Mindman MFP01 Series User manual

Flow & Pressure Sensor
INSTRUCTION
MANUAL
MFP01 series
210107

II
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Contents
Product features.....................................................................................................................................................................1
Application..................................................................................................................................................................................3
Thermal mass ow sensor principles.................................................................................................................5
Order example..........................................................................................................................................................................5
Specications...........................................................................................................................................................................6
Output type.................................................................................................................................................................................7
Pressure loss characteristics.....................................................................................................................................8
Installation instructions...................................................................................................................................................9
● Piping ..................................................................................................................................................................................................9
● Mounting bracket.................................................................................................................................................................................9
● Wiring diagrams ................................................................................................................................................................................10
● Dimensions........................................................................................................................................................................................11
● Mounting accessories........................................................................................................................................................................12
How to use.................................................................................................................................................................................13
● Names and functions of individual parts ...........................................................................................................................................13
● Internal structure ...............................................................................................................................................................................14
● Functions instructions........................................................................................................................................................................15
Functions setting mode .................................................................................................................................................................15
Measurement mode ......................................................................................................................................................................16
● Operation setting instructions............................................................................................................................................................17
[F-01] OUT1 setting selection .......................................................................................................................................................17
[F-02] OUT2 setting selection .......................................................................................................................................................19
[F-03] LCD display color selection ................................................................................................................................................19
[F-04] Response time selection.....................................................................................................................................................19
[F-05] Display refresh time selection.............................................................................................................................................20
[F-06] Unit selection ......................................................................................................................................................................21
[F-07] Flow reference standard selection......................................................................................................................................22
[F-08] Display rotation ...................................................................................................................................................................22
[F-09] Accumulated value hold selection.......................................................................................................................................22
[F-10] Flow sensor display mode selection ...................................................................................................................................23
[F-91] Power-save mode selection................................................................................................................................................23
[F-92] External input selection.......................................................................................................................................................23
[F-93] Modbus RTU setting ...........................................................................................................................................................24
[F-94] Fine adjustment setting.......................................................................................................................................................25
[F-95] Forced output function ........................................................................................................................................................26
[F-99] Reset to the default setting .................................................................................................................................................26
Pressure zero adjustment function................................................................................................................................................25
Instantaneous ow zero adjustment function ................................................................................................................................25
Reset accumulated ow function ..................................................................................................................................................27
Peak value display ........................................................................................................................................................................27
Bottom value display .....................................................................................................................................................................27
Key lock / Unlock mode.................................................................................................................................................................27
Modbus RTU instruction...............................................................................................................................................28
Error code instruction.....................................................................................................................................................30
Caution for safety................................................................................................................................................................31

1
Flow Pressure
Indicator accuracy ± 3% F.S. ± 2% F.S.
Repeatability ± 1% F.S. ± 0.2% F.S.
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Product features
Cost Control
User-friendly
High Performance
High Precision
Cost Reduction
Traditional
2 × Sensor cost
2 × Power supply
2 × Plumbing costs & Time
2 × Wiring work cost & Time
2 × Location of installation
MFP01
Total cost & Time
Reduction
COST
DRY AIR, N2
RS-485 MODBUS
CONTROL

2
Flow rate (L/min)
MFP01-005
MFP01-010
MFP01-050
MFP01-100
MFP01-500
MFP01-101
MFP01-201
Digital display Switch output Analog output Accumulated pulse output
Instantaneous ow value
Accumulated ow value
Pressure value
NPN output
PNP output
Voltage output 1~5V
Current output 4~20 mA 50ms pulse output
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Product features
2-in-1 Design
Pressure and ow rate
simultaneous monitoring
Accumulated ow rate
display at a glance
Multiple output function
Wide range of ow rates
Flow display
Pressure display
Accumulated ow rate display
0 0.5 1 5 10 50 100 200

3
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Application
Application
01 Paint and coating ow monitoring
Air ow and pressure management of paint and
coating processes.
Application
02 Laser welding monitoring
Precision management of shielding gas,
ow rate and pressure.

