Novalynx 230-601C User manual

Barometric Pressure
NovaLynx Corporation PO Box 240 Grass Valley CA 95945
DOC 230-601C IS 20210526 +1 (530) 823-7185 nova@novalynx.com
www.novalynx.com
230-601C Barometric Pressure Sensor
The 230-601C Barometric Pressure Sensor offers unparalleled
performance and flexibility for atmospheric pressure measurement. It
combines high accuracy and low power over a wide range of pressures
and temperatures.
The sensor is protected by a weatherproof polycarbonate enclosure. The
enclosure can be wall mounted or pole mounted with the addition of the
395-A-003 Mast Mounting Hardware kit.
The standard 5-conductor cable is 25' (7.6 m) long. If additional cable is
required please specify at the time of order.
BP Sensor with Enclosure
A
OUTPUT (TX)
+
PWR (7-30 VDC)
-
PWR
4-20 mA SIGNAL
B
OUTPUT (RX)
EARTH GROUND
GREEN
RED
BLACK
WHITE
BROWN
SHIELD
RX-B
I-OUT
REF
+ PWR
TX-A / SDI
Pressure Port
-
-
-
-
-
DWG 820-0004-01
230-601C
BAROMETRIC PRESSURE
DEFAULT OUTPUT
RANGE: 500-1100 hPa
SIGNAL: 4 to 20 mA
Consult RM Young
Model 61402L
instruction manual
for jumper settings to enable other options
INSULATE UNUSED WIRES
Ordering Information
230-601C Barometric Pressure Sensor, 4-20 mA output with enclosure and 25' cable
395-A-003 Mast Mounting Hardware (for use with enclosure)
330-0524 Cable 5C 24AWG Shielded, per foot

Barometric Pressure
NovaLynx Corporation PO Box 240 Grass Valley CA 95945
Phone (530) 823-7185
www.novalynx.com
Mast Mounting Bracket
(U-bolt not shown)
Polycarbonate Enclosure
90 mm

R.M. YOUNG COMPANY
2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN 49686, USA
TEL: (231) 946-3980 FAX: (231) 946-4772
PN: 61402L-90
REV: A040819
1.0 SPECIFICATIONS
Pressure 500 to 1100 hPa standard range
Digital Accuracy* 0.2 hPa (25°C)
0.3 hPa (-40°C to +60°C)
Analog Accuracy** 0.05% of analog pressure range
Analog Temperature
Dependence** 0.0017% of analog pressure
range per °C (25°C reference)
Long Term Stability: 0.08% FS per year
*
Denedas±1standarddeviationfromNIST-traceable
pressure reference in clean, dry air. Includes non-
linearity, hysteresis, repeatability, and calibration
uncertainty.
** Definedas±1standarddeviationfromidealanalog
output. Total analog output accuracy is the root sum
square of digital accuracy, analog accuracy, and analog
temperature dependence.
Output Rate 1.8 Hz (max) to 1 per minute
Current Output 4 to 20 mA or 0 to 20 mA
12-bit resolution (1 in 4000)
Selectable pressure range: 500 to
1100 hPa (standard)
Serial Output Full-duplex RS-232
Half-duplex RS-485
SDI-12
1200 to 38400 baud
Continuous ASCII text
Polled ASCII text
NMEA
0.01 hPa resolution
Supply Voltage 7 to 30 VDC
25 mA max (4-20 or 0-20 mA mode)
1.5 mA max in SDI-12 mode
8.5 mA (RS-232 or RS-485)
Case Fiber-reinforced thermoplastic
Weight 44 g (1.5 oz)
INSTRUCTIONS
MODEL 61402L
BAROMETRIC PRESSURE SENSOR

