MRC SMZ-171B User manual

c
SMZ-171
Series
Stereo
ZoomMicroscopes
Instruction
Manual
WWW.M0TIC.COM
MOTIC
INCORPORATION
LTD.
Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
MRC.VER.01-6.13

We
areconstantlyendeavoringtoimproveourinstrumentsandtoadaptthemtotherequirementsof
modernresearchtechniquesandtestingmethods.Thisinvolvesmodificationtothemechanical
structure andopticaldesignofourinstruments.
Therefore,alldescriptionsandillustrationsinthisinstructionmanual,includingallspecificationsare
subjecttochange
without
notice.
1

TABLE
OFCONTENTS
Section
Page
1.
INTRODUCTION
3
1.1Nomenclature4
2.UNPACKINGANDASSEMBLINGTHEMICROSCOPE5
2.1Unpackingthe Microscope5
2.2Assemblingthe Microscope5
3.MICROSCOPE
ALIGNMENT
ANDOPERATION6
3.1InterpupillaryDistance6
3.2Focusingthe Microscope6
3.3MagnificationandWorkingDistance7
3.4Changingthe Bulb7
4.
KNOWING
YOURMICROSCOPE8
4.1StereoBody
with
BinocularorTrinocularHead8
4.2Stand9
4.3ExternalIlluminator12
4.4EyepiecesandAuxiliaryObjectives13
4.5Other
Accessories
14
5.CLEANINGANDCARINGFORTHEMICROSCOPE16
5.1ProtectionagainstDust16
5.2ProtectionagainstWaterandMoisture16
5.3Cleaning16
5.4Movingthe Microscope17
5.5ElectricalPartsofthe Microscope17
APPENDIX
1:
SMZ-171
SPECIFICATIONS
18
APPENDIX
2:
SMZ-171
OPTICALDATA19
2

1.
INTRODUCTION
The
Motic
SMZ-171seriesstereomicroscopesarehighperformanceGreenoughdesign
stereomicroscopes
with
continuouszoomrangesof6.7:1.Themicroscopesinthisseriesallow
continuousvariablemagnificationbetween7.5X-50X,while
total
magnificationvariesfrom2.25Xto
200X
dependingontheeyepiecesandauxiliaryobjectives
used.
Withthebilateralzoomknobs,users
areabletoadjustmagnificationwhilethemicroscoperemainsperfectlyparfocal.
TheSMZ-171istheidealinstrumentforexaminingobjectsofindustrial,biological,medicalor
educationalnature.Optional
with
ESDfeature,
with
headandstand.
*Regarding
theESD
Inshort,ESDistherapidneutralizationofcharge.Weknow
that
allmatterismadeofatoms,consisting
ofelectronsandprotons.Whenthematerialgainsorloseselectron,itwillloseitsbalanceandbecome
negativelyorpositivelycharged.Positiveornegativechargeaccumulationonthesurfacewillmake
staticelectricityonobject,usuallycausingdamagetotheelectroniccomponents.
SMZ-171
usesanti-staticmaterialontheheadandstandtoachievetheESDfeature,whichcan
preventchargeaccumulation.
SMZ-171
Standard
Configuration
and
Parameters:
For
detailedspecificationsofthe
SMZ-171,
pleaserefertotheAppendix1.
Model
LensTubes Eyepieces
Auxiliary
Objectives
Zoom
factor
Magnification
Working
Distance
SMZ-171
B
Binocular
N-WF
10X
(023) Optional 1:6.7
7.5X-50X
110mm
SMZ-171T
Trinocular
N-WF
10X
(023) Optional 1:6.7
7.5X-50X
110mm
Therearetwomodelsintheseries:the
SMZ-171
B whichisequipped
with
a binocularsystem,andthe
SMZ-171T
whichisequipped
with
a trinocularsystemallowingfortheattachmentofCCDanddigital
camera
devices.
3

1.1Nomenclature
Eyepieces
Eyepiece
tube
Binocular
head
Locking
screw
Objective
Incident
illumination
Slide
stageclips
Stage
Base
Power
cord
for
incidentillumination
Column
-Zoomknob
Head
holder
Focusing
knob
Supportcollar
lockingscrew
Supportcollar
Separated
powerswitch
Lightintensitycontrol
SMZ-171BL(Fig.1)
Trinoculartube
Lightsplitknob
Armstand
SMZ-171TL
(with
ESD)(Fig.2)
4

