MP342 MUNI 4-Bay CaliCase
7
Calibration Parameter Priority
When a calibration is initiated, the CaliCse uses the parameters stored in the CaliCase itself and
overrides the values stored the MUNI configuration. Therefore, the MUNI span concentrations do not
need to match the CaliCase and UDisk values. Each time the CaliCase is turned on, the calibration
parameters automatically get loaded from the UDisk onto the CaliCase and override the existing
CaliCase parameters. Therefore, if new values are programmed onto the UDisk via PC and inserted
into the CaliCase, the CaliCase must be turned off and back on for those UDisk values to take effect.
After the first set up it is not necessary to re-configure the CaliCase directly. The direct CaliCase
procedures listed above are only needed for a few items such as the Location name and syncing the
device time and date.
4. Calibration and Bump Test Procedures
4.1 Calibration
1. Turn on 1-4 MUNI instrument(s) and place them face-down into their cradle(s).
2. If the Status LED [2] is off, turn on the CaliCase by pressing and holding the CAL button [3] until
the Status LED turns green.
3. Press the CAL button [3] to initiate calibration. The Unit LED(s) [1] should blink green for about
60 to100 seconds and then stay GREEN if passed or RED if failed.
4. A Calibration Report will be saved on the UDisk.
5. To abort a calibration, press and hold the BUMP button [4] until the Unit LED(s) turn off.
6. To power off, press CAL for about 4 seconds until the Status LED turns off.
4.2 Bump Test
1. Turn on 1-4 MUNI instrument(s) and place them face-down into their cradle(s).
2. If the Status LED [2] is off, turn on the CaliCase by pressing and holding the CAL button [3] until
the Status LED turns green.
3. Press the BUMP button [4] to initiate a test. The buzzer will be tested and the Unit LED will blink
green for about 30 seconds and then stay GREEN if passed or RED if failed.
4. A Bump Test Report will be saved on the UDisk.
5. To abort a calibration, press and hold the BUMP button [4] until the Unit LED(s) turn off.
6. To power off, press CAL for about 4 seconds until the Status LED turns off.
5. Additional Operations
5.1 Manual MUNI Configurations
The CaliCase can be used to manually set individual MUNI configurations for field measurements. See
the MUNI Manual for more information on each of the parameters including firmware upgrades.
1. Turn on 1-4 MUNI instrument(s) and place them face-down into their cradle(s).
2. Open mPower Suite and click “Search”on the bottom panel.
3. Select the MUNI from the list that appears on the left panel.
4. Adjust any configuration parameters desired and click “Write”in the bottom panel.