OTO Photonics LS-DH-1 User manual

1
1
OtO Photonics
Light Source Operation Manual
Deuterium-Halogen Light Series
(LS-DH Series)
LS-DH-1
LS-DH-2
LS-DH-3
LS-DH Series Operation Manual-801 Rev.1
www.otophotonics.com
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

2
2 2
www.otophotonics.com
Deuterium-Halogen Light Series
Operation Manual
Drawing of LS-DH series P3
Drawing of LS-DH series & Measurement platform P3
Operation Selection Features P4
Manual Mode Operation P4
Auto Mode Operation P5
I/O Extension Port Pin# Description Alt Function P6
Spectrum P7
LS-DH Series Operation Manual-801 Rev.1
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

3
3 3
www.otophotonics.com
Drawing of LS-DH Series
Drawing of LS-DH Series & Measurement Platform
LS-DH Series Operation Manual-801 Rev.1
Deuterium-Halogen Light Series
Operation Manual
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

4
4 4
www.otophotonics.com
Main Power Switch
Auto/Manual Switch
D2 Switch
HAL Switch
Shutter Switch
I/O pin 1
Operation Selection Features
Manual Mode Operation step
Step.1
After plugging the +12V power to the Deuterium
-Halogen
(DH) light source, user can turn on the main power by the
Main Power Switch. The Power Green LED will be
lightening “ON”.
Step.2
Then user can select the Auto/Manual Switch. Once the
manual mode is selected; that means the switch is at the left
position.
Step.3
After manual mode is selected, user can turn on/off the D2,
Halogen, or Shutter function by sliding the switches.
LS-DH Series Operation Manual-801 Rev.1
Deuterium-Halogen Light Series
Operation Manual
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

5
5 5
8 pin Cable
USB Cable
+12V power Cable
www.otophotonics.com
Auto Mode Operation
Step.1
Put the Spectrometer and the Deuterium
-Halogen (DH) light
source on the platform and place them at the assigned
measurement position.
Step.2
Plug in the 8pin cable to link these two devices. Please make
sure the cable is inserted correctly.
Step.3
Plug the USB cable to connect the spectrometer and the PC.
Step.4
After plugging the +12V power to the DH light source, user
can turn on the main power by the Main Power Switch. The
Power Green LED will be lightening “ON”.
Step.5
Slide the Auto mode switch to the right position. The “Auto”
Green LED will be lightening "ON”.
Step.6
Then user can use the
OtO’s software SpectraSmart or SDK
to send the command to turn on/off the D2, Halogen light
source, or Shutter function individually; the corresponding
Green LEDs will be turned on/off accordingly.
Step.7
User needs to consider the warming
-up and stable time for
DH light source.
LS-DH Series Operation Manual-801 Rev.1
Deuterium-Halogen Light Series
Operation Manual
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

6
6 6
Pin No. Direction Pin Name Function Description
1 NC NC NA
2 NC NC NA
3 NC NC NA
4 Input Shutter High: Open Shutter Low: Close
Shutter
5 Input D2_ON High: Turn on D2 LS. Low: Turn
off.
6 Input HAL_ON High: Turn on Halogen LS. Low:
Turn off.
7 NC NC No connection inside the LS-DH-1.
8 GND GND GND
I/O Extension Port Pin# Description Alt Function
There is one 8pin 2.0mm pitch connector with every DH light source.
This 8pin connector can connect SmartEngine (SE) series spectrometer
and DH light source directly to operate auto mode. Once the operation
mode is selected to Auto mode, D2, Halogen light source and Shutter
function will be controlled by the I/O level. (3.3V/5V or 0V).
www.otophotonics.com LS-DH Series Operation Manual-801 Rev.1
Deuterium-Halogen Light Series
Operation Manual
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

7
7 7
www.otophotonics.com
Spectrum
Deuterium light source on only (measured by OtO EE2061 spectrometer)
Deuterium and Halogen light source on (measured by OtO EE2061 spectrometer)
LS-DH Series Operation Manual-801 Rev.1
Deuterium-Halogen Light Series
Operation Manual
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

8
8 8
www.otophotonics.com
Software Operation Features
SpectraSmart
Spectrasmart provides Setting function at the right column window:
1. Enable / Disable shutter of each scan.
2. Turn on Halogen if lamp button as
3. Turn on Deuterium if lamp button as
(Halogen and Deuterium lamp can be controlled independently)
Setting Function
Lamp button
Software development commands
UAI_SpectrometerSetExternalPort and UAI_SpectrometerGetExternalPort
can set the operation status of shutter, halogen and deuterium lamp.
GPIO Item
GPIO 5
Halogen
GPIO 4
Deuterium
GPIO 3
Shutter
LS-DH Series Operation Manual-801 Rev.1
Deuterium-Halogen Light Series
Operation Manual
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk
This manual suits for next models
2