Newport PHAMP-3 User manual

Additional products from
Newport Electronics,Inc.
THE NEW STANDARD FOR QUALITY
In the USA and Canada: 800-NEWPORT
In Mexico 95-800-NEWPORT
Or call your local Newport Office.
This documentation may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium
or machine readable form, in whole or in part, without prior
written consent of NEWPORT ELECTRONICS, INC.
© 1995 Newport Electronics, Inc. All rights reserved.
This product may be covered by one or more
of the following patents:
United States Patents Des. 336,895; 5,274,577
France Brevet No. 91 12756
Spain 2039150
United Kingdom Patent No. 2248954
and other international patents pending
From the technical library of:
__________________________________________________________________
Clock/Timers
Counters
Flow Sensors
Frequency Meters
Multimeters
On/Off
Controllers
pH Controllers
pH Electrodes
pH Pens
PID Controllers
Printers
Process Meters
Rate Meters
Recorders
Relative Humidity
RTDs
Soldering Iron
Testers
Strain Gauge
Meter
Thermistors
Thermocouples
Thermowells
Timers
Totalizers
Transmitters
Voltmeters
Wire
PHAMP-3
pH Preamplifier
With Automatic Temperature Compensation
Operator’s Manual
M2266/N/0895 Newport Electronics,Inc.
Warranty
All Products from NEWPORT ELECTRONICS, INC. are warranted
against defective material and workmanship for a period of one (1)
year from the date of delivery.
If the unit should malfunction, it must be returned to the factory for
evaluation. NEWPORT’s Customer Service Department will issue an
Authorized Return (AR) number immediately upon phone or written
request. Upon examination by Newport, if the unit is found to be
defective it will be repaired or replaced at no charge. However, this
WARRANTY is VOID if the unit shows evidence of having been
tampered with or shows evidence of being damaged as a result of
excessive corrosion; or current, heat, moisture or vibration; improper
specification; misapplication; misuse or other operating conditions
outside of NEWPORT’s control. Components which wear or which are
damaged by misuse are not warranted. These include contact points,
fuses, and triacs.
In addition to NEWPORT’s standard warranty period, NEWPORT
ELECTRONICS will extend the warranty period for one (1) additional
year only if the warranty card enclosed with each instrument is
returned to NEWPORT.
Newport is glad to offer suggestions on the use of its various
products. Nevertheless, NEWPORT warrants only that the parts
manufactured by it will be as specified and free of defects.
NEWPORT MAKES NO OTHER WARRANTIES OR REPRESENT-
ACTIONS OF ANY KIND WHATSOEVER, EXPRESSED OR
IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED
WARRANTIES INCLUDING ANY WARRANTY OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are
exclusive and the total liability of NEWPORT with respect to this
order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the
purchase price of the component upon which liability is based. In
no event shall NEWPORT be liable for consequential, incidental
or special damages.
Every precaution for accuracy has been taken in the preparation of
this manual; however, NEWPORT neither assumes responsibility for
any omissions or errors that may appear nor assumes liability for any
damages that result from the use of the products in accordance with
the information contained in the manual.
SPECIAL CONDITIONS: Should this equipment be used in any
nuclear installation or activity, purchaser will indemnify NEWPORT
and hold NEWPORT harmless from any liability or damage
whatsoever arising out of the use of the equipment in such a manner.
Return Requests
Direct all warranty and repair requests/inquiries to the NEWPORT
Customer Service Department. BEFORE RETURNING ANY
PRODUCTS(S) TO NEWPORT, PURCHASER MUST OBTAIN AN
AUTHORIZED RETURN (AR) NUMBER FROM NEWPORT’S
CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be
marked on the outside of the return package and on any
correspondence.
The purchaser is responsible for shipping charges, freight, insurance
and proper packaging to prevent breakage in transit. NEWPORT’s
warranty does not apply to defects resulting from action of the buyer,
mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification.
FOR
WARRANTY
RETURNS,
please have the following infor-
mation available BEFORE con-
tacting NEWPORT:
1. P.O. number under which
the product was PUR-
CHASED,
2. Model and serial number of
the product under warranty,
and
3. Repair instructions and/or
specific problems relative to
the product.
