Multi-Amp TR700 User manual

I
()
I
I
J
~
I ,
,
PIN 820111/820131
SIN 8470 &
UP
910624
TRANSFORMER RATIOMETER
TR700/TR800
CAT,
NO, 820110/820130
OPERATING AND
INSTRUcnON
MANUAL

I
1
1
J
oj
(
r-
TABLE OF CONTENTS
PAGE
1.
GENERAL INFORMATION 0 0 0 0 0 0 0 0 • 0 0 0 0 0 0 0
••
••
• • • • 0 • 0 • • • • • • •
••
2
1.l
GENERAL DESCRlYI10N
.......
.
....
.
.•..•..•..•.
, . . . . . . 2
1.2
FEATURES.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . • . . . . . . . . . . . . . . 3
1.3
SPECIFICATIONS. . . . . . . . . . . . . . . . . . . . . • . . • . . . . . . . . . . . .
..
3
1.4
ACCESSORIES FURNISHED .
..
......
.
...
...
.......
....
.
..
5
1.5
ACCESSORIES AVAILABLE
...
. .
......
. .
..........
......
. 6
1.6
CHANGES.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
..
6
1.7
WARRANTY
.
..
....
...........
..
..•..•
0
••••
•
••
• • • • 0 • • • • 7
2. INSTALLATION
.........
.. ..
.
....
....
.......
.
...
..
..
.
.......
8
2.1 UNPACKING AND INSPECTION. . . . . . . . . . . . . . . . . . . . . . •. . . . 8
2.2 PREPARATION FOR USE . .
.........
..
.
..
..
. .
..
.
..
. . .
....
8
2.3
LINE SUPPLY VOLTAGE
......
. . .
..•..
00
.
........
..
.....
8
2.4
REPACKING
AND
SHIPMENT
....•..
..
...
.
..
.
..
.
....•
.
...
8
3.
OPERATING INSTRUCTIONS
.........
...
..
.
...
. 0
••
0 • • • • • • • • •
••
9
3.1 PANEL CONTROLS-AND OPERATING FUNCTIONS . .
..
.
....
. 9
3.2
GENERAL OPERATING INSTRUCTIONS USING A SELF-
CHECK PROCEDURE (Ratio 1:1)
...........
......
.
....
.
..
13
3.3
MEASUREMENT OF SINGLE-PHASE, TWO-WINDING
TRANSFORMERS . . . .
...
.
..
..
..
.
..
.
..
. .
........
. .
..
0 0 • •
17
3.4
SINGLE PHASE AUTO TRANSFORMERS AND
REGUlATORS
...
0 0 • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
21
3.5
DETERMINING TRANSFORMER CONNECTIONS USING
ONLY NAMEPLATE DATA
...........
.
.........
_. .
..
. .
..
24
3.6
lHREE
PHASE TRANSFORMERS (Two Windings) 0 • • • • • • • • •
••
26
3.7
THREE
PHASE AUTO TRANSFORMERS .
........
.
.....
. 0
••
26
3.8
INTERPRETATION AND PROPER USE OF TI:IE RATIO
DEVIATION DIAL
....
..
.
..
..........
.
..
...
0 •
•• ••
• • • 0 • •
29
3.9 CURRENT TRANSFORMERS
......................
..
....
31
3.10 INSTRUMENT VOLTAGE TRANSFORMERS . . . . . . . . . . . . . .
..
34
4.
TROUBLE SHOOTING
..............................•...
_. . . .
37
4.1 SETTING OF DEVIATION DIALS .
.....
. . 0
••••••••••
• 0 • • • •
37
4.2 QUICK REFERENCE TROUBLE CHECK . . 0 • • • •
••
0 0 • • • • • • • •
38
5. SERVICE AND MAINTENANCE
.............
.
....
.
...
.
......
. . _ 39
5.1 DOCUMENTATION . . . . .
....
. . .
..
....
.
..
0 • •
••
0
••
• 0 0 • 0 • • 39
PIN
820
111
/820
131
2

