
CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Power input voltage: Check voltage across the terminal
7 – 14 with a multimeter.
3) Input: Calculate the calibration range and check that the
result is within the usable range.
Calibration Range [W] = Measuring Wattage
(VT Ratio) × (CT Ratio)
4) DC output: Check that the load resistance meets the de-
scribed specifications.
5) Integrating pulse output: Check the load connected to
the pulse output:
Open collector: 35V DC @100mA
Power photo MOSFET relay: 120V AC/DC @100mA
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case
of regular calibration, adjust the output as explained in the
following.
■HOW TO CALIBRATE THE DC OUTPUT SIGNAL
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 10 minutes.
1) ZERO: Apply 0% input and adjust output to 0%.
2) SPAN: Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 0% input.
4) When ZERO value is changed, repeat the above proce-
dure 1) – 3).
■HOW TO SET THE PULSE UNIT
The switches S1 through S5, VT and CT ratios, and the
transducer’s calibration input range all determine the rela-
tive value of one pulse, i.e. the number of kWh per pulse.
VT Ratio × CT Ratio × Calibration Range [kW]
2.777 Hz × 3600 [s] × 10 × Pulse Unit [kWh/pulse]
= 0.XXXX × 10–X
= 0. [a1] [a2] [a3] [a4] × 10–[a5]
a1 though a4: integer, 1 though 9
a5: integer, 0 through 6
[example] VT 3300V / 110V, CT 250A / 5A, Pulse Unit 10
[kWh/pulse], Calibration Range 1000 [W]
3300 / 110 × 250 / 5 × 1 [kW]
2.777 Hz × 3600 [s] × 10 × 10 [kWh/pulse]
= 0.1500 × 10–2
S1 = 1, S2 = 5, S3 = 0, S4 = 0, S5 = 2
MAINTENANCE
Regular calibration procedure is explained below:
■CALIBRATION
Warm up the unit for at least 10 minutes. Apply 0%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the “ADJUST-
MENT PROCEDURE” explained earlier.
LIGHTNING SURGE PROTECTION
We offer a series of lightning surge protector for protec-
tion against induced lightning surges. Please contact us to
choose appropriate models.
MUWT
EM-1926 Rev.2 P. 4 / 4
MG CO., LTD. www.mgco.jp
5-2-55 Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN