PI P-736.ZRxS PInano User manual

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
P-736.ZRxS PInano® Piezo Scanner Systems

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 2 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Contents
About this Document 4
Symbols and Typographic Conventions ............................................................................................... 4
Figures.................................................................................................................................................. 4
Downloading Manuals ......................................................................................................................... 5
Other Applicable Documents............................................................................................................... 6
Safety 6
Intended Use........................................................................................................................................ 6
Safety Precautions ............................................................................................................................... 7
Electrical dangers...................................................................................................................................... 7
Mechanical dangers..................................................................................................................................8
Product Description 9
Model Overview................................................................................................................................... 9
Product View of P-736.ZRx................................................................................................................. 10
Product View of E-709........................................................................................................................ 10
Scope of Delivery................................................................................................................................ 11
Accessories......................................................................................................................................... 11
Unpacking 12
Installation 12
Connecting the P-736.ZRx to the Protective Earth Conductor .......................................................... 12
Mounting the P-736.ZRx onto a Surface............................................................................................ 14
Mounting the P-736.ZRx onto an M-545 Stage ................................................................................. 15
Mounting Holders for Petri Dishes, Microscope Slides and Accessories........................................... 16
Connecting the P-736.ZRx to the Controller ...................................................................................... 17
Start-Up and Operation 17
Starting-Up and Operating the P-736.ZRx.......................................................................................... 17
Discharging the P-736.ZRx ................................................................................................................. 17
Maintenance 18
Cleaning the P-736.ZRx ...................................................................................................................... 18

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 3 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Customer Service 18
Technical Data 19
Specifications ..................................................................................................................................... 19
Maximum Ratings .............................................................................................................................. 20
Ambient Conditions and Classifications............................................................................................. 20
Dimensions......................................................................................................................................... 21
P-736.ZRx PInano® Z Piezo Scanner........................................................................................................21
P-545.SH3 Microscope Slide Holder .......................................................................................................22
P-545.PD3 Petri Dish Holder................................................................................................................... 23
P-545.PP3 Universal Holding Plate .........................................................................................................24
P-736.AP1 Adapter Plate ........................................................................................................................ 24
Torque for Stainless Steel Screws (A2-70) ......................................................................................... 25
Old Equipment Disposal 25

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 4 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
About this Document
Symbols and Typographic Conventions
The following symbols and typographic conventions are used in this user manual:
CAUTION
Dangerous situation
If not avoided, the dangerous situation will result in minor injuries or damage to the
equipment.
Actions to take to avoid the situation.
NOTICE
Dangerous situation
If not avoided, the dangerous situation will result in damage to the equipment.
Actions to take to avoid the situation.
INFORMATION
Information for easier handling, tricks, tips, etc.
Symbol/Label Meaning
1.
2.
Action consisting of several steps whose sequential order must be
observed
Action consisting of one or several steps whose sequential order is
irrelevant
List item
p. 5
Cross-reference to page 5
RS-232
Labeling of an operating element on the product (example: socket of
the RS-232 interface)
Warning signs affixed to the product that refer to detailed
information in this user manual.
Figures
For better understandability, the colors, proportions and degree of detail in illustrations can
deviate from the actual circumstances. Photographic illustrations may also differ and must not
be seen as guaranteed properties.

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 5 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Downloading Manuals
INFORMATION
If a manual is missing or problems occur with downloading:
Contact our customer service department (p. 18).
INFORMATION
For products that are supplied with software (CD in the scope of delivery), access to the
manuals is protected by a password. Protected manuals are only displayed on the website after
entering the password.
The password is included on the CD of the product.
For products with CD: Identify the password
1. Insert the product CD into the PC drive.
2. Switch to the Manuals directory on the CD.
3. In the Manuals directory, open the Release News (file including releasenews in the file
name).
4. Find the user name and the password in the section "User login for software download"
in the Release News.
Downloading manuals
1. Open the website www.pi.ws.
2. If access to the manuals is protected by a password:
a) Click Login.
b) Log in with the user name and password.
3. Click Search.
4. Enter the product number up to the period (e.g., P-882) or the product family (e.g.,
PICMA® Bender) into the search field.
5. Click Start search or press the Enter key.
6. Open the corresponding product detail page in the list of search results:
a) If necessary: Scroll down the list.
b) If necessary: Click Load more results at the end of the list.
c) Click the corresponding product in the list.
7. Scroll down to the Downloads section on the product detail page.
The manuals are displayed under Documentation.
8. Click the desired manual and save it to the hard disk of your PC or to a data storage
medium.

