QA-ES Electrosurgical Analyzer
P/N 14025

Copyright 2001 by METRON. All rights reserved.
METRON:
USA __ FRANCE ________________ NORWAY________________
1345 Monroe NW, Suite 255A 30, rue Paul Claudel Travbaneveien 1
Grand Ra ids, MI 49505 91000 Evry, France N-7044 Trondheim, Norway
Phone: (+1) 888 863-8766 Phone: (+33) 1 6078 8899 Phone: (+47) 7382 8500
Fax: (+1) 616 454-3350 Fax: (+33) 1 6078 6839 Fax: (+47) 7391 7009
E-mail: su ort.us@metron-biomed.com E-mail: [email protected]r E-mail: su ort@metron.no
Disclaimer
METRON provides this publicatio as is without warra ty of a y ki d, either express or implied,
i cludi g but ot limited to the implied warra ties of mercha tability or fit ess for a y particular pur-
pose. Further, METRON reserves the right to revise this publicatio a d to make cha ges from time
to time to the co te t hereof, without obligatio to METRON or its local represe tatives to otify a y
perso of such revisio or cha ges. Some jurisdictio s do ot allow disclaimers of expressed or im-
plied warra ties i certai tra sactio s; therefore, this stateme t may ot apply to you.
Limited Warranty
METRON warra ts that the QA-ES Electrosurgical A alyzer will substa tially co form to pub-
lished specificatio s a d to the docume tatio , provided that it is used for the purpose for which it was
desig ed. METRON will, for a period of twelve (12) mo ths from date of purchase, replace or repair
a y defective system, if the fault is due to a ma ufacturi g defect. I o eve t will METRON or its lo-
cal represe tatives be liable for direct, i direct, special, i cide tal, or co seque tial damages arisi g
out of the use of or i ability to use the QA-ES Electrosurgical A alyzer, eve if advised of the possi-
bility of such damages. METRON or its local represe tatives are ot respo sible for a y costs, loss of
profits, loss of data, or claims by third parties due to use of, or i ability to use the QA-ES Electrosur-
gical A alyzer. Neither METRON or its local represe tatives will accept, or be bou d by a y other
form of guara tee co cer i g the QA-ES Electrosurgical A alyzer other tha this guara tee. Some ju-
risdictio s do ot allow disclaimers of expressed or implied warra ties i certai tra sactio s; there -
fore, this stateme t may ot apply to you.
ii

Table of Contents
1. Introduction........................................................................................................................................1
2. Installation..........................................................................................................................................1
3. Operating QA-ES...............................................................................................................................1
4. ESU Tests with QA-ES......................................................................................................................1
. Control and Calibration....................................................................................................................1
6. Component Functions and Parts......................................................................................................1
Appendix A – Diagrams ........................................................................................................................A-1
Appendix B – Error Report Form........................................................................................................B-1
Appendix C – Improvement Suggestion Form....................................................................................C-1
iii

This page intentionally left blank.
iv

Manual Revision Record
This record page is for recordi g revisio s to your QA-ES User & Service Manual that have bee
published by METRON AS or its authorized represe tatives. We recomme d that o ly the ma -
ageme t or facility represe tative authorized to process cha ges a d revisio s to publicatio s:
•make the pe cha ges or i sert the revised pages;
•e sure that obsolete pages are withdraw a d either disposed of immediately, or marked as
superseded a d placed i a superseded docume t file, a d;
•e ter the i formatio below reflecti g that the revisio s have bee e tered.
Rev No Date Entere Reason Signature of Person Entering Change
1.30-1 4-30-01 Ge eral Update
v