1
1
l
1
1
l
1
l )
1
-,
i
-I
1
i
I
!
!
I.
GENERAL INFORMATION
1.1
GENERAL DESCRIPTION
The
1R800
is
a portable instrument designed for the accurate ratio
measurement of
transformers
and
all devices
with
AC
ratiometric
divider
s
associated with them.
Built
into
the
TR800 is a
current
meter
for
the
measurement
of
the
ex
citing
current
of
the
transformer
under
test. Keeping a
hi
s
tory
ofexcitation
current
has
proved
to
be
a reliable way
of
detecting problems
in
transformer
s.
The phase deviation feature of the
1R800
gives the operator the phase
difference between
primary
and
secondary voltages,
which
is
useful
in
determining
if
there are shorted turns
or
whether the winding has
an
unequal
number
of
turns connected in parallel.
Three-phase switching on the
1R800
simplifies test procedures, saves time
and greatly reduces the risk of electric shock.
1.2
FEATURES
-Built-in three-phase lead switching.
-High ratio range; 1:1 to
2021
:1.
-Ratio's all types
of
C.T.s.
-Full internal protection from over-voltage and shorted output conditions.
-True four terminal measurements for high ·accuracy.
-Direct
reading
of
percent deviation
of
transformer
ratio
.
-Phase angle measurement
provided
to
give more accurate indication
of
transformer
condition.
-Durable transport case provided for longer instrument life.
-Built-in current meter for performing excitation tests.
P
/N
820
111
/820
131 3

l
l
T)
I
1.3
SPECIFICATIONS
P
IN
8201
11
/820131
Specifications:
Bridge:
25
' C
±5'
C 6 months 50
or
60Hz
±2Hz.
Transformer ratio
arm
bridge circuit with phase
sensitive
null
detector.
Bridge
Output Test
Voltage
: 120V, 1A isolated, line dependent.
12V,
1A
isolated, line dependent.
Frequency: 50
or
60Hz depending on line
frequency.
Ratio
Measurements:
(Exclusive
of
% Deviation Dial)
Multiplier
0.1
0.2
05
1.0
2.0
5.0
10.0
20.0
50.0
100.0
Accuracy:
NOTE:
%
Deviation
Dial
Resolution
0.001
O'(lO2
0.005
0.01
0.02
0.
05
0.1
0.2
05
1.0
±0.1%
of
ratio.
Maximum
Range
2.000:1
4.000:1
10.00:1
20.00:1
40.00:1
100.0:1
200.0:1
400.0:1
1000.0:1
2021.0:1
Display by
1R800
is ratio
at
test voltage (12
or
120V)
and
may
not be ratio
of
device
at
its
rated voltage
or
current.
Maximum Range: +0.5
to
-0.5
percent
of
ratio reading.
0.
02%
of
ratio reading.Resolution:
Accuracy:
±2%
of
dial reading.
4

l
1
1
,
')
l
~
II
~
j
j
L
I
!
1.3
SPECIFICATIONS (continued)
1.3
SPECIFICATIONS (continued)
PIN
8201
11
{820131
Phase
Deviation Dial
Multiplier
0.1
1.0
Accuracy:
NOTE:
Excitation
Meter
Multiplier
1.0
10.0
Accuracy
:
Protection
Resolution
O.02CR
0.2
CR
±2% reading.
Maximum
Range
±05CR
±S.OCR
1 centiradian (CR) =
34.S
minutes.
60
minutes = 1 degree.
Resolution
2rnA
20rnA
±3%
fuU
scale.
Maximum
Range
SOmA
SOOrnA
. "
XC'
Input: IS0V RMS
maximum
beforeprotection shorts
XO
to
Xl.
Energizing X side
of
transformer instead
of
H side) will
usually try
to
develop
in
excess
of
lS0V
on
X side
of
TR800 thereby enabling protection
to
short
XO
to Xl.
This prevents any damaging voltage from entering the
instrument
.
Resettable Circuit Breaker:
2A
maximum fast acting; 0.1 second tum-off time. Prevents
any circuit damage due to high currents created when
the
output supply
is
shorted
or
test transformer is connected
backwards.
S

