39 Retransmission Value Selection.....................................................................................26
Table 40 PV Status during burn out Condition...............................................................................27
Table 41 Retransmission Output during Open/Over Condition .....................................................28
LIST OF FIGURE
Figure 1 Front Panel Description ..................................................................................................................8
Figure 2 Connection Diagram of VT7S12E...................................................................................................9
Figure 3 Level-1 Flow Diagram...................................................................................................................10
Figure 4 Level-1, Level-2, Calibration & Factory Reset Mode Flow Diagram.............................................11

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 4 -
1. INTRODUCTION
Foreword
Thank you for purchasing VT7S12E - Vibration Transmitter. This manual describes the basic functions and operation methods of
VT7S12E.Please read through this user‟s manual carefully before using the product.
This is a 8 bit controller based instrument designed for vibration type. This instrument is operated by four user-friendly keys. These
keys are used for operation and programming. The manual covers all aspects of operation of the instrument. Please read
instructions carefully before altering any programming or configuration information.
VT7S12E module operate independently and also can be connected to a data Highway or remote systems communication functions
through a personal computer or a distributed control system (DCS) using RS-485 Communication.
Notice
The contents of this manual are subject to change without notice as a result of continuing improvements to the instrument‟s
performance and functions.
Every effort has been made to ensure accuracy in the preparation of this manual. Should any errors or omissions come to your
attention, however, please inform Masibus Sales office or sales representative. Under no circumstances may the contents of this
manual, in part or in whole, be transcribed or copied without our permission.
Trademarks
Our product names or brand names mentioned in this manual are the trademarks or registered trademarks of Masibus Automation
and Instrumentation (P) Ltd. (herein after referred to as MASIBUS).
Adobe, Acrobat, and Postscript are either registered trademarks or trademarks of Adobe Systems Incorporated. All other product
names mentioned in this user's manual are trademarks or registered trademarks of their respective companies.
Version Number : 1.07 , August 2015
Checking the Contents of the Package
Unpack the box and check the contents before using the product. If the product is different from that which you have ordered, if any
parts or accessories are missing, or if the product appears to be damaged, contact your sales representative.
Product Ordering Code:
Ordering Code
Model
Channel-1
Channel-2
Power Supply
Output Type
Communication o/p
VT7S12E
X
X
X
X
X
1
Accelerometer i/p
N
None
A
85 to 265 VAC
C
4-20 mA
N
None
1
Accelerometer i/p
B
18 to 36 VDC
D
0-20 mA
1
RS485
2
Proximity i/p
E
1-5V DC
F
0-5V DC
G
0-10V DC
Table 1 Product Ordering Code

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 5 -
2. INSTALLATION
Mounting method: DIN Rail mounting
To install the controller select a location where:
no one may accidentally touch the terminals
mechanical vibrations are minimal
corrosive gas is minimal
temperature can be maintained at about 25˚C to 35˚C and the fluctuation is minimal
no direct radiant heat is present
no magnetic disturbances are caused
no wind blows against the terminal board
no water splashed
no flammable materials are around
Turn off the power to the controller before installing it on the Din Rail because there is a possibility of electric shock
How to connect wires:
Before carrying out wiring, turn off the power to the controller and check that the cables to be connected are not
alive with a tester or the like because there is a possibility of electric shock.
NOTE:
oAll wiring must confirm to appropriate standards of good practice and local codes and regulations. Wiring must be suitable
for Voltage, Current and temperature rating of the system.
oProvide power from a single-phase instrument power supply. If there is a lot of noise in the power line, insert an insulating
transformer into the primary side of the line and use a line filter on the secondary side. Do not place the primary and
secondary power cables close to each other.
oUse repeater after each set of 32 instruments connected in RS-485 Communication.
oUnused terminals should not be used as jumper points as they may be internally connected, which may cause damage to
the unit.
CAUTION:
High voltage transients may occur when switching inductive loads such as some contactors or solenoid valves. Through the internal
contacts, these transients may introduce disturbances which could affect the performance of the instrument.
For this type of load it is highly recommended that a “snubber” is connected across the normally open contact of the relay switching
though load. The snubber recommended consists of a series connected resistor/capacitor (typically 15nF/100 Ohms). A snubber
will also prolong the life of the relay contacts. A snubber should also be connected across the output of a trip output to prevent false
triggering under line transient conditions.

