mPower Electronics MUNI CaliCase User manual

MUNI CaliCase
4-Bay Docking Station
MP342
User Guide
Rev 1.0
April, 2023

MP342 MUNI 4-Bay CaliCase
1
Contents
1. General Information ............................................................................................................................ 3
2. Description .......................................................................................................................................... 3
2.1 MP342 Component Diagram ..............................................................................................................................3
2.1 LED Indications....................................................................................................................................................4
3. Installation and Set Up......................................................................................................................... 4
3.1 Charging..............................................................................................................................................................4
3.2 Gas Connections..................................................................................................................................................4
3.3 Power On/Off......................................................................................................................................................4
3.4 Initial CaliCase UDisk and Software Set Up.........................................................................................................4
4. Calibration and Bump Test Procedures ................................................................................................ 7
4.1 Calibration ..........................................................................................................................................................7
4.2 Bump Test ...........................................................................................................................................................7
5. Additional Operations.......................................................................................................................... 7
5.1 Manual MUNI Configurations.............................................................................................................................7
5.2 Automated Multiple Configuration and/or Firmware Updates ..........................................................................8
6. Bump/Calibration Results Download & Certificate ............................................................................ 10
7. Replacement Parts............................................................................................................................. 11
Technical Support and mPower Contacts............................................................................................... 12

MP342 MUNI 4-Bay CaliCase
2
Read Before Operating
This manual must be carefully read by all individuals who have or will have the responsibility of
using, maintaining or servicing this product. The product will perform as designed only if it is
used, maintained and serviced in accordance with the manufacturer’s instructions. The user
should understand how to set the correct parameters and interpret the obtained results.
CAUTION!
•USE ONLY IN AREA KNOWN TO BE NON-HAZARDOUS.
•USE OF NON-mPOWER COMPONENTS WILL VOID THE WARRANTY AND CAN
COMPROMISE THE SAFE PERFORMANCE OF THIS PRODUCT
Proper Product Disposal at the End of Life
The Waste Electrical and Electronic Equipment (WEEE) directive (2002/96/EC) is
intended to promote recycling of electrical and electronic equipment and their
components at end of life. This symbol (crossed-out wheeled bin) indicates separate
collection of waste electrical and electronic equipment in the EU countries. This product
may contain one or more Nickel-metal hydride (NiMH), Lithium-ion, or Alkaline
batteries. Specific battery information is given in this user guide. Batteries must be
recycled or disposed of properly. At the end of its life, this product must undergo
separate collection and recycling from general or household waste. Please use the return
and collection system available in your country for the disposal of this product.

MP342 MUNI 4-Bay CaliCase
3
1. General Information
MP342 is the MUNI Calibration and Bump Test CaliCase. All records of calibrations and bump tests
performed on the CaliCase are stored in its USB disk. The CaliCase includes internal rechargeable
batteries and is portable for remote use.
CaliCase comes with the following standard features
•Ability to bump test, calibrate, set configurations and upgrade firmware for MUNI
•1 to 4 MUNIs can be tested at once.
•Rechargeable internal battery, up to 1500 bump tests before recharging
•USB flash for bump test and event log storage for up to 1 million bump tests
•1 Internal gas cylinder plus 2 inlets: one for extra gas, one for fresh air
•2 Buttons: one for calibration, one for bump test
This Manual describes only operation of the MUNI 4-Bay CaliCase docking station and it is assumed
that the operator is already familiar with the MUNI User’s Guide, including simple maintenance,
firmware configurations and mPower Suite software operations.
2. Description
2.1 MP342 Component Diagram
An exploded view of the MP342 main components is shown below.
1
Unit Cradle & LED x 4
2
Status LED
3
CAL Button
4
BUMP Button
5
USB Data Line Port
6
Air Inlet
7
Charging Port
8
Exhaust Port
9
Gas Inlet 2
10
Pressure Gauge
11
Thumb Screw x 2
12
Demand-flow Regulator
13
Gas Inlet 1 (C-10 fitting)
14
USB UDisk Port

