SPECIFICATIONS CLV-160
1-2
Item Specification
Air supply pressure (1) Adjustable in 3 steps (high/medium/low)
(2) Maximum pressure less than 53.9kPa
Control Setting of the air supply switch
2 Water
supply
Method Supply from the scope’s distal end by combining the scope with
the supply water tank.
Scope One touch connection
Light control cable:MH-966 Connector on the rear panel
(Light control cable for CV)
4 Connection
Light control cable:MAJ-586
Connector on the rear panel
(Light control cable for OTV)
5 Error indication
Emergency lamp When the emergency lamp breaks (including connection
failure), LED flickers to warn to replace the lamp. When the
emergency lamp is lit, LED is lit.
6 Panel
Storage of set values The setting is stored before the power switch is turned off and
when the power switch is turned off.
• Setting of manual/auto light adjustment
• Setting of air supply pressure
• Light intensity control level
1 Dimming
On removal of the scope When the scope is removed, the emergent light from the scope
output connector is dimmed by the diaphragm blade to the
minimum diaphragm position.
2 For cleaning
Panel To remove dust, dirt, and non-patient debris, wipe the unit
using a soft, lint-free cloth moistened with 70% ethyl or
isopropyl alcohol. If the unit is soiled with blood or other
potentially infectious materials, first wipe off all gross debris
using neutral detergent and then decontaminate its surface
using lint-free cloth moistened with 70% ethyl or isopropyl
alcohol.
7 Safety
3 Warning tone
Temperature switch When the temperature in the system rises over the rated
maximum value, the temperature switch automatically turns
off and the current is shut off to ensure safety. The function of
the temperature switch is detected and activates an audible
tone.