Dear TD-3140 System Owner:
Thank you for choosing TD-3140 Blood Pressure Monitoring System.
This manual provides important information to help you operate this
system correctly. Before using this product, please read the following
contents thoroughly and carefully.
Managing your blood pressure is simple with this device. It’s small
enough to take with you anywhere, and auto-uploads all of your
readings to the URIGHT smartphone App. In addition, this device
will help you and your healthcare professional, to keep your blood
pressure under control and adjust your treatments plans as needed.
If you have further questions about this system, please contact the
Customer Service for assistance.
Version 2.0 2016/04
311-3140200-003

4
IMPORTANT SAFTY INSTRUCTIONS
READ BEFORE USE
The following basic safety precautions should always be taken.
1. Use the device only for the intended use described in this manual.
2. Do NOT use the device if it is not working properly or damaged.
3. Do NOT use under any circumstances on newborns.
4. This device does NOT serve as a cure for any symptoms or diseases.
The data measured are for reference only. Always consult your doctor
to have the results interpreted.
5. Keep the equipment away from hot surfaces.
6. Do NOT apply the cuff to areas other than the place directed.
7. Use of this instrument in a dry environment, especially if synthetic
materials are present (synthetic clothing, carpets etc.) may cause
damaging static discharges. This may cause erroneous results.
8. Do NOT use this instrument in close proximity to sources of strong
electromagnetic radiation, as these may interfere with the accurate
operation.
KEEP THESE INSTRUCTIONS IN A SAFE PLACE

5
BEFORE YOU BEGIN 06
Intended Use 06
Test Principle 06
Meter Overview 07
Display Screen 08
HOW TO INSTALL AND UPDATE THE URIGHT BLOOD PRESSURE
MANAGER 09
BLUETOOTH PAIRING 10
IMPORTANT INFORMATION ABOUT PERFORMING A BLOOD
PRESSURE TEST 11
Before Measurement 11
Apply the Cuff Properly 11
Proper Measurement Position 12
HOW TO TEST YOUR BLOOD PRESSURE 13
Taking Measurement by the Device Itself 13
Taking Measurement with the TAIDOC TD3140 App 14
MAINTENANCE 17
Battery 17
Caring for your meter 18
DETAILED INFORMATION 19
SYSTEM TROUBLESHOOTING 20
Error Messages on Meter 20
Troubleshooting 21
SYMBOL INFORMATION 22
SPECIFICATIONS 23
TABLE OF CONTENTS

6
BEFORE YOU BEGIN
Intended Use
TD-3140 Blood Pressure Monitoring System is a system designed to
measure the systolic and diastolic blood pressure and pulse rate by
using a non-invasive technique in which an inatable cuff is wrapped
on the upper arm. It is intended for individuals age 18 and up, for
use in a home or clinical setting. The device is not to be used for the
diagnosis of hypertension or for testing on newborns.
Test Principle
Blood pressure is measured non-invasively at the arm based on
oscillometric method.
This device is NOT able to take measurements in the presence of
common arrhythmia, such as atrial or ventricular premature beats
oratrial brillation. It may produce reading error.

7
Meter Overview
Bluetooth Indicator
Download test results with a Bluetooth connection.
Cuff (immovable)
Display Screen
Battery Compartment
ON/OFF Button
Bluetooth Pairing Button
1
2
3
5
4
6
2
1
3
4
6
5

8
Display Screen
Low Battery Symbol
Pulse Rate Symbol
Diastolic Pressure Value
Systolic Pressure Value
Units of Blood Pressure
Systolic Pressure
Symbol
Diastolic Pressure
Symbol
Pulse Rate
1
2
3
4
5
6
7
8
4 3 2 1
5 6 7 8

9
HOW TO INSTALL AND UPDATE
THE URIGHT BLOOD PRESSURE
MANAGER
You can use the TD-3140 Blood Pressure
Monitoring System by itself or with the
URIGHT Blood Pressure Manager App.
URIGHT Blood Pressure Manager is
designed to assist you in easily monitoring,
recording, and tracking your blood pressure.
The app is simple to use and helps you to
better understand your current condition as
well as control your hypertension.
System Requirement
iOS version 7 or higher
Android version 4.3 or higher
Installing
URIGHT Blood Pressure Manager is available on the App Store/Play
Store for free download. Connect your mobile device to the internet to
download the URIGHT Blood Pressure Manager and then install on
your mobile device.
Updating
When new features are added to the URIGHT Blood Pressure
Manager, the App Store will notify you automatically. Please perform
the updates accordingly to provide you with the latest features that we
have developed in your best interest.

