Maytronics pH Drive User manual

pH Drive
Automatic pH
Controller
www.maytronics.com.au
Instruction Manual
Ensure Acid is diluted, as per
page 6 of these instructions.
pH Drive

Thank you for purchasing the Maytronics pH Drive.
The pH Drive has been designed for maximum reliability and long service life. Please be sure to
follow the instructions in this manual to get the best performance and life from your equipment.
The pH Drive is quite easy to operate, and this manual will explain each of the steps clearly.
Troubleshooting and hints are also included to allow you to get the most from your unit. If you
require assistance at any stage, please contact your Maytronics Authorised Dealer.
Page
CAUTIONS & WARNINGS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
GETTING TO KNOW YOUR pH DRIVE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
pH DRIVE DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
ACID SAFETY & DILUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
pH TEE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
pH DRIVE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
TUBE INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
pH SENSOR INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
COMPLETED INSTALLATION (ILLUSTRATION) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
CONNECT POWER LEAD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
pH DRIVE SETTINGS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Setting the Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Adding a Quick Dose of Chemical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Setting the pH Set Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Pool Size Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Enabling or Disabling the pH Alert Function . . . . . . . . . . . . . . . . . . . . . . 13
When the pH Alert is Triggered. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
pH SENSOR CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Inline Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
pH7.0 Buffer Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
CHANGING PUMP CASSETTE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
CONTACT US . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Contents
2

Cautions & Warnings
3
!Please read the instructions fully and keep
this manual on hand whenever operating
or maintaining your pH Drive.
!Always use diluted acid. Concentrated
acid will cause damage to the squeeze
tube and fittings, which are not covered
by the warranty.
See page 6 for information on the safe
handling and dilution of acid.
!DO NOT add acid into a drum that
previously contained liquid chlorine until
that drum has been rinsed three times
with fresh water. Ensure that the label on
the drum indicates that the drum contains
acid.
!As for all pool equipment and chemicals,
the pH Drive should be installed out of
reach of children. Children must not have
access to operating the pH Drive without
adult supervision.
!Although the pH Drive has a weather
resistant design, its service life will be
considerably longer if it is not exposed to
direct sunlight and rain. Wear and tear
from direct exposure to the elements is
not covered by the warranty.
!Do not allow grass or weeds to grow
around the pH Drive, or the tubing and
cables to prevent accidental damage from
garden trimming equipment.
!Any damaged cables must be replaced
immediately to prevent electrical shock.
!Unplug the pH Drive power lead before
servicing or changing the dosing pump
cassette.
!The pH Drive must be serviced only by an
authorised service agent. Please contact
your Maytronics Authorised Dealer for
det ails. Despi te being low-voltage
powered, opening the unit may still cause
an electric shock, which can cause injury or
death.

pH Drive
Flow
Getting to know your pH Drive
4
KEYPAD
Used for setup and maintenance of
your pH Drive.
2
DOSING PUMP
High quality, low maintenance
peristaltic dosing pump.
3
POWER CONNECTOR
IP65 Waterproof connector plugs
into May tronic s Ozone Swim
systems that are equipped with the
matching outlet.
4
MOUNTING CLIP
Pull the clip downwards when
mounting the pH Drive to the
mounting rail, or to unclip when
removing.
5
pH SENSOR SOCKET
Connect the pH Sensor to the BNC
pH Sensor socket.
6
pH SENSOR
pH TEE
NON-RETURN VALVE
Commercial quality pH Sensor
screws into pH Tee to firm hand
tightness.
A single pH Tee for installing the pH
Sensor and non-return injection
valve.
High quality polypropylene non-
return injection valve.
7
8
9
DISPLAY
User-friendly plain text menu and
help system.
1
5 6
4
1
2
3
7
8
9
pH Drive

pH Drive Display
5
POOL pH
SET POINT
INFORMATION LINE
This is the readout of the current pH of your pool.
The pH Drive maintains the pool pH at the programmed Set Point.
This is where pH Drive indicates helpful messages and warnings, so you know what’s happening
at all times.
pH is OK.........................................The current pH of the pool is within the Set Point. No dosing is
required at this time.
Adding Acid (or Base) ..............The pH Drive is operating the dosing pump to add acid or base to
correct the pH of the pool back to the Set Point. The default setup of
the pH Drive is to add acid.
pH Alert! Reset:Enter ..............The pH of the pool has been above the Set Point for 4 hours or more,
without returning to the Set Point. Press Enter to reset the alert timer.
24 Hr Delay..................................The user has selected the 24 Hour Delay function. This is normally
used when buffer or other chemicals have been added to the pool,
which can temporarily change pool pH. The 24 Hour Delay function
prevents excessive usage of acid or base.
Acid (or Base) Pump is Off ....The user has set the dosing pump to being totally off, regardless of
the pool pH reading.
Mixing Time Delay....................After the dosing pump has added a quantity of acid or base, the pH
Drive will switch it off for some time. This allows the added acid or
base to mix through the pool before adding more, if required.
Cool Squeeze Tube ..................The dosing pump has stopped for a short time to allow the squeeze
tube to cool down.
Startup Time Delay ..................The pH Drive has just been switched on, and is waiting for its pH
reading to become stable, before doing any automatic control.
pH is Out of Range....................The pH Drive has detected that the pool pH reading is extreme and
needs to be checked.
pH Calibration Due ...................12 months has passed since the last time that the pH Drive was
calibrated. Calibration ensures that pool pH readings are accurate.
Change Pump Cassette .........The dosing pump has operated for 50 hours or more, which is time for
the pump cassette to be replaced. This is important maintenance.
1
2
3
1
3
2

