Olympus WM-NP2 User manual

USA: CAUTION: Federal law restricts this device to sale by or on the order of a
physician.
INSTRUCTIONS
MOBILE WORKSTATION
WM-NP2


MOBILE WORKSTATION WM-NP2 1
Contents
Important Information — Please Read Before Use. . . . . . . . . . . . . . . 2
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Instruction manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Signal words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Handling precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Chapter 1 Standard Sets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.1 Standard sets of components . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Optional accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Chapter 2 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Top tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Separation transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Chapter 3 Preparation for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Castors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Shelf size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3 Fixing positions of intermediate shelves . . . . . . . . . . . . . . . . 13
3.4 Fixing position of the drawer unit . . . . . . . . . . . . . . . . . . . . . 16
3.5 Clip-in accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.6 Scope hanger assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.7 Keyboard assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.8 Equipment installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.9 Power connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.10 General information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Chapter 4 Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Chapter 5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.1 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.2 Manoeuvring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Chapter 6 Care, Storage and Maintenance . . . . . . . . . . . . . . . . . . 40
6.1 Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.2 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Chapter 7 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.1 WM-NP2 mobile workstation. . . . . . . . . . . . . . . . . . . . . . . . . 43
7.2 Separation transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Chapter 8 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Chapter 9 Technical Description . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Chapter 10 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Chapter 11 End of Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

Important Information — Please Read Before Use
MOBILE WORKSTATION WM-NP2
2
Important Information — Please Read Before Use
Intended use
The Olympus WM-NP2 mobile workstation is intended for use in medical facilities
under the direction of a trained physician, and has been designed to be used with
a range of Olympus equipment to facilitate GI endoscopy, endoscopic
ultrasound, respiratory and surgical endoscopic procedures. Do not use the
equipment for any purpose other than its intended application.
The safety and performance of an endoscopic system depends not only on the
endoscope but also on the ancillary equipment used with it. Ensure any ancillary
equipment is compatible with the endoscope and other equipment used.
Instruction manual
Please read this manual carefully before attempting to use the Olympus
WM-NP2 mobile workstation as it contains important information on the proper
care, handling and use of the equipment. In addition, read the manuals of any
other units (light source, endoscope, etc.) that form part of the system.
If you have any questions or comments regarding the information in this manual,
please contact Olympus. These instructions should be retained for reference
during the life of the product.
Signal words
The following signal words are used throughout this manual:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could
result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may
result in minor or moderate injury. It may also be used to alert against
unsafe practices or potential equipment damage.
NOTE
Indicates additional helpful information.

Important Information — Please Read Before Use
MOBILE WORKSTATION WM-NP2 3
Handling precautions
Follow the warnings and cautions given below when handling this workstation.
This information is to be supplemented by the warnings and cautions given in
each chapter.
WARNING
• The WM-NP2 mobile workstation should only be used in a medical
facility under the direction of a trained physician.
• Explosion hazard - never install or use the WM-NP2 mobile workstation
within the zone of risk of flammable gases. The workstation is fitted with
antistatic castors, none of the medical electrical equipment with which
the workstation is intended to be used is ‘AP’ or ‘APG’ rated, therefore
never install or use the WM-NP2 mobile workstation within the zone of
risk of flammable gases.
• To prevent patient shock, the combination of equipment used on the
WM-NP2 mobile workstation should never be applied directly to the
heart.
• After wiping with a piece of moistened gauze, dry the workstation and
mounted devices thoroughly before using it again. If it is used while still
wet, there is the risk of an electric shock.
• Use personal protective equipment to guard against dangerous
chemicals and potentially infectious material, otherwise, blood, mucus
and other potentially infectious material from the patient could pose an
infection risk. During cleaning, wear appropriate protective equipment,
such as eye wear, face mask, moisture-resistant clothing and chemical-
resistant gloves that fit properly and are long enough so that your skin is
not exposed.
• To prevent user or patient cross-contamination, do not hang
contaminated endoscopes on the scope hanger.
• The WM-NP2 should be used in accordance with the following.
Otherwise, the workstation will be ‘top heavy’ and may lose stability and
topple, causing equipment damage and/or personal injury.
Do not attach more than one LCD monitor arm to each side
of the workstation.
LCD monitor arms and monitors should not be installed and
used on an empty workstation.
LCD monitor arms and monitors should be installed after all
other equipment is installed on an appropriate place on the
workstation.
• (USA only): The power cord on this unit is fitted with a ‘hospital grade’
plug. Grounding reliability can only be achieved when this equipment is
connected to an equivalent receptacle marked ‘hospital grade’.