10
BLUETOOTH PAIRING
You can transmit your data from the meter to your devices via
Bluetooth. Please contact your local customer service or place of
purchase for assistance. Please note that you must complete the
pairing between meter and Bluetooth receiver before transmitting
data. Please refer to the images of quick start user guide for detailed
procedures.
Pairing with your mobile device
1. Turn on the Bluetooth function on your mobile device.
2. Press the Bluetooth button on the meter. “PCL” will be displayed
on the meter screen.
3. Follow the instruction of your APP to pair the device. (Ex. Scan to
find the meter and then add it into app.)
4. After successfully pairing the APP with the device, the Bluetooth
function of meter shall be on before transmitting the data to your APP.
Bluetooth indicator on the meter
BLUETOOTH INDICATOR STATUS
Flashing Blue The Bluetooth function is on and
waiting for connection.
Solid Blue The Bluetooth connection is
established.
NOTE:
• While the meter is in transmission mode, it will be unable to perform a test.
• Make sure your device with iOS (7 or above) or Android System (4.3 or
above) has turned on Bluetooth before transmitting the data and the meter
is within the receiving range.
• The Bluetooth functionality is implemented in different ways by the various
mobile device manufacturers, the compatibility issue between your mobile
device and the meter maybe occur.

11
IMPORTANT INFORMATION
ABOUT PERFORMING A BLOOD
PRESSURE TEST
Before Measurement
● Avoid caffeine, tea, alcohol and tobacco for at least 30 minutes
before measurement.
● Wait for 30 minutes after exercising or bathing before measurement.
● Sit or lie down for at least 10 minutes before measurement.
● Do not measure when feeling anxious or tense.
● Take a 5-10 minute break between measurements. This break can
be longer if necessary, depending on your physical condition.
● Keep the records for your doctor as reference.
● Blood pressure naturally varies between each arm. Always measure
your blood pressure on the same arm.
Apply the Cuff Properly
1. Stretch your left (right) arm in front of you with your palm facing
up. Slide and place the cuff onto your arm, the arrow should point
toward the forearm.
2. Wrap and tighten the cuff above your elbow. The red line on the
edge of the cuff should be approximately 0.8” to 1.2” (2 cm to 3 cm)
above your elbow.
3. Leave a little free space between the arm and the cuff; you
should be able to fit two fingers between them. Clothing must not
restrict the arm. Remove all clothing covering or constricting the
measurement arm.

12
4. Press the hook material firmly against the pile material. The top and
bottom edges of the cuff should be tightened evenly around your
upper arm.
Proper Measurement Position
1. Sit down for at least 10 minutes before
measuring.
2. Place your elbow on a flat surface.
Relax your hand with the palm facing
up.
3. Remain still and do not talk or move
during the measurement.
NOTE:
The cuff is fixed on the blood pressure monitor. Do NOT remove the cuff or
change to other different size cuff.
WARNING:
If the cuff is relatively lower (higher) than the heart, the obtained blood
pressure value could be higher (lower) than the actual value. A 15 cm
difference in height may result in an error around 10 mmHg.
24~43cm
9.4”~16.9”

13
2. Pump starts to inflate. The pressure in
cuff increases to working pressure.
3. The heart symbol “ ” will show
silently when pulse is detected during
the measurement. When signal of
pulse becomes strong, “ ” icon starts
to flash.
4. After the measurement, the meter
displays the systolic pressure, diastolic
pressure and pulse rate.
5. Tap ON/OFF Button to switch off. The
meter will switch off automatically after
being idle for 3 minutes.
NOTE:
When the device detects irregular heart beat, the IRB symbol is shown as
“ ”. This is for your reference only; if the problem still persists or if you
feel uncomfortable, please consult your doctor.
NOTE:
• When the “Syn” icon appears on the screen,
it indicates you have not paired the meter to
your mobile device. You need to synchronize
with your mobile device in order to adjust the date and time settings. In
addition, this measurement will not be stored in meter memory.
HOW TO TEST YOUR BLOOD
PRESSURE
Taking Measurement by the Device Itself
Always apply the pressure cuff before turning on the meter.
1. Tap ON/OFF Button. All the LCD symbols will appear with a long
“beeping”sound. Then the cuff will begin to inflate automatically.

14
Taking Measurement with the TAIDOC TD3140
App
All the screenshots displayed below come from the display screen of
the Bluetooth receiver installed with URIGHT Blood Pressure
Manager App, not from the display screen of the meter.
Single Measurement
1. Turn on the Bluetooth switch on your mobile device before doing
the test.
2. Launch URIGHT Blood Pressure
Manager App.
3. Apply the pressure cuff, and then
tap " " (in green) on the screen of
URIGHT Blood Pressure Manager
App. All the LCD symbols will appear
with a long “beeping” sound on the
device monitor. The cuff will begin to
inflate automatically until it reaches the
working pressure.
The " " (in green) will turn into "
" (in red) on the App screen during
measurement. If you want to stop the
measurement, press " " (in red)
to terminate it, and the meter will be
switched off.
4. The pulse rate is detected during the measurement.
NOTE:
If the connection is established, the meter will enter PCL mode.
• If you tap ON/OFF Button during measurement, the meter will be turned off.

