Mediana HeartOn T15 User manual

!
OPERATOR’S MANUAL
HeartOn T15®
AED Trainer
Part Number: A7427-0
Revised Date: 0814
Copyright © 2014 All rights reserved.
www.medianadefib.co.uk
Distributed by:
Reliance Medical LTD, West Avenue,
Talke, Stoke-on-Trent, Staffordshire,
ST7 1TL, United Kindom.

This page is intentionally left blank.

1
Contents
Configuration of AED Trainer........................................................................................................................2
Features of AED Trainer ................................................................................................................................2
Description of the AED Trainer .....................................................................................................................3
Components and Controls ........................................................................................................................3
Remote Control.........................................................................................................................................6
Using the AED Trainer ...................................................................................................................................8
Operating the AED Trainer .......................................................................................................................8
Custom Scenario ......................................................................................................................................9
Maintenance..................................................................................................................................................13
Specification .................................................................................................................................................13
Characteristics ........................................................................................................................................13
Environmental Conditions.......................................................................................................................13
Remote control .......................................................................................................................................13
List of Components......................................................................................................................................13
Figures
Figure 1. HeartOn T15 Components and Controls...............................................................................................3
Figure 2. HeartOn T15-G4 Components and Controls.........................................................................................3
Figure 3. Side and bottom panel of AED Trainer.................................................................................................. 4
Figure 4. Training Pads ........................................................................................................................................4
Figure 5. AED Trainer Remote Control ................................................................................................................6
Tables
Table 1. AED Trainer Components and Controls .................................................................................................4
Table 2. HeartOn T15 Action Icon ........................................................................................................................ 5
Table 3. HeartOn T15-G4 Action Icon ..................................................................................................................6
Table 4. AED Trainer Remote Control..................................................................................................................7
Table 5. Setting of External DIP Switch................................................................................................................ 9
Table 6. Scenario ............................................................................................................................................... 10
Table 7. Scenario (continued) ............................................................................................................................11

2
Configuration of AED Trainer
The following table identifies AED trainer configurations and how they are indicated. Model
name is located on the bottom of the AED trainer.
Model
Reference No.
Description
HeartOn T15
T15-G8-0E
Basic (8 Action icon)
HeartOn T15-G4
T15-G4-0E
Basic (4 Action icon)
Features of AED Trainer
!AED Trainer provides simulated shock delivery. It has no high-voltage capabilities,
ensuring safety during training.
!AED Trainer can be used with the infrared remote control. The remote control gives the
instructor the ability to alter training scenarios while in progress, to test student response.
!AED Trainer is pre-configured with 10 training scenarios that simulate realistic sudden
cardiac arrest episodes.
!The 1st scenario can be simulated at first by using the MEDIANA OPTION button of the
remote control. (HeartOn T15 only. The MEDIANA OPTION button press is ignored on
HeartOn T15-G4.)
!The shockable rhythm or non-shockable rhythm can be simulated immediately by using
the remote control.
!The PC software let you configure custom scenarios.
!Connection to a PC serial port requires USB 2.0 A-Male to Mini-B cable. .

3
Description of the AED Trainer
Components and Controls
Figure 1. HeartOn T15 Components and Controls
Figure 2. HeartOn T15-G4 Components and Controls
1
3
2
4
5
6
7
9
8
1
2
3
4
5
6
7
9
8

4
Figure 3. Side and bottom panel of AED Trainer
Figure 4. Training Pads
Table 1. AED Trainer Components and Controls
No.
Components
Description
1
Cover
Cover is used to protect the action icon, the patient mode
switch and the shock button.
2
Status indicator
Status indicator displays the AED status, the temperature
status and the battery status.
3
Pad connector
Pad connector links the pads.
4
Action icon
LED indicator flashes red LED under the relevant action
icon.
5
Patient mode
switch
After distinguishing the patient according to patient type,
select the patient mode between adult and pediatric patient
mode by pushing the patient mode switch.
6
Shock button
When preparation for electric shock is completed, press
the flashing button and then the AED delivers the
simulated shock.
7
Slide button
Slide button is used to open the cover and turn on by
pushing the slide button to right.
8
SD card port
SD card port is not used in HeartOn T15 and HeartOn T15-
G4.
9
Infrared
communication port
Infrared communication port is used to communicate with
the remote control.
10
External DIP Switch
Box
External DIP switch box is used to protect the external DIP
switch and external DIP switch in external DIP switch box
is used for changing custom scenario.
11
Training Pads
Training pads are attached to the manikin and cannot
deliver a defibrillation shock.
11
9
8
10