MP342 MUNI 4-Bay CaliCase
4
2.1 LED Indications
The interpretations of the various LED indicators are listed below.
LED
Color
Description
Status LED
Green
Power on
Green blinking
Low battery
Orange
Charging
Red blinking
1) USB Disk error
2) Date/time error
3) No pump detected
None
Power off
Unit LED
Green blinking fast
Cal/bump testing
Green blinking slow
Unit firmware upgrading
Orange
Sensor type mismatch
Green
1) Cal/bump test pass
2) Unit firmware upgrade pass
Red
1) Cal/bump test fail
2) Unit firmware upgrade fail
All LEDs
Orange blinking
CaliCase firmware upgrading
Orange
CaliCase firmware upgrade aborted
3. Installation and Set Up
3.1 Charging
The charging port is located on the right side of the CaliCase. Connect the DC adapter to an AC
110~240V power supply (wall outlet). A full charge takes approximately 3 hours and lasts for up to
1500 unit bump tests.
3.2 Gas Connections
Thread the main gas cylinder into the internal demand-flow regulator (DFR) [12] C10 inlet fitting [13].
Once tightly connected, the gauge [10] will indicate the cylinder gas pressure. If needed, a second gas
can be connected to hose-barb port [9] using a separate DFR. We recommend that any soft tubing (e.g.,
Tygon) used for port [9] be kept as short as possible to minimize potential loss of calibration gas. The
MP342 CaliCase is compatible with all the gases for sensors offered with the MUNI monitors.
3.3 Power On/Off
Press and hold the Cal/ button for about 3 seconds.
3.4 Initial CaliCase UDisk and Software Set Up
Before the first bump or calibrations can be performed, the CaliCase must be configured to match the
sensor types on the MUNIs to be tested and enter the calibration gas concentrations and expiration
date(s). The calibration gas is set up on both the CaliCase itself and the removable UDisk thumb drive.
(The MUNI span concentrations do not need to match the CaliCase and UDisk values)
IMPORTANT!!The MP342 will not operate properly without the UDisk inserted in Port [14]!!
Only the UDisk provided by mPower will function correctly.

MP342 MUNI 4-Bay CaliCase
5
Connect the CaliCase to a PC using the Type A male-to-male USB cable at port [5]. Insert the UDisk
into a USB port on the computer. Use mPower Suite software to configure the gas concentrations and
other parameters. Download mPower Suite at Software Downloads | mPower Electronics Inc.
(mpowerinc.com). Also, it is best to have the latest firmware installed in both the CaliCase and MUNI,
as earlier firmware versions may not be fully compatible with each other.
Procedure
1. Launch the mPower Suite software.
2. Click the Search button at the bottom. The CaliCase, MUNIs and UDisk should appear listed in
the left panel.
3. Click on the UDisk icon to open the CaliCase Info window and enter the calibration gas names,
concentrations and units. Make sure that the gas types match those of the MUNI sensors.
4. Enter the gas cylinder lot number and expiration date, making sure that the expiration date is past
the current date (but not more than 3 years from the current date). Be sure enter an expiration date
for Gas #2 even no Gas #2 is used.
5. Click the “Write”button on the lower right to send the information to the UDisk.
6. Click on the MUNI Info tab and click “Write”in the middle of the screen. No parameter changes
need to be made here.
7. Click on the CaliCase icon in the left panel to open its Configuration panel.
8. Enter other parameters and functions as listed below and click “Write”on the bottom bar.

MP342 MUNI 4-Bay CaliCase
6
•Hibernate Timeout –The CaliCase automatically turns off. Default is 300 seconds.
•Maximum Bump time - The maximum time allowed for a Bump Test. Can be set from 10-120
seconds. Default is 30 seconds.
•Maximum Purge time - The maximum time allowed for clear the gas tubing after
CAL/BUMP. Can be set from 0-60 seconds. Default is 30 seconds.
•Calibrate - If not selected, the calibration function will be disabled.
•Bump test - If not selected, the bump test function will be disabled.
•Update Unit Config - If selected, the CaliCase will update the MUNI configuration when
pressing the CAL or BUMP button (see Section 5.2).
•Update Unit Firmware - If selected, the CaliCase will update the MUNI firmware when long
pressing the BUMP button. The MUNI firmware must first be downloaded to the computer and
the file path selected on the UDisk MUNI Info tab (see Section 5.2). Note that MP342 is only
compatible with MUNI firmware version 1.0.9 or later.
•Soak time - The time allowed for flushing the gas tubing before CAL/BUMP is initiated. Can
be set from 0-60 seconds. Default is 0 seconds.
•Gas Inlet#1 - Set gas cylinder parameters of inlet 1 (Internal gas inlet)
•Gas Inlet#2 - Set gas cylinder parameters of inlet 2 (External gas inlet)
•Firmware Upgrade - Select computer file to upgrade the CaliCase firmware. Obtain the latest
firmware at Software Downloads | mPower Electronics Inc. (mpowerinc.com).
9. Turn off the CaliCase by holding down the Cal button. Insert the UDisk into Port [14] and then
turn on the CaliCase to automatically load the UDisk parameters onto the CaliCase.