2.UNPACKINGANDASSEMBLINGTHEMICROSCOPE
2.1
Unpacking
the
IVIicroscope
The
componentsforSMZ-171stereomicroscopesareshippeddetachedforprotection.Openthe
styrofoampacking
with
careanddonotleaveanycomponentsattachedtothepackingbeingremoved.
Do
notdiscardanyofthepackingmaterials
until
allofthecomponentshavebeenidentified.Ifany
damage
occursduringtransit,contactboththecarrierandyoursupplierimmediately.
2.2AssemblingtheMicroscope
2.2.1Whenhandlingthecomponents,especiallytheopticalparts,avoidtouchinganylenssurfaces
with
barehandsorfingersasfingerprintsandgreasestainsaffectimagequality.
2.2.2Afterunpackingthestand,putitona stableand
flat
table.Loosenthelockingscrewofthe
head
holder,adjusttheheightoftheheadholderandlockthelockingscrewagain.Make
sure
that
thesupportcollarissecured
firmly
belowtheheadholderalongtheverticalpost.
This
isimportantasthiscollarkeepsthemicroscopefromslidingdownthecolumn.
2.2.3Theheadholdershouldrestonthesupportcollarandboththeheadholderlockingscrewand
thecollarlockingscrewshouldbetightened.
2.2.4Putthestereobodyontotheheadholderandlockitusingthelockingscrewlocatedonthe
right
handsideoftheheadholder.Tomaintainthebeststability,youareadvisedtolockthe
screw
onthe
left
handsideoftheheadholderaswell.
2.2.5Afterremovingthewrappingpapersandpackingmaterialsaroundtheeyepiecesandother
opticalparts(avoidtouchingthelenssurfaces),carefullyplacetheeyepieces
into
theeyepiece
tubesandsecurethemusingthelockingscrews.Ifnecessary,loosenthelockingscrewsbefore
putting
intheeyepieces.
2.2.6Beforeturningonthepower,plugtheilluminationpowercord
into
thesocketatthetopofthe
column.
Next,plugthepowercordofthemainframe
into
thelocalpowersupply.
Note:
before
plugging
in and
turning
onthe
power,
make
sure
that
the
operation
voltage
ofthe
microscope
matches
the
voltage
ofthelocal
power
supply.
5

3.MICROSCOPE
ALIGNMENT
ANDOPERATION
3.1
Interpupillary
Distance
Adjustthetwoeyepiecetubes
until
onlyonecircularfieldcanbeseenthroughthetwoeyepieces.Iftwo
separatecirclesappear,theinterpupillarydistanceistoolarge;iftwooverlappingcirclesappear,the
interpupillarydistanceistoo
small.
Theeyepiecetubeallowsa flexibleadjustmentoftheinterpupillary
distancebetween48and75mm.
3.2FocusingtheMicroscope
To
focusthesample,usethefocusingknobslocatedonbothsidesoftheheadholder(Fig.3).By
turningtheseknobs,themicroscopecanbemovedupordowna certaindistancetofocusthesample.
This
movementisenabledbya
"rack
andpinion"mechanism.Thetensionofthefocusingknobcan
beadjustedusingthetensionknoblocatedintheinnerregionofthefocusingknobonthe
right
(Fig.4).
(Fig.3)(Fig.4)
3.2.1Usingthefocusingknob,focusthesampleusingthehighestmagnificationstrength.Ifthe
sample
cannotbebrought
into
focus,adjusttheheightofthemicroscopealongthecolumn.
Remember
totightenthelockingscrewandsupportcollarafteradjustingtheheightofthe
microscope.
3.2.2Turnthezoomtothehighestmagnification.Adjustthefocusingknob
until
a clearandsharp
imageisobtained.
3.2.3Turnthezoomtothelowestmagnification.Adjustthe
right
eyepiecediopter
until
theimage
seen
throughthe
right
eyepieceisclearandsharp.
6