Acid Dilution Procedure
1. DO NOT add acid into a drum that
previously contained liquid chlorine, until
that drum has been thoroughly rinsed
three times with fresh water.
2. Fill a 20 or 25 Litre dangerous goods
certified drum to around half way with
fresh water.
3. Add 5 Litres of concentrated Hydrochloric,
Muriatic or Non-Fuming acid to the drum.
Pour slowly to prevent spillage.
4. Top up the drum with more fresh water. Do
not overfill.
5. Replace the drum lid and swish around to
mix the acid well with the water. Keep this
drum lid fitted until the diluted acid drum is
in position.
Additional Information
!Always use acid to at least 3 parts water to
1 part acid. Damage from undiluted or
insufficiently diluted acid is not covered by
the warranty.
!Always add acid to water. Never add water
to acid.
!The acid drum must be below the level of
the non-return valve in the pH Tee when
installed. This prevents acid syphoning into
the pipeline.
!Fit the Acid Drum label supplied with your
pH Drive to your drum of diluted acid. This
will identify its contents and shows the
dilution instructions.
Acid Safety & Dilution
6
Maytronics recommends using pre-diluted acid purchased from your
reputable pool shop or service agent.
Alternatively, this section provides information on the safe handling
and dilution of acid to use with your pH Drive. Please take the time to
read and understand this section, to prevent injury or illness due to acid
spillage.
Safety Information
!Dilute acid in a well ventilated area,
preferably outdoors.
!Have a supply of fresh running water (e.g.
a garden hose) readily to hand, in case of
spillage.
!Wear protective clothing, safety goggles
and gloves when diluting acid.
!Use only a Dangerous Goods certified
drum. Your pool store can advise on this.
!If an acid spillage does occur, wash down
well with copious amounts of fresh water
immediately. Seek Medical advice if any
acid burn has occurred or if breathing is
affected.
Fresh Water
Acid
Fresh Water
Fit Acid
Drum Label HAZCHEM 2R UN 1789
PKG GRP II

pH Tee Installation
7
Install the pH Tee first, so that the glue can set while you are installing
the pH Drive unit itself.
Pre-Installation Checklist
!You will require a saw suitable for cutting
PVC pipe, PVC Primer and PVC Glue.
!Install the pH Tee after the filter, before any
sanitation equipment, but after heating
equipment.
!The pH Tee must be fitted to a horizontal
section of pipe, so that the pH Sensor is
O
within 45 of vertical (with the glass bulb
pointing downwards).
!The pH Tee should be installed before an
Ozone venturi, Salt Chlorinator cell, Electro
Oxidizing Cell or Liquid Chlorine injection
point.
!For installations into 40mm Pipe, a pair of
50x40 mm reduction bushes have been
supplied. Glue these into the pH Tee before
installing the Tee.
!Check the Water Flow direction arrows on
the pH Tee label, and ensure that it is
installed in the correct direction.
GLUE IN pH TEE
!Use PVC primer on all surfaces to be glued.
!Use suitable PVC pressure glue for joints.
!Ensure Water Flow arrows are pointing in
the correct direction.
O
!pH Sensor must be within 45 of vertical.
CUT PIPE
The pH Tee must be fitted to the return to
pool pipe, where the water is clean and the
pressure is low.
For 50mm Pipe, cut out 6cm.
For 40mm Pipe, cut out 7cm.
This allows for bushes.
Flow
These examples are all correct:
Flow
Flow
These are incorrect:
Flow
Flow

8
pH Drive Installation
Pre-Installation Checklist
!The pH Drive unit MUST be mounted
vertically, with the cables at the bottom, in
order to be weather resistant. Any water
ingress due to the unit being incorrectly
mounted is not covered by warranty.
!Check the mounting bracket of the Ozone
Swim system with which your pH Drive is
being installed. If it has an extension for
mounting the pH Drive, then you do not
need to install the mounting rail supplied
with the pH Drive.
!Choose a position that will allow the cable
from the pH Sensor to reach the pH Drive
without straining, and power lead to reach
the outlet on the Ozone Swim system. The
dosing pump tubes must also be able to
reach the chemical drum and the pH Tee.
!Ensure that the pH Drive is protected from
direct sunlight and weather. Although the
unit has a weather resistant design,
damage from long term exposure is not
covered by the warranty.
!Do not connect the pH Drive power lead to
the Ozone Swim outlet until all installation
steps are complete.
INSTALL MOUNTING PLATE
Install the mounting plate levelly in the
desired location. Screws and wall plugs are
provided. This step is not required if your
Ozone Swim system has an extended
bracket for mounting the pH Drive.
MOUNT pH DRIVE
uHook the pH Drive mounting clip onto the top rail of the mounting plate.
vPull the mounting clip down pull pushing the pH Drive against the bottom rail of the mounting
plate.
wAllow the mounting clip to lock in place behind the bottom rail.
123

