4. PARAMETERS AND CALIBRATION
4.1 Entering setting mode
To set the parameters, turn on the weighing scales and hold the TARE key during self-
checking, until the display show “F1 CAL”.
4.2 Calibration
After entering settings mode, and setting the parameter. The display will show "F1 CAL".
Press the ZERO key, display shows “UNLD”, empty the platter, after the “STABLE”indicator
turns on, press the ZERO key again. The display will show the calibrate weight. Press the
TARE key to select the appropriate calibrate weight. After selection, put the calibrate weight
on. When the “STABLE”indicator turns on, press the ZERO key to confirm. The display will
shows “pass”. The calibration will be complete.
4.3 Setting resolution
After entering setting mode, the display will show “F1 CAL”. Press the TARE key, until the
display shows “F2 RES”, press the ZERO key to enter, and press the TARE key to select
the resolution.
4.4 Changing capacity
After entering the settings mode, the display will show “F1 CAL”. Press the TARE key until
the display shows “F3 CAP”. Press the ZERO key to enter, then the TARE key to select
capacity. Press ZERO key to confirm.
4.5 Checking internal counts
After entering the settings mode, the display will show “F1 CAL”. Press the TARE key until
the display shows “F4 INP”. Press the ZERO key to enter, display will show internal counts.
Press the TARE key to return settings.
4.6 Automatic sleep-mode
In weighing mode, if the scale is unmoved, and there is no change of weight, the scale will
enter sleep mode after a minute. Increasing the weight by 4d will re-wake the scale for use.