This page intentionally left blank.
vi

1. Introduction
This chapter describes the METRON QA-ES Electrosurgical A a-
lyzer, i cludi g its features a d specificatio s.
1.1 QA-ES Descrip-
tion
The METRON QA-ES Electrosurgical A alyzer (QA-ES) is a preci-
sio i strume t desig ed to perform tests o high-freque cy electro-
surgical u its (ESU) i accorda ce with atio al a d i ter atio al
sta dards, a d is desig ed to be used by trai ed service tech icia s.
Tests i clude:
•automatic power distributio measureme t;
•crest factor measureme t;
•RF leak measureme t, a d;
•retur electrode mo itor (REM) test
Testi g is accomplished by measuri g the ESU output agai st test
loads that are set a d adjusted i the QA-ES. The QA-ES ca auto-
matically execute a power distributio test with a load resista ce
ra gi g from 10 ohms to a maximum of 5200 ohms. The automatic
measuri g of the QA-ES, comprisi g crest factor measureme ts
with a ba dwidth of 10 MHz, e sures that the test result is reliable
a d reproducible.
Test results, show i the QA-ES’s LCD display, ca be pri ted out
directly, or tra sferred to a PC via the PRO-Soft QA-ES test auto-
matio software. PRO-Soft lets you desig test protocols, remotely
co trol the QA-ES, a d store the test results.
1.2 QA-ES Specifications
Generator Output: RF LEAKAGE: From active electrode or neu-
tral late with an o en or closed load circuit.
Mo e Of Operation: Manual or user- rogrammable. Can be re-
motely controlled with PC utilizing accessory
PRO-Soft QA-ES software and RS-232 com-
munication cable connection.
Measurements: True RMS value of a lied waveform.
RMS Ban wi th: 30 Hz to 10 MHz (+3 dB).
Low Frequency Filter: 100 Hz filter to avoid low frequency distur-
bance and/or interference.
Current: 20 mA to 2200 mA.
Current Accuracy: 20 - 2200 mA ± 2% of reading.
Loa Resistance: 10 - 2500 ohms in ste s of 25 ohms (@ dc).
1

2600 - 5200 ohms in ste s of 100 ohms (@
dc).
A itional Fixe Loa : 200 ohms, 500 watt maximum.
Crest Factor: The higher of the two eak-measurements is
used for calculation.
Range: 1.4 - 16 (V eak voltage / V RMS).
Foot Switch Output: The out ut triggers the measurement after a
rogrammed delay time, defined as the time
eriod from the activation of the foot switch to
the beginning of data rocessing. The delay
time is 200 ms - 4000 ms.
Peak To Peak Voltage: 0 to 10 kV (closed load only) ± 10%. Mea-
surement is taken between the active and
dis ersive electrodes with closed load only.
Oscilloscope Output: 5 V/A uncalibrated, 100 mA RF current mini-
mum in ut.
Isolation: 10 kV isolation between measurement device
and enclosure.
1.3 General Information
Temperature Requirements:
+15°C to +35°C when o erating
0°C to +50°C in storage
Display:
Ty e LCD gra hic dis lay
Al hanumeric format 8 lines, 40 characters
Gra hics mode: 240 x 64 oint matrix
Dis lay control: 5 F-keys, enter, cancel and an encoder
Data Input/ Output (2): Parallel rinter ort (1); Bi-directional RS
-232C (1) for Com uter control
Power Source: From 115 VAC to 230 VAC, 48/66 MHz.
Mechanical Specifications:
Housing Metal case
Height 13.2 cm / 4.48 in.
Width 34.2 cm / 11.61 in.
Length 39.5 cm / 13.41 in.
Weight 9.8 kg / 21.6 lbs.
Stan ar Accessories:
QA-ES Electrosurgical Analyzer (P.N. 14010)
Power Cord (P.N. 14300)
QA-ES User and Service Manual (P.N. 14025)
A itional Accessories:
E-In ut Measuring Cable - Black (P.N. 11451)
2

E-In ut Measuring Cable - Red (P.N. 11452)
Alligator Clam - Black (P.N. 11461)
Alligator Clam - Red (P.N. 11462)
Carrying case (P.N. 14100)
PRO-Soft QA-ES software (P.N. 12200)
PRO-Soft QA-ES DEMO (P.N. 14201)
PRO-Soft QA-ES User Manual (P.N. 14225)
Storage:
Store in the carrying case in dry surroundings within the tem era-
ture range s ecified. There are no other storage requirements.
Perio ic Inspection:
The unit should be calibrated every 12 months.
3

This page intentionally left blank
4

2. Installation
This chapter explai s u packi g, receipt i spectio a d claims, a d
the ge eral procedures for QA-ES setup.
2.1 Receipt, Inspec-
tion an Return
1. I spect the outer box for damage.
2. Carefully u pack all items from the box a d check to see that
you have the followi g items:
QA-ES Electrosurgical A alyzer (PN 14010)
E-I put Measuri g Cable -- Black (PN 11411-B)
E-I put Measuri g Cable -- Red (PN 11411-R)
Alligator Clamp -- Black (PN 11412-B)
Alligator Clamp -- Red (PN 11412-R)
Power Cord (No PN)
QA-ES User and Service Manual (PN 14025)
3. If you ote physical damage, or if the u it fails to fu ctio ac-
cordi g to specificatio , i form the supplier immediately.
Whe METRON AS or the compa y’s represe tative, is i -
formed, measures will be take to either repair the u it or dis-
patch a replaceme t. The customer will ot have to wait for a
claim to be i vestigated by the supplier. The customer should
place a ew purchase order to e sure delivery.
4. Whe retur i g a i strume t to METRON AS, or the compa y
represe tative, fill out the address label, describe what is wro g
with the i strume t, a d provide the model a d serial umbers.
If possible, use the origi al packagi g material for retur ship-
pi g. Otherwise, repack the u it usi g:
a rei forced cardboard box, stro g e ough to carry the
weight of the u it.
at least 5 cm of shock-absorbi g material arou d the u it.
o abrasive dust-free material for the other parts.
Repack the u it i a ma er to e sure that it ca ot shift i the
box duri g shipme t.
METRON’s product warra ty is o page ii of this ma ual. The
warra ty does ot cover freight charges. C.O.D. will ot be ac-
cepted without authorizatio from METRON A.S or its repre-
se tative.
2.2 Setup
1. Equipme t co ectio is as show i the typical setup below
(for Power Distributio Test).
1