15
Average Measurement
This function automatically performs three (3) consecutive blood
pressure measurements and displays the average result in the end.
This function is only available when your meter is connected to your
mobile device.
5. After the measurement, the app screen
displays the systolic pressure, diastolic
pressure and pulse rate.
6. Touch the waveform area under
the pulse rate, a pop-up window
will appear to display the predicted
analysis of your result readings. Press
the upper-right icon "x" to close the
pop-up window.
1. Press the Bluetooth button. The “PCL”
icon will appear on the meter display,
indicating that the device is ready to
transmit data.
2. Launch URIGHT Blood Pressure
Manager App.
3. Set Average Mode to ON and tap
" " (in green) on the screen of
URIGHT Blood Pressure Manager
App. Then the cuff will begin to inflate
automatically.

16
5. After taking three measurements, the app screen displays the
average result with an "AVG" tag.
NOTE:
When the difference between the first and second systolic pressure is more
than 15 mmHg, the time interval for third measurement will be 40 seconds.
NOTE:
Always apply the pressure cuff before turning on the meter.
6. Touch the waveform area under
the pulse rate, a pop-up window
will appear to display the predicted
analysis of your average result. Press
the upper-right icon "x" to close the
pop-up window.
The " " (in green) will turn into " " (in red) on the App screen
during measurement. If you want to stop the measurement, press "
" (in red) to terminate it, and the meter will be switched off.
4. After the first measurement is complete, the device and URIGHT
Blood Pressure Manager app will start counting down before
the second measurement begins. The device will take three (3)
measurements consecutively with an interval of 20 seconds.

17
NOTE:
• As with all small batteries, these batteries should be kept away from small
children. If swallowed, promptly seek medical assistance.
• Batteries might leak chemicals if unused for a long time. Remove the
batteries if you are not going to use the device for an extended period (i.e.,
3 months or more).
• Properly dispose the batteries according to your local environmental
regulations.
1. The symbol appears along with
display messages: The meter is
functional and the result remains
accurate, but it is time to change the
batteries.
2. The symbol appears with E-b: The
power is not enough to do a test. You
must change the batteries immediately.
MAINTENANCE
Battery
Your meter comes with four (4) 1.5 V AAA size alkaline batteries.
Low Battery Signal
The meter will display either of the two messages below to alert you
when the meter power is getting low.
Replacing the Battery
To replace the batteries, make sure the meter is turned off.
1. Press the edge of the battery cover and lift it up to remove.
2. Remove the old batteries.
3. Replace with four 1.5 V AAA size alkaline batteries in accordance
With the mark +/- indication in the battery compartment.
4. Close the battery cover.

18
Caring for your meter
To avoid the meter from attracting dirt, dust or other contaminants,
wash and dry your hands thoroughly before use.
Cleaning
● To clean the meter exterior, wipe it with a cloth moistened with tap
water or a mild cleaning agent, then dry the device with a soft dry
cloth. Do NOT flush with water.
● Do NOT use organic solvents to clean the meter.
● Do NOT wash or iron the pressure cuff.
Meter Storage
● Storage condition: -25°C to 70°C (-13°F to 158°F), 10% to 95%
relative humidity.
● Always store or transport the meter in its original storage case.
● Avoid dropping or heavy impact.
● Avoid direct sunlight and high humidity.
• Replacing the batteries does not affect the test results stored in memory.
However, if you did not synchronize with your mobile device after the
battery replacement, the measured data will not be stored in the memory.

19
DETAILED INFORMATION
Reference Values
Clinical studies show that adult diabetes is often accompanied by
elevated blood pressure. People with diabetes can reduce their heart
risk by managing their blood pressure along with diabetes treatment*1.
Monitoring your routine blood pressure trend helps you to know
your body condition. Human blood pressure naturally increases
after reaching middle age. This symptom is a result of continuous
ageing of the blood vessels. Further causes include obesity, lack of
exercise and cholesterol (LDL) adhering to the blood vessels. Rising
blood pressure accelerates hardening of the arteries, and the body
becomes more susceptible to apoplexy and coronary infarction. The
recommended blood pressure range is as below:
Classication Systolic Pressure
(mmHg)
Diastolic Pressure
(mmHg)
Hypotension*2≤ 90 ≤ 60
Normal*3< 120 < 80
Pre-hypertension*3 120 – 139 80 – 89
Stage 1 Hypertension*3140 – 159 90 – 99
Stage 2 Hypertension*3≥ 160 ≥ 100
*1: American Diabetes Association: The Diabetes-Heart Disease Link
Surveying Attitudes, Knowledge and Risk (2002).
*2: National Heart, Lung, and Blood Institute, Diseases and Conditions.
*3: The Seventh Report of the Joint National Committee on Prevention,
Detection, Evaluation, and Treatment of High Blood Pressure. NIH
Publication. 2003. No. 03-5233.

20
SYSTEM TROUBLESHOOTING
Error Messages on Meter
MESSAGE PROBABLE
CAUSE SOLUTION
E-1 Ination or pressure
error.
Please contact local customer
service for help.
E-4 Blood pressure
measurement error.
Ret cuff tightly and
correctly. Relax and repeat
the measurement. If error
still remains, contact local
customer service for help.
E-A
E-E
Problems with the
meter.
Review the instructions and
repeat the test. If the meter
still does not work, please
contact the local customer
service for help.
E-b Battery is too low. Repeat with new batteries.
NOTE:
The information here is only displayed on the meter screen.
Table of contents