Philips Healthcare is part of
Royal Philips Electronics
How to reach us
www.philips.com/healthcare
Asia
+49 7031 463 2254
Europe, Middle East, Africa
+49 7031 463 2254
Latin America
+55 11 2125 0744
North America
+1 425 487 7000
800 285 5585 (toll free, US only)
Philips Respironics
1010 Murry Ridge Lane
Murrysville, PA 15668
Customer Service
+1 724 387 4000
+1 800 345 6443 (toll free)
Philips Respironics International
Headquarters
+33 1 47 28 30 82
+65 6882 5282
Philips Respironics Australia
+61 2 9666 4444
Philips Respironics China
+86 400 828 6665
+86 800 828 6665
Philips Respironics Deutschland
+49 8152 93 06 0
Philips Respironics France
+33 2 51 89 36 00
Philips Respironics Italy
+39 039 203 1
Philips Respironics Switzerland
+41 6 27 45 17 50
Philps Respironics United Kingdom
+44 800 1300 845
www.philips.com/respironics
© 2013 Konink lijke Philips E lectronics N .V. All right s are reserved.
BiPAP AVAPS
BiPAP S/T
MCI4105612
1106444 R02
JJB 03/20/2013
LCD display
screen
Number settings
Humidifier icon
Control wheel/Push button
In order to activate the Preheat mode, the blower must be off and a
Therapy,
then press and hold down the control wheel for 5 seconds. There
will be a single beep and the device will now be in preheat mode. The
) will illuminate during this time.
Guidelines for Ventilation Set-up
Conversion ta ble to set the Inspir ator y Time for controlled breat hs
Set bac k up breath rate
(BPM)
I/E 1/3,
Ti/Ttot 25%
I/E 1/2,
Ti/Ttot 33%
I/E 1/1,
Ti/Ttot 50%
10 1.5 s 2.0 s 3.0 s
11 1.4 s 1.8 s 2.7 s
12 1.3 s 1.7 s 2.5 s
13 1.2 s 1.5 s 2.3 s
14 1.1 s 1.4 s 2.1 s
15 1.0 s 1.3 s 2.0 s
16 0.9 s 1.3 s 1.8 s
17 0.9 s 1.2 s 1.7 s
18 0.8 s 1.1 s 1.6 s
19 0.8 s 1.1 s 1.5 s
20 0.8 s 1.0 s 1.5 s
21 0.7 s 1.0 s 1.4 s
22 0.7 s 0.9 s 1.3 s
23 0.7 s 0.9 s 1.3 s
24 0.6 s 0.8 s 1.2 s
25 0.6 s 0.8 s 1.2 s
Set the Inspirator y Time in seconds: Ti (Second) = 60/Respiratory rate x % Ti
Conversion ta ble to set the target Tidal Volu me in rel ation to the id eal weight
Height Calculated ideal
weight (i f BMI = 23)
Tar ge t Vt e
if 8 ml/kg
Tar ge t Vt e
if 10 ml/kg
1.50 m 52.0 k g 410 ml 520 ml
1.55 m 55.0 kg 440 ml 550 ml
1.60 m 59.0 kg 470 m l 590 ml
1.65 m 62.5 kg 500 ml 620 ml
1.70 m 66.5 kg 530 ml 660 ml
1.75 m 70.5 k g 560 ml 700 ml
1.80 m 74.5 kg 600 ml 740 ml
1.85 m 78.5 kg 630 ml 780 ml
1.90 m 83.0 kg 660 ml 830 ml
Calculated with an id eal Body Mass Index of 23 kg /m2(BM I = weight/he ight2)
Important: Guidelines are intende d to ser ve only as a refere nce. They shall be
used onl y in conju nction with t he instructions and/or protocol set forth by the
physician and institution in which t he assi st device is bein g used. T he guidelines are
not inte nded to supersede established medical protocols.
Initial
Settings
IPAP = 8 to 10 cm H2O, EPAP = 4 cmH2O and RR = 10 to 12 BPM
IPAP Increase IPAP if the patient wants more air, targeting patient tidal
volume at 8 ml/kg of ideal weight
EPAP without OSA S (Obst ructive Sleep Apne a Syndrome): 4 – 5 cmH2O
with OSAS: in crease EPAP to remove obst ructive apne a events
with intrinsic PEP (stable chronic COPD): 5 – 6 cmH2O
BPM Set to 2 – 3 BPM under patient’s spontaneous frequency
Rise
Time
Obstr uctive patie nts prefer short rise time:
from 1 to 4 (100 ms to 40 0 ms)
Restrictive patients prefer long ri se time:
from 3 to 6 (300 ms to 600 ms)
Ti Set Ti be tween 25% and 33% for obstr uctive patients
Set Ti be tween 33% and 50% for rest rictive patients
(refer to table opposite)
AVAP S
OHS COPD Other restrictive
diseases (NMD, etc.)
Vt Target 8 to 10 ml/kg of ideal body weight (refer to tab le opposite)
IPAP
Window
(IPAPmin
and
IPAPmax)
Allow a wide range
of pressure variation
to ensure the right
pressure at the right
time.
IPAPmin = EPAP
IPAPma x =
25 – 30 cmH
2
O
Allow a more
restrictive pressure
window to combine
IPAPmi n =
comfortable IPAP
IPAPma x =
IPAPmin + 5 cmH
2
O
Allow a more restrictive
pressure window to
combine comfor t and
safety.
IPAPmi n =
IPAPma x =
IPAPmin + 5 cmH
2
O
AVAP S
Rate
AVAPS Rate setting depends on patient needs and clinic al condition:
0.5 cmH
2
O/min to 3 cm H
2
O/min so target tidal volume is reached smoothly
3 cmH
2
O/min to 5 cm H
2
O/min so target tidal volume is reached
more rapidly
Check patient arterial blood gases (PaCO2and PaO2) and oxygen saturat ion (SpO2)
Hold entire unit at each end and slide release latch to the right.
with water.
not force door higher than the locked position.
(b) Grasp front of chamber and slide toward you.
2. (a) Gently push tab located inside round opening
on top of chamber. (b) Separate top and bottom
of chamber.
3. Fill bottom chamber with distilled water.
chamber.
Reassemble the chamber by placing the hinges on
lid over the t wo tabs on back of chamber. Close lid
until it locks in place with an audible click. Insert
setting will light up on the device. This icon lights up (different colors)
is being applied.
Orange - the heated tube is attached.
setting (0 to 5) and press. The settings represent relative humidity
levels that will remain consistent throughout the night even if room
temperature changes.
Quickstart Guide
Set-up Device
Start Therapy
StopTherapy
View Information
Comfort Settings
Menus
Guidelines forVentilation Set-up
Release latch
Bacteria filter*
Heated Tubing
(optional)
Air outlet port
Release latch
(bottom view)
Preheat Mode
*Note:
at end of tube when using the
optional heated tube.