4
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Application
Application
03 Suction operation
Detect the suction state of micro-
electronic parts by ow rate.
Application
05 Leakage detect
Container with gas and inspect
whether it leaks.
Application
04 Electronic part installation
Control the tension of gold wire for
bonding electronic parts.
Application
06 Air consumption
monitoring
Monitor air consumption of devices.

5
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Thermal mass ow sensor principles / Order example
MFP01 – 005 – 010 – R6
MP – A26
MODEL
PORT SIZE OUTPUT SPEC.
FLOW RATE
RANGE
010: 2 NPN output +
Analog output 1 ~ 5 V
011: 2 NPN output +
Analog output 4 ~ 20 mA
02 : 2 NPN output + RS485
030: 2 PNP output +
Analog output 1 ~ 5 V
031: 2 PNP output +
Analog output 4 ~ 20 mA
04 : 2 PNP output + RS485
005: 500 (ml/min)
010: 1000 (ml/min)
050: 5 (l/min)
100: 10 (l/min)
500: 50 (l/min)
101: 100 (l/min)
201: 200 (l/min)
Thermal mass wo sensor principles
Order example
Mounting accessories
Symmetric
temperature prole
(a) No ow
In the absence of ow, the heat
from the heater spreads evenly
left and right, so the temperature
distribution is like (a).
(b) Small ow
When ow begins, the inlet side
is cooled by the ow, the outlet
side is warmed by the heat of the
inlet side of the heater, and the
temperature distribution is like (b).
(c) Large ow
When the ow increases, it
becomes a distribution like (C).
Since the temperature distribution
before and after the heater is
proportional to the ow rate, the
ow rate can be determined from
the ratio.
Temperature
Sensor 1
Temperature
Sensor 1
Temperature
Sensor 1
Temperature
Sensor 2
Temperature
Sensor 2
Temperature
Sensor 2
Heating
Element
Heating
Element
Heating
Element
Skewed
temperature prole
Skewed
temperature prole
MODEL
R6: ø6 mm, for ow rate range
005, 010, 050, 100, 500.
R8: ø8 mm, for ow rate range
101, 201.
F1C: Rc1/8", with internal threads,
for ow rate range
005, 010, 050, 100, 500
F4C: Rc1/4", with internal threads,
for ow rate range 101, 201
A26: Mounting bracket
B4: Panel adapter
C4: Panel adapter + Front protective lid
OPTION PARTS
A26 C4B4

6
Model 005 010 050 100 500 101 201
Fluid Dry air, N2, Non-corrosive / Non-ammable gas
Sensor
element
Flow
Measured ow rate range
0 ~ 500 mL/min 0 ~ 1000 mL/min 0 ~ 5 L/min 0 ~ 10 L/min 0 ~ 50 L/min 0 ~ 100 L/min 0 ~ 200 L/min
Flow direction Unidirection
Pressure Rated pressure range -90 ~ 800 kPa
Display
Display 4 digital × 4 digital, 7 segment LCD display ( Red / Green / Orange )
Instant ow
rate
Display range 0 ~ 525 mL/min 0 ~ 1050 mL/min 0 ~ 5.25 L/min 0 ~ 10.50 L/min 0 ~ 52.5 L/min 0 ~ 105.0 L/min 0 ~ 210 L/min
Min. setting
scale
LPM 1 mL/min 0.01 L/min 0.1 L/min 1 L/min
CFM *1 0.01 ft3/min 0.1 ft3/min 1 ft3/min
Accumulated
Flow
Display range 99999999 mL 999999.99 L 9999999.9 L 99999999 L
Min. Setting Scale *1 1 mL 0.01 L 0.1 L 1 L
0.01 ft30.1 ft31 ft3
Pressure
Display
Display range -100 ~ 1000 kPa
Min. Setting
Scale
kPa 1
kgf/cm20.01
bar / psi 0.01 / 0.1
Accuracy
Flow
Guaranteed range 2 ~ 100 % F.S.
Indicator accuracy ± 3 % F.S. ± 1 digit *2
Analog output accuracy ± 5 % F.S. *2
Repeatability ± 1 % F.S. ± 1 digit *3
Linearity ± 3 % F.S. *3
Temp. characteristic ± 2 % F.S. ( 15 ~ 35 °C ) ; ± 5 % F.S. ( 0 ~ 15 °C, 35~ 50 °C ) ( compare with *3 )
Pressure characteristic ± 5 % F.S. ± 1 digit *4
Pressure
Guaranteed range 0 ~ 100 % F.S.
Indicator accuracy ± 2 % F.S. ± 1 digit *5
Analog output accuracy ± 2.5 % F.S. *5
Repeatability ± 0.2 % F.S. ± 1 digit *5
Linearity ± 1 % F.S. *5
Temp. characteristic ± 2 % F.S. ( compare with *5 )
Switch output
Switch output
2 NPN : open collector 2 outputs
Max. Load Current : 125 mA
Max. Supply Voltage : 28 V DC
Voltage Drop : ≤ 1.5 V
2 PNP : open collector 2 outputs
Max. Load Current : 125 mA
Max. Supply Voltage : 24 V DC
Voltage Drop : ≤ 1.5 V
Response
time
Flow 800 ms ( 50 ms, 80 ms, 120 ms, 200 ms, 400 ms, 1500 ms selectable )
Pressure 2.5 ms ( 25 ms, 100 ms, 250 ms, 500 ms, 1000 ms, 1500 ms selectable )
Output
mode
Flow Hysteresis Mode, Window Comparator Mode, Accumulated Output, Accumulated Pulse Output
Pressure One Point Set Mode, Hysteresis Mode, Window Comparator Mode
Hysteresis Adjustable
Output short circuit proterction Yes
Accumulated pulse output *1 5 mL/Pulse 10 mL/Pulse 0.05 L/Pulse 0.1 L/Pulse 0.5 L/Pulse 1 L/Pulse 2 L/Pulse
0.02 ft3/Pulse 0.04 ft3/Pulse 0.2 ft3/Pulse 0.4 ft3/Pulse 2 ft3/Pulse 4 ft3/Pulse 7 ft3/Pulse
Analog
output
Voltage output Voltage output range : 1 ~ 5 V *6 ; Output impedance : 1 KΩ
Current output Current output range : 4 ~ 20 mA *6 ; Load impedance : ≤ 300 Ω
Response time Pressure : ≤ 50 ms ; Flow : ≤ 100 ms
External input Non-voltage input, < 0.4 V, ≥ 30 ms
Communication interface RS-485 *7
Power Power supply voltage 12 ~ 24 V DC ± 10 %, Ripple ( P-P ) ≤ 10 %
Current consumption ≤ 50 mA
Environment
Withstand Pressure 1000 kPa
Enclosure IP40
Working Fluid Temp. 0 ~ 50 °C ( No condensation or freezing )
Ambient Temp. Range Operation : 0 ~ 50 °C ; Storage : -10 ~ 60 °C ( No condensation or freezing )
Ambient Humidity Range Operation / Storage : 35 ~ 85 % R.H. ( No condensation )
Insulation Resistance ≥ 50 MΩ ( 500 V DC, between case and lead wire )
Withstand Voltage 1000 V AC in 1-min ( between case and lead wire )
Vibration Total amplitude 1.5 mm or 10 G, 10 Hz ~ 55 Hz ~ 10 Hz scan for 1 minute, 2 hours each direction of X, Y and Z
Shock 100 m/s² ( 10 G ), 3 times each in direction of X, Y and Z
EMC IEC 61000-6-2, IEC 61000-6-4
Lead wire ø4 Oil-resistance cable ( PVC ) - 26 AWG ( 0.15 mm2 ) - 6 cores
Port
Size
R6, F1C
R8, F4C
Weight (with 2 meter lead wire)
Approx. 109.3 g ( ø6 port ) ; Approx. 112.7 g ( ø8 port ) ; Approx. 118 g ( Rc1/4" port ) ; Approx. 128.5 g ( Rc1/8" port )
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Specications
*1. CFM ( ft3/min*10-2 ) and ft3*10-2
*2. Condition: Inlet pressure: 300 kPa , Outlet pressure: 1 atmospheric pressure, 25 °C
*3. Condition: Outlet pressure: 1 atmospheric pressure, 25 °C
*4. -90 ~ 800 kPa, Outlet pressure: 1 atmospheric pressure, 25 °C
*5. Outlet ow rate = 0 L/min, 25 °C
*6. PWM output, corresponding to pressure sensor 0 ~ 1000 kPa
*7. This function only available for Output Specication -02 and -04