Flow Flow
DRUM WEIGHT TUBE INSTALLATION
1. Push the end of the drum weight tube
through the drum lid that was supplied with
the pH Drive.
2. Fit the tube onto the drum weight tube
fitting in the same way as described in the
pH Tee Tube Installation section above and
illustration below.
DOSING PUMP TUBE INSTALLATION
1. Cut Suction and Supply tubes to required
length from the single length of clear PVC
tube suppled with the pH Drive.
2. Fit compression nuts over the two tubes.
3. Push Suction and Supply tubes onto dosing
pump fittings.
4. Tighten the nut over the tube. Tighten firmly
with fingers.
pH TEE TUBE INSTALLATION
u Remove the nut from the non return valve
on the Tee piece. Slip the Nut over the tube.
v Push the tube firmly onto the tube fitting.
w Tighten the nut over the tube. Tighten firmly
with fingers. If the tubing can still be easily
pulled out, use a tool to tighten an extra 1/2
turn to ensure that it is securely fastened.
9
Tube Installation
2
34
Suction tube
on left side
Delivery tube
on right side
12 3
Flow

Flow
Flow
HAZCHEM 2R UN 1789
PKG GRP II
10
1. Carefully remove the protective cap from the
tip of the sensor.
2. Screw the pH Sensor into the pH Tee thread.
Tighten firmly by hand. DO NOT OVER-
TIGHTEN. DO NOT USE A TOOL.
3. After the pH Sensor is installed in the pH Tee,
connect plug on the pH Sensor cable into the
sensor socket on the pH Drive. Push the plug
into place and 1/4 turn clockwise to lock it into
place.
pH Sensor Installation
Suction tube
on left side Delivery tube on right side
Acid safety label fitted to drum.
Suction tube fed through 7.0mm
hole, drilled into acid drum lid.
Drum weight inside acid drum.
All installation steps are now complete. It is recommended that you use the Quick Dose function
with 1 Litre of fresh water to test the system for leaks before using your pH Drive with acid.
u Locate the pH Drive power socket on your
Maytronics Ozone Swim system. Remove the dust
cover. The keyway of the power socket is towards
the front of the unit. Align the pH Drive plug with
the keyway of the socket and push in.
v Screw the retaining nut of the plug clockwise until
finger tight to lock the plug into place to seal.
Connect Power Lead
Keyway
1
2
Delivery tube firmly
fixed into non return
valve in pH Tee.
Completed Installation
DO NOT fit suction
tube directly into
neat acid container.
See page 6 for Acid
Dilution information.
Hydrochloric Acid
320G/L
5 Litres
pH Drive
pH Drive

!Press Setup to enter the user-friendly menu system.
!Press ‡ and ˆ to scroll through menu choices, then press Enter to select the required option.
!Press Enter to save changes.
!Press Cancel to quit without saving changes at any screen.
!All setup screens time out if no key has been pressed for 30 seconds. Any changes will not be
saved.
Setting the Mode
!Press Setup to enter the user-friendly menu system,
then press ‡ and ˆ until Mode is selected. The current
mode is shown in brackets.
!Press Enter to go to the Mode menu.
!Press ‡ and ˆ to select between the Mode settings:
Auto Control . . The dosing pump will add acid or base
when the pH is outside the Set Point.
Standby . . . . . . The dosing pump is switched off and will
not add any acid or base automatically.
24Hr Delay. . . . The dosing pump is switched off for 24
hours, to allow buffer or other chemical
to mix and stabilise.
!Press Enter to save the selection, or press Cancel to quit
without saving.
Adding a Quick Dose of Chemical
!Use this function if you need to manually add a dose of
acid or base to the pool
!Press Setup to enter the user-friendly menu system,
then press ‡ and ˆ until Quick Dose is selected.
!Press Enter to go to the Quick Dose menu.
!Press ‡ and ˆ to select the number of litres of acid or
base to dose into the pool. The range of settings is 1 to 5
litres.
!Press Enter to start dosing, or press Cancel to quit.
!The pH Drive counts down the litres as it doses.
!Every 4 minutes of dosing pump operation, the pH Drive
will switch it off for 1 minute, to allow the squeeze tube
to cool down. This extends squeeze tube life.
pH Drive Settings
11

