Navair 17-20AQ-347 User manual

NAVAIR 17-20AQ-347
TECHNICAL MANUAL
INSTRUMENT CALIBRATION PROCEDURE
DIGITAL MULTIMETER
WAVETEK
1271
WITH OR WITHOUT
OPT12 OR 20
DESTRUCTION NOTICE - FOR UNCLASSIFIED, LIMITED DOCUMENTS, DESTROY BY ANY METHOD
THAT WILL PREVENT DISCLOSURE OF CONTENTS OR “RECONSTRUCTION OF THE DOCUMENT.”
PUBLISHED BY DIRECTION OF COMMANDER NAVAL AIR SYSTEMS COMMAND
DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U.S. GOVERNMENT AGENCIES AND THEIR
CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR OFFICIAL USE OR FOR ADMINISTRATIVE OR
OPERATIONAL PURPOSES ONLY. OTHER REQUESTS FOR THIS DOCUMENT SHALL BE REFERRED TO
MEASUREMENT SCIENCE DEPARTMENT, CORONA DIVISION, NAVAL SURFACE WARFARE CENTER,
P.O. BOX 5000, CORONA, CA 92878-5000.
AQ-347
AQ-347
1 AUGUST 2007
THIS PUBLICATION SUPERSEDES NAVAIR 17-20AQ-347
DATED 1 SEPTEMBER 1997

NAVAIR 17-20AQ-347
A
LIST OF EFFECTIVE PAGES
The purpose of this list is to identify the pages in this document affected by any technical content changes made since the
previous release of the document.
NOTE: On a revised page, with the exception of the Title, the A, and the i pages, the technical changes are indicated by a vertical
line in the outer margin of the page.
Page
No. **
Revision* Page
No. **
Revision* Page
No. **
Revision*
Title 0
A 0
i 0
1 to 17 0
19 to 21 0
*Zero in this column indicates an original page.
“R” in this column indicates a revised page.
**All pages not listed are blank pages.

NAVAIR 17-20AQ-347
i
TABLE OF CONTENTS
Section Page
1 Introduction and Description................................................................................................................... 1
2 Equipment Requirements......................................................................................................................... 3
3 Preliminary Operations............................................................................................................................ 4
4 Calibration Process.................................................................................................................................. 7
4.1 Direct Voltage Tests............................................................................................................................. 7
4.2 AC RMS Voltage Tests........................................................................................................................ 10
4.3 Resistance Tests (Option 20 Only)....................................................................................................... 15
4.4 Calibration Factors............................................................................................................................... 16
4.5 Securing The TI.................................................................................................................................... 17
Navy Calibration Checklist...................................................................................................................... 19
ILLUSTRATIONS
Figure Page
1 DC Volts Reference Setup (20 V or less)................................................................................................ 5
2 DC Voltage Tests (20 V or less).............................................................................................................. 8
3 DC Voltage Setup (100 V or More)......................................................................................................... 9
4 True RMS Voltages (1 V and Higher)..................................................................................................... 10
5 Measurement for 100 mV Range............................................................................................................. 14
6 Ohm Test ................................................................................................................................................. 15
TABLES
Table Page
1 Calibration Description............................................................................................................................ 1
2 Equipment Requirements......................................................................................................................... 3