7
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Output type
Hysteresis mode Hysteresis mode
One point set mode One point set mode
Accumulated pulse output mode
NOTE
1. In case hysteresis is set at less than or equal to 2 digits,
switch output may chatter if input pressure uctuates near the
set point.
2. When using window comparator mode, the difference
between two set points must be greater than the xed
hysteresis, otherwise will cause the switch output to
malfunction.
Window comparator mode Window comparator mode
Accumulated output mode Accumulated output mode
Flow / pressure
FH-1(FH-2)
H-1(H-2)
FL-1(FL-2)
L-1(L-2)
ON
OFF
Flow / pressure
FH-1(FH-2)
H-1(H-2)
FL-1(FL-2)
L-1(L-2)
ON
OFF
Flow / pressure
FH-1(FH-2)
H-1(H-2)
FL-1(FL-2)
L-1(L-2)
ON
OFF
Hys
Hys
Flow / pressure
FH-1(FH-2)
H-1(H-2)
FL-1(FL-2)
L-1(L-2)
ON
OFF
Hys
Hys
Pressure
P-1(P-2)
ON
OFF
Hys
Pressure
P-1(P-2)
ON
OFF
Hys
Flow
AdH1 AdL1
AdH2 AdL2
0 L Time
ON
OFF
Flow
AdH1 AdL1
AdH2 AdL2
0 L Time
ON
OFF
Time
ON
OFF
50ms
Flow range 500mL 1000 mL 5 L 10 L 50 L 100 L 200 L
Pulse output rate 5 mL 10 mL 0.05 L 0.1 L 0.5 L 1 L 2 L
Normal open mode Normal close mode

8
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Pressure loss characteristics
MFP01 - 005 - -
MFP01 - 050 - -
MFP01 - 500 - -
Pressure loss (kPa) Pressure loss (kPa)Pressure loss (kPa)Pressure loss (kPa)
Pressure loss (kPa) Pressure loss (kPa)
Flow rate (l/min)
Flow rate (ml/min)
Flow rate (l/min)
Flow rate (l/min)
Flow rate (l/min)
Flow rate (ml/min)
5
4
3
2
1
0
0.006
0.005
0.004
0.003
0.002
0.001
0
0.08
0.07
0.06
0.05
0.04
0.03
0.02
0.01
0
20
18
16
14
12
10
8
6
4
2
0
12
10
8
6
4
2
0
0.018
0.016
0.014
0.012
0.010
0.008
0.006
0.004
0.002
0
0.1 MPa
0.1 MPa
0.1 MPa
0.1 MPa
0.1 MPa
0.2 MPa
0.3/0.35/0.4
MPa
0.5 MPa
0.2 MPa
0.2 MPa
0.2 MPa
0.2 MPa
0.2 MPa
0.3 MPa
0.3 MPa
0.3 MPa
0.3 MPa
0.3 MPa
0.4 MPa
0.4 MPa
0.4 MPa
0.4 MPa
0.4 MPa
0.5 MPa
0.5 MPa
0.5 MPa
0.5 MPa
0.5 MPa
0.35 MPa
0.35 MPa
0.35 MPa
0.35 MPa
0.35 MPa
0 10 20 30 40 50
0 100 200 300 400 500
0 1 2 3 4 5
0 50 100 150 200
0 20 40 60 80 100
0 200 400 600 800 1000
MFP01 - 201 - -
MFP01 - 010 - -
MFP01 - 100 - -
MFP01 - 101 - -
Pressure loss (kPa)
Flow rate (l/min)
0.25
0.20
0.15
0.10
0.05
00 1 2 3 4 5 6 7 8 9 10
0.3 MPa
0.4 MPa
0.5 MPa
0.35 MPa