Format Description
Continuous ASCII Text
(standard)
9600 baud
16sampleaverage
1.8 Hz output
Polled ASCII Text
9600 baud
10sampleaverage
Output when polled
Marine NMEA
4800 baud
10sampleaverage
1 Hz output
SDI-12 1200 baud
Software Output parameters determined by
software settings
See wiring configurations for more information.
In POLLED ASCII mode, the unit sends data only after
receivingapollcommand.ThepollcommandisMa!where
‘a’istheaddressvalue.Thestandardaddressis‘0’(ASCII
48)forapollcommandofM0!Pleaseseesection4.3for
details on changing the poll address.
Uponreceivingaproperlyaddressedpollcommand,theunit
immediately sends data.
4.3 SOFTWARE COMMANDS
Softwarecommandssentviaserialcommunicationmaybe
used to set operational parameters. The sensor must be
configured for serial communication and connected to a PC
orothercompatibledeviceusingaprogramlikeHyperTerm.
New settings are stored in temporary memory and
must be burned to flash with command CMD420 to be
retained.
NotethatallP1jumpers(A,B,C)mustberemovedfor
SOFTWAREformattochangebaudrate,sampleaverage
count, and output rate parameters.
Althoughthesensormayreceivecommandsatany
time, measurement output may be paused by sending
three ESC (ASCII 27) characters or CMD300 1. While
paused, COMMAND mode is indicated by a “>” prompt.
All commands must be terminated with a carriage return
CR (ASCII 13).
Use“?”tolistavailablecommandsasshownbelow.
>?
YOUNG 61402L SERIES BAROMETER
------------------------------------------------
CMD100 n OUTPUT PERIOD (0-60 sec)
CMD105 n AVG SAMPLE COUNT (0-32)
CMD110 n OUTPUT FORMAT
1=ASCII 2=POLLED 3=NMEA 4-RAW 5-DIAG)
CMD120 n MULTIPLIER (10000 = 1.0000)
CMD125 n OFFSET (hPA x 100)
CMD130 n SEA LEVEL CORRECTION (hPA x 100)
CMD140 n IOUT RANGE HI (500-1100)
CMD142 n IOUT RANGE LO (500-1100)
CMD144 n IOUT SCALE (1=0-20mA 4=4-20mA)
CMD150 c POLL ADDR CHAR
CMD 1 60 n BAUD RATE (38400,19200,9600,4800,2400,1200)
CMD300 n 1=STOP, 2=RUN
CMD310 n FORCE IOUT (0-20000 FULL SCALE)
CMD400 n 1=REPORT TEXT, 2=REPORT CODE ONLY
CMD410 GET CAL TABLE
CMD420 STORE SETUP PARAMETERS
>
CMD100 n sets output period in seconds. When set to zero,
the output rate is 1.8 per second.
CMD105 n setstheaveragesamplecount.Highersample
count yields more stable readings but will
lengthen the startup delay.
CMD110 n sets output format.
CMD120 n sets the output multiplier. 10 00 0 repres en ts
1.0000
CMD125 n sets the output offset in hPa x 100
CMD130 n setsthesealevelcorrectioninhPax100
=[Punadjust *( Mult / 10000)] +[(Offset /100)+(Correction Sea Lvl /100)]
2.0 INTRODUCTION
MODEL61402LBarometricPressureSensorisaversatile
electronic barometer featuring high accuracy, low power,
wide operating temperature range, and calibrated 4-20 mA
output, with SDI-12 and RS-232/485 serial data I/O.
The4-20mAoutputandRS-485serialI/Omakethe
61402L ideal for commercial and industrial applications that
commonly use these signals.
Full scale current output may be set to span any portion of
the 500 to 1100 hPa operating range. Serial output options
include continuous or polled ASCII text output, NMEA marine
output or SDI-12. The sensor is supplied with the following
standard configuration:
Current Output 4-20 mA = 500 to 1100 hPa
RS-232 Continuous ASCII text output 9600 baud
16samplesaveraged
1.8 Hz update rate
3.0 INSTALLATION
Thesensoroperatesoveratemperaturerangeof-40°Cto
+60°C and must remain dry. Electrical connections are made
with the 5-position terminal plug included with the sensor.
Jumper settings may be used to select operating options.
Access jumpers by loosening two screws in the sensor
enclosureandremovingthecover.Seelastpageforwiring
details and jumper settings.
4.0 OPERATION
Operation begins approximately 1 second after power is
applied. Output will appear after a delay determined by
thenumberofsamplesaveragedandtheoutputrate.With
defaultvalues(1.8Hzoutputand10samplesaveraged),
output begins after an initial 5 second delay then remains
continuous at the 1.8 Hz rate. Parameters for output rate
andnumberofsamplesaveragedmaybechangedusing
software commands.
4.1 CURRENT OUTPUT
Calibrated4-20mAcurrentoutputisalwaysactive.The
connection scheme appears in the Wiring configurations
section.
The current output scale may be mapped to any part of the
500 to 1100 hPa pressure range using the IOUT HI and
IOUT LO software parameters. The standard setting maps 4
to 20 mA with 500 to 1100 hPa. With this setting:
hPa = 37.5 * mA + 350
Effectiveresolutionisabout0.15hPa
Greatereffectiveresolutionmaybeachievedbymapping
the current output to a narrower range of pressure with the
IOUT HI and IOUT LO parameters. For example, with IOUT
LO set to 950 hPa and IOUT HI set to 1050 hPa:
hPa = 6.25 * mA + 925
Effectiveresolutionisabout0.025hPa
Please see section 4.3 below for details on changing the
IOUT HI and LO parameters.
4.2 SERIAL OUTPUT
The61402Lmaybeconguredforfull-duplexRS-232,half-
duplex RS-485, or SDI-12 by using the lower jumper pair. In
addition, the upper jumpers must be set as follows:
Lower Upper Output
RS-232 (A,B,C) Continuous ASCII Text
RS-485 (B,C) Polled ASCII Text
SDI-12 (C only) SDI-12
Standardbaudrateis9600butseveralbaudratesbetween
1200and38400areavailable.
SDI-12baudrateisxedat1200.