NAVAIR 17-20AQ-347
1
SECTION 1
INTRODUCTION AND DESCRIPTION
1.1 This procedure describes the calibration of the Wavetek 1271 Digital Multimeter with or without OPT12 or 20. The
instrument being calibrated is referred to herein as the TI (Test Instrument).
1.2 All comments concerning this procedure should be directed to the Measurement Science Department, Corona Division,
Naval Surface Warfare Center, P.O. Box 5000, Corona, CA 92878-5000.
1.3 This procedure includes tests of essential performance parameters only. Any malfunction noticed during calibration,
whether specifically tested for or not, should be corrected.
1.4 The TI is not calibrated to 1 MHz above 10 V and not calibrated over 500 V. It is calibrated using the manufacturer’s
12 month specifications. This is an NSWC Measurement Science Department (MSD) assigned tolerance.
1.5 It should be noted that the specifications cited below, in Table 1, are for the Fluke 8505/8506 series digital multimeters
(unless noted otherwise). Since the Fluke specifications were the “procurement specifications” for the Wavetek 1271 (TI), a
SPECIAL CALIBRATION tag shall be affixed to the TI, upon completion of calibration, indicating that the TI was calibrated to
the (12 month) Fluke 8505/8506 specifications. This is an NSWC/MSD assigned specification.
Table 1. Calibration Description
TI
Characteristics Performance
Specifications Test
Method
Direct voltage Ranges: 100 mV, 1, 10, 100 and 1000 V dc
Tolerances:
100 mV: ±(0.00403% iv + 90 counts)
1 V: ±(0.0024% iv + 9 counts)
10 V: ±(0.0019% iv + 9 counts)
100, 100 V: ±(0.00297% iv + 9 counts)
Tested with a voltage
measurement system
containing a voltage divider,
reference divider, and DC
voltage standard.
Resistance Ranges: 10, 100 Ω, 1, 10, 100 kΩ, 1, 10,
100 MΩ
Tolerances:
10 Ω: ±(100 ppm + 20 counts)
100 Ω: ±(60 ppm + 1.4 counts)
1, 10, 100 kΩ, 1 MΩ:
±(60 ppm + 0.8 counts)
10 MΩ: ±(398 ppm + 0.8 counts)
100 MΩ: ±(1004 ppm + 1 count)
Compared with a known
resistance.
AC RMS Voltage Tolerances:
100 mV range:
10 to 40 Hz: ±0.152% iv
40 to 20 kHz: ±(0.035% iv + 5 μV)
20 to 50 kHz: ±0.0825% iv
50 to 100 kHz: ±0.308% iv
Compared with known direct
voltages by thermal voltage
conversion.
10 V range:
10 to 40 Hz: ±0.152% iv
40 to 20 kHz: ±0.025% iv
20 to 50 kHz: ±0.0825% iv
50 to 100 kHz: ±0.308% iv
100 to 200 kHz: ±0.689% iv
200 to 500 kHz: ±2.04% iv
500 kHz to 1 MHz: ±4.49% iv

NAVAIR 17-20AQ-347
2
TI
Characteristics Performance
Specifications Test
Method
AC RMS Voltage
(continued) 100 V range:
10 to 40 Hz: ±0.152% iv
40 to 20 kHz: ±0.025% iv
20 to 50 kHz: ±0.0825% iv
50 to 100 kHz: ±0.308% iv
1000 V range:
10 to 40 Hz:
iv%
600
V
x)02.0152.0( 2
in
⎥
⎥
⎦
⎤
⎢
⎢
⎣
⎡
⎟
⎟
⎠
⎞
⎜
⎜
⎝
⎛
+±
40 Hz to 20 kHz:
iv%
600
V
x)02.0025.0( 2
in
⎥
⎥
⎦
⎤
⎢
⎢
⎣
⎡
⎟
⎟
⎠
⎞
⎜
⎜
⎝
⎛
+±
20 to 50 kHz:
iv%
600
V
x)02.00825.0( 2
in
⎥
⎥
⎦
⎤
⎢
⎢
⎣
⎡
⎟
⎟
⎠
⎞
⎜
⎜
⎝
⎛
+±
50 to 100 kHz:
iv%
600
V
x)02.0308.0( 2
in
⎥
⎥
⎦
⎤
⎢
⎢
⎣
⎡
⎟
⎟
⎠
⎞
⎜
⎜
⎝
⎛
+±
NOTES
1. Volt-Hertz product must be <107for the Wavetek 1271.
2. The 500 V range voltage coefficient error is ppm200x
600
V2
in ⎟
⎟
⎠
⎞
⎜
⎜
⎝
⎛
3. Slow filter must be used for AC RMS voltage tests below 100 Hz.