9
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Installation instructions
Installation instructions
●Piping
Install the pipe by following the arrow indication that shows
the air ow direction on the product.
Use straight piping 8cm or longer to connect the Piping Port
(Inlet side).
If straight piping is not installed, the accuracy may vary by ±
2% F.S..
* Straight Piping: The pipe is without bending and the cross
sectional areas of the pipe keeps the same.
Blow the air to ush out the foreign matters, dust and etc.
before installing the pipe.
Uncleaned air may cause malfunction or damage to the
product.
Piping for the One-Touch Fitting, insert the tube rmly into the
tting and make sure it cannot be pulled out.
Also using the proper tube cutter is recommended to ensure
square edge tube.
●Mounting Bracket / Optional Parts
The LCD display may be difcult to see at certain angles.
The sensor can be installed horizontally or vertically, but
the ow rates
may change because of the installation way of the product
or piping.
The tightening torque for screws should be under 0.5 ±0.1
N.m.
* The tightening torque for bracket mounting screws should
be under 0.5 ± 0.1 N.m.
Flow direction
straight piping
8cm or longer
Piping Port
(Inlet side)
Piping Port
(Outlet side)
Horizontal mounting (by Through-Hole)
Bracket mounting
Mounting screw
Mounting screw
Mounting screw
P type
3×6L

10
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 Wiring diagrams
●Wiring diagrams
NPN Output, Analog Output and External Input
● NPN Output / Analog Voltage Output / External Input
● NPN Output / Analog Current Output / External Input
● NPN Output, RS-485 Modbus
PNP Output, Analog Output and External Input
● PNP Output / Analog Voltage Output / External Input
● PNP Output / Analog Current Output / External Input
● PNP Output, RS-485 Modbus
Brown (Power supply +)
1kΩ
Yellow (External input)
Load
Main Circuit
Load
Black (Output 1)
White (Output 2)
Blue (Power supply –)
Orange
(Analog output)
1kΩ
Brown (Power supply +)
Yellow (External input)
Load
Main Circuit
Load
Black (Output 1)
White (Output 2)
Blue (Power supply –)
Orange
(Analog output)
Brown (Power supply +)
Yellow (External input)
Main Circuit
Load
Load
Load
Black (Output 1)
White (Output 2)
Blue (Power supply –)
Orange
(Analog output)
Brown (Power supply +)
Main Circuit
Load
Load
Orange
RS-485 (B+)
Yellow RS-485 (A-)
Black (Output 1)
White (Output 2)
Blue (Power supply –)
Brown (Power supply +)
Main Circuit
Load
Load
Orange
RS-485 (B+)
Yellow RS-485 (A-)
Black (Output 1)
White (Output 2)
Blue (Power supply –)
Brown (Power supply +)
Yellow
(External input)
Main Circuit
Load
Load
Load
Black (Output 1)
White (Output 2)
Blue (Power supply –)
Orange (Analog output)
Pin. no. Line color Content Content (RS-485 Modbus)
1Brown Power supply (DC 12 to 24V DC) Power supply (DC 12 to 24V DC)
2Orange Analog voltage output: 1 ~ 5V
Analog current output: 4 ~ 20mA RS-485 (B+)
3Yellow External input RS-485 (A-)
4Black Output 1 (Max. load current: 125mA) Output 1 (Max. load current: 125mA)
5White Output 2 (Max. load current: 125mA) Output 2 (Max. load current: 125mA)
6Blue 0V (GND) 0V (GND)
1
2
3
4
5
6
NOTE
Procedure to wiring RS485 products : To prevent product damage due to short circuit,
MUST do RS485 line conneciton BEFORE power line connection.