Important Information — Please Read Before Use
MOBILE WORKSTATION WM-NP2
4
WARNING
• (RoW only): The power cable on this unit is terminated in flying leads,
for a suitable plug to be fitted according to the area of intended use. The
cores of the cable are coloured in accordance with the following code:
BROWN - LIVE (‘HOT’)
BLUE - NEUTRAL
GREEN/YELLOW - EARTH (GROUND)
Only suitably qualified personnel should fit the power plug to
this cable.
CAUTION
• Do not exceed the load capacity detailed in Chapter 7.
• Do not lean or stand on any part of the mobile workstation or accessory
as failure of the product may result.
• When assembling equipment onto the mobile workstation, the
installation should meet the requirements of EN IEC 60601-1-1:20015,
the safety standard for medical electrical systems.
• Before commissioning an installation including a separation transformer,
ensure its rating will not be exceeded by the equipment connected to it.
• When configuring a system, the user should ensure that there is
sufficient spare equipment available that in the event of a fault in any
part of the system, the endoscopy procedure may be completed or
terminated without endangering the patient.
• When installing the mobile workstation in the procedure or operating
room, ensure that the mains cable has sufficient slack that it cannot be
pulled taut as the mobile workstation is moved around the working area.
If necessary, relocate the mains connector to a nearer supply socket.
• Section 5.2 Manoeuvring gives detailed instructions and precautions
which must be read and understood before attempting to move the
mobile workstation.
• Spray type medical agents, such as lubricants, anaesthetic or alcohol,
should be used away from the mobile workstation.
• Keep open containers of liquid away from electrical equipment used on
the mobile workstation to prevent operator electrical shock and
equipment damage. Do not use the system if liquid has entered the
equipment.
• To minimise risk of injury, it is recommended that heavy items are placed
on the mobile workstation by two people working together. Ensure
electrical items used on the mobile workstation are positioned centrally
on the shelves.
• Never place any items on top of a monitor (if fitted) as these may topple
when the mobile workstation is moved.
• The electrical resistance of the conductive antistatic castors should be
tested at regular intervals in accordance with ISO 2878 to assess
antistatic effectiveness.
• Avoid contact of the castor wheels with oil and avoid a build up of floor
cleaning wax and polish as these will impair the antistatic effectiveness
of the castors.

Important Information — Please Read Before Use
MOBILE WORKSTATION WM-NP2 5
CAUTION
• Store and use the WM-NP2 mobile workstation within the environmental
conditions described in Chapter 7; failure to do so may lead to
equipment malfunction or failure.
• Before attempting to move the workstation outside the procedure room,
ensure that the endoscope and fluid containers are removed.
• There are no operator-serviceable components on the WM-NP2 mobile
workstation. Refer servicing to qualified service personnel.
• Before commencing shelf adjustment, ensure equipment is
disconnected and removed from the shelf.
• To minimise the risk of injury and/or damage, do not fit accessories other
than those described in Section 1.1 to the WM-NP2 mobile workstation.
• The maximum storage load of each drawer is 5kg, otherwise stability
may be compromised.
• Ensure each drawer is locked when moving the workstation.
• Take care not to trap fingers when opening/closing or repositioning the
drawer unit.
• When installing a keyboard on the keyboard tray, ensure the cable
passes around the outside of the sliding runner, otherwise the cable may
be pinched when operating the sliding tray.
• Check all equipment connectors on a regular basis to ensure there are
no loose connections.

