Peraqua iQntrol DOS-ECO Instruction Manual

www.peraqua.com
DOS-ECO
INSTRUCTIONS AND USER MANUAL

Instructions and user manual DOS-ECO NOT041ENV01_FP1
2
SOMMAIRE
1WARNING AND ADVICES .......................................................................................... 4
1.1 Guarantee and Warning.....................................................................................................4
1.2 Shipping and transporting..................................................................................................5
1.3 Proper use of the pump .....................................................................................................5
1.4 Risk .....................................................................................................................................5
1.5 Toxic and/or dangerous liquid dosage...............................................................................6
1.6 Assembling and dismantling the pump..............................................................................6
2DOS-ECO DESCRIPTION ............................................................................................ 7
2.1 Parts description .......................................................................................................................7
2.2 Buttons description...................................................................................................................8
2.3 DOS-ECO plugs description .......................................................................................................9
3RECOMMENDATIONS AND INSTALLATION........................................................ 9
3.1 Recommandations ....................................................................................................................9
3.2 Hydraulic and electrical installation principle.........................................................................10
3.3 Device Dimensions..................................................................................................................11
3.4 Device installation...................................................................................................................11
4ADVANCED DEVICE SETTINGS............................................................................. 23
4.1 pH setpoint..............................................................................................................................23
4.2 Redox setpoint ........................................................................................................................23
4.3 Activation type ........................................................................................................................24
4.4 DOS-ECO settings ....................................................................................................................25
4.5 Duo standby mode..................................................................................................................29
4.6 DOS-ECO priming ....................................................................................................................29
4.7 Display backlight .....................................................................................................................29
5SOLUTIONS TO PROBLEM..................................................................................... 29
5.1 Incorrect measure and procedure for increasing the TAC .....................................................29
5.2 Alarms and displayed symbols................................................................................................30
6MAINTENANCE ......................................................................................................... 31
6.1 General....................................................................................................................................31
6.2 Chlorine injection valve maintenance ....................................................................................31

Instructions and user manual DOS-ECO NOT041ENV01_FP1
3
6.3 Installation of a long probe (120 mm) ....................................................................................32
6.4 Peristaltic tube replacement...................................................................................................33
6.5 Wintering ................................................................................................................................33
7AFTER SALE SERVICE ............................................................................................. 34
8SPARE PARTS LIST .................................................................................................. 35

Instructions and user manual DOS-ECO NOT041ENV01_FP1
4
1WARNING AND ADVICES
Please read the warning notices very carefully because they provide important information regarding safety in
installation, maintenance and use of the pump.
•Keep this manual in a safe place, so that it will always be available for further consultation.
•The device complies with EMC: 2004/108/EEC “electromagnetic compatibility” and LVD: 2006/95/EC “ low
voltage ” as also the subsequent modifications.
N.B. : the pump has been constructed in accordance with best practice. Both its life and its electrical and mechanical
reliability will be enhanced if it is correctly used and subjected to regular maintenance.
1.1 Guarantee and Warning
1.1.1 General equipment
The material is guaranteed 24 months parts and labour. The return address in our center is the following :
PERAQUA Professional Water Products GmbH
Handelsstraße 8
4300 St. Valentin - Austria
The shipping costs are charged of the customer. If the repair is carried out as part of the warranty, return is
supported by Peraqua. The wear parts excluded of the warranty are: pH or redox probe (except for the validation
of the warranty extension), injection valve, foot filter and peristaltic hose.
WARNING : any intervention or repair to the internal parts of the pump must be carried out by
qualified and authorized personnel. The manufacturers decline all responsibility for the
consequences of failure to respect this rule.
1.1.2 Warranty extension for the sensor
Applicable conditions : The user's purchase must be declared on the PERAQUA website (www.peraqua.com) within
one month. The warranty runs for 12 months from the date of. The warranty consists of the direct replacement of
the sensor identified as being faulty by our Technical Department, or its repair, after examination by our Technical
Department, excluding any other compensations of whatever form.
Exclusion conditions : This warranty does not cover damage caused by an accident, improper installation or use, a
lack of reasonable care, attempted falsification or attempted repair performed by an unauthorised person.
Return conditions :
•The probe's serial number must always be on its cable and readable as at the time of purchase.
•A photocopy of the purchase receipt must be sent together with the return package.
•The sensor must be sent in its original packaging with the extremity protected by its cap and filled with
water.
The transport costs to our company must be paid by the customer. However, for sensors under warranty, the return
transport costs will be paid by PERAQUA.