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 6 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Other Applicable Documents
The devices which are mentioned in this user manual are described in their own manuals.
Product
Document
E-709 digital piezo controller PZ222E user manual
Safety
Intended Use
The P-736.ZRx PInano® Z piezo scanners and the E-709.PRG digital piezo controllers are
laboratory devices as defined by DIN EN 61010-1. They are intended to be used in interior
spaces and in an environment which is free of dirt, oil and lubricants.
Based on their design and realization, the Z piezo scanners are intended to position in
Z direction. They are designed to position a sample holder with suitable dimensions (e.g.,
M-687.AP1) that can carry an microscope slide or a Petri dish.
The Z piezo scanners can carry a maximum load of 0.5 kg. They are intended to be used with
M-545 microscope stages or with other suitable microscope stages.
The Z piezo scanners are to be mounted horizontally.
The intended use of the Z piezo scanner is only possible in combination with the E-709.PRG
digital piezo controller. The E-709.PRG controller provides the required operating voltages. To
ensure proper performance of the servo-control system, the controller is able to read out and
process the signals from the piezoresistive sensors.

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 7 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Safety Precautions
Electrical dangers
CAUTION
Dangerous voltage and residual charge on piezo actuators!
The P-736.ZRx is driven by piezo actuators. Temperature changes and compressive stresses can
induce charges in piezo actuators. After being disconnected from the electronics, piezo
actuators can also stay charged for several hours. Touching or short-circuiting the contacts in
the connector of the P-736.ZRx can lead to minor injuries from electric shock. In addition, the
piezo actuators can be destroyed by an abrupt contraction.
Do notopen the P-736.ZRx.
Discharge the piezo actuators before installation:
Connect the P-736.ZRx to the switched-off PI controller.
Do not pull the connector out of the controller during operation.
For piezo scanners with sub-D connector:
Touching the contacts in the connector can lead to an electric shock (max. 130 V DC) and minor
injuries.
Do not touch the contacts in the connector.
Secure the connector of the piezo scanner with screws against being pulled out of the
controller.
CAUTION
Risk of electric shock if the protective earth conductor is not connected!
If a protective earth conductor is not or not properly connected, dangerous touch voltages can
occur and there is a risk of electric shock. In the event of malfunction or failure of the system,
touching the P-736.ZRx can result in minor injuries.
Connect the P-736.ZRx to a protective earth conductor (p. 12) before start-up.
Do not remove the protective earth conductor during operation.
If the protective earth conductor has to be temporarily removed (e.g., for modifications),
reconnect the P-736.ZRx to the protective earth conductor before starting it up again.
NOTICE
Destruction of the piezo actuators by electric flashovers!
The use of the P-736.ZRx in environments that increase the electrical conductivity can lead to
the destruction of the piezo actuators by electric flashovers. Electric flashovers can be caused
by moisture, high humidity, liquids and conductive materials such as metal dust. In addition,
electric flashovers can also occur in certain air pressure ranges due to the increased
conductivity of the air.
Avoid operating the P-736.ZRx in environments that can increase the electric conductivity.
Only operate the P-736.ZRx within the permissible ambient conditions and classifications
(p. 20).

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 8 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
NOTICE
Destruction of the piezo actuators by continuously high voltage!
The constant application of high voltage to piezo actuators can lead to leakage currents and
flashovers that destroy the ceramic.
If the P-736.ZR
x is not used, but the controller is to remain switched on to ensure temperature
stability:
Set the piezo voltage to 0 V on the controller.
NOTICE
Unsuitable cables!
Unsuitable cables can damage the controller.
Only use cables from PI for connecting the P-736.ZRx to the controller.
INFORMATION
Extended cables can affect the performance of the P-736.ZRx.
Do not use cable extensions. If you need longer cables, contact our customer service
department (p. 18).
Mechanical dangers
NOTICE
Mechanical overload due to high torque and high loads!
High torque and high loads can overload the motion platform of the P-736.ZRx. Mechanical
overload can cause damage to the piezo actuators, sensors, and flexures of the P-736.ZRx and
lead to loss in accuracy.
Only hold the P-736.ZRx externally by the base body.
Do not exceed a push force of 50 N on the motion platform.
Do not exceed a pull force of 30 N on the motion platform.
Avoid torques >0.5 Nm on the motion platform.
Pay attention to the torque range (p. 25) given for the screws used during installation.
Do not exceed the maximum permissible stress and load capacities according to the
specifications (p. 19).
NOTICE
Screws that are too long!
The P-736.ZRx can be damaged by screws that are too long.
Note the depth of the mounting holes in the motion platform (p. 21).
Only use screws of the correct length for the respective mounting holes.