5
Table 2. HeartOn T15 Action Icon
Action icon
Voice prompt
“Check for response. Are you right?”
“Call for help.”
“Open the airway.”
“Check breathing.”
“Remove clothes from the patient’s chest. Place pads exactly as
shown in the picture. Press pads firmly to patient’s bare chest.”
or
“Plug in pads. Insert connector firmly.”
“Adult pads” or “Pediatric pads”
“Analyzing heart rhythm. Do not touch the patient.”
“Shock advised. Charging. Do not touch the patient.”
or
“Analyzing heart rhythm. Do not touch the patient.”
“No shock advised.”
“Press the red flashing button now. Deliver shock now.”
“Shock delivered.”
or
“Press the red flashing button now. Deliver shock now.”
“Shock cancelled.”
or
“Press the red flashing button now. Deliver shock now.”
“Shock button not pressed.”
“It is safe to touch the patient.”
“Begin CPR.”
or “If needed, Begin CPR.”
{[Beep] “Give two breaths.”}
{} Repeat 5 times
“Stop CPR.”

6
Table 3. HeartOn T15-G4 Action Icon
Action icon
Voice prompt
“Remove clothes from the patient’s chest. Place pads exactly as
shown in the picture. Press pads firmly to patient’s bare chest.”
or
“Plug in pads. Insert connector firmly.”
“Adult pads” or “Pediatric pads”
“Analyzing heart rhythm. Do not touch the patient.”
“Shock advised. Charging. Do not touch the patient.”
or
“Analyzing heart rhythm. Do not touch the patient.”
“No shock advised.”
“Press the red flashing button now. Deliver shock now.”
“Shock delivered.”
or
“Press the red flashing button now. Deliver shock now.”
“Shock cancelled.”
or
“Press the red flashing button now. Deliver shock now.”
“Shock button not pressed.”
“It is safe to touch the patient.”
“Begin CPR.”
or “If needed, Begin CPR.”
{[Beep] “Give two breaths.”}
{} Repeat 5 times
“Stop CPR.”
Remote Control
Figure 5. AED Trainer Remote Control

7
Table 4. AED Trainer Remote Control
Symbols
Description
The LED flashes when the button is pressed.
Select one of the 10 standard scenarios.
Select one of 3 custom scenarios.
Adjust speaker volume.
The last setting will be stored when the AED cover is closed.
Shockable rhythm button
Non-shockable rhythm button
Poor pad condition recovery button
Motion artifact button
Poor pad condition button
Not used in HeartOn T15 and HeartOn T15-G4.
Low battery condition button
Error condition button
Resumption button
Freeze button
MEDIANA OPTION button
(HeartOn T15 only. The MEDIANA OPTION button press is
ignored on HeartOn T15-G4.)

8
Using the AED Trainer
Operating the AED Trainer
1. Push the Slide Button to open the AED cover.
2. If the AED is turned on by pushing the Slide Button, the first scenario is started
automatically.
3. To change the scenario during other scenario simulation, press the desired scenario
number on the remote control then a selected scenario will be started automatically.
4. The volume can be controlled by 10 steps. The last setting will be stored and maintained
even if the AED has been turned off.
Function of remote control
The shockable rhythm is simulated to deliver the shock immediately. After
deliver the shock, the next scenario is simulated.
The non-shockable rhythm is simulated. The CPR scenario is simulated
without shock delivery.
The poor pad condition is recovered to the good pad condition. After then,
the AED Trainer resumes on the analyzing rhythm scenario.
The motion artifact is simulated.
When the button is pushed once, the poor pad condition is simulated and
push again, the changing pad condition is simulated. When the pad
disconnects and connects again, the changing pad condition is recovered
to normal condition.
Not used in HeartOn T15 and HeartOn T15-G4.
Simulate the Low battery condition and indicate Low battery condition on
the status indicator. When the button is pushed again, the AED Trainer
displays real battery level on the status indicator.
Pressing one time, the AED is operated normally, current temperature is
indicated low, and temperature icon is flashing. Pressing once again, error
condition is simulated, the X mark is displayed on the status indicator and
temperature icon is lit. Finally, pressing once again, the AED will not be
operated, unless the AED cover is closed and then reopened.
Resume the audible activity from the point at which it was paused.
Pause all audible activity in the AED Trainer. If button is not pushed,
the AED Trainer does not play any audible activity during pushing any
button.
The starting point of the standard scenario can be selected by user.
MEDIANA OPTION button is toggles between starting point from the
checking patient’s consciousness and starting point from the beginning
part. (HeartOn T15 only. The MEDIANA OPTION button press is ignored
on HeartOn T15-G4.)
Note: When the pads are not connected during scenario simulation, the audio prompt
sounds repeatedly to ask the pads is inserted.