NAVAIR 17-20AQ-347
3
SECTION 2
EQUIPMENT REQUIREMENTS
NOTES
Minimum use specifications are the principal parameters required for performance of the calibration, and are
included to assist in the selection of alternate equipment, which may be used at the discretion of the using
laboratory. Satisfactory performance of alternate items shall be verified prior to use. All applicable
equipment must bear evidence of current calibration.
The instruments utilized in this procedure were selected from those known to be available at Navy calibration
facilities and the listing by make or model number carries no implication of preference, recommendation, or
approval for use by other agencies. It is recognized that equivalent equipment produced by other
manufacturers may be capable of equally satisfactory performance in this procedure.
Table 2. Equipment Requirements
Item Minimum Use
Specifications Calibration
Equipment
2.1 Direct voltage
measurement system Range: 10 mV to 1000 V
Uncertainties:
100 mV: ±0.003% iv
1 V: ±0.000825% iv (at 1 V)
10 V: ±6 ppm (at 19 V)
100 V: ±0.001% iv (at 100 V)
1000 V: ±0.001% iv (at 1000 V)
Fluke 7105A
Required Components:
2.1.1 DV source Fluke 332B or 335A
2.1.2 Voltage divider Fluke 720A
2.1.3 Reference divider Fluke 750A
2.1.4 Null detector Note: Not required if item 2.1.1 is a Fluke 335A
(p/o Fluke 7105) Fluke 845AB
2.1.5 DC transfer standard
(ERS) Fluke 732AAN, 731B
2.2 Alternating voltage source
(AV) Voltage range: 0 to 1000 V ac
Uncertainty: based upon a comparison of the
output voltage by the TTS (item 2.10) with the
direct voltage measurement system (item 2.1)
Frequency range: 10 Hz to 1 MHz
Uncertainty: based upon a comparison of the
output voltage by the TTS (item 2.10) with the
direct voltage measurement system (item 2.1)
Fluke 5200A and 5205A
2.3 Resistor, standard Resistance: 10 Ω, ±0.007% Leeds and Northrup 4025B
2.4 Resistor, standard Resistance: 100 Ω
Uncertainty: NA (actual value is measured) Leeds and Northrup 4030B
2.5 Resistor, standard Resistance: 1 kΩ
Uncertainty: NA (actual value is measured) Leeds and Northrup 4035B
2.6 Resistor, standard Resistance: 10 kΩ
Uncertainty: NA (actual value is measured) Leeds and Northrup 4040B

NAVAIR 17-20AQ-347
4
Item Minimum Use
Specifications Calibration
Equipment
2.7 Resistor, standard Resistance: 100 kΩ
Uncertainty: NA (actual value is measured) Electro-Scientific Industries
SR1-100K
2.8 Resistor, standard Resistance: 1 MΩ
Uncertainty: NA (actual value is measured) Electro-Scientific Industries
SR1-1M
2.9 Resistor, standard Resistance: 100 MΩin 10 MΩsteps
Tolerance: ±0.01% iv Electro-Scientific Industries
SR1-1050-10M STEP
2.10 Thermal transfer standard
(TTS) Certified with a Report of Calibration from NPSL Fluke 540B
2.11 Resistor, standard Resistance: 1 Ω±0.003% Leeds and Northrup 4020B
SECTION 3
PRELIMINARY OPERATIONS
3.1 Ensure that all power switches are set to off, and set all auxiliary equipment controls as necessary to avoid damage to the
equipment and so that dangerous voltages will not be present on output terminals when the power switches are turned on.
3.2 Ensure that the following equipment controls are set as follows:
Reference Divider (item 2.1.3)
STD CELL CIRCUIT switch to OPEN
INPUT VOLTAGE switch to 100 V
OUTPUT VOLTAGE switch to 100 V
ERS (item 2.1.5)
Voltage-Output controls to Vref (its 10 V Assigned-Reference-Output voltage)
Voltage Divider (item 2.1.2)
Voltage Dials set to one-twentieth of proceeding
20
V
:V ref
ref
Null Detector (item 2.1.4
OPR-ZERO switch to ZERO
RANGE switch 300 μV (to be changed as required later)
DV Source (item 2.1.1)
POWER switch to STDBY/RESET
VOLTAGE RANGE switch to 100
OUTPUT DIALS for 22 V
VOLTAGE TRIP vernier to fully CW
METER switch to VOLTAGE
CURRENT LIMIT control to approximately ¼ of fully CW
position (about 15 mA)

NAVAIR 17-20AQ-347
5
3.3 Connect the applicable equipment as shown in Figure 1, the reference setup for low voltage tests. Turn the power
switches on and allow the setup to warm-up for 1 hour. (During warm-up time, proceed with steps 3.4 through 3.10.)
3.4 Turn the TI power switch to ON.
3.5 Allow the TI to warm-up for 4 hours. Ensure that the TI CAL switch (TI rear panel) is set to DISABLE.
DV SOURCE
NULL
DETECTOR
+
-
VOLTAGE DIVIDER
HI LO GND
ERS
+
-
1.1 1.0
+
-
GUARD OUTPUT
Figure 1. DC Volts Reference Setup (20 V or less)
3.6 Make a copy of the following table. Using the Report of Calibration for the TTS (item 2.10), fill in the values of the ±%
AC-DC difference (column 2) that are applicable to the voltage ranges and frequency values listed. For frequencies below 1 kHz,
use the 1 kHz AC-DC difference value.
2.1.5
2.1.2
2.1.4
2.1.1