12
Setting the pH Set Point
The default pH Set Point of 7.6 is suitable for most pool
systems. You should only change this setting if advised by
a pool professional.
!Press Setup to enter the user-friendly menu system,
then press ‡ and ˆ until pH Set Point is selected. The
current Set Point is shown in brackets.
!Press Enter to go to the pH Set Point menu.
!Press ‡ and ˆ to set the required pH Set Point.
!Press Enter to save the new setting, or press Cancel to
quit without saving.
Pool Size Setting
The pH correction demand and pool turnover time is
different from pool to pool, even when two pools are the
same size. The Pool Size setting will be suitable for the
majority of pools. However, it may be necessary to adjust
this during the first few days or weeks of operation, as
detailed in this section.
!Press Setup to enter the user-friendly menu system,
then press ‡ and ˆ until Pool Size is selected. The
current Pool Size setting is shown in brackets.
!Press Enter to go to the Pool Size menu.
!Press ‡ and ˆ to set the required Pool Size.
!Press Enter to save the new setting, or press Cancel to
quit without saving.
NOTES
If the pH Drive does not keep up with acid or base demand,
and the Alarm sounds, increase the Pool Size setting by
one step and check in 2 days.
If the pH Drive overdoses too much acid or base, reduce
the Pool Size setting by one step and check in 2 days.
New concrete pools have an extremely high acid demand
for the first few months. It may be necessary to enter a
Pool Size up to 50% more than the true size of the pool if
your pH Drive is not keeping up with acid demand during
this initial period. The setting can be reduced again 1 to 3
months later.
pH Drive Settings

13
pH Drive Settings
Enabling or Disabling the pH Alert Function
The pH Drive will sound an alarm and display an alert if the
pH has been outside the set point for 4 hours or more.
If this function is disabled, the alarm will never be
triggered and the pH Drive may over dose acid or base.
Only advanced users should change this setting, and only
as a temporary measure.
!Press Setup to enter the user-friendly menu system,
then press ‡ and ˆ until pH Alerts is selected. The
current Set Point is shown in brackets.
!Press Enter to go to the pH Alerts menu.
!Press ‡ and ˆ to select between On and Off.
!Press Enter to save the selection, or press Cancel to quit
without saving.
When the pH Alert is Triggered
When the pH Alert is triggered, the pH Drive will display the
alert message and sound the beeper.
!The Dosing Pump will now be disabled, and will not
operate again until the Alert has been reset.
!Press Enter to reset the pH Alert.
!Refer to the list of possible causes and fixes on this
page, to rectify the condition that caused the alert to be
triggered.
Possible causes and solutions for
the pH Alert being triggered:
Cause ....Very high acid demand in
new concrete pool.
Fix ..........Increase Pool Size Setting
to a higher value until the
acid demand settles to
normal amounts.
Cause ....Acid or Base drum empty.
Fix ..........Refill drum with water and
a c i d o r b a s e a s p e r
instructions on page 6.
Cause ....Pool Size Setting is too low
or too high.
Fix ..........S e e p a g e 1 2 f o r
information on how to
change Pool Size Setting.
Cause ....pH Reading is inaccurate.
Fix ..........Re-calibrate pH Sensor.
Cause ....pH S e nso r b r oken o r
faulty.
Fix ..........If re-calibration is not
possible, then replace pH
Sensor.
Cause ....Pump is not pumping acid
correctly.
Fix ..........Check all tube fittings for
leakage. Re-fit or replace
as necessary.
Check squeeze tube inside
pump cassette for leakage
and replace if necessary.

pH Sensor Calibration
14
The pH Drive will display a reminder to
calibrate the pH Sensor approximately every
12 months:
The pH Sensor should be calibrated to ensure
that the readout is correct. Calibration is
simply the process of placing the pH Sensor
into a solution of known pH, and adjusting
the readout to that value.
The pH Drive can be calibrated in one of two
ways:
1. Inline Calibration
This is done by obtaining an accurate pH
measurement of the pool water and
adjusting the readout to that value.
2. pH7.0 Buffer Calibration
This involves removing the pH Sensor
from the pH Tee, placing it into a pH7.0
Buffer solution, and adjusting the readout
to 7.0.
Inline Calibration
1. Measure the pH of the Pool
Set the pH Drive to Standby or 24Hr Delay
mode an hour or more before measuring
the pH of the pool. Do not change back to
Au to Control mode until af ter pH
calibration is completed.
Ensure that the pool filtration pump has
been running and no acid has been added
for at least 1 hour prior to doing an Inline
Calibration. This is to ensure that the pH
level is stable throughout the whole pool.
Measure the pH of the pool with a good
quality, calibrated pH meter. Alternatively,
take a sample of pool water to a pool shop
which has an electronic pool balance
measurement system. A standard pool pH
test kit is NOT accurate enough to be
relied upon for pH calibration.
2. Calibration Procedure
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Calibrate pH Sensor is selected.
!Press Enter to go to the Calibration screen.
The current reading is displayed in
brackets. The flashing “X” indicates that
the reading is not yet stable. DO NOT
calibrate until the “X” changes to a “ü”
!Press ‡ and ˆ to set the “Cal:” pH value to
the pH of the pool, as measured in step 1.
!Press Enter to calibrate the pH Sensor, or
press Cancel to quit without calibrating.
!Set the pH Drive back to Auto Control
mode.
Calibration Failure
If pH calibration fails, the pH Drive will display
this message:
!C l e a n t h e p H S e n s o r a s p e r t h e
troubleshooting section, then retr y
calibration.
!If the calibration failure persists, the pH
Sensor will most likely need to be replaced.
pH7.0 Buffer Calibration is on the next page.