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 6 -
3. HARDWARE SPECIFICATION DETAIL
3.1 Input specification
No. of Input (1)
: 1 , 2 & 4 (Ref. Table 1)
Accelerometer Input
: Remote ICP piezoelectric Accelerometer
Sensitivity: 100 mV/g ±10%
Dynamic range: 80 g pk
RTD Input
: Range: -199.9 to 850.0C
Resolution: 0.1C
Accuracy: +0.25% of full range + 1 digit
Current Source: 1mA
Wiring Resistance: for RTD Max.15 Ohm/ Wire
RPM Input
: Input Type : Proximity sensor NPN/PNP, Encoder, Magnetic Sensor
Input Frequency: 0 –166.666 Hertz Max
Range: 0-9999
Resolution: 1 RPM
Accuracy:±0.015% of Full range
Pulse per rotation : 1 to 60 (user set)
Max pulse input: RPM x Slots(pulse/rotation) <= 600,000
Input signal level : 0-24V DC, min on pulse width 100
uSec.
Input High : >1 Volt
Input Low : < 0.2 Volt
Liner Input
: Range: -1999 to 9999
Resolution: 1 Count
Accuracy: +0.1% of Full range + 1 digit
Sensor Excitation Current
: 4 mA, Voltage: 21 VDC
-3 dB Cut-off frequency
High Pass Filter : 2.5Hz, 5 Hz, 10Hz
Low Pass Filter :1KHz, 2.5KHz, 10KHz
Filter Accuracy
: ± 10% of Cut-off Frequency
Measuring parameter
: Acceleration , Velocity , Displacement
Measuring Range
: 1) Acceleration 0-50 g, RMS/peak(derived)
2) Velocity 0-100 mm/s, RMS/peak (derived)
3) Displacement 0-2000 micron, peak to peak (derived)
Accuracy Input-display
: ± 2% of full span (Input to Display)
Channel No.
Input Types
CH-1
Vibration
CH-2
Vibration, RTD, RPM, Not Used
CH-3(2)
Temperature, Linear, Not Used
CH-4(2)
Temperature, Linear, Not Used
Table 2Input Type Selection Table
(1) Factory Set
(2) When Temperature and liner Input Type is Selected Serial Communication will not Available. Temperature is
measured by an Accelerometer with Temperature output ,10mV/°C, Range +2° to +120°C.
3.2 Power supply specification
Rated Voltage
: 85 - 265VAC 50/60 Hz / 100-300VDC Or 18-36VDC
Power Consumption
: Max 12VA (85-265VAC) and Max.7VA (18-36VDC)
3.3 Display specification
Display
: 1- 0.3" Green seven segment For Channel Display
4-0.3" Red seven segment For Data Display
Keys
: 4 Keys (ENT, UP, DOWN & ESC)
Scan Time
: <=150 mSecond
3.4 Mechanical specification
Mounting
:DIN Rail
Dimensions
: 70 mm(W) x 75 mm(H) x 110 mm(D)
Material
: ABS Plastic
Weight
: 350 gm
Terminal Cable Size
: 2.5mm2
Mounting
: IP20