NAVAIR 17-20AQ-347
6
Frequency TTS Range
(1)
±% AC-DC
Difference
(2)
Correction Factors
CF=0.01X(Column 1)X(Column 2)
(3)
10 Hz 1 V
50 Hz 1 V
1 kHz 1 V
20 kHz 1 V
50 kHz 1 V
100 kHz 1 V
1 MHz 1 V
10 Hz 10 V
50 Hz 10 V
1 kHz 10 V
20 kHz 10 V
50 kHz 10 V
100 kHz 10 V
1 MHz 10 V
10 Hz 100 V
50 Hz 100 V
1 kHz 100 V
20 kHz 100 V
50 kHz 100 V
100 kHz 100 V
10 Hz 1000 V
50 Hz 1000 V
1 kHz 1000 V
20 kHz 1000 V
50 kHz 1000 V
100 kHz 1000 V
3.7 To complete the chart, fill in the CF (Correction Factor) values by multiplying the following three items: the TTS voltage
range, the ±% AC-DC difference, and the value 0.01 (to convert % to decimal). Fill in CF values to 1 μV resolution. For
example, if at a TTS Range of 100 V and at a frequency of 20 kHz, a +0.001% AC-DC difference is listed, the CF value is
(100)(+0.001)(0.01) = +0.001000 V; if the % AC-DC difference is a -0.001%; the CF value would be (100)(-0.001)(0.01) =
-0.001000 V.
3.8 Set all other calibration equipment controls, as necessary, to avoid damage to the equipment and so that dangerous
voltages will not be present on output terminals when the power switches are turned on. Connect the equipment to their
appropriate power sources, turn all power switches on, and allow a sufficient warm-up time for the equipment.
3.9 Set the null detector (item 2.1.4) OPR-ZERO switch to ZERO. Perform the mechanical or electrical zero adjustments for
the null detectors, as required. Electrically zero adjust the null detectors up to their 3 μV ranges; then set null range selectors, as
necessary, to avoid damage.
3.10 Measure items 2.4 (100 Ω) through 2.8 (1 MΩ) as described in NAVAIR 17-20AR-29 to establish the nominal values of
the standard resistors. Record the measured values as the reference values in step 4.3.6.

NAVAIR 17-20AQ-347
7
SECTION 4
CALIBRATION PROCESS
NOTE
Unless otherwise specified, verify the results of each test and take corrective action whenever the test
requirement is not met, before proceeding.
4.1 DIRECT VOLTAGE TESTS
4.1.1 Zero Tests
4.1.1.1 Connect a low thermal short to the TI HI and LO terminals. Depress the TI ZERO pushbutton. Wait until the TI
BUSY light is extinguished. Depress the TI DCV pushbutton, and depress the RANGE pushbutton to select from the TI RANGE
listed in the following table, one range at a time.
4.1.1.2 Verify that the TI indication is within the tolerance limits listed in the following table.
TI RANGE Nominal
(V) Tolerance Limits
(V)
100 mV 0.0000-3 -0.0090-3 to +0.0090-3
1 V 0.000000 -0.000009 to +0.000009
10 V 0.00000 -0.00009 to +0.00009
100 V 0.0000 -0.0009 to +0.0009
1 kV 0.000 -0.009 to +0.009
4.1.1.3 Disconnect the low thermal short from the TI INPUT HI and LO terminals.
4.1.2 Establish Reference Voltage. Referring to Figure 1, set the DV source power switch to OPR. Adjust the controls of the
DV source and the null detector, as necessary, to obtain a reference null on the null detector. The DV source output indication at
null should be approximately 22 V.
4.1.3 Taking precautions, disconnect the ERS (item 2.1.5) and the null detector from the reference setup. Connect the TI as
shown in Figure 2.

