MRC PWR-3030 User manual

PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
MRC.VER.01-2.12
Operation Manual for
Power Supply
PWR-3030

1
Packing list
3:5:
- 1x PWR-3030 Power Supply
- 1x Power Cord
- 1x PWR-3030 Power Supply Instruction Manual
MRC is liable for all missing or damaged parts / accessories
within 7 days after customer receives this instrument package. Please
contact MRC immediately regarding this issue. If no response is
received within such a time period from consignee party, it will be the
consignee party’s whole responsibility.

2
Table of contents
Packing list................................................................................................1
Warning ....................................................................................................3
Product Specifications ..............................................................................6
Product Description ..................................................................................7
Introduction...............................................................................................8
Overview............................................................................................8
Controls and Features.......................................................................9
Installation Instructions...........................................................................10
Operation Instructions.............................................................................10
Display Screen ................................................................................11
Constant Setup................................................................................11
Programming Setup.........................................................................14
Limiting Parameter Setting ..............................................................17
Troubleshooting Guide............................................................................18
Maintenance...........................................................................................21
Ordering Information...............................................................................21
Warranty .................................................................................................22

3
Warning
MRC PWR-3030 Power Supply has been tested and found to comply
with the limits for the CE regulation.Also, PWR-3030 Power Supply series is
RoHS compliant to deliver confident product which meets the environmental
directive. These limits are designed to provide reasonable protection against
harmful interference when the instrument series is operated in a commercial
environment. This instrument series used together with power supply unit
generates, uses, and can radiate radio frequency energy, and if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this instrument series in a
residential area is likely to cause harmful interference in which case the user
will be required to correct the interference at their expense. Changes or
modifications not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment. It is strongly
recommended for the user to read the following points carefully before
operating this equipment.
1. Read and follow the manual instructions carefully.
2. Do not alter the equipment. Failure to follow these directions could result
in personal and/or laboratory hazards, as well as invalidate equipment
warranty.
3. Use a properly grounded electrical outlet with correct voltage and current
handling capacity.
4. Disconnect from power supply before maintenance and servicing. Refer
servicing to qualified personnel.
5. Never use this instrument series without having the safety cover correctly
in position.
6. Do not use the unit if there is any sign of damage to the external tank or
cover. Replace damaged parts.
7. Do not use in the presence of flammable or combustible material; fire or
explosion may result. This device contains components which may ignite
such materials.
8. Refer maintenance and servicing to qualified personnel.
9. Ensure that the system is connected to electrical service according to
local and national electrical codes. Failure to make a proper connection
may create fire or shock hazard.
10. Use appropriate materials and operate correctly to avoid possible hazards
of explosion, implosion or release of toxic or flammable gases arising from

4
overheated materials.
11. The unit shall be operated only by qualified personnel.
Safety Information
Use high level of precaution against any electrical device. Before connecting
the electrical supply, check to see if the supply voltage is within the range
stated at the rating label, and see to it that the device be seated firmly. Place
the unit in a safe and dry location; it must NOT touch the surrounding. Follow
the safety precautions for chemicals / dangerous materials. If needed, please
Environmental Conditions
Ensure the instrument is installed and operated strictly in the following
conditions:
1. Indoor use only
2. ≤95% RH
3. 75 kPa – 106 kPa
4. Altitude must not exceed 2000 meters
5. Ambient to 40°C operating temperature
6. Pollution degree: 2
7. Mains supply voltage fluctuations up to ±10% of the normal voltage
Avoiding Electrical Shock
Follow the guidelines below to ensure safe operation of the unit.
PWR-3030 power supply has been designed to use with shielded wires thus
minimizing any potential shock hazard to the user. MRC recommends
against the use of unshielded wires.
To avoid electrical shock:
1. In the event of solution accidentally spilling into the instrument, it must be
dried out for a period of time (at least 2 hours) and restored to NORMAL
CONDITION before each operation.
2. Never connect or disconnect wires loading from the power jacks when the
red indicator light of power switch is on.
3. WAIT at least 5 seconds after stopping a run before handling output leads
or any connected apparatus.
4. ALWAYS make sure that your hands, work area, and instruments are clean
and dry before making any connections or operating the power supply.
5. ONLY connect the power cord to a properly groundedAC outlet.