,
I
1
1
1
1
1
f)
1
1.4
1.5
Physical:
Instrument and accessories supplies with portable foam-lined
canying
casC.
Case
has
hinged opening from above
with
a
lockable latch.
Sizes:
Instrument
:
Case: 400 x
220
x 250mm (IS x 9 x
10
inches)
480 x
530
x 370mm (19 x
21
x
14
inches)
Weight: Instrument: 9kg (20 lbs)
12kg (27 lbs)Shipping:
ACCESSORIES FURNISHED
The TR700 Transfonner Ratiometer
is
supplied complete with:
820111 Instruction Manual
820112 Set of 10m leads
Transport Case
AC Line Cord
The TR800 Transformer Ratiometer
is
supplied complete with:
820131 Instruction Manual
820132 Set of 10m Leads
.Transport·Case
AC Line Cord
ACCESSORIES AVAIlABLE
820111
820112
820113
820116
820118
820131
820132
820133
820134
820135
820136
Operating and Instruction Manual, TR700
10M Test Leads
25M Test Leads
Range Extension Transformer, 5:1/2:
1,
Step-up
CCYT Ratio Test Extension Adaptor
Operating and Instruction manual, TR800
10M Test Leads
25M Test Leads
Invertor, 12VDC to 120VAC
10M
Test Lead Extension
25M Test Lead Extension
1.6
CHANGES
P{ N 820
111
/820131
Please Dote that this instrument is subject to continuous development,and
improvement. This instrument may therefore incorporate minor changes in
detail from the information contained herein.
6

I-
I
I
1
1
1
I
lC)
1
1
I
1
1
I
Iu
I
I
1.7
PIN 820
111
/82
0131
WARRANTY
Multi-Amp warrants this instrument sold by us or our authorized agent
s,
to
be
free from defects in material and workmanship for a period of
12
months
from date of shipment. During the warranty period, Multi-Amp will,
at
our
option, repair
or
replace the Instrument or part thereof which proves to be
defective providing:
(1)
The
Instrumentisreturned properlypackedand transportationprepaid
with
prior
authorization
from
us
or
our
appointed agent,
(2)
The
Instrument has not been altered, modified or repaired
by
unauthorized personnel
and,
(3) That
our
examination discloses
to
our
satisfaction
that
any
improper
operation
or
failure
was
the result of defective material
or
workmanship
and
was not the result
of
improper use, negligence
or
accident, exceeding environmental limits,
or
connecting
the
instrument
to incompatible equipment.
This warranty
is
exclusive
and
is
given
and
accepted in lieu of all other
warranties,
express
or
implied,
and
constitutes
flllfil1ment
of
all
our
liabilities
to the purchaser. Multi-Amp specifically disclaims the implied warranties of
merchantability and fitness for a specific purpose. We assume no liability, in
any event,
for
consequential damages, for anticipated or lost profits,
for
personal
injury
due
to
use
or
accident,
for
incidental damages
or
loss
of
time
or
otber losses incurred by the purchaser
or
any third party in connection with
instruinents covered by this warranty
or
otherwise.
7

- '
I
1
-I
-I
1
f )
1-
1
I
1
1
I
1
l)
I
I
2.
INSTALlATION
2.1
2.2
2.3
UNPACKING AND INSPECTION
Prior to shipment this instrument
was
electrically tested and mechanically
inspected and found
to
meet specifications and be free ofmechanical defects.
After unpacking the instrument, visually inspect the instrument and
accessories
for
damage
. H evidence of
damage
is
present YOU
must
contact
the carrier who transported the unit and file a claim
in
writing. The shipp
ing
container and packing material should be retained for inspection
by
the
carrier's agent. Electrical operation per section 3 should be checked
as
soon
as
possible after shipment.
PREPARATION FOR USE
INSURE
THAT1HE
APPARATUS
TO
BE TESTED
IS
CLEARED AND
DE-ENERGIZEDPRIOR
TO
TEST.
It
is
highly
recommended that the user
familiarize himselfwith the controls, functions and features detailed
in
se.ction
3 prior to use.
UNE
SUPPLY VOLTAGE
This instrument
is
shipped from the factory for operation on either
120V,
60Hz line or
220V,
50Hz. Voltage
is
noted on the test report.
To
prevent
damage to the instrument, ensure that the voltage
on
the test 'report matches
your particular line voltage.
2.4
REPACKING AND SHIPMENT
To insure proper shipment
of
thi
s instrument
it
is
recommended that
the
original reusable container and packing material
be
retained.
If
being
returned
for
calibration
or
service,
please
attach
a
card
to
the
instrument
specifying the owner, model and serial number and service required.
P
IN
820111/820131
8

