Instructions for use
12 CLIMACELL EVO_np_en_1401_mmm_V1.01
The devices are not designed for environment
with risk of flammable or explosive substances
(e.g. anaesthetics).
If the device is used with CO2, it is necessary to
protect the premises in case of excessive gas
leakage (well ventilated room, sensors of CO2
with alarm, etc.).
Any assemblage or disassembly of parts of the
device may only be performed after the device
disconnection from electric network by pulling
the feeding cable from the socket!
In case of temperatures below 0 °C or above
60 °C there is a risk of injury caused by metal
handles on inner glass doors. Use protective
aids.
When operating a device with pre-set
temperature above 70 °C be extremely careful
when opening the door. As the pre-set RH is kept,
hot steam develops in the chamber and it may
cause scalding.
The protection of the temperature case, its
surroundings and processed material against
prohibited temperature exceeding is arranged by a
protective thermostat according to EN 61010-2-010.
The network connection must not get in touch with
hot parts of the device.
You can find all and any installation conditions of the
device in tables, see chapter 4.
The cases stand on the floor, they are space saving,
equipped with castor wheels. It is necessary for the
cooling incubator to stand on flat floor (because of
the device functionality). And we recommend a waste
canal to be located in the room for any case of water
leakage.
After the device placement to the desired location,
turn and set the brake of the castor wheels. The front
one in the front direction and the rear ones in the rear
direction, in parallel with sides of the device in such a
way so as the device gets maximal possible stability
in forward and backward directions.“
Always place the material on screens in the device,
never put it on the bottom of the device!
If you want to start the chamber decontamination, it
must be empty (only the RH sensor may stay).
7 PUTTING THE
DEVICE INTO
OPERATION
7.1 UNPACKING AND CONTROL
After unpacking please check whether the device and its
accessories are complete and undamaged.
Any possible damage must be immediately reported
to the carrier! Do not lift up the case for the handrail or
door in the course of handling (e.g. case lifting, etc.). Lift
the case up using the enclosed hooks. When removing the
case from the palette take care of the device bottom. Lift the
case up using the hooks and move the palette completely
out. Then put the case carefully on the wheels designed for
local movements on smooth floor – not for any prolonged
transport. Pay attention mainly to door steps and entries
to lifts. In such cases it is also necessary to lift the case up
using the hooks. In the course of any handling of the case,
the door must be closed and the case must be disconnected
from power mains. When moving the case be always very
careful and handle the case in such a way so as not tip it
over. In case of sizes 404, 707 and 1212 it is necessary to
protect the case against possible tipping over by fastening it
to the wall see chapter 7.3.
7.2 IMPORTANT WARNINGS
CLIMACELL is designed and manufactured in compliance
with requirements of EU regulations No. 2006/95/EC and
2004/108/EC and it is individually tested according to
EN 61010-1.
Let the installed device stand still in operation position
for at least two hours before putting it into operation for
the first time.
Read these Instructions for Use carefully before
starting the use of the device!
The load capacity of the floor for the device
placement must correspond with the weight of
the device, taking into consideration the maximal
charge weight (see chapter 4.1).
It is not allowed to place the device on a mat that
may cause a risk of fire or smouldering in case of
hot objects to fall out of the device.
It is not allowed to insert any flammable,
explosive or poisonous substances or materials
that could release such substances to the device!