5
Avoiding Damage to the Instrument
1.Do not attempt to operate the device if damage is suspected.
2.Protect this unit from physical damage, corrosive agents and extreme
temperatures (direct sunlight, etc.).
3.For proper ventilation and safety concerns, keep at least 10 cm of space
behind the instrument, and at least 5 cm of space on each side.
4.Use high level of precaution against the damages on the unit.
5.Do not operate the unit out of environmental conditions addressed above.
6.Prior to applying any cleaning or decontamination methods other than
manufacturer’s recommendation, users should check with the
manufacturer’s instruction to see if the proposed method will damage the
equipment.
7.When the PWR-3030 power supply is taken into a cold room environment,
the unit can be operated immediately. However, when the PWR-3030
power supply is removed from the cold room environment, let the unit
equilibrate to room temperature for a minimum of 2 hours before using it
.
Equipment Operation
Follow the guidelines below to ensure safe operation of the unit:
1. Check the displayed figures to see if the unit is functioning correctly before
using this unit.
2. NEVER access dangerous chemicals or other materials to prevent possible
hazard of explosion and damage.
3. Do not operate the unit without lids or covers to prevent possible hazards.
4. Atemporary conductivity caused by condensation might occur even though
this series is rated Pollution Degree 2 in accordance with IEC 664.
Symbols
The symbols used on PWR-3030 Power Supply are explained below.
Indicates an area where a potential shock hazard may exist.
Consult the manual to avoid possible personal injury or instrument
damage.
Indicates disposal instruction.
DO NOT
throw this unit into a municipal trash bin when this unit has
reached the end of its lifetime. To ensure utmost prote
ction of the
global environment and minimize pollution, please recycle this unit.
Max. voltage: 300V
Max. Current: 3000mA
Max. Watt: 300W

6
Product Specifications
Output Voltage / Inc.5 - 300V / 1V
Output Current / Inc.10 - 3,000mA / 10mA
Max. Watt / Inc.300W / 1W
Operating Mode Voltage or current or power
Control Microprocessor controller
Program Storage 30 programmed files
Program Multi-Step Up to 6 steps
Terminal Pairs 4 pairs
Display 2.6” LCD
Timer Constant mode: 1-9999 min. with alarm, continuous
Programmable mode: 1-999 min with alarm,
continuous
Safety Device No Load detection
Leakage detection
Sudden load change detection
Over temperature protection
Over current detection
Over voltage detection
Shrouded plugs and sockets
Crossover Yes
Operating Temperature Ambient - 40°C
Unit Dimension 190 x 305 x 95 mm (W x L x H)
Construction Material Flame retardant ABS faceplate and aluminum
Rated Voltages 100 - 240V; 47/60 Hz
Input Rating 360W
Weight approx. 2.5 kg

7
Product Description
The MRC PWR-3030 Power Supply series is a
microprocessor controlled and designed to meet most electrophoresis needs.
This manual describes the setup and operation of the PWR-3030 Power Supply
including important information on safety and maintenance of the unit. The PWR-3030
power supply is capable of running horizontal & vertical electrophoresis,
SDS-PAGE, native PAGE application, two-dimensional electrophoresis, and
electro-blotting. It comes equipped with a timer and alarm function as well as a
pause function. Furthermore, the powerful specifications plus four terminal
pairs can be used for multiple electrophoresis units simultaneously.
MRC PWR-3030 power supply provides Constant Voltage or Constant
Current or Constant Power to instruments used in electrophoresis. 4 pairs of
terminals and the powerful specification equipment enable more capability of
the PWR-3030 power supply than other existing similar product on the market.
PWR-3030 also has a 2.6” LCD screen where many parameters are displayed at
the same time, which provides a better aspect of the user friendly concept to
the user.
Features of 3:5 power supply:
- Compact size
- Advanced capacity: 300W, 3000mA, 300V
- Microprocessor controller
- Constant voltage, constant current, and constant power
- Four outlet terminal pairs
- LCD display
- Timer with alarm function
- Advanced safety devices
- Stackable
- Wide applications for DNA, RNA and protein electrophoresis



10
Installation Instructions
The MRC PWR-3030 Power Supply is actually a pre-installed instrument.
As long as it is placed on a sturdy and level surface in a safe and dry place,
and is further connected with a well-prepared electrophoresis system, it is
ready for operation.
Operation Instructions
*Note: To operate under constant voltage or constant current modes, adjust
the other parameter to the maximum value. For example, to operate under
constant voltage, adjust current to max before running using constant voltage,
and vice versa.
The PWR-3030 Power Supply is designed to operate under two modes: Constant
Mode or Programming Mode, depending on your electrophoresis conditions.
Use the Constant Voltage / Current / Power Operation for applications that
require only one specific voltage limit, current limit, and power limit
continuously during the entire duration of electrophoresis.

11
Display Screen
Switch on the power, the screen will first show the version and model name.
After the welcome page you will be able to select the operation mode:
Constant or Program mode (Screen shown below)
Constant Setup
Program Setup
Constant Setup
This section gives an instruction for constant mode operation. Constant
Voltage / Current / Power Mode allow you to specify a voltage limit or a
current limit to be used continuously during the entire duration of
electrophoresis. Review the guidelines provided in this manual before
starting electrophoresis using PWR-3030 Power Supply. We recommend reading
the guidelines provided in this manual for best results before starting an
operation.
1. Place the PWR-3030 Power Supply on a sturdy and level surface in a safe and
dry place, away from laboratory traffic.
2. Ensure that the AC power switch is OFF, and then plug the three-pronged
power cord into a grounded three-prong AC outlet of the appropriate voltage
(110V to 240V as indicated on the rating sticker near the AC cord on the
back of the unit).
3. Connect the DC output jacks from the electrophoresis unit; insert the red
lead (+) into the red output jack, and the black lead (-) into the black output
jack.

12
4. Turn on the power switch. The version and model name will appear on the screen.
5. Use Key and Key to select , and then press
the Key or to enter the next screen: .
6. Use , , , or to move cursor to the parameters such
as voltage (V) or current (mA) or power (W) or time (Min.). Press the
key to set the specified parameter.
7. Use or to set the appropriate value, and then press .
Move to the next parameter until all the parameters are set in the same
operation method.
8. Press Key to start electrophoresis. the LED will light up,
and the screen will show the real time parameter values, .
Press or to see the following screen, . Press
or to go back to
9. Press Key to temporarily interrupt the power to the ongoing

13
electrophoresis without terminating the operation. , the LED is
flashing. Press Key to restart the run.
10.Press the Key again to stop electrophoresis.
11.When the run is completed, operation stops with alarm and
is shown on the screen. Press Key to terminate a timed run, and
turn the AC power OFF by the switch on the rear.
12.To change limits of electrophoresis in progress
If you need to make changes to the current running limits, you must stop
electrophoresis by pressing the Key. Press Key to enter the
setting screen, . Enter the changes and then press
Key once again to restart your operation.
Note: After stopping and restarting an operation, the timer resets to selected
time and does not take into account the time that electrophoresis was in
progress before it was stopped.

14
Programming Setup
Instructions for operating PWR-3030 Power Supply in the Programming
Operation are provided in this section. The Programmable Mode allows you
to vary levels in voltage (V), current (mA), and power (W) during specified time
periods as discrete changes (STEP) or as gradients (RAMP) for up to 6 steps,
depending upon your electrophoresis needs. The PWR-3030 power supply is
capable of having storage room for 30 different program files for the user’s
convenience. We recommend reading the guidelines provided in this manual
for best results before starting an operation.
1. Place the PWR-3030 Power Supply on a sturdy and level surface in a safe
and dry place, away from laboratory traffic.
2. Ensure that the AC power switch is OFF, and then plug the three-pronged
power cord into a grounded three-prong AC outlet of the appropriate
voltage (110V to 240V as indicated on the rating sticker near the AC cord
on the back of the unit).
3. Connect the DC output jacks from the electrophoresis unit; insert the red
lead (+) into the red output jack, and the black lead (-) into the black output
jack.
4. Turn on the power switch from the back panel. TheYHUVLRQDQGPRGHOQDPH
will appear on the screen.
5. Use Key and Key to select , and then press
Key or to enter the next screen, .
6. Press Key first and then use or to select appropriate

15
file number, and then press Key to enter the following screen,
.
7. Use , , , or to move cursor to the parameters such
as voltage (V), current (mA), power (W), or time (min.). Press Key to
set the specified parameter.
8. Use or to set the appropriate value, and then press ,
and move to the next parameter until all the parameters are set in the same
operation method.
9. Use Key to move down to for setting Step 4 – 6.
Press Key back to Step 1 – 3 screen, .
10.Press Key to start electrophoresis. , the LED is lit, and the
screen will show the real time parameter values, . Press
or to see the following screen, . Press
or Key back to .
11.Press Key to temporarily interrupt power to ongoing run without