Instructions and user manual DOS-ECO NOT041ENV01_FP1
5
1.2 Shipping and transporting
L’expédition No matter what the means of transport employed, delivery of
the pump, even when free to purchasers or the addressee’s domicile, is
always at the purchaser’s risk. Claim for any missing materials must be
made within 10 days of arrival, while claims for defective materials will be
considered up to the 30th day following receipt. Return of pump or other
materials to us or the authorized distributor must be agreed beforehand
with the responsible personnel. Material should be disassembled and
packed with care. All parts in contact with the chemical should be purged
and rinsed for everyone’s safety. If the pump can not be purged, fasten the
input with the output using a tube (see figure below). In the case of non-
compliance with this guideline, we reserve not to support the materials and
to return it at your expense and any damage created by the chemical on
the material will undergo a repair quote.
1.3 Proper use of the pump
The pump should be used only for the purpose for which it has been expressly designed namely the dosing of
liquids. Install the pump in a dry and not floodable place and well away from sources of heat, and in any case, at
environmental temperature not exceeding 40°C. The operating minimum temperature depends of the liquid to be
pumped, bearing in mind that it must always remain in a liquid state. Any different use is to be considered
dangerous. The pump should not therefore be used for applications that were not allowed for in its design. In case
of doubt please contact our offices for further information about the characteristics of the pump and it proper use.
The manufactures can not be held responsible for damage deriving for improper, erroneous or unreasonable use
of the pump.
1.4 Risk
After unpacking the pump, make sure that it is completely sound. In case of doubt, do not use the pump and contact
qualified personnel. The packing materials (plastic bags, polystyrene, etc.) should be kept out of the reach of
children: they constitute potential sources of danger.
Before you connect the pump make sure that the voltage ratings correspond to your particular voltage supply. You
will find this value on the rating plates attached to the pump.
The electrical installation to which the pump is connected must comply with the standards and good practice rule
in force in the country under consideration.
Use of electrical equipment always implies observance of some basic rules. In particular:
- Do not touch with wet or damp hands or feet.
- Do not operate the pump with bare feet (typical place: swimming pool )
- Do not leave the equipment exposed to the atmospheric agents
- Do not allow the pump to be used by children or unskilled individuals without supervision
The device must be supplied through a circuit with a residual current differential device (RCD) rated differential
operating current not exceeding 30 mA
In case of breakdown or improper functioning of the pump, switch off, but do not touch. Contact our technical
assistance for any necessary repairs and insist on the use of original spares. Failure to respect this condition could
render the pump unsafe to use.

Instructions and user manual DOS-ECO NOT041ENV01_FP1
6
In case of damage of power supply cord please ask the repair to our technical assistance or qualified and authorized
personnel, to avoid any hazard.
When you decide to make no further of an installed pump, make sure to disconnect it from the power supply.
Before carrying out any service on the item, check:
1) Disconnect the plug from the mains.
2) Relieve all the pressure from the pump head or the injection tube.
3) Drain or flush all dosing liquid from the pump head. This operation can also be done with the pump disconnected
from the plant by turning the pump upside-down for 10 seconds and without connecting the tubing to the nipples.
In event of possible losses on the hydraulic system of the pump (breakage of the O-ring, the valves or the hoses),
the pump should immediately be brought to a stop, emptying and depressurizing the delivery hos while taking all
due the safety precautions (gloves, goggles, overalls, etc.).
1.5 Toxic and/or dangerous liquid dosage
To avoid risk from contact with hazardous liquids or toxic fumes, always adhere to the notes in this instruction
manual:
• Follow the instructions of the dosing liquid manufacturer
• Check the hydraulic part of the pump and use it only if it is in perfect condition
• Use only the correct materials fir the tubing, valves and seals to suit the liquid to be dosed : where possible shield
the tubing with PVC conduit.
• Before disconnecting the metering pump, make sure to flush out and neutralize the pump head with the proper
reagent liquid.
1.6 Assembling and dismantling the pump
1.6.1 Assembly
See paragraph 3 RECOMMeNDATIONS and INSTALLATION page 9
1.6.2 Dismantlement
Proceed as follows when you dismantle the pump or before performing any operation on it :
•Disconnect the plug from the mains.
•See paragraph 1.2 Shipping and transporting page 5

