Mindman MPG-60 Series User manual

Unit:mm
We reserve the right to change the specification without prior notice.
PR-0246D 2021/11 Printing
9
6
15.5
22.5
20
20
30
30
37
21.8
Rated pressure range
Withstand pressure
Applicable fluid
kgf/cm2
bar
psi
Pressure
resolution
Ambient humidity range
Vibration
Shock
Temperature characteristic
Weight
MPa
A. SPECIFICATIONS
TYPE
Battery
Battery life
Low-power indicator
Battery replaceable
Turn-on interval
Sampling rate
Repeatability
Display
Indicator accuracy
B. ORDERING INFORMATION
F1 : R1/8", M5
F2 : NPT1/8", #10-32 UNF
F3 : G1/8"(BSPP), M5
F4 : R1/4", M5
F5 : NPT1/4", #10-32 UNF
F6 : G1/4”(BSPP), M5
Optional Parts
C. PANEL DESCRIPTION
Battery Meter
Pressure Display
Setting Button
Pressure Unit
Display Section
D. DIMENSIONS
E. BATTERY STATUS / INSTALLATION
F1~F3:Hex flat:12
F4~F6:Hex flat:14
F1 : R1/8", M5
F2 : NPT1/8", #10-32 UNF
F3 : G1/8"(BSPP), M5
F4 : R1/4", M5
F5 : NPT1/4", #10-32 UNF
F6 : G1/4”(BSPP), M5
1
4
78
MPa
MPa
6"+"
"-"
1.5MPa
0.000 ~ 1.000MPa
Filtered air, incombustible and non-corrosive gases
0.01
0.01
0.1
0.001
Operation/Storage: 35 ~ 85% RH ( No condensation)
Total amplitude 1.5mm or 10G, 10Hz-55Hz-10Hz scan
for 1 minute, two hours each direction of X, Y and Z
100m/s2(10G), 3 times each in direction of X, Y and Z
Approx. 40g
Yes
Yes
2 Hz (2 times/sec.)
3 ½ digital, 7 segment LCD display ( Black )
±2% F.S. ±1 digit (ambient temperature: 25 ±3 °C)
Operation: 0 ~ 50 ºC, storage: -10 ~ 60 ºC
( No condensation or freezing)
±2% F.S. of detected pressure (25°C) at temp.
range of 0~50°C
Ambient temp. range
Enclosure
Environment
*1. When the pressure is -0.1MPa, LCD displays MPa.
*2. Air tube must be installed to maintain IP65. (See section K)
5
2 3
MPG-60 Series
For your safety, please read the following before using.
1
2
3
4
5
6
FOLLOW THE STEPS BELOW
●Using a flat screwdriver size 1.4 ~ 3.0 mm insert into the back of the
pressure gage as pictured. (See figure )
● Turn the screwdriver horizontally 90° toseparate the case and the base.
Repeat this process again for the other side. (See figure )
● Carefully remove the outer cover (See figure ), and follow the diagrams
below to replace battery and reassembly. Polarity must be observed,
incorrect battery installation will cause malfunction.
2
3
4
(Positive)(Vacuum)
MPG-60V- MPG-60P-
MPG-60VL- MPG-60PL-
mmHg
CR 2032 lithium
IP65 *2
Pressure Range Pressure Port
V : Vacuum(10 ~ -101kPa)
P : Positive (-0.100 ~ 1.000MPa)
0 ~ -101kPa
-0.100 ~ 1.000MPa *1
10 ~ -101kPa
Display pressure range
300kPa
kPa
-
-
0.01
0.1
1
1
-
-
Display turn off after 60 sec.
(Vacuum) (Positive)
psi bar mmHg kPa psi bar kgf MPa
Port size F1:R1/8”, M5; F2:NPT1/8”, #10-32 UNF; F3:G1/8”(BSPP), M5
F4:R1/4”, M5; F5:NPT1/4”, #10-32 UNF; F6:G1/4”(BSPP), M5
3 years
(5 times/day)
3 years
(5 times/day)
1 year
(5 times/day)
1 year
(5 times/day)
Pressure unit psi, bar, kgf/cm2and MPa
user selectable
psi, bar, mmHg and kPa
user selectable
Back light Yes YesNo No
Back Light
Blank : Back light unavailable
L : Back light available
Do not use corrosive or flammable gas or liquid with this product.
Please use within the rated pressure range. Do not apply pressure
beyond recommended maximum withstand pressure, permanent
damage to the pressure gage may occur.
Do not drop, hit or allow excessive shock. Even if pressure gage
body appears undamaged, internal components may be broken and
can cause malfunction.
Do not use in environment containing steam or oil vapor.
This product is not explosion-proof rated. Do not use in atmosphere
containing flammable or explosive gases.
Using filter can extend pressure gage’s service life.
MP-A5 : Mounting bracket (BT-5)
MP-A6 : Mounting bracket (BT-6)
MP-B2 : Panel adapter (PA-C)
MP-C2 : Panel adapter + Front protective lid (PA-D)
±0.2% F.S. ±1digit
±1% F.S. ±1digit
MPG-60PL-F1
Ø 2mm MAX
31×31 mm
+0
- 0.4
MP-A6
1
2
M3x0.5
M3x0.5
Mounting bracket
MP-A5
Mounting bracket
MP-A6
Mounting bracket
Panel mounting
Panel adapter
Panel adapter
Front protective lid
Unit:mm
F. OPTIONAL PARTS DIMENSIONS
J. ZERO POINT RESET
Under measure mode, press button
for 2 ~ 5 seconds.
I. PRESSURE UNIT SETTING MODE
Press button for more than 5seconds.
Measure mode
Use button to set desired pressure unit.
Unit setting
Measure mode
MPa
MPa
MPa
Measure mode
MPa
Measure mode
MPa
Zero point setting
MPa
kgf
bar
psi
MPa
MPa
G. PRESSURE UNIT CONVERSION TABLE
K. DUST AND SPLASH ENVIRONMENT
To maintain IP65, please insert an air
tube (O.D.: 4mm) into atmospheric
release port and pipe the other end away
from dust / spattering liquid environment.
Use proper air tube dimension of
O.D.: 4mm, I.D.: 2.5mm, and the length
is depending on the user’s environment. Atmospheric release port
●
●
Air tube
Do not press button for 3 seconds to return to
measure mode.
L. ERROR CODE INSTRUCTION
Residual
pressure error
During zero reset, ambient
pressure is over ±3% F.S.
Change input pressure to ambient
pressure and perform zero reset again.
Applied
pressure error
Supply pressure is exceeds the
upper limit of pressure setting. Adjust the pressure within operating
pressure range.
Error Type
Error code
Error Condition Troubleshooting
Battery error
When battery level is under
2.40V, the battery symbol
and will flash 4 times
then enter power save mode.
Replace battery.
From To
1 kPa
1 MPa
6.895 0.006895 0.07031 51.7157 10.06895
0.010000
10
0.145038
145.038
7.500616
7500.616
0.010197
10.197
0.001000
1
1
1000
100.0000 0.100000 1.01972 750.062 14.5038 1
0.98066514.2233735.5591
0.0980665
98.0665
0.00133320.01933610.00135950.0001330.13332
1 kgf/cm
2
1 mmHg
1 psi
1 bar
kPa MPa
kgf/cm
2
mmHg psi bar
H. POWER-SAVE MODE
Measure mode Power-Save mode
MPa
MODEL:MPG-60P/MPG-60PL
kPa
mmHg
bar
psi
MODEL:MPG-60V/MPG-60VL
Supply pressure is exceeds the
lower limit of pressure setting.
◎The pressure gage will enter power saving mode if the button is not
pressed within 60 sec.
◎In measure mode, press the button to extend display for 120 sec.
no operation for 60 or 120 sec.
Front protective lid
Release the button when “ ” is shown.
Panel adapter
MP-A5

