1 Introduction
10
Electric Hazards
WARNING
Electric shock hazard: Do not open the monitor housing. Refer all servicing to qualified service
personnel.
• Always use the supplied power cord with the earthed mains plug to connect to an earthed AC
mains socket. Never adapt the mains plug from the fetal monitor to fit an unearthed AC mains
socket.
• Do not use AC mains extension cords or multiple portable socket-outlets.
• The protective earth conductor is required for EMC purposes. It has no protective function
against electric shock. Double and/or reinforced insulation protects this device against electric
shock.
• Do not use a device in the patient vicinity if it does not comply with IEC/EN 60601-1 A1: 2012.
The whole installation, including devices outside of the patient vicinity, must comply with IEC/
EN 60601-1 A1: 2012. Any non-medical device, including a PC running an OB TraceVue/
IntelliSpace Perinatal system, placed and operated in the patient's vicinity, must be powered via a
separating transformer (compliant with IEC/EN 60601-1 A1: 2012) that ensures mechanical
fixing of the power cords and covering of any unused power outlets.
• Do not connect any devices that are not supported as part of a system.
• Any non-medical device placed and operated in the patient’s vicinity must be powered via an
approved isolation transformer that ensures mechanical fixing of the power cords and covering of
any unused power outlets.
• Do not use USB devices with own power supplies unless an appropriate separation device is used
(either between USB interface and device or between device and power).
• Only the power cables provided with the system may be used. For reasons of safety, power (mains)
extension cables, or adapters shall not be used.
• Observe ESD (electrostatic discharge) precautions when working within the unit.
• The use of accessories, transducers, and cables other than those specified may result in increased
electromagnetic emissions, or decreased electromagnetic immunity of the device.
• The fetal/maternal monitor is NOT intended for use during defibrillation, electro-surgery, or
MRI. Remove all transducers, sensors, and accessories before performing electro-surgery,
defibrillation, or MRI, otherwise harm to the patient or the user can result.
• Do not touch the charging contacts for the cableless transducers at the Avalon base station while
you are touching the patient
Leakage currents: If several items of equipment used to monitor a patient are interconnected, the
resulting leakage current may exceed allowable limits.