Instructions and user manual DOS-ECO NOT041ENV01_FP1
7
2DOS-ECO DESCRIPTION
The DOS-ECO is characterized by its ability to measure the pH (0 to 14 pH) with its probe placed in the filtration
circuit which analyzes the water of your pool.
Depending on the measured value, the DUO will inject the amount of product needed to maintain the water balance
of the pool thanks to its dosing pump.
The DUO operates as dosing acid (pH-) or basic (pH+) depending on its configuration.
2.1 Parts description
The complete list of accessories delivered with the equipment is given below. All the parts will be necessary for the
correct operation of your unit.
A1 : pH4 / pH7 calibration solutions (kit pH)
A2 : 475mV calibration solution (kit redox)
B : Probe holder (x2)
C : Guide collar (x2)
D : Clip (x2)
E : Foot strainer
F : Drill bit
G : Probe’s o-ring (x4)
H : Probe holder’s o-ring (x2)
I : 4x6 PVC tube (4m) (x2)
J1: pH probe
J2: redox probe
K : Probe holder’s screw (x8)
L : Probe holder’s nut (x8)
M : Wall mounting screw (x2)
N : Wall mounting plug (x2)
O : Spacers for Ø63 pipe (x8)
P : Spacer for long
probe 120mm (x2)
Q : Water direction flux arrow
Optionnal : Sucction lance

Instructions and user manual DOS-ECO NOT041ENV01_FP1
8
2.2 Buttons description
Buttons
Type d’appuis
Actions
Short press
- Decrease a numerical value
- Scroll a list of selections
Long press - Decreases a numerical value quickly
Short press
- Increase a numerical value
- Scroll a list of selections
Long press - Increases a numerical value quickly
Short press
- Switch from one display to an other
- Exit current menu
Long press - Give access to the installation configuration
with - SET &
▲
Increase the value of setpoint
with - SET &
▼
Decrease the value of setpoint
Short press
- OK allows the user to validate the value changes or validate
a selection in a menu
- Switch from one display to an other
Long press
- CAL allows the user to enter the automatically-guided
calibration phase

Instructions and user manual DOS-ECO NOT041ENV01_FP1
9
2.3 DOS-ECO plugs description
3RECOMMENDATIONS AND INSTALLATION
3.1 Recommandations
1.1.1 Nameplate
3.1.2 Recommended chemicals
•It is recommended to use sulfuric acid, which is 100% compatible with the Santoprene tube.
•It is recommended to use chlorine diluted to 5% (maximum 10%)
3.1.3 Unrecommended chemicals
•It is not recommended to use hydrochloric acid which can reduce the peristaltic tube's lifetime to a few
weeks and oxidise the pump's s metallic parts. In this case, the warranty is voided
•The stabilizer rate must be less than 15 ppm in order not to disturb the reading of the redox.
•The use of chlorine in hard water may clog the injection valve by the formation of a white deposit on the part
in contact with water. Please periodically check that the passage of the valve is free. In case of blockage, the
peristaltic tube may be damaged.
Power supply (230V ~)
Supply frequency (50 Hz)
Power (10W)
Max flow rate (1,5 l/h)
Max pressure (1500 hPa = 1,5 bars)
pH probe plug
Chlorine probe
pH low level
Chlorine low level
230V power supply cable
Filtration detection cable 230V

Instructions and user manual DOS-ECO NOT041ENV01_FP1
10
3.2 Hydraulic and electrical installation principle
Description des repères
A- DOS-ECO connected to a permanent power supply and filtration detection
B-H 2 in 1 probe holder :
B- pH probe holder: supports the pH probe and the injection valve
H- Probe holder for Rx: supports the redox probe and the injection valve
C- Chemicals tanks
D- Electrical control box for the control of the swimming pool circulation pump
E- Pool water circulation pump
F- Pool
G- Pool filtration
i- Electrical cable for wire detection connected in electrical box D
Warning :
-Under no circumstances will "i" be connected in parallel to the power supply terminals of the filter pump "E" in
the electrical box "D".
- The 2 in 1 probe holder "B" or "H" can have a maximum inclination of +/- 45° with respect to the vertical axis.
-For an optimal lifetime of your peristaltic tube, it is preferable not to exceed a pressure of 1 bar and in no case
exceed a pressure of 1.5 bar.
-The pH probe must be located before (upstream) the redox probe

Instructions and user manual DOS-ECO NOT041ENV01_FP1
11
3.3 Device Dimensions
Height: 260 mm
Width: 220 mm
Depth: 130 mm
3.4 Device installation
STEP 1 : Mounting the housing
220 mm
260 mm