9
Custom Scenario
The default custom scenario is simulated if the custom scenario setting is not changed by
user.
Changing Custom Scenario
1. Open the external DIP switch box at bottom panel.
2. Place the 5 and 6 pin of the External DIP Switch to down position to activate custom
scenario function.
3. Connect the USB mini B cable.
4. Run the custom scenario software (option) on the PC.
5. Set up the serial port (COM port) on the PC.
6. Select the desired scenario and then perform the download to the custom scenario 1, 2,
or 3.
7. After completion of downloading, turn off the AED by closing the AED cover and
disconnect the USB mini B cable.
8. Set the External DIP Switch to normal operation.
9. Turn on the AED by opening the AED cover again and verify the setting of custom
scenario by pressing the Custom Scenario Button on the remote control.
Table 5. Setting of External DIP Switch
Setting of External DIP Switch
Description
Setting available
Normal operation
Standard use
Changing custom
scenario
Standard use
Changing audible
list
Not use without approval by
service personnel

10
Table 6. Scenario
Scenario 1
Scenario 2
Scenario 3
Scenario 4
Scenario 5
1*
Checking patient’s
consciousness
Checking patient’s
consciousness
Checking patient’s
consciousness
Checking patient’s
consciousness
Checking patient’s
consciousness
2*
Calling to EMS
Calling to EMS
Calling to EMS
Calling to EMS
Calling to EMS
3*
Open the airway
Open the airway
Open the airway
Open the airway
Open the airway
4*
Checking breath
Checking breath
Checking breath
Checking breath
Checking breath
5
Attaching pad
Attaching pad
Attaching pad
Attaching pad
Attaching pad
6
Warning: Do not
touch with the
patient
Warning: Do not
touch with the
patient
Poor Pads
connecting
Warning: Do not
touch with the
patient
Warning: Do not
touch with the
patient
7
Analyzing rhythm
Analyzing rhythm
Warning: Do not
touch with the
patient
Analyzing rhythm
Analyzing rhythm
8
Simulating
Shockable rhythm
Simulating
Shockable rhythm
Analyzing rhythm
Simulating
Shockable rhythm
Simulating non-
shockable rhythm
9
Delivering shock
Delivering shock
Simulating
Shockable rhythm
Delivering shock
CPR(30:2,
5CYCLE)
10
CPR(30:2,
5CYCLE)
CPR(30:2,
5CYCLE)
Delivering shock
CPR(30:2,
5CYCLE)
11
Warning: Do not
touch with the
patient
Repeat 3 times
No. 6 to 10
CPR(30:2,
5CYCLE)
Warning: Do not
touch with the
patient
12
Analyzing rhythm
Warning: Do not
touch with the
patient
Warning: Do not
touch with the
patient
Analyzing rhythm
13
Simulating non-
shockable rhythm
Analyzing rhythm
Analyzing rhythm
Simulating non-
shockable rhythm
14
CPR(30:2,
5CYCLE)
Simulating non-
shockable rhythm
Simulating non-
shockable rhythm
CPR(30:2,
5CYCLE)
15
CPR(30:2,
5CYCLE)
CPR(30:2,
5CYCLE)
Warning: Do not
touch with the
patient
16
Analyzing rhythm
17
Simulating
Shockable rhythm
18
Delivering shock
19
CPR(30:2,
5CYCLE)
* HeartOn T15 only. The MEDIANA OPTION button press is ignored on HeartOn T15-G4.
Note: When the standard scenario is selected, AED trainer starts the simulation from ‘Analyzing rhythm’ step.
To start the simulation from ‘Checking patient’s consciousness’ step, select the desired standard
scenario after pressing the MEDIANA OPTION button on the remote control. (HeartOn T15 only.
The MEDIANA OPTION button press is ignored on HeartOn T15-G4.)