2. If you are usi g a oscilloscope, attach the BNC cable to the
Scope Output co ector, located o the fro t of the QA-ES.
3. If PRO-Soft QA-ES is bei g used, attach a RS-232 ( ull mo-
dem/data tra sfer co figured) cable to the 9-pi D-sub outlet
port located at the rear of the QA-ES. Do ot attach the pri ter
cable to the QA-ES. See below. However, if you are ot usi g
PRO-Soft QA-ES, a d are se di g directly to a pri ter for pri t-
outs, attach the pri ter cable to the 25-pi outlet port.
2.3 PRO-Soft QA-ES
PRO-Soft QA-ES is a fro t-e d test automatio a d prese tatio
tool for METRON's ESU Performa ce A alyzer. It allows you to
co duct the same tests, but by remote co trol via a IBM-compati-
ble PC/XT with MS Wi dows (Versio 3.1 or later). Additio ally,
the program has features to e ha ce your QA-ES’s performa ce.
Each of the QA-ES tests ca be ru i depe de tly from PRO-Soft i
the “Ma ual” test mode. Results are show o the PC scree duri g
testi g, a d the user is prompted to set the tested equipme t accord-
i gly. At the co clusio of tests, the user may pri t a report, store
the test a d results o disk, or both. Combi atio s of tests ca be
created a d stored as “Test Seque ces.” The program mai tai s a li-
brary of these seque ces. I this way you ca store a d retrieve se-
CAUTION
The QA-ES uses a 3-wire
ower cord and lug for
the user's safety. Use this
ower cord in conjunction
with a ro erly grounded
electrical outlet to avoid
electrical shock.
2

que ces that are appropriate for each ESU bei g tested at your facil-
ity.
Seque ces ca the be used i depe de tly, or ca be attached to a
checklist, writte procedure, a d equipme t data i the form of a
test “Protocol.” The equipme t data ca be e tered ma ually i to
the protocol, or it may be retrieved by PRO-Soft from a database
program, or other equipme t files. Protocols ca be created easily
for each ESU i your i ve tory, a d stored for use. Test protocols
with results ca be pri ted, or stored o disk, a d the results of test-
i g ca be se t back to the equipme t database to close a work order
a d update the service history.
NOTE
PRO-Soft QA-ES has its
own user manual, which
contains all the information
concerning the rogram. If
you order a demonstration
version of the rogram you
also receive the manual.
3

This page intentionally left blank
4

3. Operating QA- S
This chapter explai s the operati g co trols, switches a d me us
of the QA-ES, a d details how to use them i ESU testi g.
3.1 Control Switches
an Connections
Front Panel
1. Power Switch Turns the ower on and off.
2. Enco er Sets values according to the s ecified range and
choose between different o erations/ measurement
ranges.
3. Enter Admits newly s ecified information.
4. Cancel Cancels a new value and returns to reviously cho-
sen value.
5. LCD Display Shows messages, test results and function menus.
6. Function Keys Fl - F5 are used to select the functions shown on the
bottom line of the LCD dis lay, i.e., for selecting the
function that is directly above the key.
7. RF-Detect Indicates when the ESU is activated.
8. Remote Indicates that REMOTE CONTR. (F4) has been
ressed.
9. Scope Output
Connector
BNC-cable connector for attenuator signal in real
time. (When oscillosco e out ut is desired.)
1