This manual suits for next models

4

Other Mindman Measuring Instrument manuals

Mindman MP30 Series User manual

Mindman

Mindman MP30 Series User manual

Mindman MPG-60 Series User manual

Mindman

Mindman MPG-60 Series User manual

Popular Measuring Instrument manuals by other brands

ABB Navigator 550 quick start guide

ABB

ABB Navigator 550 quick start guide

Fluke T90 instruction sheet

Fluke

Fluke T90 instruction sheet

Omega DFG207 user guide

Omega

Omega DFG207 user guide

Extech Instruments 407790 user manual

Extech Instruments

Extech Instruments 407790 user manual

ECOTEST BDBG-09S operating manual

ECOTEST

ECOTEST BDBG-09S operating manual

CWP Accurate pH 3 operating instructions

CWP

CWP Accurate pH 3 operating instructions

Extech Instruments VB400 user guide

Extech Instruments

Extech Instruments VB400 user guide

Thermo Scientific TSQ 9610 Hardware manual

Thermo Scientific

Thermo Scientific TSQ 9610 Hardware manual

Nagano Keiki GC31 instruction manual

Nagano Keiki

Nagano Keiki GC31 instruction manual

FILL-RITE 901 Owner's operation & safety manual

FILL-RITE

FILL-RITE 901 Owner's operation & safety manual

TFA 30.5046 instruction manual

TFA

TFA 30.5046 instruction manual

Lovibond Tintometer SensoDirect Con110 instruction manual

Lovibond

Lovibond Tintometer SensoDirect Con110 instruction manual

TECALEMIT FMT 3 operating manual

TECALEMIT

TECALEMIT FMT 3 operating manual

Fluke 771 instruction sheet

Fluke

Fluke 771 instruction sheet

ADA INSTRUMENTS AngleRuler operating manual

ADA INSTRUMENTS

ADA INSTRUMENTS AngleRuler operating manual

MOULDPRO Flosense 3.0 Installation quick guide

MOULDPRO

MOULDPRO Flosense 3.0 Installation quick guide

NAVIS WR3-B manual

NAVIS

NAVIS WR3-B manual

Triax MCT 049B user manual

Triax

Triax MCT 049B user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.