NAVAIR 17-20AQ-347
8
DV SOURCE
VOLTAGE DIVIDER
HI LO GND
GUARD
Ω
GUARD
1.1 1.0
+
-
INPUT
TI
HI LO
I-
I+
GUARD OUTPUT
Figure 2. DC Voltage Tests (20 V or less)
4.1.4 Depress the TI RANGE controls to select the following TI RANGE. Set the applicable equipment controls for the
settings indicated, and verify that the TI indications are within the tolerance limits listed in the following table.
TI RANGE Divider Setting Nominal
(V) Tolerance Limits
(V)
100 mV 0.0005000 10.0000-3 9.9906-3 to 10.0094-3
100 mV 0.0050000 100.0000-3 99.9869-3 to 100.0131-3
1 V 0.0050000 0.100000 0.099989 to 0.100011
1 V 0.0500000 1.000000 0.999967 to 1.000033
10 V 0.0250000 0.50000 0.49990 to 0.50010
10 V 0.0375000 0.75000 0.74990 to 0.75010
10 V 0.0700000 1.40000 1.39988 to 1.40012
10 V 0.1300000 2.60000 2.59986 to 2.60014
10 V 0.2550000 5.10000 5.09981 to 5.10019
2.1.2
2.1.1

NAVAIR 17-20AQ-347
9
TI RANGE Divider Setting Nominal
(V) Tolerance Limits
(V)
10 V 0.4000000 8.00000 7.99976 to 8.00024
10 V 0.5000000 10.00000 9.99972 to 10.00028
10 V 0.7000000 14.00000 13.99964 to 14.00036
10 V 0.9500000 19.00000 18.99955 to 19.00045
Reverse output polarity of DV source.
10 V 0.9500000 -19.00000 -18.99955 to -19.00045
4.1.5 Set the controls of the DV source for minimum output. Disconnect the test setup.
4.1.6 Connect the equipment as shown in Figure 3.
TI INPUT
+
- OUT
DV
SOURCE
+
- +-
NULL DET STD CELL
HI
OUTPUT
LO
HI
INPUT
LO
REFERENCE
DIVIDER
NULL DETECTOR
INPUT
COMM
GND
+
-
GND
DC TRANSFER STD
(ERS)
Figure 3. DC Voltage Setup (100 V or More)
4.1.7 Set the ERS output controls for its certified 1 volt reference output voltage and set the reference divider standard cell
dials for the value of the 1 volt reference.
4.1.8 Zero the null detector on its 10 μV range; then, set its OPR-ZERO switch to OPR. Set the DV source power switch to
OPR and the ERS (item 2.1.5) power switch to ON.
4.1.9 Set the reference divider STD-CELL switch to MOMENTARY long enough to establish a reference null indication
within ±5 μV on the null detector, by adjusting the DV source voltage output controls and the reference divider coarse and fine
controls, as required. Increase the null range sensitivity of the null detector, as necessary, during the adjustment for the reference
null. The DV source output indication at null should be approximately 100 V.
4.1.10 Set the reference divider STD-CELL switch to LOCKED. Adjust the DV source voltage-output controls and the
reference divider input adjust controls, as required, to maintain the reference null on the null detector.
2.1.4 2.1.5
2.1.1
2.1.3

NAVAIR 17-20AQ-347
10
4.1.11 Perform the measurements tabulated below as follows:
4.1.11.1 Set the null detector OPR-ZERO switch to ZERO and set the equipment controls, as required, to obtain the settings
listed in the following table for the TI RANGE, the reference divider Input-Output switches, and the DV source. Depress the TI
RANGE pushbutton to select the TI 100 V or 1000 V RANGE, as required.
4.1.11.2 Set the null detector OPR-ZERO switch to OPR. Adjust the DV source voltage output controls and the reference
divider coarse and fine adjust controls to maintain a null on the null detector 10 μV range, taking precautions as required to avoid
meter pegging.
TI RANGE
(V) Input
(V) Output
(V) DV Source
(V) Nominal
(V) Tolerance Limits
(V)
100 100 100 100.000 100.000 99.9961 to 100.0039
1000 1000 1000 1000.00 1000.00 999.961 to 1000.039
4.1.11.3 Verify that the TI indication is within the tolerance limits listed.
4.1.12 Set the DV source power switch to STDBY/RESET, and disconnect the test setup.
4.2 AC RMS VOLTAGE TESTS
4.2.1 Connect the equipment as outlined in Figure 5 (line power input from a common source).
DV SOURCE
+
-
GUARD SEN OUT
TI
TTS
DC AC
AV SOURCE
GUARD
SEN OUT
I+
LO
GUARD
Ω
GUARD
HI
I-
Figure 4. True RMS Voltages: 1 V and Higher
4.2.2 Set the following equipment controls:
4.2.2.1 TI controls:
FUNCTION ACV
2.2
2.1.1
2.10