MP342 MUNI 4-Bay CaliCase
7
Calibration Parameter Priority
When a calibration is initiated, the CaliCse uses the parameters stored in the CaliCase itself and
overrides the values stored the MUNI configuration. Therefore, the MUNI span concentrations do not
need to match the CaliCase and UDisk values. Each time the CaliCase is turned on, the calibration
parameters automatically get loaded from the UDisk onto the CaliCase and override the existing
CaliCase parameters. Therefore, if new values are programmed onto the UDisk via PC and inserted
into the CaliCase, the CaliCase must be turned off and back on for those UDisk values to take effect.
After the first set up it is not necessary to re-configure the CaliCase directly. The direct CaliCase
procedures listed above are only needed for a few items such as the Location name and syncing the
device time and date.
4. Calibration and Bump Test Procedures
4.1 Calibration
1. Turn on 1-4 MUNI instrument(s) and place them face-down into their cradle(s).
2. If the Status LED [2] is off, turn on the CaliCase by pressing and holding the CAL button [3] until
the Status LED turns green.
3. Press the CAL button [3] to initiate calibration. The Unit LED(s) [1] should blink green for about
60 to100 seconds and then stay GREEN if passed or RED if failed.
4. A Calibration Report will be saved on the UDisk.
5. To abort a calibration, press and hold the BUMP button [4] until the Unit LED(s) turn off.
6. To power off, press CAL for about 4 seconds until the Status LED turns off.
4.2 Bump Test
1. Turn on 1-4 MUNI instrument(s) and place them face-down into their cradle(s).
2. If the Status LED [2] is off, turn on the CaliCase by pressing and holding the CAL button [3] until
the Status LED turns green.
3. Press the BUMP button [4] to initiate a test. The buzzer will be tested and the Unit LED will blink
green for about 30 seconds and then stay GREEN if passed or RED if failed.
4. A Bump Test Report will be saved on the UDisk.
5. To abort a calibration, press and hold the BUMP button [4] until the Unit LED(s) turn off.
6. To power off, press CAL for about 4 seconds until the Status LED turns off.
5. Additional Operations
5.1 Manual MUNI Configurations
The CaliCase can be used to manually set individual MUNI configurations for field measurements. See
the MUNI Manual for more information on each of the parameters including firmware upgrades.
1. Turn on 1-4 MUNI instrument(s) and place them face-down into their cradle(s).
2. Open mPower Suite and click “Search”on the bottom panel.
3. Select the MUNI from the list that appears on the left panel.
4. Adjust any configuration parameters desired and click “Write”in the bottom panel.

MP342 MUNI 4-Bay CaliCase
8
5.2 Automated Multiple Configuration and/or Firmware Updates
The CaliCase can be used to simultaneously update 1-4 MUNI configurations and/or firmware. To do
so, the path to a saved configuration and/or firmware must be entered onto the UDisk MUNI Info tab.
1. Insert the UDisk into the PC, Open mPower Suite, click “Search”and click on the UDisk in the
left panel.
2. Click “Load” to read a previously saved MUNI configuration template from the UDisk.
3. Make any parameter changes desired.
4. For firmware updates, browse the computer to select the firmware path
5. Click “Write” on the middle panel to save the MUNI configuration template and firmware to
the UDisk.

MP342 MUNI 4-Bay CaliCase
9
6. Click on the CaliCase Info Tab and make sure that the Update Unit Config and/or Update Unit
Firmware boxes are checked. If desired, also enter CaliCase Station Config and Firmware
Paths. Click on Write in the Station Firmware Upgrade box.
7. Insert the UDisk memory back to the CaliCase while it is off. Then power it on to transfer the
UDisk information to the CaliCase. If a CaliCase firmware path was entered, it will update
itself automatically.
8. Turn on 1-4 MUNI instrument(s) and place them face-down into their cradle(s).
9. The configuration of all the MUNIs will be automatically updated when a Calibration or Bump
Test is initiated.
10. Long press the BUMP button and the CaliCase will check and upgrade the firmware on all
MUNI units.
NOTES:
1. If the CaliCase has a low battery charge (Status LED is flashing in GREEN), the CaliCase will
skip the firmware upgrade when it is powered on, to avoid the risk of firmware corruption.
2. The UDisk must be inserted back to the CaliCase before powering it on.

MP342 MUNI 4-Bay CaliCase
10
6. Bump/Calibration Results Download & Certificate
1. Remove the UDisk from the MP342 MUNI CaliCase and insert it into a computer. Using
mPower Suite, select the UDisk when it appears in the left panel after several seconds.
2. Select the tabs for CaliCase Info and Reports Information. If desired, choose a Filter and/or Last
number of Days to limit the types of Calibration or Bump events shown.
3. Select a Cal/Bump event in the middle panel. The right panel shows the event overview followed
by the detailed results for each sensor.
4. Click on Summary for a list of filtered event results in tabular form.
5. Click on Certificate to complete and print a calibration certificate.
UDisk Reports Information MP342 MUNI CaliCase

MP342 MUNI 4-Bay CaliCase
11
7. Replacement Parts
Description
Part Number
List Price*