CMD140 n sets the IOUT Pressure Range HI
CMD142 n sets the IOUT Pressure Range LO IOUT
Pressure Range HI must be greater than IOUT
Pressure Range LO
CMD144 n sets the IOUT Scale:
1 = 0 to 20mA, or 4 = 4 to 20 mA.
CMD150 c sets the poll address character for ASCII and
SDI-12 modes. Any ASCII character from ‘0’
(ASCII 48) to ‘z’ (ASCII 122) may be used. For
SDI-12 limit the characters from 0 to 9.
CMD160 n setsthebaudrate.Acceptablevaluesare
38400, 9600, 4800, 2400, and 1200. Note
that the baud rate change occurs immediately
afterthecommandisreceived.Therefore,in
order to continue communicating, the serial
communication program that issued the
command must also change to match.
CMD300 n causes the sensor to pause or resume
measurement. 1 = Stop, 2 = Run.
CMD400 n
causes the sensor to issue a report
summarizing parameter settings.
1=Text(parametervaluesandlabels)
2=CodeOnly(parametervaluesonly).
CMD420 burns the current parameter settings to flash
memory. This step musttakeplacetoretain
changes to settings otherwise parameters will
reverttopreviousvaluesatthenextpowerup.
4.4 SDI-12 COMMANDS
SDI-12 stands for serial data interface at 1200 baud. It is
often used to interface battery powered data recorders with
microprocessorbaseddevicesdesignedforenvironmental
data acquisition. SDI-12 is used typically when low power
isaconcern.TheSDI-12devicenormallyremainsina
low power, standby state until it is polled at which time
ameasurementissent.SDI-12caneffectivelyaddress
multiple sensors on the same cable.
Model 61402L uses the SDI-12 (V1.4) serial communication
protocol to initiate measurements and set sensor operation
parameters. The default sensor address is 0 (zero), and
canbechangedtoanyvalidsingle-charactervalueif
needed. Additional details about the SDI-12 protocol may
be found at www.sdi-12.org.
After initial power-up with 12 VDC, the sensor is in a low-
power standby state with a quiescent current of 1.60 mA. A
validandproperlyaddressedSDI-12commandwakesthe
sensortoinitiateameasurement,setorcheckoperating
parameters. After command processing has finished, the
sensor returns to the low-power standby state.
SDI-12 ‘M’ or ‘C’ commands initiate a measurement. The
sensor response message indicates the maximum time
neededbeforethemeasurementisready,anddatavalues
willbeavailable.Themaximumtimerangesfrom1to5
seconds depending on the Sample Count.
If an ‘M’ command initiates a measurement, the sensor
sendsaServiceRequestwhenthemeasurementisready
toberetrieved.Whenthepollingdevicereceivesthe
ServiceRequestitthensendsa‘D’commandtorequest
measurement results from the sensor. With ‘C’ commands,
noServiceRequestissent,andthepollingdevicemust
wait the full delay time before sending the ‘D’ command to
request the measurement.
SDI-12 Measurement commands and responses:
COMMAND RESPONSE DESCRIPTION
aM! attts<CR><LF> Take a measurement
aMC! attts<CR><LF> Take a measurement with CRC
checksum
aC! atttss<CR><LF> Take a concurrent measurement
aCC! atttss<CR><LF> Take a concurrent measurement
with CRC checksum
aD0! a+tttt.t+bbbb.b<CRC><CR><LF>
where:
a= Sensor address
ttt = Delay time (seconds)
s/ss = Number of samples to collect
ttt.t = Temperature (selected units)
bbbb.b = Barometric Pressure (selected units)
<CRC> = CRC checksum (only where requested)
<CR><LF> = Carriage return, line feed (ASCII 13, 10)
SDI-12 NON-MEASUREMENT COMMANDS:
The SDI-12 protocol includes standard commands for
identifying the sensor and changing its address.
CMD RESPONSE DESCRIPTION
?! a<CR><LF> Address Query
a! a<CR><LF> Acknowledge Active
aI!
a13YOUNGC092000vvvnnnnnn<CR><LF>
Send Identification: vvv = Firmware Version
nnnnnn = Serial Number
aAb! b<CR><LF> Change Address:
a = Sensor address
b = New sensor address
aV! a0000<CR><LF> Start Verification
EXTENDED COMMANDS:
The SDI-12 command set may be customized with
Extended Commands to accommodate manufacturer
settings and other functions. Extended Commands are
listed below. Where two responses are shown, one is for a
validcommand,theotherisforaninvalid(ERR)command.
COMMAND RESPONSE DESCRIPTION
aXB! aXB, OK<CR><LF>
Save parameter settings to flash.
Cycle power after saving parameters.
aXP!
aXP,V=v,OP=o,ASC=s,MP=m,OS=op,SC=sc<CR><LF>
v = voltage (120 = 12.0 volts)
o =
s =
m =
op = offset (hPa x 10)
sc = sea level correction (hPa x 10)
aXC±nnnn! aXC,SC=±nnnn<CR><LF>
Sea Level Correction (-5000 to +5000 hPa)
aXF±nnnn! aXF,OP=±nnnn<CR><LF>
Offset (-5000 to +5000 hPa)
aXM aXM, M=nnnnn<CR><LF>
Multiplier (10000 = 1.0000)
aXO aXO, O=nn<CR><LF>
Output Period (0 – 60)
aXS aXS, S=nn<CR><LF>
Sample Count (1 – 32)
5.0 MAINTENANCE
The MODEL 61402L barometer requires no regular
maintenance. Periodic calibration certification, if required,
isavailablefromthefactory.
6.0 WARRANTY
This product is warranted to be free of defects in materials
and construction for a period of 12 months from date of
initial purchase. Liability is limited to repair or replacement
ofdefectiveitem.Acopyofthewarrantypolicymaybe
obtained from R. M. Young Company.

