Mirion Technologies ACCURAD PRD User manual

1
QUICKSTART GUIDE
PERSONAL RADIATION DETECTOR

2
INTRODUCTION

3
The AccuRad™ PRD is a Personal
Radiation Detector (PRD) designed
for law enforcement, fire and rescue,
and other emergency responders
to detect and interdict nuclear and
radioactive materials. It also provides
dose measurement and alarming
capabilities for event response.

4
Exceptional Radiological Performance
The AccuRad PRD features accurate and wide dose
rate measurement from background to 1000 Rem/h;
mitigation of false alarms with Varying Background
Suppression (VBS). Cumulative dose measurements
are in compliance with industry dosimetry standards.
Smooth User Experience
Dual display screens and large buttons optimize
operation for belt-worn or handheld applications;
the AccuRad PRD quickly alerts the operator when
a threat is detected through eective loud audible,
attentive vibration, clear visual display and bright LED
alarm indicators; one-touch alarm acknowledgment;
automatic data recording.
Radar Indication Leading Straight
to the Source
An intuitive radar screen provides directional indica-
tion to help localize a thread in a wide or crowded
environment.
Discreet Mode for Covert Operations
Discreet mode silences alarms for stealth applica-
tions. USB-C headphone connectivity and silent
vibrations add a greater level of stealth operation.

5
Robust Protection
The AccuRad PRD is resistant to dust, sand and
moisture, drop tested at 1.5 m, and rated IP-67.
The PRD body has superior protection providing
shock resistance and easy handling. The AccuRad
PRD can be used in temperatures from -4 °F (-20 °C)
to 140 °F (60 °C).
>900 Hours Continuous Operation
The AccuRad PRD has outstanding battery life and is
powered by two commercially available AA batteries.
Simplified Data Transmission via
Near Field Communication (NFC) and
Bluetooth Low Energy (BLE)
Smartphone application (Android, iOS) enables
reachback, remote display and alarms,
personalization of AccuRad PRDs by batch.

6
A. Front Display
B. Back Button
C. Enter Button
D. Arrow Up
E. Arrow Down
ACCURAD PRD OVERVIEW
A
C
D
E
B
A
C
F
G
H
I
J
L
K
M
D
E
B

7
A
C
F
G
H
I
J
L
K
M
D
E
B
F. Battery Cover Tab
G. Battery Compartment Cover
H. Belt Clip
I. Power Button
J. USB-C Port
K. USB LED Indicator
L. Alarm LED Indicator
M. Top Display

8
Installing Batteries
2.
GETTING STARTED
STEP 1
1. Lift and left
quarter turn of
battery cover
tab
2. Remove
battery cover
STEP 3
1. Install battery
cover
2. Turn the
battery cover
ring quarter
turn clockwise
to lock
STEP 2
1. Replace the two (2) AA
alkaline batteries and
install in correct orientation
3.
1.

9
1. Press the
power button
1. HOLD the
power button
3. Background
update
sequence
2. Start-up with
AccuRad PRD
details and self
test sequence
2. Wait for the
end of the
countdown
(3-2-1-OFF)
4. Measurement
data is
displayed and
ready to go!
Power ON The AccuRad PRD
Power OFF The AccuRad PRD
96%
96%
96%
96%

10
Correct position to wear the AccuRad PRD:
USING THE ACCURAD PRD
Top Display
Using the PRD in Discreet Mode
The Discreet mode can be used when the operator does not want the PRD alarms
to be visible to others. By default, the audible and visual alarms (LEDs) in Discreet
mode are deactivated while the vibration and display messages are active.
Wearing the PRD
The reference direction of the measurement is orthogonal to the front display
to guarantee an optimal measurement of both detectors when the AccuRad
PRD is carried. The reference direction of the measurement and the distance
measurement have an impact of the quality measurement.
STEP 1
Press the enter (O)
button to enter the
menu
STEP 2
Use the arrow buttons
to select ‘Yes’
STEP 3
The PRD is now in
Discreet Mode
Note: the speaker icon
indicates the sound is
muted
96%
96%
96%