NAVAIR 17-20AQ-347
11
RANGE 1 V
FILTER 10 Hz
REMOTE GUARD REM Ω(light on)
MEASUREMENT RESOLUTION 5 digit
NOTES
To select ACV filter:
Depress TI CONFIG pushbutton
Select Filt function
Select 10 Hz filter
Depress TI ACV pushbutton
To select REMOTE GUARD:
Depress the TI INPUT pushbutton
Select REM Ωfunction
Depress TI ACV pushbutton
4.2.2.2 AV source (item 2.2) controls:
Voltage Range to 1 V
Voltage output setting 1 V
Frequency controls to 50 Hz
Vernier Voltage Error to 0%
Voltage Error % SW X.1
Sense switch to EXT
Mode switch to OPER
Phase Lock (5200) OFF
4.2.3 Perform the DC to AC Transfer measurements tabulated below as follows:
4.2.3.1 Set the DV source output controls for a voltage value equal to the difference between 1 volt and the CF value
(correction factor) that is applicable to the frequency of the measurement listed in the following table. For example, if the
measurement frequency is 20 kHz and the CF at 20 kHz (from step 3.6) is -0.000030, set the DV source output controls for
1.000030 V (because: 1-(-0.000030) =1.000030); if CF is +0.000020, set DV source for 0.999980 V.
4.2.3.2 Set the TTS mode switch to the DC Transfer position, selecting the correct range so that the percent input meter will
indicate in the green area. Establish a galvanometer null at its most sensitive range while maintaining an onscale meter indication
by adjusting the applicable reference adjust control when switching to a more sensitive range.
NOTE
Allow for stabilization time. The Fluke 540B may initially take as long as 10 minutes; however, it is
considered stabilized when the galvanometer indication drifts <2 small divisions per minute in its most
sensitive position. Re-adjust the reference adjust controls to re-establish the galvanometer null.

NAVAIR 17-20AQ-347
12
AV Source
DV Source
Output Frequency % Error Indication Tolerance Limits
(V)
1-CF 10 Hz* 10 Hz 0.99848 to 1.00152
1-CF 50 Hz* 50 Hz 0.99975 to 1.00025
1-CF 1 kHz 1 kHz 0.99975 to 1.00025
1-CF 20 kHz 20 kHz 0.99975 to 1.00025
1-CF 50 kHz 50 kHz 0.99918 to 1.00082
1-CF 100 kHz 100 kHz 0.99692 to 1.00308
1-CF 1 MHz 1 MHz 0.95510 to 1.04490
*Select 10 Hz ACV filter.
4.2.3.3 Set the TTS GALV switch to open and the sensitivity switch to low.
4.2.3.4 Set the TTS mode switch to AC search, selecting the correct range switch setting so that the percent input meter
indicates in the green area; then switch the mode switch to the AC transfer position.
4.2.3.5 Vary the % error mode controls of the AV source, as required, to re-establish the galvanometer null indication. Do
not adjust the TTS reference controls while in the AC MOD.
4.2.3.6 Obtain a null indication successively in the “medium” and “high” positions of the TTS sensitivity switch by adjusting
the % error controls of the AV source. Make note of the AV source % error indication.
4.2.3.7 Set the TTS GALV switch to open, sensitivity switch to low and polarity switch to the opposite setting as in
step 4.2.3.3.
NOTE
If the TTS polarity switch has been disconnected, skip to step 4.2.3.12.
4.2.3.8 Repeat steps 4.2.3.2 through 4.2.3.6.
4.2.3.9 Set the TTS GALV switch to open and the sensitivity switch to low.
4.2.3.10 Set the AV source % error control to midway of the two settings noted in steps 4.2.3.6 and 4.2.3.8.
4.2.3.11 Verify that the TI indication is within the tolerance limits listed in the following table.
4.2.3.12 Record the AV source ±% error indications on a separate piece of paper or fill in the values on a duplicate copy of
the tabulated measurements. These values will be referred to during the calibration of the TI 100 mV range in step 4.2.11.1.
4.2.4 Set the AV source error controls for 0% error indication and initially, set the equipment controls, as necessary, for the
DC to AC transfer measurement values listed in the following table. Perform step 4.2.3.2 to establish a DC reference null.
DV and AV Source
Output
TTS and TI
RANGE
(V) (V) (Freq)
Characterized
V ac Equivalent
(V) Tolerance Limits
(V)
10 10.0000 10 Hz* 9.9848 to 10.0152
10 10.0000 50 Hz* 9.9975 to 10.0025
10 10.0000 1 kHz 9.9975 to 10.0025
10 10.0000 20 kHz 9.9975 to 10.0025
10 10.0000 50 kHz 9.9918 to 10.0082