7.0 EMC COMPLIANCE
ThisdevicecomplieswithPart15oftheFCCRules.Operation
is subject to the following two conditions: (1) this device may
notcauseharmfulinterference,and(2)thisdevicemustaccept
anyinterferencereceived,includinginterferencethatmaycause
undesired operation.
This equipment has been tested and found to comply with the
limits for a ClassA digital device, pursuant to part 15 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment
is operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to
cause harmful interference in which case the user will be
required to correct the interference at his own expense.
ThisISMdevicecomplieswithCanadianICES-001.
Cet appareil ISM est conforme à Ia norme NMB-001 du Canada.
EN55011/CISPR11,Group1,ClassBdevice.
Class B equipment is suitable for use in domestic establishments
andinestablishmentsdirectlyconnectedtoalowvoltagepower
supply network which supplies buildings used for domestic
purposes.
8.0 CE COMPLIANCE
This product has been tested and complies with European
CErequirementsfortheEMCDirective.Pleasenotethat
shielded cable must be used.
1200 BAUD
EXAMPLE:
900 ohms max load resistor = ( 24 V - 6 ) / 0.02A
supply
( V
- 6 )
0.02
load resistor (ohms) =
out
Maximum I
SET LOWER JUMPER PAIR
TO RS-485 OR RS-232
9600 BAUD
16 SAMPLE AVERAGE COUNT
1.8 HZ
WIRING CONFIGURATIONS
Table of contents
Other Novalynx Security Sensor manuals
Popular Security Sensor manuals by other brands

LEGRAND
LEGRAND CP Electronics EBDHS-KNX installation guide

Bircher Reglomat
Bircher Reglomat EasyMotion operating instructions

Federal Signal Corporation
Federal Signal Corporation 191XL Series Installation and maintenance manual

netvox
netvox RA0708 user manual

SILENTRON
SILENTRON DualTech HT manual

AJAX Systems
AJAX Systems DoorProtect user manual