11
Using the PRD in Detection Mode
STEP 1
Wearing the PRD in
Detection Mode
Dose rate is automatically
provided on both
displays
STEP 2
PRD alarms when it
detects a source
The PRD activates the
audible buzzer and
vibration
Investigate with
Caution
Move Away and
Investigate with
Caution
(Note: actions above are suggestions only and should not
replace operator procedures and CONOPS)
Move Away!
STEP 3
Press any button to
acknowledge the alarm
Respond according to
the alarm level
DANGERHIGH ALARMLOW ALARM

12
Enabling Search Mode
STEP 1
Press enter (O) button
to access the menu
STEP 1
Press either arrow
on side of device to
enable Search mode
STEP 2
Press the enter
(O) button again to
begin Search mode
STEP 3
AccuRad PRD
assists with locating
the source
(See next page)
STEP 2
AccuRad PRD
assists with locating
the source
(See next page)
Search Mode via Menu
Enable Search Mode with Trend
96%
96%
96%
96%
96%

13
Using the Radar in Search Mode
STEP 1
Turn Left or Right
to build the radar sectors.
STEP 3
Go in the direction of the black sectors and arrows to localize the
source.
STEP 2
Sectors indicate source intensity
versus direction.
Black sectors/arrows indicate the
direction.

14
Exceptional
Radiological
Performance
NFC & BLE
Data
Transmission
Robust
Protection
Smooth User
Experience
Directional
Source
Localization
>900 Hour
Continuous
Operation
Discreet Mode for
Covert Operations

15
“I keep this thing on me 24/7
when I’m working...It’s like carrying
the gun; you know it’s there, but you
don’t have to think about it.”
SGT. TIM PRICE
COBB COUNTY BOMB SQUAD

16
Dose Equivalent
Quantity
• Hp(10) – Personal dose equivalent rate
Units • Sv, rem, unit-less (0-9), cps (search mode)
Detectors
• CsI(Tl) scintillation detector with temperature
compensated SiPM
• Sensitive volume: 10 mm (0.4 in.) diameter by
35 mm (1.4 in.) long
• Silicon diode for integrated dose and high dose rate
Alarm Response
And False Alarm
Rate
• According to ANSI/IEC standards*
• Detects 50 μR/h (0.05 µSv/h) increase within 2 seconds
(Am-241, Cs-137, Co-60)
False Alarm Rate
• Less than two alarms per 8 hours in stable background
(typically zero alarm per 24 hours)
• VBS mitigates or suppresses alarms due to varying
background
Type Of Radiation • Photons (Gamma, X)
Energy Response • 25 keV to 3 MeV
Count Rate
(Search Mode)
• 0 cps to 160 000 cps
Dose Rate
• ANSI N42.32: 1 µrem/h (0.01 µSv/h) - 10 rem/h (1 mSv/h)
• ANSI N42.49A: 1 µrem/h (0.01 µSv/h) - 1000 rem/h (10 Sv/h)
• Accuracy: ±20%
Dose • Accuracy: ±20%
Stabilization • Warm-up time: 35 seconds
• No source required
*IEC 62401 and ANSI N42.32
TECHNICAL
Nuclear

17
Audible Sound • >85 dB(A) at 30 cm
Batteries
• AA (LR6) 1.5 V alkaline
• Lithium 1.5 V batteries (non-rechargeable)
• NiMH 1.2 V batteries (rechargeable)
Power Supply • USB-C (recharges NiMH)
Please inquire on accurad.mirion.com for test reports and certificates of
conformity (ANSI N42.32, IEC 62401, CE).
Certificate provided with purchase of the AccuRad PRD.
*IEC 62401 and ANSI N42.32
Physical, Electrical & Environmental
Functional
Dimensions • Weight: 200 g (7 oz), including batteries and clip
• Size: 4.25 x 2.4 x 1.4 in. (108 x 61 x 36 mm), without clip
Power
• Two AA alkaline batteries for more than 900 hours of
continuous operation
• Tool-less battery cover
Ingress
Protection
• IP-67 - Dust and 1 m water immersion
Temperature
• -4 °F to 140 °F (-20 °C to 60 °C)
• Temperature shocks and relative humidity according to
ANSI/IEC standards*
Drop • 1.5 m (4 ft 11 in.) on concrete
• Vibration, impact according to ANSI/IEC standards*
Find more information at:
ACCURAD.MIRION.COM