16
terminating the operation. , the LED is flashing. Press Key
to restart the run.
12.When electrophoresis is completed, is shown on the screen.
Press the Key again to stop electrophoresis.
13.To change limits of electrophoresis in progress.
If you need to make changes to the current running limits, you must stop
electrophoresis by pressing the Key. Press Key to enter the
setting screen, . Enter the changes and then press
Key once again to restart your operation.
Note: After stopping and restarting an operation, the timer resets to selected
time and does not take into account the time that electrophoresis was in
progress before it was stopped.

17
Limiting Parameter Setting
The PWR-3030 Power Supply is capable of operating under limiting voltage,
limiting current, or limiting power in both Constant Setup mode or
Programming Setup mode. We use Programming Setup mode here as an
example.
Voltage Limiting
1. Use Key, Key, Key, Key, and Key to set
Maximum Current (3A) and Maximum Power (300W), and on the real time
screen “Volt” will be shaded, for instance, .
Current Limiting
1. Use Key, Key, Key, Key, and Key to set
Maximum Voltage (300V) and Maximum Power (300W), and on the real time
screen “mA” will be shaded, for instance, .
Power Limiting
1. Use Key, Key, Key, Key, and Key to set
Maximum Voltage (300V) and Maximum Current (3A), and on the real time
screen “Wat” will be shaded, for instance, .

18
Troubleshooting Guide
Many operating problems may be solved by carefully reading and following the
instructions in this manual accordingly. Some suggestions for troubleshooting
are given below. Should these suggestions not resolve the problem, contact
our SERVICE DEPARTMENT or a distributor in your region for assistance. If
troubleshooting service is required, please include a full description of the
problem.
Problem Cause Solution
No Display / lights
No AC power Check if PWR power supply is unplugged, or
AC power source problem
AC power cord is not
connected
Check AC power cord connections at both
ends. Use the correct cords.
The fuse has blown Replace the fuse
Repeated fuse
broken Hardware failure Contact MRC service department
Operation stops with
alarm: The screen
displays
“”
Electrophoresis leads are
not connected to the
power supply or to the
electrophoresis unit(s), or
there is a broken circuit in
the electrophoresis cell
Check the connections to the power
supply
and on your electrophoresis
cell to make
sure the connection is
intact; check
condition of wires in electrophoresis unit.
Close the circuit by reconnecting the
cables.
Press START/STOP to restart the run.
High resistance due to
tape left on a pre-cast
gel, incorrect buffer
concentration, or
incorrect buffer volumes
in the electrophoresis cell
Correct the condition by making
sure the
tape is removed from the pre-
cast gel,
buffers are prepared
correctly, and the
recommended volume of buffer is
added to
the electrophoresis unit.
Operation stops with
alarm:
Display shows
Bad connections for
terminal connectors or
damaged wires or
damaged platinum wires
Check all the connections to terminators,
cables, wires, and gel tanks

19
Operation stops with
alarm:
Display shows
Circuit is interrupted
- Verify that the running buffer is correct.
- Verify the all cables are attached correctly
- Turn the Power switch off and
on again;
restart application.
- If you cannot restart the instrument, turn off
the power, disconnect the power cord from
the outlet, and contact Technical Service.
Operation stops with
alarm:
Display shows
Circuit is interrupted
- Verify that the running buffer is correct.
- Verify the all cables are attached correctly
- Turn the Power switch off and
on again;
restart application.
- If you cannot restart the instrument, turn off
the power, disconnect the power cord from
the outlet, and contact Technical Service.
Operation stops with
alarm:
Display shows
Ground leak detected
during run
Check the electrophoresis system for
improper grounding. Restart the power
supply by turning the Power
switch off and on.
Operation stops with
alarm:
Display shows
Power supply is
overheating
- Turn off power supply. Check for sufficient
airflow around the power supply fan. After
cooling down, restart the power supply by
turning the Power switch to the on
position.
- If you cannot restart the instrument, turn off
the power, disconnect the power cord
from the outlet, and contact Technical
Service.
Operation stops with
alarm:
Display shows
and countdown
Power supply is getting
restart ON
- Verify that the electrophoresis condition
and system is correct.
- Pay attention to the own safety.
Table of contents
Other MRC Power Supply manuals