3.2.4Repeattheprocedureforthe
left
eyepiece.Next,checktheimagefocusfortheentirezoom
range;itshouldnowbeperfectlyparfocal.
3.3
Magnification
and
Working
Distance
3.3.1Selectthedesiredmagnificationstrengthbyadjustingthezoomknob.Changetheoptional
eyepieces
and/oraddanoptionalauxiliaryobjective,forotherrangeofmagnification.
3.3.2Totalmagnificationusedcanbecalculatedusingthefollowingequation:
Totalmagnification= EyepiecemagnificationX ZoommagnificationX
Objectivelensmagnification.
3.3.3Workingdistancevaries
from
301mm(whenusinga 0.3Xobjectivelens)to38.6mm(whena 2X
objectivelensisused).Normalworkingdistanceforstandardconfiguration(IXobjectivelens)is
110mm.
3.4
Changing
theBulb
3.4.1Beforechangingthe
light
bulb,makesure
that
thepowerisswitchedoffandthepowercordhas
been
disconnected
from
themainpowersupply.
3.4.2Forincident
light,
removethelampcollectorpiecebyAllenkeysupplied
with
theinstrument,
unplugtheoldLEDcircuitboard
from
thesocketandcarefullypluginthenewLEDcircuitboard.
Screw
thecollectorpiecebackinafterchangingtheoldone.
7

3.4.3Fortransmitted
light,
turn
themicroscopeoverso
that
thebottomplateofthestandfaces
towardstheuser.Removethebottomplate
with
thesuppliedAllenkey,unplugthelamp
collectorpiece,removetheoldLEDcircuitboardfromthesocketandcarefullypluginthenew
LED
circuitboard.Finally,
firmly
securethebottomplateafterchangingtheoldone.
3.4.4Nevertouchtheglasssurfaceofthe
light
bulb
with
barehands.Anygreaseonthe
light
bulbwill
affectheatdissipation,greatlyshorteningthelifespanofthe
light
bulb.Ifthesurfaceofthebulb
has
beenaccidentallytouched,clean
with
alcoholandtissue.
4.
KNOWING
YOURMICROSCOPE
4.1
Stereo
Body
with
Binocularor
Trinocular
Head
For
theSMZ-171stereomicroscopes,thebinocularortrinoculartubesare
built
into
thestereobodyto
forma single
unit
calleda
"Stereo-head".
TheSMZ-171BL(Fig.1)isequipped
with
a binoculartube
whiletheSMZ-171TL(Fig.2)isequipped
with
a trinoculartube.
4.1.1StereoBody
•TheStereobodyisthekeycomponentofthemicroscope.ItincludestheGreenoughstereo-zoom
system
with
a continuouszoomrangeof6.7:1.Italsoincludesseparate
left
and
right
non-telescopic
opticssystems.
•Withthissystem,usersareabletoenjoyexcellentdepthoffieldandstereoeffects.Withthehelpof
precisionopticsfrom
M5(Y,
perfectparfocalityismaintainedthroughouttheentirezoomrange.
•Thezoomknobsarelocatedonbothsidesofthemicroscopeand
scales
areprintedontheknobsto
displaythecuYrrentmagnification.Adjusttheseknobstochangethemagnificationoftheimage.Ifthe
microscope
hasbeenproperlyadjusted,theimageshouldremaininfocusevenwhenthe
magnificationischanged(parfocal).Foradjustmentprocedures,refertosection
3.2
ofthismanual.
•Thestereobodyismountedontothecircularmountoftheheadholderandislocked
into
place
usingthelockingscrewonthe
right
handsideoftheholder.Whileoperatingthemicroscope,this
lockingscrewshouldalwaysbelockedtomaintainutmoststability.
8

4.1.2
Binoculartubeforthe
SMZ-171B
•Theinterpupillarydistancecanbeadjustedbymovingthetwoeyepiecetubeshorizontally.For
properinterpupillarydistanceadjustment(Fig.6),refertosection
3.1
ofthismanual.
•N-WFeyepiece,higheye-point10X(023),diopteradjustable,interchangeable
with
biological
Eyepieces.
(Fig.6)
4.1.3
Trinoculartubeforthe
SMZ-171T
•Theproceduresforadjustingtheinterpupillarydistanceandsecuringtheeyepiecesarethesamefor
thetrinoculartubeastheyareforthebinoculartube.
•Byturningtheknobatthe
left
sideofthetrinoculartube,allthelightfromthe
right
eyepiecetubewill
be
deflectedintothephototubefortheattachedimagingdevice.
•Atthetopofthetrinoculartube,thereisa lockingscrew
that
isusedtosecuretheadapterforthe
imagingdevice.After
fitting
theadapter,thislockingscrewshouldbetightened.
•N-WFeyepiece,higheye-point10X(023),diopteradjustable,interchangeable
with
biological
eyepieces
4.2
Stand
Three
differentstandsareavailableforthe
SMZ-171:
4.2.1LEDstand
with
Armversion(Fig.1)
•Newstand
with
reflectordesigntoreachmorehomogeneousilluminationandlowertemperature;the
angle
ofthereflectorcanbeadjustedbyturningthereflectorcontrolonthe
left
handsideofthe
base.
(Fig.7)
9

(Fig.7)
•LED(Fig.8a)orfiberopticalillumination(Fig.8b)canbeselected
with
a switchonthestand.There
is
a fiber
access
atthebackofthestand(Fig.9).Whenfiberilluminationis
used,
pullthedraw-rod
backwardsthe
left
handofthe
base.
(Fig.8b)
(Fig.9)
10

•Optionalstandinblackcolor
with
ESDfeatureisavailable.
•Theheadholderisfixedonthearmandcannotberemovedfromthestand.Thearmversionisnot
as
versatileasthepoleversion.
•Thisstandisinstalled
with
built-intransmittedlight(3WLED)andincidentlight(3WLED)
illuminators.Itdeliversalmostnoheattothesampleandisidealforbiologicalandheat-sensitive
specimen
applications.
•To
turn
thestandpoweron,themainswitch(locatedatthebackside)shouldbeswitchedtothe"on"
position.
(Fig.10)
•Separatepowerswitchesfortransmittedandincidentilluminationarelocatedonthe
right
hand
side.
Users
areabletoselecttransmittedorincidentlightorboth
with
thesetwopowerswitches.(Fig.11)
•Lightintensitycanbeadjustedbyturningtheknobatthebottomofthestandonthe
right
hand
side.
This
knobgovernsthelightintensityforthetransmittedandincidentlights.
(Fig.
12)
•Theilluminatingangleoftheincidentlightcanbeadjustedbydirectlyturningthelampcollector
piece.
1
(Fig.10)
IncidentlightTransmittedlight
intensitycontrolintensitycontrol
J
(Fig.11) (Fig.12)
11

•Besidestineblackandthewhitestageplate,a frostedglassstageplateisprovidedfortransmitted
light.
4.2.2LEDstand
with
Pole
version(Fig.2)
•Basicallythesameasthe"Armversion"standexcept
that
theheadholdercanbemovedfreely
alongtheverticalpostandberemovedcompletelyfromthestand.Thediameterofthepostis32mm.
4.2.3Plainstand
•Thisindustrialstandallowsforextremeflexibilityinpositioning.Itincludesa heavybasetomaintain
goodstability.
4.3
External
Illuminator
4.3.1Allkindsofexternalilluminatorscanbeusedasincidentilluminatorsrangingfromsimpledesk
lampstospecializedringilluminators.
4.3.2A speciallydesignedcold
light
sourceemployinga 12V/150Whalogenilluminatoris
recommendedasa
light
source.Sucha
light
sourceallowsuserstobendand
twist
thearmin
anydirectioninordertoachievethebestpossibleillumination.Aswell,
with
a cold
light
source,
noheatistransmittedtothespecimenmakingitidealforuseinbiologicalresearchand
anatomy.
12

4.3.3A 3WLEDring
liglit
illuminatorforMoticSMZ-171stereomicroscopesalsoisavailable.
4.4Eyepiecesand
Auxiliary
Objectives
4.4.1ThereareN-WFeyepieces
with
higheye-pointofdifferentmagnificationstochoosefrom
including10X,12.5X,15Xand20X.Standardconfigurationisa pairof10X(023),diopter
adjustable,interchangeable
with
biologicaleyepieces
4.4.2Tochangetheeyepieces,unscrewthelockingscrew,removetheoriginaleyepieces,replace
with
thenewpairofeyepiecesandsecuretheeyepiecesusingthelockingscrew.
4.4.3Thereareadditionalauxiliaryobjectivesofdifferentmagnificationstochoosefromincluding
0.3X,
0.5X,0.63X,0.75X,1.5Xand2X(Fig.13).Usersarerecommendedtoselectadditional
objectivesaccordingtoworkingdistanceandmagnificationrequirements.
Please
refertothe
appendix(2)fordetails.
(Fig.13)
4.4.4Toaddanadditionalobjectivetothemicroscope,screwitontothebottompartofthestereo-
head.
Theheightofthemicroscopemustbere-adjustedastheworkingdistancewillchange
whenanadditionalobjectiveisattached.
13