Risk of electric shock. Any work on powered equipment must be performed
by qualified and authorized personnel.
Sources with ionizing radiation present are dangerous for the user if proper
protective measures are not strictly applied. Although our equipment is built
in compliance with the strictest safety standards, ionizing radiation sources
present a danger when the user is not qualified or not suciently informed of
the danger.
Radioactive sources must be handled exclusively by qualified and authorized
personnel. Consequently, all precautions must be taken to prevent any non-
authorized or non-qualified personnel from using this equipment, potentially
endangering others and themselves.
Prior to any handling, those qualified and authorized to use this equipment
shall acquire information on the protective measures set forth by the national
standards in force.
Abandonment or destruction of equipment containing a radioactive source
is forbidden. If no longer required, the user must inform Mirion Technologies
who will arrange to take the source back (according to the contract) and
establish a certificate stating that the source has been taken back. In the
event that the source is lost or stolen, the user must inform the appropriate
authorities as soon as possible.
DISPOSAL: At the end of the product’s useful life, please dispose of it
according to local legislation. (Directive 2002/96/EU of the European
parliament and of the council of January 2003 on waste electrical and
electronic equipment (WEEE).) If you have questions about disposal in Europe,
CAUTION: The AccuRad PRD is not certified for explosive atmospheres.
CAUTION: When inserting the batteries, make sure to respect the polarities.
Always use batteries of the types given on page 17.
CAUTION: Earphones. Do not use at high volume during extended periods.
High acoustic pressure. Risk of hearing impairment.
SAFETY INFORMATION

In partnership with our customers in nuclear power plants, military
and civil defense agencies, hospitals, universities, national labs, and
other specialized industries, Mirion Technologies strives to deliver
cutting-edge products and services that constantly evolve based on
the changing needs of our customers.
Combining state-of-the-art technology with exceptional customer
service, Mirion is dedicated to providing an unmatched experience in
radiation detection and instrumentation.
ABOUT MIRION TECHNOLOGIES
Contact Us
Mirion Technologies (MGPI), Inc.
Detection & Measurement Division
5000 Highlands Parkway, Ste 150
Smyrna, GA 30082
Phone: 1.770.432.2744
Inquiries: accurad.mirion.com
Mirion Technologies (MGPI), S.A.S
174 Route d’Eyguières
FR 13113 Lamanon
France
Phone: +33 (0) 4 90 59 59 59
Inquiries: accurad.mirion.com
Copyright ©2019 Mirion Technologies, Inc. or its aliates. All rights reserved. Mirion, the Mirion
logo, and other trade names of Mirion products listed herein are registered trademarks or
trademarks of Mirion Technologies, Inc. or its aliates in the United States and other countries.
Third party trademarks mentioned are the property of their respective owners.
The publication, translation or reproduction of this document, either in part or in its entirety, is
not allowed without direct written consent from Mirion Technologies.
PRD-44 12/ 2019

20
Learn More:
ACCURAD.MIRION.COM
Other manuals for ACCURAD PRD
2
Table of contents
Other Mirion Technologies Security Sensor manuals
Popular Security Sensor manuals by other brands

Fike
Fike FIK-W-PHOTO Installation and maintenance instructions

SICK
SICK KTX-WB9114125A Series manual

Airthings
Airthings Wave operating manual

PCB Piezotronics
PCB Piezotronics ICP 109B01 Installation and operating manual

WITTUR
WITTUR USG-25P operating instructions

Inovonics
Inovonics EN1751 installation instructions