Standard Sets
MOBILE WORKSTATION WM-NP2
22
Chapter 1 Standard Sets
The Olympus WM-NP2 mobile workstation has been designed to accommodate the
Olympus video system center and associated ancillary devices as a practical mobile
workstation to facilitate GI endoscopy, endoscopic ultrasound, respiratory and surgical
endoscopic procedures.
Please ensure familiarity with all important information prior to workstation operation.
Refer to the following table to check that all items in the set ordered are present, and contact
Olympus if any parts are damaged or missing.

Standard Sets
MOBILE WORKSTATION WM-NP2 23
1.1 Standard sets of components
Standard set
123456
Japan
USA
Europe
UK
ROW
USA
Japan
USA
Europe
UK
ROW
Japan
USA
Japan
USA
Europe
UK
ROW
Europe
UK
AWM-NP2(incl.Transformer) 111111
B MAJ-1641 LCD Monitor Arm 1 1 1 1 1
C MAJ-1640 Keyboard Tray 1 1 1 1 1
D MAJ-1638 Scope Pole 1 1
E MAJ-1653 CO2 Cylinder Holder Double 1 1
F MAJ-1643 Drawer Unit 1
GBlisterpackToolkit 111111
H Blister pack Drawer key 1
I IEC lead set 1 11111
JMAJ-150adaptorcable(USA/Japanonly)11111
A
B
C
D
E
F
(supplied fitted)
(supplied fitted)
(supplied fitted)
(supplied boxed)
(supplied boxed)
(supplied boxed)

Standard Sets
MOBILE WORKSTATION WM-NP2
24
1.2 Optional accessories
The following products may be used as optional accessories.
Code Description
MAJ-150 Adaptor cable
MAJ-1662 Nurses control panel arm - short
MAJ-1663 EUS Arm mount kit
MAJ-1664 Nurses control panel arm - long
MAJ-1635 EUS connector holder
MAJ-1638 Scope pole kit
MAJ-1639 Single CO2holder (up to Ø140mm)
MAJ-1640 Keyboard tray
MAJ-1641 LCD monitor arm
MAJ-1642 IV Pole
MAJ-1643 Drawer unit
MAJ-1650 Single CO2holder (Ø140mm to Ø205)
MAJ-1653 Double CO2holder
MAJ-1654 Equipotential terminal strip
MAJ-1655 OEV261H Monitor PSU holder
MAJ-1656 Footswitch holder
MAJ-1657 Keyboard arm - side mounted
MAJ-1658 Universal stowage container
MAJ-1659 Equipment support rail
MAJ-1660 Cable retaining hook
MAJ-1659 Equipment support rail

Warnings and Cautions
MOBILE WORKSTATION WM-NP2 25
Chapter 2 Warnings and Cautions
Follow the warnings and cautions given below when handling this workstation. This information is
to be supplemented by the warnings and cautions given in each chapter.
WARNING
• The WM-NP2 mobile workstation should only be used in a medical
facility under the direction of a trained physician.
• To prevent patient shock, the combination of equipment used on the
WM-NP2 mobile workstation should never be applied directly to the
heart.
• After wiping with a piece of moistened gauze, dry the workstation and
mounted devices thoroughly before using it again. If it is used while still
wet, there is the risk of an electric shock.
• Use personal protective equipment to guard against dangerous
chemicals and potentially infectious material, otherwise, blood, mucus
and other potentially infectious material from the patient could pose an
infection risk. During cleaning, wear appropriate protective equipment,
such as eye wear, face mask, moisture-resistant clothing and chemical-
resistant gloves that fit properly and are long enough so that your skin is
not exposed.
• To prevent user or patient cross-contamination, do not hang
contaminated endoscopes on the scope hanger.
• The WM-NP2 should be used in accordance with the following.
Otherwise, the workstation will be ‘top heavy’ and may lose stability and
topple, causing equipment damage and/or personal injury.
- Do not attach more than one LCD monitor arm to each side
of the workstation.
- LCD monitor arms and monitors should not be installed and
used on an empty workstation.
- LCD monitor arms and monitors should be installed after all
other equipment is installed on an appropriate place on the
workstation.
• (USA only): The power cord on this unit is fitted with a ‘hospital grade’
plug. Grounding reliability can only be achieved when this equipment is
connected to an equivalent receptacle marked ‘hospital grade’.