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 7 -
3.5 Environmental specification
Ambient Temperature
: 0 to 55 °C
Humidity
: 30 to 95% RH non-condensing
Storage Temperature
: 0-85°C
Warm-Up Time of Instrument
: 15 Min.
3.6 Output relay specification
No of Relays
: Four
Purpose
: Alarm or trip
Output Signal
: 1 Form C Configuration
Delay for relay
: 05 - 50 sec to avoid false tripping.
Relay contact Rating
: 2A @250VAC/30VDC & 5A @250VAC (optional)
3.7 Analog output specification
No. of Analog Output
: Two
Output Signal
: 0-20 mA, 4-20 mA or 0-5 V, 1-5 V, 0-10 V DC (Voltage or
current outputs are factory settable)
Load Resistance
: For current output, 500 ohms Max. For Voltage output,
3000 ohms Minimum.
Output Accuracy
: ±0.25 % of Full Range(Display to output)
3.8 Rs-485 communication specification (optional)
No. of Communication Port
: 1- RS485
Protocol
: Modbus RTU.
Parity
: 2-Stop bit Parity none, 1-Stop bit Parity Even and 1-Stop
Bit Parity Odd.
Baud Rate
: 9600,19.2K
Maximum No. of Unit
: 32
Communication Error Detection
: CRC Check
3.9 Buffer output specification
No. of output
: Two
Output Impedance
: <100 Ohms
Frequency Range
: 0.5 Hz to 10KHz
Accuracy
: 0.25% of Full Range
3.10 Signal isolations and insulation specification
Isolation Rating
: Withstanding Voltage:
1) Between primary terminals(3) and secondary terminals(4)
1500VAC for 1 minute
2) Between secondary terminals 500V AC for 1 minute
Insulation Register
: > 20 Mohms at 500V DC
Signal Isolation
: As Specified below Table:2
Sr
No
Signals
Signal Isolation
1
Power Input
Isolated from other input/output terminals and internal circuit
2
Analog Inputs
Not isolated from other analog input terminals and from the internal
circuit. But isolated from other input/output terminals.
3
RS-485 Communication
Isolated from other input/output terminals and internal circuit
4
Relay contacts
Isolated between contact output terminals and from other
Input/output terminals and internal circuit
5
Analog Output
Isolated from other input/output terminals and internal circuit
Table 3 Signal Isolation Specification
(3) Primary terminals indicate power terminals and relay output terminals
(4) Secondary terminals indicate analog input signals, Digital Contact output terminals, communication Terminal.

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 8 -
4. FRONT PANEL DESCRIPTION
Figure 1 Front Panel Description
Name of Part
Function
Process Value
Display(DATA window)
Displays Process Value.
Display Parameter Name When You Set Parameter.
Displays Error Message When An Error Occurs.
Channel No. Display
(CHANNEL window)
Displays Channel Number in run mode. Also it will display relay number (01 –04)
in set mode (i.e.R.1,R.2,R.3,R.4)
It will also display Group no(GP 1,GP 2/GP 1,GP 2,GP 3,GP 4) in set mode
Relay Status LED (RL1, RL2, RL3,
& RL4) OR
Open Sensor Indicator LED
When Respective Relay LED Lits (In Red).
OR
When Channel is OPEN (Channel no. is corresponding to Relay no.)
Auto/Manual status LED (A/M)
If LED is on, it indicates Manual mode and if LED is off Auto Mode.
Communication Status LEDs (R,T)
When Communication on, two LEDs (In Red) blink.
Input Type Status LEDs (A, V, D, T)
Depending upon the channel Input type LED is on.
Table 4 Front Panel Description
5. KEY FUNCTION DESCRIPTION
MENU/ENTER KEY:
It is used to enter in the sub menu (various levels) and save the parameters to nonvolatile memory, when user setting a proper data
by Increment and shift key for parameter configuration.
ESCAPE KEY / AUTO-MANUAL KEY:
It is used to come out from any sub menu (various levels) to the run mode. It is used for come out of the manual mode to auto mode
too.
INCREMENT KEY:
It is used to increment the parameter for selection. Value of parameter can be incremented by pressing this key. When first time
increment key is pressed, DP (decimal point) in PV display blink, so user can modify the value with increment key. It is used to
increment the value in particular digit. Value can be incremented from 0- 9 and from „9‟ again it rollovers to „0‟. The function
Auto/Manual is integrated with this key. This key is used to enter into manual mode while the unit is in run mode.
SHIFT KEY/DECREMENT KEY:
It is used to Shift the digit to set the parameter as describe in increment key when DP (decimal point) started to blink. Menu key is
used to go forward to show next parameter and Shift key is used to go backward to show previous parameter. It is also used to
acknowledge the trip relay.

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 9 -
6. CONNETION DIAGRAM
Figure 2 Connection Diagram of VT7S12E

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 10 -
7. MENU LAYOUT
7.1 Flow Diagram
Vibration Transmitter - VT7S12E has a number of software parameters which may or may not be required depending on your
particular applications.
Figure 3 Level-1 Flow Diagram

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 11 -
LVL2
pwd
0000
0000
inPt
Pv-H
Pv-L
dP
rLFn
rLdL
rLoS
rLLG
rLmP
+
rLtP
LVL3
pwd
0000
0000
SkiP
rLLH
rLGP
SCAn
Srno
bAud
PrSt
tout
rtos
rttP
rtdr
rtmp
CAL
pwd
0000
0000
CALZ
CALS
rtrZ
rtrS
Procedure for changing/setting
different values to different channels
in all levels is shown in the level 1
Hysteresis setting. “see Hys settings”
rLoP
FLTR
rtrd
SPwd
+
slot
FrSt
pwd
0000
0000
LdEF
CLAL
CLAL
PArA
ALL
Figure 4 Level-1, Level-2, Calibration & Factory Reset Mode Flow Diagram