11
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
8
MFP01 Dimensions
Internal Thread
Rc1/4”, Rc1/8”
Quick tting ø6, ø8
2×ø2.5×5.5 depth
2×ø2.5×5.5 depth
36.7
36.7
9.5
9.5
18
28
28
18
55
55
24
24
3
38
50.250.2 1717
5252
A
B
Code
Tube I.D. A (mm)
ø8 92.2
ø6 83
Code
Thread B (mm)
Rc1/8" 76.2
Rc1/4" 76.2
Port size
F1C (Rc1/8")
F4C (Rc1/4")
Port size
R6 (ø6)
R8 (ø8)

12
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
33.6
26
2-ø3.5
4×3.5
4×6
18
8
30
28
5
67.3
27 6.2 27 49.5
32.6
22
3.3
65
21.8
25.3
5.9
258.5
59.8
40
MFP01 Mounting accessories
Mounting bracket
Panel type
MP-A26
Front protective lid Panel adapter Panel adapter Panel adapter
Front protective lid + Panel adapter MP-C4
Panel thickness 1~4.5
Panel cut dimensions
Panel adapter MP-B4
22.5±0.3
34 or over
60.5±0.3
4×R3 or less
Mounting screw
P type 3×6

13
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 How to use –Names and functions of individual parts
Through-Hole
Piping Port
(OUT side)
Piping Port
(IN side)
Connector
Output 2 Indicator
Output 1 Indicator
Analog Signal Indicator
Pressure symbol
Pressure display
Pressure Unit
Display Section
Flow Unit
Display Section
Flow Symbol
Analog Signal Indicator
Output 1 Indicator
Output 2 Indicator
Flow Display
Setting button
3 color LCD display
Button
8.8.8.8 8.8.8.8
kgf/cm3 bar psi kPa Pulse l 3/min mL3/min
Ag OUT1 OUT2 Ag OUT1 OUT2
Panel description
Body description
●Names and Functions of Individual Parts
How to use

14
MFP01 How to use –Internal structure
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
No. Part name (Material)
1Case (PC + ABS)
2Electronic circuit board
3Sensor (SiC)
4Liquid crystal cover (PC)
5Liquid crystal
6Button (Silicone)
7Port Filter (Stainless Steel)
No. Part name (Material)
8Resin Body (PBT)
9O-ring (Viton)
10 O-ring (NBR)
11 Quick-Fitting
12 O-ring (Viton)
13 Internal thread (SUS303)
Port size
F1C (Rc1/8")
F4C (Rc1/4")
Port size
R6 (ø6)
R8 (ø8)
1
7 8 9 11
2 3 4 5 6
10
1
7 8 9 13
2 3 4 5 6
12

15
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 How to use –Functions instructions
Function
code Item Default setting Explanation
F-01
oUt1 OUT1 setting oUt1
Select output 1 corresponding to
ow sensor or pressure sensor.
Set the ow rate or pressure
value to switch ON / OFF.
oUt1 OUT1 sensor correspondence FLoy
FLoy OUT1 output mode KY5
oUt1 OUT1 output type no
FL-1 OUT1 set value input
50 % of maximum measured ow rate
005 : 250 mL/min, 010 : 500 mL/min
050 : 2.50 L/min, 100 : 5.00 L/min
500 : 25.0 L/min, 101 : 50.0 L/min
201 : 100 L/min
FK-1 OUT1 set value input
60 % of maximum measured ow rate
005 : 300 mL/min, 010 : 600 mL/min
050 : 3.00 L/min, 100 : 6.00 L/min
500 : 30.0 L/min, 101 : 60.0 L/min
201 : 120 L/min
F-02
oUt2 OUT2 setting oUt2
Select output 2 corresponding to
ow sensor or pressure sensor.
Set the ow rate or pressure
value to switch ON / OFF.
oUt2 OUT2 sensor correspondence FLoy
FLoy OUT2 output mode KY5
oUt2 OUT2 output type no
FL-2 OUT2 set value input
50 % of maximum measured ow rate
005 : 250 mL/min, 010 : 500 mL/min
050 : 2.50 L/min, 100 : 5.00 L/min
500 : 25.0 L/min, 101 : 50.0 L/min
201 : 100 L/min
FK-2 OUT2 set value input
60 % of maximum measured ow rate
005 : 300 mL/min, 010 : 600 mL/min
050 : 3.00 L/min, 100 : 6.00 L/min
500 : 30.0 L/min, 101 : 60.0 L/min
201 : 120 L/min
F-03
CLor LCD display setting CLor
Select back light color and
display mode.
d15p LCD display corresponding to output oUt1
d15p LCD display color selection 5oG
F-04
rE5P Response time selection rE5P Select the response time for
analog output.
Pressure sensor: 2.5 ms ~ 1500 ms
Flow sensor: 50 ms ~ 1500 ms
rE5P Flow / Pressure sensor selection FLoy
FLoy Response time selection 800 (ms)
F-05
UPdR Display refreshtime selection UPdR
Display refresh cycle can be set
in 200 ms, 500 ms or 1000 ms.
UPdR Flow / Pressure sensor selection FLoy
UPdR Display refresh time of ow sensor
selection 500 (ms)
F-06
Un1t Unit selection Un1t
Select the UNIT of pressure /
ow sensor.
FLoy Flow unit selection LPA
PrE5 Pressure unit selection PR
Function setting mode