j-
II
~
~
~
l
~
~
~
,
!
/
~
~
~
~
~
~
A~
l
)
II
3.
OPERATING INSlRUCTIONS
3.1
PANEL
CONlROLS
AND OPERATING FUNCTIONS
PIN
8201ll/82OlJl
This section details and describes the operating features of the
lR700
and
lR800.
(Refer
to
Figures 1 and 2). .
1.
Line Input Socket -Standard input for line cord provided.
2.
Instrument ON -Flipping this switch upwards activates the instrument,
immediately sending voltage to the
"H"
output terminals. This
switch
is also a
fast-acting
circuit
breaker,
therefore, protecting
the
instrument
by
shutting itself
off.
To reset, simply flip switch upwards.
3. Power
ON
Indicator -This red
LE.D.
is
illuminated when instrument
is
energized.
4.
5.
6.
7.
8.
9.
Detector Phasing Switch -State of bridge balance of ratio or phase
deviation. Null meter (17)
is
indicator for balance condition.
Detector Sensitivity -
The
sensitivity of
null
meter (17)
is
controlled
with this adjustment. As a general rule, the higher ratio's require
higher sensitivity.
ego
A
1:1
ratio, therefore would require the lowest
gain and a
2000:1
ratio would require a high sensitivity setting.
Phase Multiplier Switch -The approximate range of phase deviation
is
determined with this switch.
Phase Deviation Dial -The precise value of phase deviation
is
determined by this setting in conjunction with the phase
mu
ltiplier
switch (6).
Ratio Multiplier Switch -The approximate range of the ratio being
tested
is
determined with this switch.
Ratio Switches -These switches determine the "Ratio" of the test
transformer, when used in conjunction with ratio multiplier (8).
9

1
-,
f)
1
1
I
1
1
-'1
-I
{_I
I
3.1
PANEL
CONTROLS AND OPERATING FUNCTIONS (continued)
3.1
PANEL
CONTROLS
AND
OPERATING FUNCTIONS (continued)
P
IN
820111/820131
10.
Ratio Deviation
Dial-
Once thecalculated transformer ratio
is
entered
into the ratio switches (9) and ratio multiplier (8), this dial gives a
direct reading
of
percent deviation
of
the test transformer versus the
entered ratio. When balancing ratio switches, this dial should
ALWAYS
be
in
"0"
position.
11.
"X"
Input-This connector couples with the provided leads marked
"XO
to
X3".
12
.
"H"
Input-This connector couples with the provided leads marked
"HO
to
ID
It
,
13.
14
.
15.
16.
17.
18.
19.
20.
21.
"X" Selector -This switch selects input to be measured from
"X
Lead"
connections.
"H" Selector -This switch selects input to
be
measured from
"H
Lead"
connections. .
Polarity -This switch reverses iile polarity of measured ratio and also
acts as a polarity indicator.
Test Voltage Selector -This switch applies 12V
or
120V to the test
transformer. 12V is generally used for C.T.s where cores saturate at
low voltages.
Null Meter -Indicator for balance
of
ratio
and
phase. Detector
phasing switch (4) determines which
is
being balanced.
Excitation Ammeter -
This
meter indicates the amount
of
current
required to excite the test transformer
at
the selected test voltage (16)
under no load conditions.
Excitation Range Switch -
The
appropriate range
of
excitation current
(18)
is
determined with this switch.
"x" Input -These binding posts couple with the provided leads. .
"H" Input -These binding posts couple with the provided lead
s.
10

1
1
1
1
1
I
I
I
I
[( i
[
'-.-
,
I
3.! PANEL CONTROLS AND OPERATING FUNCTIONS (continued)
~
/
N
U)
1ll
/
C20131
~.,
.
..
~
__
.J."
/J-~",
•.
'."
PANEL CONTROLS AND OPERATING FUNCTIONS -TR800
FIGURE!
11

•
( )
3.1 PANEL CONTROLS
AND
OPERATING FUNCTIONS (continued)
Hx
-
uo
uO
--~
-
PANEL CONTROLS AND OPERATING FUNCTIONS· TR700
FIGURE 2
12