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 12 -
7.2 Menu Parameters
Run Mode:
Following parameters can view or change during run time.
Immediately after powering, unit will run in Auto Mode. In auto mode channel will scan automatically according to scan
time selection (1-250 second).
By pressing INCREMENT Key in run mode, Channel number scanning on display is stopped. By pressing increment key
again, we can change channel number manually.
By pressing SHIFT/DECREMENT Key in run mode, relay acknowledgement can be done.
By pressing ESCAPE Key after going into the Manual Mode unit comes back to run mode.
Other Modes:
For entering in Menu press Enter key in RUN mode. There are total Five modes in menu. By pressing increment or decrement
key mode can be changed. To enter into any mode press enter key again.
1. Level –1
2. Level –2
3. Level –3
4. Calibration Mode
5. Factory Reset Parameter
The table below show the description and menu detail of the all the modes. In general to enter into any menu or to edit sub
menu, press enter key, to change the submenu parameter press increment or decrement key. To come out from any menu
press esc key.
1. Level –1 :
Pressing MENU key DATA window shows LVL1 (LvL1) message. Press MENU key again DATA window shows pwd(PWD)
message, press increment key twice to select password and then press MENU key to enter into Level-1. DATA window shows
SP.1 (SP.1) message and by pressing increment key, DATA window shows Set Point-1 Value. Use INC and SHIFT key to
modify value. OR press MENU key again to change Set-point 1 for Channel 2. ESCAPE KEY is used to come out SP.1
LEVEL 1
Parameter
(DATA window)
Setting name and description
Default value
Shows only
if
Symbol
Name
Pwd
(PWD)
Level-1 Password
0 to 9999
0000
-
Sp.1
(SP.1)
Target Set point-1
SetPoint-1 for All Channel
50.0
-
SP.2
(SP.2)
Target Set point-2
SetPoint-2 for All Channel
50.0
Relay group
2 is
selected
SP.3
(SP.3)
Target Set point-3
SetPoint-3 for All Channel
50.0
Relay group
1 is
selected
SP.4
(SP.4)
Target Set point-4
SetPoint-4 for All Channel
50.0
Relay group
1 is
selected
Hys
(HYS)
Hysteresis
Hysteresis for All Channel
1.0
-
Table 5 Level –1 Parameter description
2. LEVEL –2 :
Pressing MENU key DATA window shows LvL2 (LvL2) message. Press MENU key again DATA window shows pwd (PWD)
message, press increment key twice to select password and then press MENU key to enter into Level-2. Following parameters
can be configured in LEVEL –2.
LEVEL 2:
Parameter
(DATA Window)
Setting name and description
Default value
Shows only
if
Symbol
Name
pwd
(PWD)
Level-2
Password
0 to 9999
0000
-