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 9 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
NOTICE
Lubricants, dirt, condensation!
Dirt, oil, lubricants and condensation will render the motor/drive inoperable.
Ensure that the piezo actuators of the P-736.ZRx do not come into contact with lubricants.
Keep the P-736.ZRx free from dirt and condensation.
NOTICE
Warping of the P-736.ZRx due to mounting onto uneven surfaces!
Mounting the P-736.ZRx onto an uneven surface can warp the P-
736.ZRx. Warping reduces the
accuracy.
Mount the P-736.ZRx onto an even surface. The recommended evenness of the surface is
≤10 µm.
For applications with large temperature changes: Only mount the P-736.ZRx onto surfaces
that have the same or similar thermal expansion properties as the <Produktname> (e.g.,
surfaces made of aluminum).
NOTICE
Uncontrolled oscillation!
Oscillations can cause irreparable damage to the P-736.ZRx. Oscillations are indicated by a
humming and can result from the following causes:
The load and/or dynamics of operation differ too much from the calibration settings.
The P-736.ZRx is operated near its resonant frequency.
If you notice oscillations, stop the P-736.ZRx immediately.
Product Description
Model Overview
The following PInano® piezo scanner systems are available as combinations of Z piezo scanner
and digital piezo controller:
System Z piezo scanner Digital piezo controller
P-736.ZR1S P-736.ZR1
PInano® Z piezo slide scanner, 100 µm,
clear aperture for microscope slides,
piezoresistive sensors
E-709.PRG
Digital piezo controller, 1 channel, -30
to 130 V, piezoresistive sensor, bench-
top
P-736.ZR2S
P-736.ZR2
PInano® Z piezo slide scanner, 200 µm,
clear aperture for microscope slides,
piezoresistive sensors
E-709.PRG
Digital piezo controller, 1 channel, -30
to 130 V, piezoresistive sensor, bench-
top

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 10 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Product View of P-736.ZRx
Figure 1: Example view of P-736.ZRx
1Motion platform
2Sample holder mounting plane of the motion platform
3Protective earth connection
4Cable exit
5Base body
Z arrow: Positive direction of motion
Product View of E-709
Figure 2: Example view of E-709 (see the E-709 user manual for details)

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 11 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Scope of Delivery
A PInano® piezo scanner system is delivered with the following components:
Product number
Description
P-736.ZRx PInano® Z piezo slide scanner according to order (p. 9)
E-709.PRG Digital piezo controller (p. 9)
C-501.24050H Wide-range-input power supply, 24 V / 50 W
3763 Power cord
000036360 USB cable (3 m, USB-A (m)/USB Mini-B (m)) for PC connection
E709B0002
RS-232 adapter HD sub-D 26 (m) to sub-D 9 (m) for PC connection via
RS-232
C-815.34
Null-modem cable for PC connection via RS-232
E-709.04 Adapter cable for analog I/O (1 m, HD sub-D 26 / 2 × BNC)
Documentation, consisting of:
PZ240EK Short instructions for piezo positioning systems
PZ222E User manual for E-709 digital piezo controllers
E-709.CD
CD with PC software and documentation of the E-709 digital piezo
controller
A000T0032
Technical note on updating the PI Software using the PI Update Finder
Accessories
Product number Description
P-545.PD3 Petri dish holder, 35 mm, for PInano® piezo stages
P-545.SH3
Microscope slide holder for PInano® piezo stages
P-545.C18
Coverslip holder with opening for 18 mm × 18 mm coverslips, suitable for
P-545.SH3
P-545.C22
Coverslip holder with opening for 22 mm × 22 mm coverslips, suitable for
P-545.SH3
P-545.C25 Coverslip holder with opening for 25 mm × 25 mm coverslips, suitable for
P-545.SH3
P-545.PP3 Universal holding plate for PInano® piezo stage accessories
P-736.AP1 Adapter plate, P-736 PInano® Z piezo slide scanner to M-545 XY stage
To order, contact our customer service department (p. 18).