3.2
l)
PIN
820
lll
/8
20
m
GENERAL· OPERATING INSlRUCTIONS USING A SELF·CHECK
PROCE
DURE
(Ratio
1:1)
lR800
(Refer to Figure I)
1.
Insure that instrument power switch (2)
is
in down position.
2.
3.
4.
5.
Connect high side leads(H markings on alligator clips) to
"H"
input(12). Connect low side leads (X markings on alligator clips) to
"X"
input(ll)
.
Short all alligator ends as follows:
HO
to
XO
HI
to
XI
H2 to X2 H3 to
X3
If
line cord is not plugged into instrument, do so now.
Set the
lR800
instrument controls as follows
Detector phasing switch: Ratio.
Detector sensitivity: Midway.
Phase multiplier switch:
0.1.
Phase deviation dial
:·
O.
Ratio multiplier switch:
0.1.
Ratio switches:
10
..
0 .
O.
Percent deviation dial:
O.
"X"
Selector
switch
:
X3
-
X2
.
"H"
Selector switch: H3 -
H2
.
Polarity switch: Normal.
Test voltage selector: 12V.
Excitation range switch: x
10.
6. Ensure that personnel are kept clear ofalligator clips. Clips should.be
clear of ground or any metal objects, as they will be at a potentia
lly
lethal voltage.
7.
Tum
power switch ON.
8.
Observe "Null Detector" needle.
It
should be pointing towards
"0".
If
not, obtain centre balance
by
adjusting "Percent Deviation Dial". Dial
should remain at approximately
"0"
(±I division) with needle balanced
at centre
"0"
position.
9. The excitation meter should read approximately "
0"
.
If
not, just above
the
"0"
line.
13

I·
•
·f)
1
1
3.2 GENERAL OPERATING
INSTRUcnONS
USING A SELF-CHECK
PROCEDURE (continued)
PIN
820
111
/820131
10.
Set 'Test Voltage Selector" to
"120V".
Re-adjust "Percent Deviation
Dial". Itshould still balance at approximately
"0".
If
needle movement
is
too sensitive, set "Detector
Sensitivity"
back until needle movement
is
comfortable to operator (this does not affect accuracy).
If
all of the above goes accordingly, the instrument has been balanced to a
ratio of
1.000
:1.
11.
Switch detector phasing switch to "Phase".
12.
Observe "Null Detector" needle. It should point towards centre
"0"
,
if
not, obtain centre balance
by
adjusting phase deviation dial (7). Dial
should remain at approximately
"0"
(±I
division) with needle balanced
at
centre
"0"
position.
If
all ofsteps
11
and
12
go
accordingly, the instrument has been balanced to
a phase deviation of O.O.CR (centiradians) which indicates the internal
reference transformer
is
okay. The following steps will verify the correct
operation of the
"X"
and
"H"
selector switches
(13
&
14)
and the
corresponding H and X leads.
13.
Turn instrument OFF.
14.
Set "R" selector to "
HZ
-HI". Set
"X
" selector to
"X2
-XI".
15.
Turn instrument ON.
16.
"Null Detector" needle should be balancedwith detector phasingswitch
on "Ratio" and "Phase.
17.
Repeat steps
13
to
15.
Exchanging the following:
"-
HZ
-
HI
to
HI
-
ID.
X2 -
XI
to
XI
-
X3,
then,
b.
HI
-
ID
to
HI
-
HO.
XI
-
X3
to
XI
-
XO,
then,
c.
HI
-
HO
to
HZ
-
HO
.
XI
-
XO
to X2 -
XO,
then,
d.
HZ-HOtoID
-
HO.
X2-XOtoX3-XO
.
This should
fully
complete a self-test of the instrument
as
well
as
the test
leads.
14

1
L!
I
I
3.2 GENERAL OPERATING INSTRUCTIONS USING A SELF-CHECK
PROCEDURE (continued)
3.2
PIN
820
111
/82013
1
GENERAL OPERATING INSTRUCTIONS USING A SELF-CHECK
PROCEDURE (continued)
TR700 (Refer to Figure 2)
1.
Insure that instrument power switch (2) is in down position.
2.
Connect high side leads ("H" markings on alligator clips) to
"H"
input
(2
1)
(HlC
, HIP, H2C and H2P).
3. Connect low side leads ("X" markings on alligator clips) to
"X"
input
(20)
(Xl
and X2).
4. Short all clip ends as follows:
HI
to
Xl
H2
to X2
5.
If
line
cord
is
not plugged into
instrument
,
do
so
now
.
6.
7.
Set the TR700 instrument controls as follows
Detector phasing switch: Ratio.
Detector sensitivity: Midwa
y.
Phase multiplier switch: 0.
1.
Phase deviation dial: O.
Ratio multiplier switch: 0.
1.
Ratio switches:
10.
- 0 -
O.
Percent deviation dial:
O.
Test voltage selector:
l2V
.
Ensure that personnel are kept clear ofalligator clips. Clips should be
clear of ground or any metal objects, as they will be
at
a potentially
lethal voltage.
B.
Turn power switch ON.
9.
Observe "Null Detector" needle.
It
should be pointing towards
"0"
.
If
not, obtain centre balance
by
adjusting "Percent Deviation Dial".
Dial
should remain at approximately
"0"
(±l
divi
sion) with needle balanced
at
centre "
0"
position.
15