NAVAIR 17-20AQ-347
13
DV and AV Source
Output
TTS and TI
RANGE
(V) (V) (Freq)
Characterized
V ac Equivalent
(V) Tolerance Limits
(V)
10 10.0000 100 kHz 9.9692 to 10.0308
10 10.0000 1 MHz 9.5510 to 10.4490
100 100.000 10 Hz* 99.848 to 100.152
100 100.000 50 Hz* 99.975 to 100.025
100 100.000 1 kHz 99.975 to 100.025
100 100.000 20 kHz 99.975 to 100.025
100 100.000 50 kHz 99.918 to 100.082
100 100.000 100 kHz 99.692 to 100.308
NOTE
Set the AV source controls for minimum output; disconnect the low voltage output connections of
the AV source from the TTS AC input terminals and connect the AV source power output (high
voltage) terminals to where the low voltage (100 V and less) were previously connected.
1000** 500.000 10 Hz* 499.240 to 500.760
1000** 500.000 50 Hz* 499.875 to 500.125
1000** 500.000 1 kHz 499.875 to 500.125
1000** 500.000 20 kHz 499.875 to 500.125
1000** 200.000 50 kHz 199.835 to 200.165
1000** 100.000 100 kHz 99.692 to 100.308
* Select 10 Hz ACV filter.
** 6 digit resolution required.
4.2.5 Set the TTS GALV switch to open, and the sensitivity switch to low.
4.2.6 Set the TTS mode switch to AC search, select the correct range so that the percent input meter indicates in the green
area and then switch the mode switch to the AC transfer position.
4.2.7 Adjust the AV source output voltage controls (not the % error controls) to re-establish the galvanometer null indication.
Do not adjust the TTS reference control while in the AC Mode.
4.2.8 For each transfer measurement, perform the following:
4.2.8.1 Verify that the difference between the TI indication and the applicable CF (correction factors to compensate for the
TTS % AC-DC difference) is within the tolerance limits listed in the following table.
4.2.8.2 If more than one TI is to be calibrated, record the AV source final output voltage indication on a separate piece of
paper (or on a duplicate copy of the preceding tabulated measurements). These characterized voltage values may be used
(re-dialed) to calibrate other TIs when connected directly to the AV source within 24 hours.
4.2.9 Set equipment output controls for minimum output and disconnect the test setup.
4.2.10 Connect the equipment as outlined in Figure 6. Both pieces of equipment should be connected to the same power
source.

NAVAIR 17-20AQ-347
14
TI
AV SOURCE
GUARD
SOURCE SENSE
CHASSIS HI
LO HI LO
I+ I-
GUARD ΩGUARD
Figure 5. Measurement for 100 mV Range
4.2.11 Select the AV source 1 V range and perform the following measurement.
4.2.11.1 Depress the TI controls to select the 100 mV RANGE. Set the AV source controls to the indicated voltage value, to
the indicated frequency, and to the ±% error setting applicable to the frequency of measurement as obtained in step 4.2.3.8.
4.2.11.2 Verify that the TI indication is within the tolerance limits listed in the following table.
AV Source
TI RANGE
(mV) Output Volts Frequency Tolerance Limits
100 0.100000 10 Hz* 99.848-3 to 100.152-3
100 0.100000 50 Hz* 99.960-3 to 100.040-3
100 0.100000 1 kHz 99.960-3 to 100.040-3
100 0.100000 20 kHz 99.960-3 to 100.040-3
100 0.100000 50 kHz 99.918-3 to 100.082-3
100 0.100000 100 kHz 99.692-3 to 100.308-3
*Select 10 Hz ACV filter.
4.2.12 Set the equipment controls for minimum output and disconnect the test setup.
2.2