FOR
NON-WARRANTY
REPAIRS,
consult NEWPORT
for current repair charges. Have
the following information avail-
able BEFORE contacting NEW-
PORT:
1. P.O. number to cover the
COST of the repair,
2. Model and serial number of
product, and
3. Repair instructions and/or
specific problems relative to
the product.

PHAMP-3
pH Preamplifier TABLE OF
CONTENTS PHAMP-3
pH Preamplifier
TABLE OF
CONTENTS PHAMP-3
pH Preamplifier
TABLE OF
CONTENTS
General Description
The PHAMP-3 is a unity gain
preamplifier which converts the
high impedance mV signal of a
pH electrode to a low impedance
signal which has been compensated
for variations in temperature. The
signal can travel 1,000 feet over
ordinary wire and connectors to
process meters and controllers (such
as the IDP and INFC).
The PHAMP-3 requires an external
power source between 10-40 volt
DC. It can be powered by the
excitation of a process meter (such
as the NEWPORT IDP) or the
optional PSU-12V power supply.
The PHAMP-3 uses a l000 ohm RTD
sensor for automatic temperature
compensation. (RTD not included.
Fixed resistors must be connected
for applications not using an RTD).
The electronics are entirely
encapsulated in a compact epoxy
filled enclosure designed for easy
mounting in a manufacturing
process. The PHAMP-3 offers
redundant outputs from either a
BNC connection, terminal strip or
both at the same time. PHAMP-3 Accessories
¥ BNC-(M) to BNC-(M) 3 ft
extension cable (part number
BNC-C-36/N)
¥ 12 Volt dc power supply (part
number PSU-12 V/N).
Specifications
Output Offset: 3 mV typical; (.05 pH)
Input: -1 to 15 pH
Input Impedance: 1013 Ohms
Output Impedance: 20K Ohms
Output Voltage: 472mV to -472mmV
Operating
Temperature: 0 to 60°C
Power: 10 to 40 Vdc
Dimensions: 38.1 H x 76.2 W x 38.1mm D
(1.5" x 3" x 1.5")
INPUT
OUTPUT from Electrode
to Meter
+
+--
12345
Power
supply
Input
pH Output
to Meter
ATC
Operating Instructions
Connect as shown below:
Power supply connects to
Terminals 1&2. pH input
from electrode connects to
BNC connector. pH output
to meter(s) can be connected
to Terminals 3&4 or BNC.
RTD (1K ohm) connnects
to Terminals 4&5. (Terminal
#4 is common for pH output
& ATC).
The table (below) shows how the
millivolt output of a pH electrode
and the resistance of a 1k ½ RTD
varies with temperature. The
PHAMP-3 compensates temperature
to 25¡C.
Temperature Compensation
Temp. 0°C 25°C 40°C 50°C 70°C 90°C 100°C
pH mv. mv. mv. mv. mv. mv. mv.
0 +379.3 +414.0 +434.9 +448.8 +476.6 +504.4 +518.2
1 +325.1 +354.9 +372.8 +384.7 +408.5 +432.3 +444.2
2 +270.1 +295.8 +310.7 +320.6 +340.5 +360.3 +370.2
3 +216.8 +236.6 +248.5 +256.5 +272.4 +288.2 +296.1
4 +162.6 +177.5 +186.4 +192.4 +204.3 +216.2 +222.1
5 +108.4 +118.3 +124.2 +128.2 +136.2 +144.1 +148.1
6 +54.19 +59.15 +62.13 +64.12 +68.09 +72.05 +74.03
70000000
8 –54.19 –59.15 –62.13 –64.12 –68.09 –72.05 –74.03
9 –108.4 –118.3 –124.2 –128.2 –136.2 –144.1 –148.1
10 –162.6 –177.5 –186.4 –192.4 –204.3 –216.2 –222.1
11 –216.8 –236.6 –248.5 –256.5 –272.4 –288.2 –296.1
12 –270.1 –295.8 –310.7 –320.6 –340.5 –360.3 –370.2
13 –325.1 –354.9 –372.8 –384.7 –408.5 –432.3 –444.2
14 –379.3 –414.0 –434.9 –448.8 –476.6 –504.4 –518.2
1KΩ
RTD 1000Ω1097Ω1155Ω1194Ω1271Ω1347Ω1385Ω
Automatic Temperature
Compensation