.,
'1
'1
1
-I
-I
1
f
-I
1
I
1
1
1
l(
1
1
'!
3.2
GENERAL OPERATING
INSTRUcnONS
USING A SELF-CHECK
PROCEDURE (continued)
3.2
P
IN
820111/820131
GENERAL OPERATING
INSTRUcnONS
USING A SELF-CHECK
PROCEDURE
(continued)
10.
Set "Test Voltage Selector" to
"120V".
Re-adjust "Percent Deviation
Dial".
It
should still balance at approximately
"0".
If
needle
m~vement
is
too sensitive, set "Detector Sensitivity" back until
needle
movement
is
comfortable to operator (this does not affect accuracy).
If
all of the above goes accordingl
y,
the instrument has been balanced to a
ratio of 1.000:
1.
11.
Switch detector phasing switch to "Phase".
12
. Observe "Null Detector" needle.
It
should point towards centre
"0"
, if
not, obtain centre balance
by
adjusting phase deviation dial (7). Dial
should remain
at
approximately
"0"
(±1 division) with needle balanced
at centre
"0"
position.
If
all of steps
11
and
12
go
accordingly, the instrument has
been
balanced to
a phase deviation of 0.0
CR
(centiradians) which indicates the internal
reference
transformer
is okay.
This should fully complete a self-test of the instrument as well as the test
leads.
16

I
J-
3.3
I
1
1
1
1
1
1
]C
)
1
1
J
PIN 820111/820131
MEASUREMENT
OF
SINGLE-PHASE,
TWO-WINDING
TRANSFORMERS (Refer to Figures 1 and 2)
1.
Insure that instrument power switch (2)
is
in down position.
2.
Connect high side leads
("H"
marking on alligator clips) to "
H"
input
(12 & 21).
3.
Connect l
ow
side leads
("X"
markings on alligator clips) to
"X"
input
(11
& 20).
4. Assuming the transformer under test has the following markings:
FIGURE
3
Connect the
TRSOO
leads to the corresponding markings:
"H"
Leads -H1 to
Hl.
"X"
Leads -
Xl
to
Xl.
-H2 to H2. -
X2
to
X2.
5. Ensure that the remaining leads (H3,
HO,
X3,
XO)
are free and
clear of the test transformer and any grounded points, as they
may
accidentally
be
energized.
6.
Set the instrument controls as follows:
Detector phasing switch: Ratio.
Detector sensitivity: Midway.
Phase multiplier switch:
0.1.
Phase deviation dial:.
O.
Ratio multiplier switch:
0.1.
Ratio switches:
S.
- 0 -
O.
Percent deviation dial
O.
"X"
Selector switch:
X2
-
Xl.
"H" Selector switch: H2 -
Hl.
Polarity switch: Normal.
Test voltage selector:
12V.
Excitation
range
switch:
x
10.
17

I
l
i
1
3.3
3.3
PIN 820111j820131
MEASUREMENT
OF
SINGLE-PHASE
,
TWO-WINDING
TRANSFORMERS (continued)
MEASUREMENT
OF
SINGLE-PHASE, TWO-WINDING
TRANSFORMERS (continued) .
It
should be noted that the test voltage,
12V
is
used in this application to
achieve
an
approximate
balance,
and
therefore
ratio,
unless test specimen
is
a current transformer where upon
12V
is
used to achieve final balance
as
well
7.
Ensure that personnel are kept clear of alligator clips and test
transformer.
8.
9.
10
.
Turn power switch ON.
Observe
"Null
Detector" needle (17).
It
will be pointing to the left
side of the meter.
(If
not, check polarity switch (normal) and lead
connections
("H"
on high side and
"X"
on low
side»
.
While still observing
"Null
Detector",
tum
ratio multiplier
switch
clockwise until null detector (17) swings to the right. Turn ratio
multiplier
switch
counter-clockwise one position.
ego
Needle
swings
right on
"10"
multiplier. Switch should be placed
in
"5"
multiplier before continuing
to
next
step.
11.
With the ratio in approximate range (needle pinned to left), rotate left
most ratio switch (9)
cloc1cwise
(start on
"8")
until null indicator
swings
right. Turn ratio switch back one position.
12.
Rotate middle ratio switch clockwise (start
on
"0")
until null indicator
swings right. Turn ratio switch back one position.
13.
Rotate right most ratio switch clockwise until needle either swings
right or comes to a balance in the centre of the meter.
14.
If
null detector needle
is
still settled either left or right of centre,
rotate percent deviation dial (10) until needle balances to zero.
15.
Switch detector phasing switch to "Phase".
16. Adjust phase deviation dial (7) until "Null Detector" (17) balances
to
centre
"0".
18