16
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
MFP01 How to use –Functions instructions
Function
code Item Default setting Explanation
F-07 rEFE Flow reference standard selection rEFE Select the ow value is shown
under standard (ANR) or normal
condition ( NOR ).
rEFE Flow reference standard selection Anr
F-08 RnG Analog output selection AnG Select the analog corresponding to
pressure or ow sensor.
FLoy Analog output selection FLoy
F-09 EEPr Accumulated value hold selection EEPr To save the last accumulated ow
value every 2 or 5 minutes.
EEPr Accumulated value hold selection oFF
F-10 d1s Flow sensor display mode selection d1s Select to display Instantaneous
Flow or Accumulated Flow Mode.
dsp Flow sensor display mode selection 1nS
F-91 ECo Power-Save mode selection ECo Select if turn on powersave
mode to reduce power consumption.
ECo Power-Save mode selection no
F-92 1nP External input selection 1nP Select for Accumulated ow
rate zero clear, Auto-Shift or Auto-
Shift zero.
1nP External input selection r_r
F-93
AbUS Modbus RTU setting AbUS
Set ID number, baud rate and
transmission format.
1d ID number setting 1
rAt Baud rate setting 96 (9600 Bd)
For Transmission format setting n81
trR Communications protocol setting rtU
F-94 F1nE Fine adjustment setting F1nE The displayed value can be
adjusted slightly.
F1nE Fine adjustment setting oFF
F-95
FoUt Forced output function FoUt
To turn the analog ON / OFF
forcibly.
oUt1 Forced output function oFF
oUt2 Forced output function oFF
F-99 rESt Reset to the default setting rESt
Return to the factory default setting.
rSt Reset to the default setting oFF
Item Explanation
Pressure display Display pressure value.
Flow display Display instantaneous ow rate.
Accumulated ow rate display Display accumulated ow rate.
Pressure zero setting The displayed pressure value can be adjusted to "0".
Instantaneous ow rate zero setting The displayed instantaneous ow rate value can be adjusted to "0".
Accumulated ow rate zero clear The accumulated ow rate can be set to "0".
Peak value display The maximum pressure or instantaneous ow can be detected when the power is supplied for
a period.
Bottom value display The minimum pressure or instantaneous ow can be detected when the power is supplied for
a period.
Key lock / unlock mode To prevent errors occurring due to unintentional changes of the set values.
Measurement mode