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 13 -
Inp.t
(inP.t)
PV Input Type
Follow Table 22(Input type for
All Channel)
Acceleration(for all
channel)
-
Pv-h
(PV-H)
Process value
range high setting
(PV high > PV
low)
Follow Table 22(Range High
for All Channel)
40.0
-
Pv-l
(PV-L)
Process value
range lower
setting
Follow Table 22(Range Low
for All Channel)
0.0
-
dp
(dP)
Decimal Point
Setting Only
applicable for
Linear input type
is selected
0 to 3(For All Channel)
0
-
FLTR
(FLtr)
Digital Filter
0 to 9
1
Rllg
(rLLG)
Relay
Logic(Applicable
for All Relay)
nl / fs
(Normal / Fail Safe)
0:Noraml
1:Fail Safe
Normal(for All Relay)
-
RloP
((rLOP)))
Relay Operation
Al / Co
(Alarm / Control Output)
0:Alarm
1: Control Output
Alarm
-
Rlfn
(rLFn)
Relay
Function(Applicab
le for All Relay)
AL / TR
(Alarm / Trip )
0:ALARM
1:TRIP
Alarm(for All Relay)
Rl.oP
is set to Al
Rldl
(rLdL)
Relay
Delay(Applicable
for All Relay)
1 to 99 seconds
1 second(for All
Relay)
-
Rlos
(rLoS)
Relay Open
sensor(Applicable
for All Relay)
up / Dn
(UP / Down)
0:DOWN
1:UP
Up Scale(for All
Relay)
-
Rlmp
(rlmp)
Relay mapping
(Applicable for All
Channel)
See Relay Configuration
Refer Note:3
Rl.oP
is set to Al
Rltp
(rltp)
Relay Group
Type
See Relay Configuration
Refer Note:2
-
Slot
No. of Slot For
RPM Input Type
1 to 60
1
Inpt
is set to rpm
Table 6 Level –2 Parameter Description
7.2.1 Relay Configuration
Relay configuration depends on selection of Relay group i.e. Relay group 1 or Relay group 2 or Relay group 4 in Level-3.
Relay Group - 1:
If Relay group –1 is selected, there will be only one group of relay. That group has four relays. (G-1). G-1 means RELAY 1,
RELAY 2, RELAY 3 and RELAY 4
Table 7 RelayGroup –1 Configuration Detail
Example:-
Note:- None means no group is selected for particular channel.
Relay Group - 2:
CHANNEL NO
NONE
G-1
Relay Group
Relay Type
1
G - 1
Low ON (L) or High ON (H)
2
3
Relay Type can be selected as shown below:
4

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 14 -
If relay group –2 is selected, there will be two group of relay. Each group has two relays. (G-1 and G-2). G-1 means relay 1,
relay 3 & G-2 means relay 2, relay 4
Example:
Table 8 RelayGroup –2 Configuration Detail
Note : - 1) All Groups cannot be selected for single Channel .
2) None means no group is selected for particular channel.
Relay Group - 4:
If relay group –4 is selected, there will be four group of all 4 relays. Each group has one relay. (G -1, G -2, G -3 and G -4). G -
1 means relay 1, G -2 means relay 2, G -3 means relay 3, G -4 means relay 4
Example:
Table 9 RelayGroup –4 Configuration Detail
Note: -1) More than one Group cannot be selected for single Channel .
2) None means no group is selected for particular Channel.
3) If rL.oP is set to Co then All relay will be assigned channel wise.
Relay as Control Output:
Whenever rL.oP is set to Co then All relay will be assigned channel wise and all assignments are remain fixed. And Relay
type
Example:
Table 10 Relayas Control Output Configuration Detail
For Relay functionality Refer Relay outputs (Chapter –8).
3. LEVEL - 3 :
Pressing MENU key DATA window shows LvL3 (LvL3) message. Press MENU key again DATA window shows pwd (PWD)
message, press increment key twice to select password and then press MENU key to enter into Level-3.Following parameters can
be configured in LEVEL –3.
LEVEL 3:
Parameter
(DATA Window)
Setting name and description
Default value
Shows only
if
Symbol
Name
pwd
(PWD)
Level-3
Password
0 to 9999
0000
-
CHANNEL NO
NONE
G-1
G-2
Relay Group
Relay Type
1
G –1
High/ Very High (H-VH) or
Very Low /Low (VL-L) or
Low/High (L-H)
2
G –2
High/ Very High (H-VH) or
Very Low /Low (VL-L) or
Low/High (L-H)
3
4
CHANNEL
NO
NONE
G -1
G -2
G -3
G -4
Relay Group
Relay Type
1
G - 1
Low ON (L) or High ON (H)
2
G - 2
Low ON (L) or High ON (H)
3
G - 3
Low ON (L) or High ON (H)
4
G –4
Low ON (L) or High ON (H)
CHANNEL
NO
NONE
G -1
G -2
G -3
G -4
Relay Group
Relay Type
1
G - 1
Low ON (L) or High ON (H)
2
G - 2
Low ON (L) or High ON (H)
3
G –3
Low ON (L) or High ON (H)
4
G –4
Low ON (L) or High ON (H)

