Instructions Guide for G3 WorkStation
Procedure Setup
G3 WorkStation
ENT
E0499049_Folder_Gyrus_G3_A5_2.indd 1 22.03.2010 12:59:18 Uhr

2G3 WORKSTATION
The G3 WorkStation provides precise delivery and control of PK
Technology for head and neck surgical procedures. In addition
to PlasmaCision, the G3 supports TCRFTM (Temperature-
Controlled Radio Frequency) for the Somnoplasty procedures.
As a result, one technology platform can be used for a growing
variety of surgical procedures, providing enhanced value and
versatility, now and in the future.
G3 WORKSTATION
1 2 4
3 6 7
5
1Return Electrode Connection
2 TCRF Handpiece Cable Connection
3 TCRF Set Target Temperature/
PlasmaCision, PlasmaBlend Settings
4 Standby/Power Button
5Mode/ Menu
6 TCRF Set Target
7 PlasmaCision Handpiece Connection
E0499049_Folder_Gyrus_G3_A5_2.indd 2 22.03.2010 12:59:20 Uhr

G3 WORKSTATION PROCEDURE SETUP GUIDE
Using Temperature Controlled Radio Frequency (TCRF)
Initial Setup:
A Connect the power cord and footswitch on the rear panel of the WorkStation.
B Turn the machine on at the back of the unit.
C Press the Stand-By button on the center of the front panel (the “apple” switch).
D Place the grounding pad on the patient and connect the cord to the front of the
controller.
E CAREFULLY connect the handpiece (Green or Gray) cable to the front of the controller
(Use the green cable for turbinate, tonsil, and dual-channel tongue-base and the gray
cable for palate and single-channel tongue base).
F Set the temperature using arrows on the left side of the controller.
G Set the Joules using arrows on the right side of the controller.
H See the charts below for temperature and Joules settings for each procedure.
Rapid Lesion Mode:
A Press the temperature arrow until 85ºC is reached.
B Release and press once more to reach the RAPID LESION menu.
C The temperature is now set for rapid lesion.
D Maximum Joule output for single channel instruments using rapid lesion is 300.
E Maximum Joule output for dual channel instruments using rapid lesion is 600.
F Do not use rapid lesion for palate reduction.
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
E0499049_Folder_Gyrus_G3_A5_2.indd 3 22.03.2010 12:59:22 Uhr

4G3 WORKSTATION
Begin Treatment:
A Press and release the BLUE foot pedal OR press the MODE button (located in
the center of the controller with a round arrow) ONCE (This will set the system to RF
SENSING OFF).
B Press and release the BLUE foot pedal OR press the Mode button a second time.
(This will set the system to RF SENSING ON, Impedance values are recognized here).
C Press and release the BLUE foot pedal OR press the Mode button a third time.
(This initiates the energy delivery sequence).
Pause: Press the BLUE foot pedal•
Restart: Press BLUE foot pedal once more•
Reset and Additional treatment:
A After each completed lesion, press the BLACK mode foot pedal once to reset the
system to RF SENSING OFF.
B Press and release the BLUE foot pedal will set the system to RF SENSING ON
(Impedance values are recognized).
C Press the BLUE foot pedal once more to initiate energy delivery to the additional
site.
D Audible tone is recognized every 50 Joules.
Settings using TCRF for turbinate:
Temperature: 85ºC•
Joules: 350-500•
Rapid lesion mode:
Temperature: Set to Rapid Lesion Mode•
Joules: Single channel instruments: 300 peak•
Impedance: 110 - 160 OHMS•
E0499049_Folder_Gyrus_G3_A5_2.indd 4 22.03.2010 12:59:24 Uhr

Settings using TCRF for palate reduction:
Temperature: 85ºC•
Joules: 600 to 750 for midline lesion•
Joules: 300 to 350 for lateral lesions•
Impedance: 110 - 160 OHMS•
IMPORTANT NOTE: Rapid Lesion is not recommended for soft palate.
Settings using TCRF for base of tongue:
Temperature: 85ºC•
Joules: Single channel instrument: 600 - 750 per site•
Joules: Dual channel instrument: 1200 - 1500 per site•
Rapid lesion mode:
Temperature: Set to Rapid Lesion Mode•
Joules: Single channel instruments: 300 peak•
Impedance: 110 - 160 OHMS•
Settings using TCRF for base of tongue:
Temperature: 85ºC•
Joules: 800 - 1000•
Rapid lesion mode:
Temperature: Set to Rapid Lesion Mode•
Joules: Dual channel instrument: 600 peak•
Impedance: SETTINGS USING TCRF FOR BASE OF TONGUE: 120 - 170 OHMS•
E0499049_Folder_Gyrus_G3_A5_2.indd 5 22.03.2010 12:59:26 Uhr

6G3 WORKSTATION
Using Plasmacision
Initial Setup:
A Connect the power cord and footswitch on the rear panel of the WorkStation.
B Turn the machine on at the back of the unit.
C Press the Stand-By button on the center of the front panel (the “apple” switch).
D Connect the PlasmaKnife cable to the purple “PC” plug on at the far right side of
the G3 WorkStation.
E The WorkStation will automatically identify the PlasmaKnife and set the appropriate
default PlasmaBlend setting.
F To change the PlasmaBlend CUT/COAG ratio, press the BLACK mode foot pedal
for 1 second until a beep is heard.
G Press the YELLOW pedal to increase the cut component in 5% increments, or the
BLUE pedal to increase the COAG component in 5% increments.
H Press the BLACK mode foot pedal to select the new PlasmaBlend ratio.
I Alternatively, use the “up/down” PlasmaBlend buttons on the left side of G3
WorkStation to adjust the CUT/COAG to desired settings.
J Press YELLOW pedal for PlasmaBlend CUT output.
K Press BLUE pedal for COAG output.
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
E0499049_Folder_Gyrus_G3_A5_2.indd 6 22.03.2010 12:59:28 Uhr

Cleaning the cutting surface:
Keep the PlasmaKnife clean for optimum performance.•
Wipe the instrument tip frequently with a wet 4 x 4 gauze.•
Periodic dipping in a container of saline may help to keep the suction channel •
patent.
If the WorkStation gives a “clean tip” message; clean the instrument tip and reset •
the WorkStation by pressing the BLACK mode foot pedal.
EG70353001 G3 WorkStation
EG70353002 Footswitch for G3 WorkStation
EG720100901 Powercord USA
EG70353003 Tonsil PlasmaKnife (4-pack)
EG70353005 Dissector PlasmaKnife (4-pack)
EG70353006 jPlasmaKnife (5-pack)
EG1120411005 TCRF Turbinate Handpiece (5-pack)
EG1010211005 TCRF Palate Handpiece (5-pack)
EG2220221005 TCRF Dual Channel Tongue Base Handpiece (5-pack)
EG1200211005 TCRF Single Channel Tongue Base Handpiece (5-pack)
EG2420132805 TCRF Tonsil Handpiece (5-pack)
EG7201013 RC2 Cable for 1010 and 1200 (20 reuses)
EG7201017 RC20 Cable for 1120, 2220 and 2420 (20 reuses)
EG7201010 Bending Tool
EG7201001 Return Electrode (5-pack)
Ordering Information
Cat. No. Description
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
E0499049_Folder_Gyrus_G3_A5_2.indd 7 22.03.2010 12:59:30 Uhr