NAVAIR 17-20AQ-347
15
4.3 RESISTANCE TESTS (OPTION 20 ONLY)
4.3.1 Connect the 10 Ωstandard resistor (item 2.3) to the TI as shown in Figure 7.
STANDARD RESISTOR
TI
HI LO
I-I+ TI
Figure 6. Ohm Test
4.3.2 To measure the 10 Ωstandard resistor, perform these steps:
Depress the TI OHMS pushbutton
Depress the TI CONFIG pushbutton
Select the CHG function
Select the Tru Ωfunction
Select the 10 Ωrange
Depress the TI CONFIG pushbutton
Select the Resl function
Select 5 digit resolution
Depress the TI OHMS pushbutton
4.3.3 Connect a short across the 10 Ωstandard resistor.
4.3.4 Depress the TI ZERO pushbutton. Wait until the TI BUSY light is extinguished.
4.3.5 Remove the short from across the 10 Ωstandard resistor.
4.3.6 Using Figure 7 for the test setup, connect each of the standard resistors required for the measurements tabulated in the
following table. Set the TI ohms range manually (using the correct pushbuttons) to the range indicated. For each measurement,
verify that the TI indication is within the tolerance limits listed.
2.3 THROUGH 2.9

NAVAIR 17-20AQ-347
16
NOTES
The nominal values applicable for the 100 Ω through 1 MΩmeasurements in the following table are the
reference values that were measured in step 3.10 as described in NAVAIR 17-20AR-29.
To select the 100 MΩrange,perform these steps:
Depress TI CONFIG pushbutton
Select the CHG function
Select the HI Ωfunction
Select the 100 MΩrange
Depress the TI OHMS pushbutton
Test Resistance
Nominal Value TI RANGE TI
Tolerance Limits
10 Ω10 Ω9.9970 to 10.0030 Ω
100 Ω100 ΩRef ±0.0074%
1 kΩ1 kΩRef ±0.0068%
10 kΩ10 kΩRef ±0.0068%
100 kΩ100 kΩRef ±0.0068%
1 MΩ1 MΩRef ±0.0068%
10 MΩ10 MΩ9.9959 to 10.0041 MΩ
100 MΩ100 MΩ99.899 to 100.101 MΩ
4.3.7 Disconnect the TI from the 100 MΩstandard resistor.
4.4 CALIBRATION FACTORS
4.4.1 If all ranges and quantities are within tolerance, skip to step 4.5.1. Otherwise, enter the TI calibration factors as
follows. Perform only those steps required for the quantities that tested out-of-tolerance.
4.4.2 To begin setting the calibration factors engage the CAL switch on the rear panel of the TI.
4.4.3 Perform an INTERNAL CHARACTERIZATION. This is accomplished by performing the following (in the order
shown): press CAL, select EXT, QUIT, and then TRIG.
4.4.4 Upon completion of the internal characterization, perform the FAST TEST. This is accomplished by selecting TEST
and then selecting FAST.
4.4.5 If the FAST TEST was satisfactorily completed, then perform the FULL TEST. The FULL TEST is accomplished by
selecting TEST and then selecting FULL. If the FAST TEST was unsatisfactory, then consult the manufacturer’s manual to
troubleshoot the specific problem.
4.4.6 Upon satisfactory completion of the full test, proceed to enter the CALIBRATION DATE. This is accomplished by
performing the following (in the order shown): press CAL, select EXT, QUIT, QUIT, enter the date, and then select ENTER. If
the FULL TEST was unsatisfactory, then consult the manufacturer’s manual to troubleshoot the specific problem.
4.4.7 Once the calibration date has been entered, then disengage the CAL switch on the rear of the TI.

NAVAIR 17-20AQ-347
(18 blank) 17
4.4.8 Repeat the sections of the TI calibration that had previously failed. If these sections are completed satisfactorily, then
proceed to Section 4.5. Otherwise, consult the manufacturer’s manual to troubleshoot the problems encountered.
4.5 SECURING THE TI
4.5.1 Any steps that produced an out-of-tolerance reading must be repeated at this point. Use the TI SECOND RUN
checklist column to repeat the necessary steps and verify that these steps are now within the given tolerance limits.
4.5.2 Affix a SPECIAL CALIBRATION label to the TI indicating the following: “TI calibrated to Fluke 8505/8506 12
months specs (MSD assigned tolerance).” The date of calibration may by added to the TI memory, if desired, by following the
instructions in the manufacturer’s manual.
4.5.3 Unless other measurements are to be performed, set all power controls to off or standby, and disconnect the TI from the
setup.
Table of contents
Popular Multimeter manuals by other brands

Major tech
Major tech MT1885 instruction manual

Extech Instruments
Extech Instruments MiniTec Series user manual

Gossen MetraWatt
Gossen MetraWatt METRACLIP 86 user manual

VOLTCRAFT
VOLTCRAFT VC 220 manual

Robin
Robin AR6006 operating instructions

Gossen MetraWatt
Gossen MetraWatt METRACAL MC operating instructions