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 15 -
Skip
(skip)
Channel skip/
Unskip selection.
yes / no
0:NO
1:YES
0(for all channel)
-
RllH
(rLLH)
Relay Latch
on / off
0:OFF
1:ON
0
Rl.oP
is set to Al
RlGp
(rLGP)
Relay Group
RGP.4 / rGP.2
0:Relay Group-4
1:Relay Group-2
2:Relay Group-1
1
Rl.oP
is set to Al
SCAn
(SCAn)
Scan Time
1 to 250 seconds
1
-
Srno
(Srno)
Unit ID
1 to 247
1
-
BAUd
(baud)
Communication
Baud rate
9 6 0 0 /19.2K
0:(9 6 0 0 ) –9600 bps
1:(1 9 . 2 K ) –19.2 Kbps
9600 bps
PrSt
(PrSt)
Parity/Stop bit
selection
PnS1/PnS2/PoS1/ PES1
0:(P n S 1 )-parity none-stop bit-1
1:(P n S 2 )-parity none-stop bit-2
2:(P o S 1 )-parity odd-stop bit-1
3:(P e S 1 )-parity even-stop bit-1
parity none-stop bit2
TouT
(tout)
Timeout for
display back to
Run Mode
10 to 100 Seconds
60
-
RToS
(rtos)
Retransmission
Open sensor
up / DoWn
0:DOWN
1:UP
1(For Both Output
Card)
-
Rttp
(rttp)
Retransmission
Output Type
0-20/4-20/0-5v/1-5v/
0-10v
0:(0 - 20) –0-20mA
1:(4 - 20) –4-20mA
2:(0-5) –0 –5volt
3:(1-5) –1 –5volt
4:(0 - 10) - 0 -10volt
1(For Both Output
Card)
-
Rtdr
(rtdr)
Retransmission
direction
Dir / rev
1:(dir)
0: (rev)
1(For Both Output
Card)
-
Rtmp
(rt mp)
Retransmission
mapping
See Table 8
Rtrd
(rtrd)
Retransmission
Channel Value
Max / Min
1:(Max)
0: (Min)
1(For Both Output
Card)
-
SPwd
(SPWD)
Password
Set password to
lock selected
level
0 to 9999
0
-
Table 11 Level –3 Parameter Description
Note:-
1) None means no AO(Analog Output)(Retransmission Output) is assigned to particular channel.
2) AO O/p number = 0 represents “None”. AO O/p Number = 1 and 2 represents “Retransmission No. 1 and Retransmission No.
2.
3) In “Retransmission mapping” parameter, maximum 2 numbers of Analog Outputs can be assigned to each channel. I.e. each
channel can have maximum two numbers of analog outputs simultaneously. See the example below.
Example 1: If Retransmission mapping for channel 1 = 0102 H, it means Retransmission output number 1 and output number
2 are assigned to channel 1. And output of channel 1 can be taken on analog output (retransmission output) number 1 and
number 2.
Example 2: If Retransmission mapping for channel 2 = 0001 H, it means Retransmission output number 1 is assigned to
channel 2. And output of channel 2 can be taken on analog output (retransmission output) number 1.

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 16 -
Retransmission Output number
(Decimal)
Retransmission Output
number (Hex)
Setting Name and Description
0
0 x 00
none
(None)
1
0 x 01
Rt.01
(Retransmission Output-1)
2
0 x 02
Rt.02
(Retransmission Output-2)
Table 12 AO (Analog Output)(Retransmission Output) description
Note: -By default For Channel-1 Retransmission output-1 and For Channel-2 Retransmission output- 2 is set.
4. Calibration mode :
Pressing MENU key, DATA window shows CAL (CAL) message. Press MENU key again, DATA window shows pwd (PWD)
message, press increment key twice to select password and then press MENU key to enter into Calibration.
Calibration:
Parameter
(DATA Window)
Setting name and description
Default
value
Shows only if
Symbol
Name
pwd
(PWD)
Password
0 to 9999
0000
-
CAL.Z
(CAL.Z)
All Input Type Zero
Calibration
Depending on PV sensor type
selected
-
-
CAL.S
(CAL.S)
All input type Span
Calibration
Depending on PV sensor type
selected
-
-
Rtr.Z
(rtr.Z)
Retransmission
voltage and current
Zero calibration
Depending on Retransmission
type selected
-
-
Rtr.S
(rtr.S)
Retransmission
voltage and current
Span calibration
Depending on Retransmission
type selected
-
-
Table 13 Calibration Mode Parameter Description
5. Factory Reset Parameters :
Pressing MENU key, DATA window shows FRST (FrST) message. Press MENU key again, DATA window shows pwd (PWD)
message, press Increment key twice to select password and then press MENU key to enter into Factory Reset.
Factory Reset Mode:
Parameter
(DATA window)
Setting name and description
Default
value
Shows only if
Symbol
Name
Pwd
(Pwd)
Password
0 to 9999
-
-
Ldef
(LdEF)
LOAD Default
C A L \P A R A \all
(CAL)\(PARA)\(ALL)
C A L -Only calibration set to
default value
P A R A -All parameters excluding
calibration will set to default
value
A L L -Calibration and parameters
will set to default value
-
-
Table 14 FactoryReset Mode Parameter Description
Note: - Factory reset will load default parameters, as mention in MENU LAYOT (Default value). Once this function applies, user has
to switch off the instrument and again switch on the instrument to work according to Default value.