Warnings and Cautions
MOBILE WORKSTATION WM-NP2
26
WARNING
• (RoW only): The power cable on this unit is terminated in flying leads,
for a suitable plug to be fitted according to the area of intended use. The
cores of the cable are coloured in accordance with the following code:
BROWN - LIVE (‘HOT’)
BLUE - NEUTRAL
GREEN/YELLOW - EARTH (GROUND)
Only suitably qualified personnel should fit the power plug to
this cable.
• To avoid risk of electric shock, this equipment must only be connected to
a supply mains with a protective earth.
• Do not modify this equipment without authorisation of the manufacturer.
CAUTION
• Do not exceed the load capacity detailed in Chapter 7.
• Do not lean or stand on any part of the mobile workstation or accessory
as failure of the product may result.
• When assembling equipment onto the mobile workstation, the
installation should meet the requirements of the safety standard for
medical electrical systems, as described in section 4.10.
• Before commissioning an installation including a separation transformer,
ensure its rating will not be exceeded by the equipment connected to it.
• When configuring a system, the user should ensure that there is
sufficient spare equipment available that in the event of a fault in any
part of the system, the endoscopy procedure may be completed or
terminated without endangering the patient.
• When installing the mobile workstation in the procedure or operating
room, ensure that the mains cable has sufficient slack that it cannot be
pulled taut as the mobile workstation is moved around the working area.
If necessary, relocate the mains connector to a nearer supply socket.
• Spray type medical agents, such as lubricants, anaesthetic or alcohol,
should be used away from the mobile workstation.
• Keep open containers of liquid away from electrical equipment used on
the mobile workstation to prevent operator electrical shock and
equipment damage. Do not use the system if liquid has entered the
equipment.
• To minimise risk of injury, it is recommended that heavy items are placed
on the mobile workstation by two people working together. Ensure
electrical items used on the mobile workstation are positioned centrally
on the shelves.
• Never place any items on top of a monitor (if fitted) as these may topple
when the mobile workstation is moved.
• The electrical resistance of the conductive antistatic castors should be
tested at regular intervals in accordance with ISO 2878 to assess
antistatic effectiveness.

Warnings and Cautions
MOBILE WORKSTATION WM-NP2 27
CAUTION
• Avoid contact of the castor wheels with oil and avoid a build up of floor
cleaning wax and polish as these will impair the antistatic effectiveness
of the castors.
• Store and use the WM-NP2 mobile workstation within the environmental
conditions described in Chapter 7; failure to do so may lead to
equipment malfunction or failure.
• Before attempting to move the workstation outside the procedure room,
ensure that the endoscope and fluid containers are removed.
• There are no operator-serviceable components on the WM-NP2 mobile
workstation. Refer servicing to qualified service personnel.
• Before commencing shelf adjustment, ensure equipment is
disconnected and removed from the shelf.
• To minimise the risk of injury and/or damage, do not fit accessories other
than those described in Section 1.1 to the WM-NP2 mobile workstation.
• The maximum storage load of each drawer is 5kg, otherwise stability
may be compromised.
• Ensure each drawer is locked when moving the workstation.
• Take care not to trap fingers when opening/closing or repositioning the
drawer unit.
• When installing a keyboard on the keyboard tray, ensure the cable
passes around the outside of the sliding runner, otherwise the cable may
be pinched when operating the sliding tray.
• Check all equipment connectors on a regular basis to ensure there are
no loose connections.