4.5
Other
Accessories
4.5.1For
the
SMZ-171there
are
variousaccessoriesdesigned
for
variousapplications:
Darkfield
Condenser
•Must
be
used
with
transmittedlight.
•
By
puttingthisaccessoryonto
the
transmittedlightoutlet
(with
the
frostedglass
or
stageplate
removed)
a
darkfieldeffect
is
created.
•
Itis
especiallyuseful
for
analyzingjewelry
and
specialtechniques
in
Bio-Medapplicationsincluding
"In-situ
silver
gain
staining"
and
embryoobservation.
Polarizing
kit
•Must
be
used
with
transmittedlight.
•This
kit
(includingboth
a
polarizerandanalyzer)
is
alsoplacedonto
the
transmittedlightoutlet
(with
thestageplateremoved).
The
sample
canbe
placedbetween
the
polarizerandanalyzer
to
perform
polarized
lightmicroscopy.
•Useful
for
analyzingjewelry
andthe
study
of
sectionedrock
and
syntheticfibers.
Jewelry
Clamp
(Fig.14)
•Designed
to
holdgems
or
jewelryunder
the
microscopewhileperformingobservation.
(Fig.14)
Photo
adapter
SY10
•Attaches
tothetopofthe
trinoculartube,allowing
any
SLRcamera
tobe
connected
tothe
microscope
for
imagingpurposes.
•
The
SY10adapter
forthe
selectedcamerahas
tobe
screwedinto
the
adapterbeforeconnecting
to
thecamera.ThisSY10adaptercan
be
obtainedfrom
any
camerastore
in
yourarea.
•Canonly
be
used
with
the
SMZ-171T.
14

C-mount
or
CS-mount
•Attachestothetopofthetrinoculartube,allowinganyCCDcameraorimagingdevicetobe
connectedtothemicroscope.
Select
eithertheC-mountorCS-mountaccordingtotheCCDcameratobeused
•Canonlybeused
with
theSMZ-171T.
Improved
industrial
boom
stands
•Withnewslot/ groovedesignforbetterlockingofmicroscope.
•Grubscrewlocks
into
slot/ groove,stopping
tilt.
Microscopeispositionedverticaltobase
with
no
Slant.
•Addpositionforhand-carryingaroundfoursides
•Newlydesignedstandscanbeusedinwide-rangeofourSMZseriesmicroscopeswhichallows
observinglargerviewingsamples
•OnlyoneH3Allenkeyisnecessaryforadjustingtheboomstand
•Aluminumsupportcollar
with
elasticplasticgaskettoprotectandavoidscratchesonthecolumn.
15

5.CLEANINGANDCARINGFORTHEMICROSCOPE
To
keepthemicroscopeingoodworkingorder,avoid
dust
and
water.
Ifanydustorwaterhappensto
getintothemicroscope,funguswillgrow,damagingthemicroscope.
Please
note
that
oncefungushas
grown,evenaftercleaning,theproblemmayreoccur.
Grease
stains
and
fingerprints
affectimagequality;avoidallowingfingerstocomeintocontact
with
thesurfaceofopticalcomponents.
5.1
Protection
against
Dust
Iftheinstrumentisnottobeusedfora longperiodoftime,coverit
with
thedustcoverprovided.Never
leave
theeyepiecetubeexposed.Eitherleavetheeyepieceinthetube(recommendedifthe
microscope
isfrequentlyused)orcoverit
with
wrappingpaperora coveringcap.
Eyepieces
andother
opticalcomponents
that
willnotbeusedfora relativelylongperiodoftimeshouldbestoredina dry
cardboardbox,preferably
with
a desiccatingagentadded,toshieldagainstdustandmoisture.
5.2
Protection
against
Water
and
IVIoisture
The
instrumentshouldbekeptawayfromallwatersources,includingpipesandsinks.Humidityinthe
roomwheretheinstrumentlocatedshouldbekeptaslowaspossible(relativehumidityshouldbekept
below70%).Itisrecommended
that
opticalcomponentsbekeptina dryboxwhennotinuse,
preferably
with
a desiccatingagentadded.Theuseofdehumidifierand/or24-hourairconditioningis
highlyrecommendedifthesurroundingareaishumid.
5.3Cleaning
5.3.1Ifdustisfoundontheopticalsurface,removebyusinganairblowerorcompressedair.
5.3.2Forfingerprint,greasestainsordustwhichcannotberemovedusingair,twopossiblemethods
arerecommended:
•Breathelightlyontheglasssurfaceandwipe
with
a cleanpieceofcloth,lenspaperorcottonswab.
Please
note
that
smallcottonfibersmayremainonthesurfaceofthelensifa cottonswabis
used.
•Usea cottonswaborlenspaperdippedina smallamountofpurealcoholandcleanthelenssurface
carefully.Nootheraggressivesolventsshouldbe
used.
16