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 17 -
8. RELAY OUTPUTS
Following function can be set for Relay outputs.
Relay Logic (Direction):
Relay Logic means Relay contact can be changed from Open to Close OR Close to Open. If relay logic is selected Normal,
when Fault occur Relay contact will change from Close to Open. If relay logic is selected Fail Safe, when Fault occur Relay
contact will change from Open to Close.
Relay Function: Relay function can be selected as ALARM or TRIP.
If relay is selected as ALARM, when abnormal condition occur Relay will ON, once normal condition after abnormal condition
occur Relay will OFF.
If relay is selected as TRIP, when abnormal condition occur Relay will ON, once normal condition after abnormal condition
occur Relay will ON. Relay will be off through Acknowledge.
Relay Delay: A time delay can be provided for the actual output.
Relay Open Sensor: Open sensor upscale or down scale can be selected for each relay output.
Relay Mapping: Refer Menu layout LEVEL - 2
Relay Types: Various relay operations are shown in the reference figure. (High, Low, Very High- High, Low-Very Low, High-
Low). For relay types selection Refer Menu layout LEVEL –2.
Control Output (ON/OFF Control ) :-
ON/OFF Controller is the simplest form of temperature control device. The output from the device is either on or off, with no
middle state. An on-off controller will switch the output only when the temperature crosses the set point. For heating control,
the output is on when the temperature is below the set point, and off above set point. Since the temperature crosses the set
point to change the output stage, the process temperature will be cycling continually, going from below set point to above, and
back below. In cases where this cycling occurs rapidly, and to prevent contactors and valves from getting damaged, an on-off
differential, or “hysteresis,” is added to the controller operations. This hysteresis assures, if temperature exceed set point by a
certain amount before then only output will turn off or on again. On-Off hysteresis prevents the output from “chattering” or
making fast, continual switches if the cycling above and below the set point occurs very rapidly.
Relay type High (H-ON):
For High type of set value, once process value reaches up to set point + Hysteresis value,
relay will be ON after few seconds (as per relay delay) and it will be ON until process value goes down to Set point.
Relay type Low (L-ON):
For Low type of set value, once process value reaches down to set point - Hysteresis value relay will be ON after nearly
few seconds (as per relay delay) and it will be ON until process value goes up toward Set point.
NOTE:-
VT7S12E has both Control Logic (ON-OFF) & Alarm Logic. If Control Logic (ON-OFF) is required, rloP in lvl2 must
be selected as Co.
Whenever Control Output (ON-OFF) logic is selected, all relays are mapped as shown in chapter no.6. As well as, relay
latch is off and relay group is assigned as relay per group 1. In this logic, user cannot change relay mapping, relay
latch and relay group and user cannot acknowledge relays.

