This is a quick guide. Please see the Instructions for Use that come with NovoPen®6 or NovoPen Echo®Plus for full information.
NovoPen®6 and NovoPen Echo®Plus are registered trademarks of Novo Nordisk A/S.
© Novo Nordisk Allé, DK-2880 Bagsværd, Denmark.
The N Mark is a trademark or registered trademark of NFC Forum, Inc. in the United States and in other countries.
HQ19NP600012; Approval date: October 2019.
NovoPen®6 or NovoPen Echo®Plus must be
prescribed by your healthcare provider. NovoPen®6
& NovoPen Echo®Plus are intended for patients who
have been prescribed Tresiba®or Fiasp®.
NovoPen®6 offers a 60-unit maximum dose in 1-unit
dose increments. NovoPen Echo®Plus offers a 30-unit
maximum dose in 0.5-unit dose increments.
Keep a personal injection log
If you have a smartphone, tablet, personal computer,
blood glucose monitor or another device that supports
Near Field Communication (NFC), you can easily transfer
your insulin dosing information from your pen to your
device. Just make sure to download a compatible app
or other software that can read and display
this information.
How to check your last dose
Pull out the dose button
Make sure the dose counter shows “0” before you start.
Push the dose button back in
The dose memory will now show the details of your last dose.
Read your dose memory
The example above shows 12 units injected 1 hour,
52 minutes and 3 seconds ago.
Insert an insulin cartridge
Pull off the pen cap. Twist off the cartridge holder. The piston
rod may stick out of your pen. If so, push it in until it stops.
Attach a new needle
Tear off the paper tab. Push the needle straight onto
the pen. Turn until it is on tight. Pull off both needle caps.
Save the outer needle cap.
Slide a new insulin cartridge into the cartridge holder with
the threaded end first. Twist the cartridge holder back on.
Prepare your NovoPen®6 or NovoPen Echo®Plus
ALWAYS test the insulin flow
Select 2 units and press the dose button until the dose
counter shows “0”. The insulin flow test is complete
when you see insulin squirt and there is no longer a
gap between the piston rod head and the piston.
Select your dose
Pull out the dose button. Make sure the dose counter
shows “0” before you start. Turn the dose button to
select the dose you need.
Inject your dose
Insert the needle. Press the dose button until the dose
counter shows “0”. Count slowly to 6 before you remove
the needle from your skin. Safely remove and throw
away the needle. Put the pen cap back on.
Give your injection
Wait and
count slowly
1-2-3-4-5-6