Under
nocircumstancesshoulduserscleananylenssurface
with
a dry
cotton
swab,dry
cloth
ordrylens
paper
as
this
will
scratchthelenssurfacecausing
irreparable
damage.
Water
isnot
recommended
forcleaninglensesasit
will
leave
water
stains
onthelenssurface,
possibly
leading
tofungus
growth
andcausing
irreparable
damage.
5.4
Moving
theMicroscope
5.4.1Themicroscopeshouldbemovedaroundas
little
aspossible.
5.4.2Ifitisnecessarytomovethemicroscope,usersshouldensure
that
theeyepiecesarefirmly
secured
intheeyepiecetubes,themicroscopefirmlysecuredtotheverticalpostandthe
supportcollarfirmlysecuredbeforemoving.
5.4.3Whenmovingthemicroscope,usebothhands,onehandholdingthebottomofthestandand
theotherhandholdingthetopoftheverticalpostoftheheadholderofthemicroscope.
5.4.4Themicroscopeshouldalwaysbekeptuprightwhilemoving.
5.5Electrical
Parts
oftheMicroscope
5.5.1Beforepluggingthepowercordintothepowersupply,makesure
that
thesupplyvoltage
matches
theoperationvoltageoftheequipment.
5.5.2Turntheequipmentoffbeforepluggingthepowercordintothepowersupply.
5.5.3Itisrecommended
that
users
turn
downtheilluminationbeforeturningofftheequipment.
5.5.4Donot
turn
thepoweronagainimmediatelyafterithasbeenturnedoffasthiswillshortenthe
lifespanofthelightbulbandmaycausedamagetotheelectricalsystem.
5.5.5
Users
shouldobservealllocalsafetyregulations.WhiletheequipmentisCEsafetyapproved,
users
areexpectedtousetheequipmentina safeandresponsiblemanner.
17

APPENDIX
1:SMZ-171
SPECIFICATIONS
Model
SMZ-171BL
SMZ-171TL
Opticalsystem
Greenough
Observationangle
45°/
60°
45°
Magnificationrange
(standard)
0.75X~5X
Zoom
ratio
1:6.7
Eyepiece
N-WF,
higheye-point10X(ct)23),Diopteradjustable
interchangeable
with
biologicaleyepieces
Eyepiece
N-WF
12.5X(018),15X(ct)16),20X(O13)
for
optional
Interpupilary
adjustment 48mm-75mm
Height
of
eye
point
405mm
Working
distance(standard) 110mm
Weight 5.95kg(head
with
1.25kg)
6.2kg(head
with
1.5kg)
C-mountadapters ~ Trinocularheadonly
C-mountadapters —
0.5X,
0.65X,
IX
adaptersavailable
AuxiliaryESDobjectives
0.3X
[WD= 301mm]
0.5X
[WD=
191.8
mm]
0.63X
[WD= 142.7mm]
0.75X[WD
= 128.6mm]
1.5X
[WD= 56.3mm]
2.0X
[WD= 38.6mm]
Max.
workingdistance
(with
0.3X
auxiliary
objective) 301mm
Stand
option
•Stablepole
andarm
basestandavailable
-
Plain
stand
-
3W
LEDincident
and
transmitted
light
stand
with
reflectordesign
•Optionallyseveralboomstands
for
industrial
useare
available
18