Right Side Panel
10. Terminal
RED an BLACK
Connection for the electrode out uts of the VAR.
LOAD ESU. Active electrode to the red terminal,
and neutral electrode to the black terminal.
11. Terminal
BLUE an BLUE
Additional fixed load resistance of 200 ohms
FIXED LOAD 500 watt for serial connection dur-
ing leakage test.
12. Terminal
GREEN an GREEN
Foot switch out ut switch can be used to trigger
the ESU.
Rear Panel
13. RS-232 Serial Port 9- in D-sub
14. Printer Outlet Port 14-25 in D-sub
15. Main Connector 3- in SCHUCO- lug
16. Voltage Selector 115 VAC/230 VAC
17. Fuses T 200 mA @230 VAC / T 400 mA @ 11 5 VAC
2

3.2 QA-ES Menu an
Function Keys
The QA-ES uses a display, programmable fu ctio keys a d a set-
ti g regulator to provide flexibility a d co trol over the operatio s.
The upper part of the scree displays messages, status a d results.
The me u bar is at the bottom of the display. The fu ctio keys are
umbered from Fl to F5. A fu ctio is selected by pressi g the key
located directly u der the Me u Item displayed i the me u bar.
3.3 LCD Display Menu/
Messages (Overview)
3

3.4 LCD Display
Menu/ Messages (Detail)
1. Startup Screen. The followi g scree will be displayed for 2
seco ds after the QA-ES has bee switched o .
2. Main Menu
a. First Me u Bar (Page 1)
b. Seco d Me u Bar (Page 2)
3. SHOW CHOICES (F1). This fu ctio is activated whe you
see a asterisk (*) i the status field u der ‘Mode.’ Choose a
test fu ctio by pressi g UP (F2) or DOWN (F3). (The e -
coder ca also be used for choosi g a test fu ctio ) Press
ENTER (F ) to save it u der Mode i the STATUS field. Press
CANCEL (F4) to u do.
4. KNOB PARAM. (F2). With this fu ctio , you ca choose be-
twee ‘Mode,’ ‘Load’ a d ‘Delay’ i the STATUS field. (*
marks the active item). If you choose ‘Load’, use the e coder to
set the load from 10 ohms to 5200 ohms i steps of:
25 ohms from 50 ohms to 2500 ohms.
100 ohms from 2500 ohms to 5200 ohms.
Save the selected load i ‘Mode’ u der the STATUS field by
pressi g ENTER (F ). Press CANCEL (F4) to u do. If you
choose ‘Delay’, use the setti g regulator to set the delay from
200 ms to 4000 ms i steps of:
50 ms from 200 ms to 1000 ms.
4

100 ms from 1000 ms to 4000 ms.
Save the chose delay i ‘Delay’ u der the STATUS field by
pressi g ENTER (F ). Press CANCEL (F4) to u do.
5. START (F3). Whe you press o START, the test procedure
will begi , a d the text i the field ‘Oper.’ will cha ge from
‘Ready’ to ‘Measuri g’. If the u it is set to the positio for a
REM test, this text will cha ge from ‘Ready’ to ‘l cr'. res.’
Press STOP (F3) to stop the test procedure.
6. SETUP (F4). Here you ca set the power distributio level for
start, stop a d step i ohms.
Choose the ‘Start load’ by usi g KNOB PARAM (F4). (see
stars). Use the e coder to set the level. Save the level by press-
i g ENTER (F ). Press CANCEL (F4) to u do. Go to ‘E d
load’ a d ‘Step Size’ a d repeat the same procedure.
Pwr. distr. Start load is the first load to be used duri g the
measureme ts; it ca be set from 10 ohms to 2100 ohms, with
steps of 25 ohms starti g at 25 ohms o wards.
Pwr. distr. End load is the last load used i the measureme ts;
R ca be set from 525 ohms to 5200 ohms, with steps of 25
ohms from 525 to 2500 ohms a d step of 100 ohms from 2500
ohms to 5200 ohms
Pwr, distr. Step Size is the load set with steps of 25, 50, 100,
a d 200 ohms
Press QUIT MENU (F ) to retur to the mai me u.
7. PRINT HEADER (F3). Writes a headi g for a ew test proto-
col.
8. REMOTE CONTR. (F4). E ables you to co trol the QA-ES
through a PC. Required software: PRO-Soft QA-ES.
5

3.5 Printout
Press PRINT HEADER (F3) before pri ti g out a page if you wa t
it to have a ew headi g. The QA-ES automatically pri ts out the
test results via the pri ter output after every measureme t. See ex-
ample below.
3.6 Foot Switch Output
A Foot Switch Output is activated by use of relays (K11), a d locat-
ed o the right side of the u it. This is used to trigger the foot switch
i put o the ESU bei g tested.
3.7 Main Switch On/Off
The QA-ES has to be tur ed off for at least 5 seco ds before tur i g
it o agai to allow the reset circuit to u load.
6