Instructions and user manual DOS-ECO NOT041ENV01_FP1
12
STEP 2 : Prepare the sensor for start-up
Do the same for the Rx probe.
STEP 3 : Drill a hole in the pipe using tool F
Positioning on a horizontal pipe
1) Remove the sensor's
protective cap and
keep it to re-use it for
the winter shutdown
period
2) Leave the sensor submerged
in tap water for at least 20
minutes, before proceeding
with the calibration.
We therefore recommend that
you start your installation with
this step.
Min 15°
Min 15°
F
Sensor holder pH + sensor holder RX

Instructions and user manual DOS-ECO NOT041ENV01_FP1
13
Positioning on a vertical pipe

Instructions and user manual DOS-ECO NOT041ENV01_FP1
14
STEP 4 : Preparation of the probe holder
Mount the O-ring (H) under the support flange (B).
Remove the cap and store
1
2

Instructions and user manual DOS-ECO NOT041ENV01_FP1
15
STEP 5 : Position the support flange by inserting the injection pipe in the direction of flow.
1
2
Q
3

Instructions and user manual DOS-ECO NOT041ENV01_FP1
16
STEP 6 : Attachment to the pipe
Ø50 pipe
Ø63 pipe
C
L
K
K
C
L
O

Instructions and user manual DOS-ECO NOT041ENV01_FP1
17
STEP 7 : Electrical connection
Filtration sensing cable and power cable
1° Plug the cable i (without plug) for filtration detection into the electrical box D on your pool:
- Either on the terminals of an auxiliary contact to the relay of the filtration pump
- Either on terminals A1 and A2 in parallel with the coil of the filter pump relay
The voltage to be applied to this cable is 230V~. This means that the unit will only be supplied with power when
the filter pump is switched on.
2° Connect the power cable with its domestic electrical plug. It is preferable that the 230 Vac power supply come
from the electrical box D of your pool just after the main switch.
STEP 8 : Connect the pH and Rx probes to the corresponding plug of the instrument.
Insert the plug into the bayonet socket
Turn the plug ¼ to close the connection.

Instructions and user manual DOS-ECO NOT041ENV01_FP1
18
STEP 9 : OPTIONAL Connecting the suction lances to the unit
2 protective covers for Jack sockets
Remove the 2 protective covers
from the sockets Jack
Plug the Jack plug into the socket
If you remove the suction lances for a long period of time :
replace the protective covers on the jacks.

Instructions and user manual DOS-ECO NOT041ENV01_FP1
19
INSERT PROBE IN BUFFER 7pH, MOVE, LET INSIDE DON’T TOUCH IT
press OK
CALIBRATING pH 7
7.25pH
*
(-90s)
QUALITY 100%
4.0Ph press OK
INSERT PROBE IN BUFFER 4pH, MOVE, LET INSIDE DON’T TOUCH IT
press OK
CAUTION : before soaking the sensor in the
pH4 solution, rinse well using clean water
CALIBRATING pH 4
4.25pH
*
(-90s)
*: 7.25 is an example of instantaneous
measurement before calibration (values are
commonly between 6.5 and 7.5)
↓6.9pH
SET CAL
↓6.9pH 720mV↑
720mV ↑
SET CAL
Tip
PRESS THE ESC KEY AT ANY TIME TO
EXIT THE CALIBRATION PROCEDURE
After pressing OK, the unit
continues operating normally
*: 4.25 is an example of instantaneous
measurement before calibration (values are
commonly between 3.5 and 4.5)
STEP 10 : Calibrating the pH probe
Remove the probe from its soaking water in STEP 2 and remove the excess water.

Instructions and user manual DOS-ECO NOT041ENV01_FP1
20
INSERT PROBE IN BUFFER 475 mV, MOVE, LET INSIDE DON’T TOUCH IT
press OK
CALIBRATING 475mV
462mV*(-90s)
QUALITY 100%
475mV press OK
720mV↑
SET CAL
Tip
PRESS THE ESC KEY AT ANY TIME TO EXIT THE
CALIBRATION PROCEDURE
↓6.9pH 720mV↑
After pressing OK, the unit
continues operating normally
*: 462mV is an example of instantaneous
measurement before calibration
STEP 11 : Calibration of redox sensor
Remove the sensor from its soaking water indicated in STEP 2 and remove any excess water..
Table of contents
Popular Controllers manuals by other brands

FUTABA
FUTABA GYC440 instruction manual

DK System
DK System Master 500 operating instructions

Telaire
Telaire Ventostat 8000 Series Operation & installation manual

Remote Control
Remote Control RCEL005X instructions

Woodward
Woodward 505 Installation and operation manual

Munters
Munters TRIO User and installation manual