pH7.0 Buffer Calibration
The pH Drive is powered from your Ozone
Swim unit. The pH Drive will need power for
calibration, but with the pH Sensor removed
from the pH Tee, the pool pump must not be
running. You will therefore need to unplug
the pool pump from the Ozone Swim unit.
1. Prepare pH7.0 Calibration Buffer
! A sachet containing 2 ziplock
bags of salt was supplied with
your pH Drive. Simply pour
BOTH sachets into the 250mL
bottle and fill to the 250mL
mar k w i t h d eio nis e d o r
distilled water. Shake well
until all salts have dissolved.
!This solution will last up to 6
months after it has been
made, providing it is not
contaminated and is stored in
a cool, dark place
2. Remove the pH Sensor from the Tee
! Disconnec t the pH Sensor to avoid
straining the cable and plug when
removing the sensor from the pH Tee.
Unscrew the sensor completely from the
Tee.
! Reconnect the pH Sensor to the pH Drive.
! Rinse the sensor well in clean water. De-
ionized or distilled water is ideal, but tap
water will suffice.
3. Place pH Sensor into pH7.0 Buffer
!Decant a small amount of
pH7.0 buffer into a clean cup.
Place the pH Sensor into this
buffer, so it is immersed to
15mm or more, as per the
diagram on the right.
!DO NOT place the pH Sensor
directly into the 250mL
bottle of pH7.0 buffer.
4. Calibration Procedure
! Switch the pH Drive on by activating the
pool pump output of the Ozone Swim unit.
! Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Calibrate pH Sensor is selected.
!Press Enter to go to the Calibration screen.
The current reading is displayed in
brackets. The flashing “X” indicates that
the reading is not yet stable. DO NOT
calibrate until the “X” changes to a “ü”
!Press ‡ and ˆ to set the “Cal:” pH value to
exactly 7.0.
!Press Enter to calibrate the pH Sensor, or
press Cancel to quit without calibrating.
5. Re-fit pH Sensor and Power Connections
! Disconnect the pH Sensor from the pH
Drive to avoid straining the cable and plug
when screwing into the pH Tee. Re-fit into
the pH Tee. Re-connect the pH Sensor to
the pH Drive when done.
! Re-connect the pool pump to the Ozone
Swim power outlet.
! Switch your pool equipment back on.
Calibration Failure
If pH calibration fails, the pH Drive will display
this message:
!C l e a n t h e p H S e n s o r a s p e r t h e
troubleshooting section, then retr y
calibration.
!If the calibration failure persists, the pH
Sensor will most likely need to be replaced.
15
pH Sensor Calibration
pH Buffer Powder
250mL Solution Powder
O
pH 7.00 @ 25 C
50
100
150
200
250
pH7.0 Buffer Calibration
The pH Drive is powered from your Ozone
Swim unit. The pH Drive will need power for
calibration, but with the pH Sensor removed
from the pH Tee, the pool pump must not be
running. You will therefore need to unplug
the pool pump from the Ozone Swim unit.
1. Prepare pH7.0 Calibration Buffer
! A sachet containing 2 ziplock
bags of salt was supplied with
your pH Drive. Simply pour
BOTH sachets into the 250mL
bottle and fill to the 250mL
mar k w i t h d eio nis e d o r
distilled water. Shake well
until all salts have dissolved.
!This solution will last up to 6
months after it has been
made, providing it is not
contaminated and is stored in
a cool, dark place
2. Remove the pH Sensor from the Tee
! Disconnec t the pH Sensor to avoid
straining the cable and plug when
removing the sensor from the pH Tee.
Unscrew the sensor completely from the
Tee.
! Reconnect the pH Sensor to the pH Drive.
! Rinse the sensor well in clean water. De-
ionized or distilled water is ideal, but tap
water will suffice.
3. Place pH Sensor into pH7.0 Buffer
!Decant a small amount of
pH7.0 buffer into a clean cup.
Place the pH Sensor into this
buffer, so it is immersed to
15mm or more, as per the
diagram on the right.
!DO NOT place the pH Sensor
directly into the 250mL
bottle of pH7.0 buffer.
4. Calibration Procedure
! Switch the pH Drive on by activating the
pool pump output of the Ozone Swim unit.
! Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Calibrate pH Sensor is selected.
!Press Enter to go to the Calibration screen.
The current reading is displayed in
brackets. The flashing “X” indicates that
the reading is not yet stable. DO NOT
calibrate until the “X” changes to a “ü”
!Press ‡ and ˆ to set the “Cal:” pH value to
exactly 7.0.
!Press Enter to calibrate the pH Sensor, or
press Cancel to quit without calibrating.
5. Re-fit pH Sensor and Power Connections
! Disconnect the pH Sensor from the pH
Drive to avoid straining the cable and plug
when screwing into the pH Tee. Re-fit into
the pH Tee. Re-connect the pH Sensor to
the pH Drive when done.
! Re-connect the pool pump to the Ozone
Swim power outlet.
! Switch your pool equipment back on.
Calibration Failure
If pH calibration fails, the pH Drive will display
this message:
!C l e a n t h e p H S e n s o r a s p e r t h e
troubleshooting section, then retr y
calibration.
!If the calibration failure persists, the pH
Sensor will most likely need to be replaced.
pH7.0 Buffer Calibration
The pH Drive is powered from your Ozone
Swim unit. The pH Drive will need power for
calibration, but with the pH Sensor removed
from the pH Tee, the pool pump must not be
running. You will therefore need to unplug
the pool pump from the Ozone Swim unit.
1. Prepare pH7.0 Calibration Buffer
! A sachet containing 2 ziplock
bags of salt was supplied with
your pH Drive. Simply pour
BOTH sachets into the 250mL
bottle and fill to the 250mL
mar k w i t h d eio nis e d o r
distilled water. Shake well
until all salts have dissolved.
!This solution will last up to 6
months after it has been
made, providing it is not
contaminated and is stored in
a cool, dark place
2. Remove the pH Sensor from the Tee
! Disconnec t the pH Sensor to avoid
straining the cable and plug when
removing the sensor from the pH Tee.
Unscrew the sensor completely from the
Tee.
! Reconnect the pH Sensor to the pH Drive.
! Rinse the sensor well in clean water. De-
ionized or distilled water is ideal, but tap
water will suffice.
3. Place pH Sensor into pH7.0 Buffer
!Decant a small amount of
pH7.0 buffer into a clean cup.
Place the pH Sensor into this
buffer, so it is immersed to
15mm or more, as per the
diagram on the right.
!DO NOT place the pH Sensor
directly into the 250mL
bottle of pH7.0 buffer.
4. Calibration Procedure
! Switch the pH Drive on by activating the
pool pump output of the Ozone Swim unit.
! Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Calibrate pH Sensor is selected.
!Press Enter to go to the Calibration screen.
The current reading is displayed in
brackets. The flashing “X” indicates that
the reading is not yet stable. DO NOT
calibrate until the “X” changes to a “ü”
!Press ‡ and ˆ to set the “Cal:” pH value to
exactly 7.0.
!Press Enter to calibrate the pH Sensor, or
press Cancel to quit without calibrating.
5. Re-fit pH Sensor and Power Connections
! Disconnect the pH Sensor from the pH
Drive to avoid straining the cable and plug
when screwing into the pH Tee. Re-fit into
the pH Tee. Re-connect the pH Sensor to
the pH Drive when done.
! Re-connect the pool pump to the Ozone
Swim power outlet.
! Switch your pool equipment back on.
Calibration Failure
If pH calibration fails, the pH Drive will display
this message:
!C l e a n t h e p H S e n s o r a s p e r t h e
troubleshooting section, then retr y
calibration.
!If the calibration failure persists, the pH
Sensor will most likely need to be replaced.
pH7.0 Buffer Calibration
The pH Drive is powered from your Ozone
Swim unit. The pH Drive will need power for
calibration, but with the pH Sensor removed
from the pH Tee, the pool pump must not be
running. You will therefore need to unplug
the pool pump from the Ozone Swim unit.
1. Prepare pH7.0 Calibration Buffer
! A sachet containing 2 ziplock
bags of salt was supplied with
your pH Drive. Simply pour
BOTH sachets into the 250mL
bottle and fill to the 250mL
mar k w i t h d eio nis e d o r
distilled water. Shake well
until all salts have dissolved.
!This solution will last up to 6
months after it has been
made, providing it is not
contaminated and is stored in
a cool, dark place
2. Remove the pH Sensor from the Tee
! Disconnec t the pH Sensor to avoid
straining the cable and plug when
removing the sensor from the pH Tee.
Unscrew the sensor completely from the
Tee.
! Reconnect the pH Sensor to the pH Drive.
! Rinse the sensor well in clean water. De-
ionized or distilled water is ideal, but tap
water will suffice.
3. Place pH Sensor into pH7.0 Buffer
!Decant a small amount of
pH7.0 buffer into a clean cup.
Place the pH Sensor into this
buffer, so it is immersed to
15mm or more, as per the
diagram on the right.
!DO NOT place the pH Sensor
directly into the 250mL
bottle of pH7.0 buffer.
4. Calibration Procedure
! Switch the pH Drive on by activating the
pool pump output of the Ozone Swim unit.
! Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Calibrate pH Sensor is selected.
!Press Enter to go to the Calibration screen.
The current reading is displayed in
brackets. The flashing “X” indicates that
the reading is not yet stable. DO NOT
calibrate until the “X” changes to a “ü”
!Press ‡ and ˆ to set the “Cal:” pH value to
exactly 7.0.
!Press Enter to calibrate the pH Sensor, or
press Cancel to quit without calibrating.
5. Re-fit pH Sensor and Power Connections
! Disconnect the pH Sensor from the pH
Drive to avoid straining the cable and plug
when screwing into the pH Tee. Re-fit into
the pH Tee. Re-connect the pH Sensor to
the pH Drive when done.
! Re-connect the pool pump to the Ozone
Swim power outlet.
! Switch your pool equipment back on.
Calibration Failure
If pH calibration fails, the pH Drive will display
this message:
!C l e a n t h e p H S e n s o r a s p e r t h e
troubleshooting section, then retr y
calibration.
!If the calibration failure persists, the pH
Sensor will most likely need to be replaced.