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 18 -
Alarm 1
Momemtary alarm
(when in abnormal condition ack not pressed)
CONDITION
NORMAL
ABNORMAL
UP (O/S)
DOWN (O/S)
ACK **
NORMAL *
ACK ***
High
Alarm
Latch
Yes
Lamp
Off
Steady
Steady
Off
Steady
Off
Relay
Off
On
On
Off
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Steady
Off
Off
Off
Relay
Off
On
On
Off
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Off
Relay
Off
On
Off****
Off****
On
Off
Low
Alarm
Latch
Yes
Lamp
Off
Steady
Off
Steady
Steady
Off
Relay
Off
On
Off
On
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Off
Steady
Off
Off
Relay
Off
On
Off
On
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Off
Relay
Off
On
Off****
Off****
On
Off
Vlow
Alarm
Latch
Yes
Lamp
Off
Steady
Off
Steady
Steady
Off
Relay
Off
On
Off
On
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Off
Steady
Off
Off
Relay
Off
On
Off
On
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Off
Relay
Off
On
Off****
Off****
On
Off
Alarm al2
Momemtary alarm
(when in abnormal condition ack not pressed)
CONDITION
NORMAL
ABNORMAL
UP (O/S)
DOWN (O/S)
ACK **
NORMAL *
ACK ***
Vhigh
Alarm
Latch
Yes
Lamp
Off
Steady
Steady
Off
Steady
Off
Relay
Off
On
On
Off
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Steady
Off
Off
Off
Relay
Off
On
On
Off
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Off
Relay
Off
On
Off****
Off****
On
Off
High
Alarm
Latch
Yes
Lamp
Off
Steady
Steady
Off
Steady
Off
Relay
Off
On
On
Off
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Steady
Off
Off
Off
Relay
Off
On
On
Off
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Off
Relay
Off
On
Off****
Off****
On
Off
Low
Alarm
Latch
Yes
Lamp
Off
Steady
Off
Steady
Steady
Off
Relay
Off
On
Off
On
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Off
Steady
Off
Off
Relay
Off
On
Off
On
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Off
Relay
Off
On
Off****
Off****
On
Off

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 19 -
Alarm al1
Maintained alarm
(when in abnormal condition ack is pressed)
CONDITION
NORMAL
ABNORMAL
UP (O/S)
DOWN
(O/S)
ACK **
NORMAL *
ACK ***
High
Alarm
Latch
Yes
Lamp
Off
Steady
Steady
Off
Steady
Steady
Off
Relay
Off
On
On
Off
Off
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Steady
Off
Steady
Off
Off
Relay
Off
On
On
Off
Off
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Steady
Off
Relay
Off
On
Off****
Off****
On
On
Off
Low
Alarm
Latch
Yes
Lamp
Off
Steady
Off
Steady
Steady
Steady
Off
Relay
Off
On
Off
On
Off
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Off
Steady
Steady
Off
Off
Relay
Off
On
Off
On
Off
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Steady
Off
Relay
Off
On
Off****
Off****
On
On
Off
Vlow
Alarm
Latch
Yes
Lamp
Off
Steady
Off
Steady
Steady
Steady
Off
Relay
Off
On
Off
On
Off
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Off
Steady
Steady
Off
Off
Relay
Off
On
Off
On
Off
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Steady
Off
Relay
Off
On
Off****
Off****
On
On
Off
Alarm al2
Maintained alarm
(when in abnormal condition ack is pressed)
CONDITION
NORMAL
ABNORMAL
UP (O/S)
DOWN
(O/S)
ACK **
NORMAL *
ACK ***
Vhigh
Alarm
Latch
Yes
Lamp
Off
Steady
Steady
Off
Steady
Steady
Off
Relay
Off
On
On
Off
Off
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Steady
Off
Steady
Off
Off
Relay
Off
On
On
Off
Off
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Steady
Off
Relay
Off
On
Off****
Off****
On
On
Off
High
Alarm
Latch
Yes
Lamp
Off
Steady
Steady
Off
Steady
Steady
Off
Relay
Off
On
On
Off
Off
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Steady
Off
Steady
Off
Off
Relay
Off
On
On
Off
Off
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Steady
Off
Relay
Off
On
Off****
Off****
On
On
Off
Low
Alarm
Latch
Yes
Lamp
Off
Steady
Off
Steady
Steady
Steady
Off
Relay
Off
On
Off
On
Off
Off
Off
Alarm
Latch
No
Lamp
Off
Steady
Off
Steady
Steady
Off
Off
Relay
Off
On
Off
On
Off
Off
Off
Trip
Lamp
Off
Steady
Off****
Off****
Steady
Steady
Off
Relay
Off
On
Off****
Off****
On
On
Off

VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 20 -
NOTE:
* means normal condition after abnormal has occurred
** means ack pressed in abnormal condition
*** means ack pressed in normal condition after abnormal has already occurred.
**** means it remains in the previous state. If previous state is ON then it will remain ON and the same case for OFF condition.
Pressing Shift/Decrement key for 3 seconds, acknowledgement will be given for alarm and trip relay in abnormal condition.
Alarm Latch function applicable only for ALARM, there is no affect when TRIP Selected as a relay function LEVEL –2.
Table of contents