i
()
(-
I
3.3
MEASUREMENT
OF
SINGLE-PHASE,
TWO-WINDING
TRANSFORMERS (continued)
3.3
MEASUREMENT
OF
SINGLE, TWO-WINDING TRANSFORMERS
(continued)
PIN 820111/820131
17
.
If
balance is not possible, switch phase multiplier switch to "I" and re-
attempt step
16.
18.
Once balance
is
achieved, set ''Test Voltage Selector" to
"120V"
and re-
balance. (Adjustments should be minimal). 120V should only be used
for C.T.s that have sufficient core to handle this voltage. Otherwise,
12V
is
used on
ITs.
EXAMPLES
OF
DIAL INTERPRETATION
After the ratiometer is balanced, the ratio of the transformer under test
is
determined as follows:
IT
ratio
deviation
dial
is
not
used, i.e.
set
·
at
0,
then;
Ratio = Multiplier x Reading on Ratio Dials
Examples:
1.
If
multiplier
is
0.1 and ratio dials read
17
.32, then:
Ratio = 0.1 x 17.32 = 1.732 or 1.732:1
2.
If
multiplier is
0.5
and ratio dials read 10.02, then:
Ratio = 0.5 x 10.02 = 5.01
or
5.01:1
19

I
I
I
I
f j
J
J
I
J
.I
J
leI
I
3.3
MEASUREMENT
OF
SINGLE-PHASE,
TWO-WINDING
TRANSFORMERS (continued)
PIN 820
1I1
f82()13l
If
ratio deviation dial (10)
is
used for balancing, determine ratio
as
on
previous page and multiply obtained ratio
by
a factor
1000
+ D
to
obtain
exact value.
1000
Examples:
1.
If
ratio dial reads
10.00,
multiplier
is
2.0 and deviation dial reads +
1.5,
then:
Ratio = 2 x
10.00
x
1000
+
1.5
1000
= 20.00 x
1001,5
1000
=
20
.
03
The
ratio deviation
dial
is
most
useful
when
percent deviation
from
a
predetermined value has to be determined.
If
the deviation dial reads
positive then the ratio of the transformer under test
is
higher than the preset
value. Similarly,
if
the ratio
of
the transformer under test is lower
than
the
preset value, then the deviation dial will read a negative value.
Examples:
Name plate ratio 4800 volts
to
120
volts.
Ratio = 40.00
Set multiplier and ratio dials
to
obtain a ratio of 40.
Two
possibilities exist: M = 2, R = 20.00 (19 + 1 +
0)
or,
M =
5,R=8.0(8
+ 0 + 0)
Now
balance only with ratio deviation dial. Deviation dial reads in
mils
(1
mil = 1/1000 or 0.1%)
as
to
how
high
or
l
ow
the ratio
is
from nameplate.
When obtaining ratio readings, the second and third ratio dials
may
exceed
9.
If
that occurs they should be read in the following manner:
8 +
10
+ 9 =
9.09
19
+
11
+
10
= 20.20
10
+ 0 +
10
=
10.10
20
This manual suits for next models
3
Table of contents
Popular Transformer manuals by other brands

Zenith
Zenith 122C-A - Heath - Lock-Nut Transformer installation instructions

Goobay
Goobay 65581 user manual

marklin
marklin 6646 user guide

ProMariner
ProMariner ProSafe 21036 Owner's manual and installation guide

AquaStar
AquaStar LST100 Installation, operation and service manual

music first audio
music first audio MC Step-up owner's manual

HAMPTON BAY
HAMPTON BAY DIY-900PS Use and care guide

Keysight
Keysight 16060A Operation and service manual

Simpson Electric
Simpson Electric 186 Specifications

VOLTCRAFT
VOLTCRAFT SPAS-2100 operating instructions

Coilcraft
Coilcraft Planar Transformer Prototyping Kit manual

Delta Electronics
Delta Electronics Current Sense Transformers TCE0502H Specification sheet