!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
Changing Pump Cassette
16
pH Drive
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
Suction tube
on left side
(From Drum)
Supply tube
on right side
(To pH Tee)
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
The pH Drive monitors the running time of
the dosing pump and will display a reminder
to replace the pump cassette when this time
exceeds 50 hours.
!The pH Drive’s dosing pump squeeze tube
is contained in a complete replaceable
cassette, so replacing the squeeze tube is
a simple matter of replacing the cassette.
!Use the Quick Dose function to pump 1
Litre of fresh water through the system.
This will ensure that all acid or base has
been flushed out and will make it safe to
handle.
!Switch off the pH Drive and ensure that the
pool pump is also off.
!Removing the clear PVC
Suction and Supply tubes
is a reversal of the tube
fitting instructions on
Page 9.
!Turn the pump cassette anti-clockwise
around 1/8 of a turn. When you feel it stop,
pull the cassette off the base.
!Push the replacement dosing pump
cassette onto the pump base. Rotate it a
few degrees either way to make the axle
in the pump base align with the hole in the
back of the cassette. Once it has engaged,
push the cassette onto the base at the
angle shown at the top, right of this page.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.
!Turn the cassette clockwise around 1/8 of
a turn until it clicks into place.
!Refit the clear PVC Suction and Supply
tubes. Suction is on the left and Supply is
on the right . Refer to tube fit ting
instructions on page 9.
!Switch the pH Drive back on.
Resetting the Pump Cassette Reminder
!Press Setup to enter the user-friendly
menu system, then press ‡ and ˆ until
Maintenance is selected.
!Press Enter, then select Cassette from the
Maintenance menu. The total pump
operating hours is displayed on the right.
!At the Change Pump Cassette screen,
press Enter to reset the Pump Cassette
hours to zero. Press Cancel to exit if the
pump cassette was not replaced.

