Performance check device
screening tool
Performance check allows you to
quickly evaluate a therapy
device remotely. The check
operates the blower and screens
the device for any operation
errors. The screen then indicates
whether the device passed the
check (displays a green check
mark) or should be returned for
service (displays a red X).
A troubleshooting dashboard
can be populated from Encore
software, either via direct upload
to EncoreAnywhere (when a
modem is installed) or by asking
the patient to read the ve codes
o the Performance check screen
over the phone, and decode them
on Encore (Pro / Basic).
Sleep progress
The device provides summary
information about your patient’s
therapy use each time the therapy
is turned o. The screen displays
the “Three Night Summary.”
It shows your patient’s nightly
usage for the last 3 sleep sessions
(measured in 24 hour periods,
ending at noon each day). The
most recent session is displayed
in the right hand bar, labeled
with the number of hours slept.
A green bar indicates that your
patient slept more than 4 hours,
and a yellow bar indicates less
than 4 hours of use.
Additional features
F
Check mask t
This feature allows patients to
check mask t prior to starting
therapy.
The device will deliver a test
pressure while the screen counts
down 40 seconds. A green bar
indicates good t, while a red bar
indicates improvement is needed.
After the test, normal therapy will
start, and the screen will either
display a green checkmark or a
red “X”. The green checkmark
indicates that the leak found
allows for optimal performance of
the device. The red “X” indicates
that the leak may aect device
performance. However, the device
will remain functional and deliver
therapy.