APPENDIX
2:SMZ-171
OPTICAL
DATA
CD
Eyepiece
Wlag.(X)
Standard
Objectives
Auxiliary
Objectives
Eyepiece
Wlag.(X)
Standard
Objectives
0.3X 0.5X
0.63X 0.75X
1.5X
2X
Eyepiece
Wlag.(X)
WD110mm
WD
301mm
WD
191,8mm
WD
142.7mm
WD
128,6mm
WD
56,3mm
WD
38,6mm
Eyepiece
Wlag.(X)
Mag.
(X)
FD
(mm)
Mag,
(X)
FD
(mm)
Mag,
(X)
FD
(mm)
Mag,
(X)
FD
(mm)
Mag,
(X)
FD
(mm)
Mag.
(X)
FD
(mm)
Mag,
(X)
FD
(mm)
10X/23*
0.75 7.5 30.67 2.25 102,22 3,75 61,33 4,725 48,68
5,625
40,89 11,25 20,44 15 15.33
10X/23*
1
10 23
3
76,67
5
46 6,3 36,51 7,5 30,67 15 15,33 20 11,5
10X/23*
2
20 11.5 6 38,33 10 23 12.6 18,25 15 15,33 30 7,67 40 5,75
10X/23*
3
30
_j
7.67 9 25,56 15 15,33 18,9 12,17 22,5 10,22 45 5,11 60 3,83
10X/23*
4 40 5.75 12 19,17 20 11,5 25,2 9,13 30 7,67 60 3,83 80 2,875
10X/23*
5
50 4.6 15 15,33 25 9,2 31,5 7,30 37,5 6,13 75 3.07 100 2.3
12.5X/18*
0.75
9.375
24 2.8125 80 4,6875 48 5,91 38,10 7,03 32 14.06 16 18,75 12
12.5X/18*
1
12.5 18 3,75 60 6,25 36
7,875
28,57
9,375
24 18.75 12 25 9
12.5X/18*
2
25 9 7.5 30 12,5 18 15,75 14,29 18,75 12 37.5 6 50 4,5
12.5X/18*
3
37,5 6 11,25 20 18,75 12 23.625 9,52 28,125 8 56.25 4 75
3
12.5X/18*
4 50 4.5 15 15 25 9 31,5 7,14 37,5 6 75
3
100 2,25
12.5X/18*
5
62.5 3.6 18,75 12 31,25 7,2 39,375 5,71 46,875 4,8 93,75 2,4 125
1,8
15X/16*
0.75 11.25 21.33
3.375
71,11
5,625
42,67 7,0875 33,86 8,4375 28,44 16,875 14,22 22,5 10,67
15X/16*
1
15 16 4,5 53,33 7,5 32 9,45 25,40 11,25 21,33 22,5 10,67 30 8
15X/16*
2
30 8 9 26,67 15 16 18,9 12,70 22,5 10,67 45 5,33 60 4
15X/16*
3
45 5.33 13,5 17,78 22,5 10,67 28,35 8,47 33,75 7,11 67,5 3.56 90 2,67
15X/16*
4 60 4 18 13,33 30 8 37,8 6,35 45 5,33 90 2,67 120
2
15X/16*
5
75 3.2 22,5 10,67 37,5 6,4 47,25 5,08 56,25 4,27 112,5 2,13 150
1.6
20X/13
0.75 15 17.33 4,5 57,78 7,5 34,67 9,45 27,51 11,25 23,11 22,5 11,56 30 8,67
20X/13
1
20 13 6 43,33 10 26 12,6 20,63 15 17,33 30 8,67 40 6,5
20X/13
2
40 6.5 12 21,67 20 13 25,2 10,32 30 8,67 60 4,33 80 3,25
20X/13
3
60 4.33 18 14,44 30 8,67 37,8 6,88 45 5,78 90 2,89 120 2.17
20X/13
4 80 3.25 24 10,83 40 6,5 50.4 5.16 60 4,33 120 2,17 160 1,625
20X/13
5
100 2.6 30 8,67 50 5,2 63 4,13 75 3,47 150 1,73 200 1.3
Note:
"WD"= WorkingDistance
"Mag."
= Magnification"FD"= FieldDiameter
Note:
* - Higheyepointeyepiece
This manual suits for next models
1