11
Table 7. Scenario (continued)
Scenario 6
Scenario 7
Scenario 8
Scenario 9
Scenario 10
1*
Checking patient’s
consciousness
Checking
patient’s
consciousness
Checking patient’s
consciousness
Checking patient’s
consciousness
Checking patient’s
consciousness
2*
Calling to EMS
Calling to EMS
Calling to EMS
Calling to EMS
Calling to EMS
3*
Open the airway
Open the airway
Open the airway
Open the airway
Open the airway
4*
Checking breath
Checking breath
Checking breath
Checking breath
Checking breath
5
Attaching pad
Attaching pad
Attaching pad
Attaching pad
Attaching pad
6
Warning: Do not
touch with the
patient
Warning: Do not
touch with the
patient
Poor Pads
connecting
Warning: Do not
touch with the
patient
Simulating motion
artifact
7
Analyzing rhythm
Analyzing rhythm
Warning: Do not
touch with the
patient
Analyzing rhythm
Warning: Do not
touch with the
patient
8
Simulating
Shockable rhythm
Simulating
Shockable
rhythm
Analyzing rhythm
Simulating
Shockable rhythm
Analyzing rhythm
9
Delivering shock
Delivering shock
Simulating
Shockable rhythm
Delivering shock
Simulating
Shockable rhythm
10
CPR(30:2,
5CYCLE)
CPR(30:2,
5CYCLE)
Delivering shock
CPR(30:2,
5CYCLE)
Delivering shock
11
Warning: Do not
touch with the
patient
Warning: Do not
touch with the
patient
CPR(30:2,
5CYCLE)
CPR(30:2,
5CYCLE)
12
Analyzing rhythm
Analyzing rhythm
Warning: Do not
touch with the
patient
Warning: Do not
touch with the
patient
13
Simulating
Shockable rhythm
Simulating
Shockable
rhythm
Analyzing rhythm
Analyzing rhythm
14
Delivering shock
Delivering shock
Simulating
Shockable rhythm
Simulating non-
shockable rhythm
15
CPR(30:2,
5CYCLE)
CPR(30:2,
5CYCLE)
Delivering shock
CPR(30:2,
5CYCLE)
16
Warning: Do not
touch with the
patient
Warning: Do not
touch with the
patient
CPR(30:2,
5CYCLE)
Low Battery
17
Analyzing rhythm
Analyzing rhythm
Warning: Do not
touch with the
patient
18
Simulating non-
shockable rhythm
Simulating non-
shockable rhythm
Analyzing rhythm
19
CPR(30:2,
5CYCLE)
CPR(30:2,
5CYCLE)
Simulating non-
shockable rhythm
20
Warning: Do not
touch with the
patient
CPR(30:2,
5CYCLE)

12
Scenario 6
Scenario 7
Scenario 8
Scenario 9
Scenario 10
21
Analyzing rhythm
22
Simulating
Shockable
rhythm
23
Delivering shock
24
CPR(30:2,
5CYCLE)
25
Warning: Do not
touch with the
patient
26
Analyzing rhythm
27
Simulating non-
shockable rhythm
28
CPR(30:2,
5CYCLE)
* HeartOn T15 only. The MEDIANA OPTION button press is ignored on HeartOn T15-G4.
Note: When the standard scenario is selected, AED trainer starts the simulation from ‘Analyzing rhythm’
step. To start the simulation from ‘Checking patient’s consciousness’ step, select the desired
standard scenario after pressing the MEDIANA OPTION button on the remote control. (HeartOn
T15 only. The MEDIANA OPTION button press is ignored on HeartOn T15-G4.)

13
Maintenance
To clean the AED Trainer, wipe the AED trainer with a soft cloth that has been dampened by
one of the following:
!Soapy water
!70% Isopropyl alcohol
Do not immerse any part of the AED Trainer in water or any type of fluid. Contact with fluids
may seriously damage the device or cause fire or shock hazard.
Do not clean the AED Trainer with abrasive materials, cleaners or solvents.
Specification
Characteristics
!Power
1) DC 10.8V / 2200mAh 3cell Li-ion rechargeable Battery
2) DC 15V / DC adapter 15V, 1.5A
!Static Current : < 5uA
!Maximum Current : < 500mA
!Size: 240mm ×294mm ×95mm
!Weight: 1.44kg (including the battery)
Environmental Conditions
Operation
!Temperature: 0 to 43°C (32 to 109.4°F)
!Relative Humidity: 5 to 95% Non-condensing
!Altitude: 0 to 4,575 m
Storage
!Temperature: -20 to 60°C (-4 to 140°F)
!Relative Humidity: 5 to 95% Non-condensing
!Altitude: 0 to 12,192 m
Remote control
!Power: DC 3V / Coin cell battery
!Static Current : < 5uA
!Maximum Current : < 10mA
List of Components
No.
Description
Q’ty
Unit
1
HeartOn T15 (including the battery)
1
pc
HeartOn T15-G4 (including the battery)
2
Remote control
1
pc
3
Training pad for adult/pediatric
(It can use 100 times)
1
pc
4
Operator’s manual
1
pc
5
Storage bag (Nylon Bag)
1
pc
6
DC Adapter
1
pc
7
AED Trainer Software CD (Option)
1
pc
This manual suits for next models
1
Table of contents