Symbols
MOBILE WORKSTATION WM-NP2
28
Chapter 3 Symbols
The following symbols are used on the WM-NP2.
3.1 Top tray
Refer to instructions for use
3.2 Separation transformer
Warning to observe rules
regarding the use of multiple
socket-outlets
Potential equalisation terminal
Input Output
Not exceeding Ta=40°C Maximum ambient
temperature
Mains power connected Breaker
Heavy unit Remote switch
Non-inherently short circuit
proof separation
transformer
Class A/B
Insulation temperature rating
of transformer according to
IEC 60085:2004
Unit complies with relevant
European directives
(MAJ-1648 only)
Disposal conditions see
Chapter 11 End of Life
(MAJ-1648 only)
Consult instructions for use
Indicates the unit has been
independently assessed for
compliance with UL and CSA
standards as a Recognized
Component of the workstation
(MAJ-1647 only).

Symbols
MOBILE WORKSTATION WM-NP2 29
NOTE
MAJ-1648 (220-240V) is shown. Other versions are similar.
View of the separation transformer rear panel

Preparation for Use
MOBILE WORKSTATION WM-NP2
30
Chapter 4 Preparation for Use
4.1 Castors
1. The workstation has two diagonally located braked castors. Ensure both brakes are applied
prior to use and released before moving the workstation.
NOTE
Brakes should always be applied when the workstation is not being
manoeuvred.
2. To lock and unlock the castor brake, refer to Figure 4.2.
Figure 4.1
Figure 4.2
Brake castor
Brake castor

Preparation for Use
MOBILE WORKSTATION WM-NP2 31
4.2 Shelf size
Refer to Figure 4.3 for the size of equipment that can be mounted on the product's shelves.
Figure 4.3
470
409
512
473
325
439
470
617
479
480
Dimensions in mm
Top shelf
Middle shelf
Base shelf

Preparation for Use
MOBILE WORKSTATION WM-NP2
32
4.3 Fixing positions of intermediate shelves
The product columns contain holes for the intermediate shelves. You can reposition the shelves to
suit the size of the equipment being mounted.
On delivery
The product is delivered with the shelves fitted in the positions shown in Figure 4.4.
Figure 4.4
Shelf fixing holes
25mm apart
292mm
35mm
35mm
215mm
275mm
Drawer unit not shown

Preparation for Use
MOBILE WORKSTATION WM-NP2 33
Repositioning of shelves
CAUTION
Before commencing shelf adjustment, ensure equipment is
disconnected and removed from the shelf.
Use the following steps to reposition the intermediate shelves.
1. Remove any equipment that is on the shelf.
2. Position the workstation on a flat surface and apply the brakes on two diagonally opposite
castors.
3. Remove the two cable ducts that are fitted on the rear of the intermediate shelf and the cable
tie that is fitted at the same height as the shelf. You can remove the cable tie by inserting a
flat bladed screwdriver behind the stud in the middle of the cable tie and pulling it towards
you, as shown in Figure 4.5.
4. Check that nothing has been placed on the shelf and, referring to Figure 4.6, loosen the two
M8 screws that are securing the shelf about half a turn using the supplied 6mm Allen key –
do not remove.
Figure 4.5
Cable duct
Stud
Cable tie

Preparation for Use
MOBILE WORKSTATION WM-NP2
34
5. Take a firm hold of the shelf and, referring to Figure 4.6, unlock the two spring plungers fitted
on each side. Pull the spring plunger head towards you, twist 90° and release. The spring
plunger remains extended and is unlocked.
NOTE
The spring plungers cannot be removed from the shelves.
6. Move the shelf to the required height. Pull each spring plunger head, twist 90° and release.
The spring plunger head will spring back to its original position when aligned with the column
holes.
CAUTION
After moving the shelf, check that the four spring plunger heads are not
protruding. If any spring plunger head is protruding, the shelf may tilt
and get damaged due to insufficient engagement with the column holes.
7. Check that the shelf is not tilting and firmly tighten the two M8 screws with the Allen key.
Figure 4.6
Spring plunger
M8 screw
Table of contents
Other Olympus Desktop manuals