Symptom Possible Causes and Remedies
Unstable or inaccurate
pH readings
pH reading cannot be
calibrated
pH reads around pH7
without changing
pH reads around pH4
to 5 without changing
Unit does not power
up
Unit is not adding acid
when dosing pump is
operating
pH Sensor is dirty.
!Disconnect pH Sensor from the pH Drive, then unscrew sensor from
the pipe.
!Wipe the glass bulb with a soft tissue soaked with methylated spirits.
!Soak the tip of the sensor in diluted Hydrochloric acid (for example,
from your pH Drive diluted acid drum).
!Rinse the pH Sensor thoroughly in clean water.
!Screw sensor back into pipe and re-connect sensor to pH Drive.
pH Sensor is dirty.
!Follow the pH Sensor cleaning procedure above.
Short circuit in pH Sensor BNC Plug or Socket
!Disconnect pH Sensor plug from socket. Dry out with a hair dryer or
allow to air dry. Reconnect and re-check.
!If problem persists, return pH Sensor to have connector replaced, or
replace the pH Sensor.
Glass bulb or internal stem of pH Sensor is cracked.
!Replace pH Sensor.
Faulty Unit
!Faulty pH Drive main unit. Return for repair.
Alarm sounds due to
pH being too low
Acid Pump turns, but
acid is not flowing.
Pool Size Setting is set too high
!Reduce the Pool Size Setting by one level and check in 2 days.
Leaking or worn squeeze tube and/or rollers.
!Tighten Acid Pump fittings further or replace fittings.
!Replace dosing pump cassette as needed (see page 16).
!Check that the pH Drive diluted acid drum is not empty.
!Check that all suction and output tubes are fitted correctly, with no
leakage.
!Check that squeeze tube inside pump cassette is not leaking.
WARNING : IF ANY ACID HAS LEAKED OUT OF TUBES, WASH DOWN THE
AREA WITH FRESH WATER IMMEDIATELY. TAKE THE LID OFF THE PUMP,
REMOVE SQUEEZE TUBE AND WASH OUT WITH FRESH WATER IF
SQUEEZE TUBE WAS LEAKING.
pH Sensor is faulty.
!Replace pH Sensor.
pH Sensor needs to be re-calibrated.
!Clean pH Sensor as per procedure above.
!Re-calibrate pH Sensor as per details on pages 14 and 15.
Pool water is very pure and has low conductivity.
!Add one bag of pool salt per 50,000 Litres of pool water.
IF YOU NEED TO RETURN YOUR pH DRIVE FOR SERVICE OR REPAIRS AT ANY TIME, PLEASE USE THE QUICK DOSE
FUNCTION (PAGE 11) TO DOSE 1 LITRE OF FRESH WATER THROUGH THE SYSTEM BEFORE SENDING BACK. THIS
WILL STOP ACID LEAKING FROM THE PUMP DURING TRANSPORT (WHICH WOULD DAMAGE YOUR UNIT OR THE
PACKAGING, VOIDING WARRANTY), AND IS ALSO A SAFETY HAZARD FOR OUR TECHNICIANS.
Troubleshooting
17
?