17
●Operation setting instructions
At Measurement Mode, press button for more than 3 sec.
to display [F-01]. Press or button to select other setting
functions.
Press for 3 sec. at Function Setting Mode to return to
Measurement Mode.
●[F-01 ]OUT1 setting selection
Setting corresponding sensor and operating mode of OUT1.
Flow sensor setting
Press or button at Function Setting Mode to display
[ F-01 ] [ oUt1 ].
MFP01 How to use –Operation setting instructions
Selected
Output
Sensor
Selected
sensor
Selected
mode
Hysteresis
mode
Window
comparator
mode
Accumulated
output mode
Accumulated
pulse
output mode
Flow sensor Pressure sensor
Selected
Mode
Fixed
Hysteresis
Mode
NO mode NC mode
Set
Value
Set
Value
Selected
Output
Selected
Mode
Measurement Mode
Enter in each function setting
Press button for
more than 3 sec.
Press button
Press button
Press button
Press button
Press button to return to
Function Selection Mode
Press button
Press button
Function selection mode
Sensor selection
Output mode setting
Set Value Setting
Fixed Hysteresis Setting
Output Type Setting
Press or button to select
ow sensor of OUT1.
Press or button to select
output mode of OUT1.
Out1 set value setting
Press or button to adjust the
set value.
Fixed hysteresis setting
Press or button to adjust
xed hysteresis value.
Hysteresis mode [ HYS ]: [ FL-1 ], [ FH-1 ]
Window comparator mode [ y1n ]: [ FL-1 ], [ FH-1 ]
Accumlated output mode [ ADD ]: [ AdL1 ], [ AdH1 ]
OUT1 type setting
Press or button to select
OUT1 type.
NOTE
Type setting will not display when accumulated pulse output mode
is set.
NOTE
Set value setting will not display when accumulated pulse output is set.
NOTE
Fixed hysteresis setting will not display when hysteresis mode,
accumulated output Mode and accumulated pulse output mode is set.
NOTE The Accumulated Pulse Output Mode can only be set in OUT1,
and OUT2 does not have this setting.
f-01 oUt1
L/minkPa
Ag
oUt1 FLoY
L/minkPa
Ag
oUt1 no
L/minkPa
Ag
FL-1 05.00
L/minkPa
Ag
HYS 0.03
L/minkPa
Ag
FL-1 oUt1
L/minkPa
Ag
oUt1 nC
L/minkPa
Ag
oUt1 no
L/minkPa
Ag
FLoy HYS
L/minkPa
Ag
oUt1 PrES
L/minkPa
Ag
oUt1 FLoy
L/minkPa
Ag
FLoy ADD
L/minkPa
Ag
FLoy PLS
L/minkPa
Ag
FLoy Yjn
L/minkPa
Ag
FLoy HYS
L/minkPa
Ag
f-02 oUt2
L/minkPa
Ag
f-99 rE5t
L/minkPa
Ag
0 0.00
L/minkPa
Ag
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )

18
MFP01 How to use –Operation setting instructions
●[F-01 ]OUT1 setting selection
Setting corresponding sensor and operating mode of OUT1.
Pressure sensor setting
Press or button at Function Setting Mode to display
[ F-01 ] [ oUt1 ].
Selected
Output
Sensor
Selected
Mode
Sensor
Hysteresis
mode
Window
comparator
mode
One point
set mode
Flow sensor Pressure sensor
Mode
Fixed
Hysteresis
Mode
NO Mode NC Mode
Selected
Set Value
Set
Value
OutputSelected
Mode
Press button
Press button
Press button
Press button
Press button to return to
Function Selection Mode
Press button
Sensor Selection
Output Mode Setting
Set Value Setting
Fixed Hysteresis Setting
Output Type Setting
Press or button to select
pressure sensor of OUT1.
Press or button to select
output mode of OUT1.
Out1 set value setting
Press or button to adjust
the set value.
Fixed hysteresis setting
Press or button to adjust
xed hysteresis value.
Hysteresis mode [ HYS ]: [ L-1 ], [ H-1 ]
Window comparator mode [ y1n ]: [ L-1 ], [ H-1 ]
One point set mode [ oPS ]: [ P-1 ]
OUT1 type setting
Press or button to select
OUT1 type.
NOTE
Fixed hysteresis setting will not display when hysteresis mode is set.
oUt1 FLoy
L/minkPa
Ag
no oUt1
L/minkPa
Ag
500 L-1
L/minkPa
Ag
3 HYS
L/minkPa
Ag
FL-1 oUt1
L/minkPa
Ag
nC oUt1
L/minkPa
Ag
no oUt1
L/minkPa
Ag
KY5 PrES
L/minkPa
Ag
oUt1 PrES
L/minkPa
Ag
oUt1 FLoy
L/minkPa
Ag
yjn PrES
L/minkPa
Ag
oPS PrES
L/minkPa
Ag
KY5 PrES
L/minkPa
Ag
FLOW & PRESSURE SENSOR ( INSTRUCTION MANUAL )
This manual suits for next models
7
Table of contents
Other Mindman Accessories manuals