Alert Messages
Reminder Messages
General Messages
Hardware Fail Messages
Possible Causes and Remedies
Action Required
Explanation of Message
Explanation of Message
pH has been outside of Set Point for more than 4 hours of operation.
!Please refer to page 13 for a complete list of possible causes and
remedies when the pH Alert message is flashing.
!The pH Drive will not attempt any further acid addition until the
alarm is cleared. Press Enter to clear the alarm after rectifying the
issue.
pH reading is above pH12 or below pH2.
!This alert is displayed immediately if the pH reading goes outside
the range of pH2 to pH12. A beeper also sounds. This alert can only
be cleared when the pH returns to normal range.
!Check the actual pH of the pool with an independent test.
!Attempt to calibrate the pH Drive (see pages 14 and 15).
!If calibration fails and the issue persists, the pH sensor may require
replacement.
!This reminder is displayed 12 months after the last successful pH
Calibration. Calibrate the pH Sensor as per pages 14 and 15.
!This reminder is displayed when the peristaltic dosing pump has
exceeded 50 hours of operation since the last time the peristaltic
pump cassette was changed.
!Change the Pump Cassette as per page 16.
!This message is displayed each time the pH Drive is switched on.
!If you need to contact us for technical support, we may ask you the
software version number (“v2.0” in this example).
!This message is displayed each time the pH Drive is switched on,
after the version screen above.
!On this screen, the pH Drive displays the accumulated dosing pump
operating time, and number of days since the last pH calibration.
!The pH Drive has not been able to obtain data from the internal
clock chip.
!Press Cancel to restart, or switch the pH Drive off for 5 seconds,
then back on again. If problem persists, return the unit for repair.
!The pH Drive has not been able to obtain data from the pH input.
!Press Cancel to restart or switch the pH Drive off for 5 seconds, then
back on again. If the problem persists, return the unit for repair.
18
Troubleshooting
?

v2.0b, 20-Sep-2021
Warranty
19
Contact Us
Maytronics Australia Pty Ltd
Unit 2 / 19 Rudd St
Oxley QLD 4075
AUSTRALIA
Tel .........................1300 693 657
Email.......................infoau@maytronics.com
techau@maytronics.com
Web . . . . . . . . . . . . . . . . . . . . . . . www.maytronics.com.au
ABN........................87 148 058 495
Maytronics Australia Pty Ltd (“Maytronics”) guarantees the pH Drive Controller (including
inbuilt peristaltic pump) to be free from defects in material and workmanship when
subjected to normal use and service. This is a replacement guarantee, whereby the faulty
device is returned to Maytronics, or Authorised Dealer, freight prepaid within two years
from the date of purchase. A further one year back to base warranty is provided, in addition
to this two year period, whereby the faulty device is returned to Maytronics for repair,
freight prepaid. The faulty device will be repaired and returned, free of charge.
Maytronics provides the same replacement guarantee for the pH Sensor, but limited to a
period of one year from the date of purchase.
The squeeze tube inside the peristaltic pump, valves and fittings are not covered by this
guarantee.
Acid must be diluted as instructed. Wear and damage caused by the use of undiluted acid is
not covered by this warranty.
There are no expressed or implied warranties which extend beyond the face hereof, and
Maytronics is not liable for any incidental or consequential damages arising from the use or
misuse of this product. This limited warranty does not apply to any injury, loss, damage,
defect or malfunction of the product or failure to function resulting from any failure to
operate the product in accordance with the directions contained in the operating
instructions, failure to function resulting from any accidents, acts of God, tampering, abuse,
acts, omissions, or negligence by anyone other than Maytronics, including but not limited to
such damage or injuries resulting from improper installation. Damage from excessive
concentration of one or more chemicals is not covered by this warranty.
This limited warranty shall apply only to the Customer as an original purchaser. It is the
customer's responsibility to follow safety regulations and laws regarding electrical
installation. Shipping damage is not covered by this warranty.
No claims will be recognised without the proof of purchase. This warranty becomes invalid if
unauthorised person or persons attempt modifications or repairs.
Any dispute between customer and Maytronics must be conducted in Queensland,
Australia.

Table of contents
Popular Controllers manuals by other brands

Servomech
Servomech CLA 30 Installation, operation and maintenance manual

MADRIX
MADRIX STELLA Technical Manual & Quick Start Guide

Intiel
Intiel INT0193 user manual

Weathermatic
Weathermatic SoloRain 8024 DC Series owner's manual

Speed Passion
Speed Passion Reventon wireles Bluetooth Unit user manual

Mitsubishi Electric
Mitsubishi Electric MELSEC Q Series user manual