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 12 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Unpacking
NOTICE
Mechanical overload due to incorrect handling!
An impermissible mechanical load on the motion platform of the P-736.ZRx can cause damage
to the piezo actuators, sensors, and flexures of the P-736.ZRx as well as loss of accuracy.
Only hold the P-736.ZRx externally by the base body.
1. Unpack the P-736.ZRx with care.
2. Compare the contents with the items listed in the contract and the packing list.
3. Inspect the contents for signs of damage. If parts are missing or you notice signs of
damage, contact PI immediately.
4. Keep all packaging materials in case the product needs to be returned.
Installation
Connecting the P-736.ZRx to the Protective Earth Conductor
INFORMATION
In the case of P-736.ZRx stages with sub-D connectors, ground loops can occur when the stage is
grounded via its protective earth connection or the mounting holes as well as by the shield of the
connecting cable for the controller.
If a ground loop occurs, contact our customer service department (p. 18).
INFORMATION
Observe the applicable standards for mounting the protective earth conductor.
The P-736.ZRx features a protective earth connection (p. 10) which is marked with the symbol
for the protective earth conductor .
Requirements
You have read and understood the safety precautions (p. 7).
The P-736.ZRx is not connected to the controller.

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 13 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Tools and accessories
Suitable protective earth conductor: Cross-sectional area of the cable ≥0.75 mm2
M4 screw which is preassembled on the protective earth connection of the P-736.ZRx
Alternatively, if there is enough installation space:
Other electrically conductive M4 screw of suitable length
Two flat washers
Two safety washers
Suitable screwdriver
Figure 3: Connecting the protective earth conductor (profile view)
1
Base body of the P-736.ZRx
2
Flat washer
3
Safety washer
4
Screw
5
Cable lug
6
Protective earth conductor
Connecting the P-736.ZRx to the protective earth conductor
1. If necessary, fasten a suitable cable lug to the protective earth conductor.
2. Use the M4 screw to affix the cable lug of the protective earth conductor to the
protective earth connection of the P-736.ZRx as shown in the profile view. If possible,
use suitable washers and self-locking washers.
3. Tighten the M4 screw with a torque of 1.2 Nm to 1.5 Nm.
4. Make sure that the contact resistance at all connection points relevant for mounting
the protective earth conductor is <0.1 Ω at 25 A.

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 14 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Mounting the P-736.ZRx onto a Surface
Figure 4: Mounting holes in the base body of the P-736.ZRx
Requirements
You have read and understood the safety precautions (p. 7).
The P-736.ZRx is not connected to the controller.
You have provided a suitable surface:
-Four suitable M3 holes are present. See “Dimensions” (p. 21) for details.
-The evenness of the surface is ≤10 µm.
-For applications with large temperature changes: The surface should have the
same thermal expansion properties as the P-736.ZRx (e.g., surface made of
aluminum).
You have accounted for the space required to route cables without bending and
according to regulations
Tools and accessories
Four M3x10 screws (included with P-736.ZRx)
Suitable screwdriver
Mounting the P-736.ZRx onto a surface
1. Align the P-736.ZRx on the surface so that the mounting holes in the P-736.ZRx (see
arrows in the figure) overlap with the the corresponding holes in the surface.
2. Insert four M3 screws into the mounting holes in the base body of the P-736.ZRx.
3. Tighten each of the four screws with a torque between 0.8 Nm and 1.1 Nm.
4. Check that the P-736.ZRx is affixed firmly to the surface.

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 15 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Mounting the P-736.ZRx onto an M-545 Stage
Figure 5: Mounting of the P-736.ZRx onto an M-545 stage using the P-736.AP1 adapter plate
Requirements
You have read and understood the safety precautions (p. 7).
The P-736.ZRx is not connected to the controller.
Tools and accessories
P-736.AP1 adapter plate (see “Accessories”, p. 11)
Four M6x8 pan head screws (included with P-736.AP1)
Four M3x10 socket head cap screws (included with P-736.ZRx)
Suitable screwdrivers
Mounting the P-736.ZRx onto an M-545 stage
Mount the P-736.ZRx onto the M-545 stage as depicted in the figure.

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 16 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Mounting Holders for Petri Dishes, Microscope Slides and
Accessories
Figure 6: M2 threaded holes in the sample holder mounting plane (p. 10) of the motion platform
Requirements
You have read and understood the safety precautions (p. 7).
The P-736.ZRx is not connected to the controller.
Tools and accessories
P-545.PD3, P-545.SH3 or P-545.PP3 holder (p. 11)
Four M2 screws of suitable length (included with the P-545.xx3 holder)
Suitable screwdriver
Mounting a holder
1. Align the holder so that the holding clamps (P-545.PD3 and P-545.SH3) are pointing
upwards or the larger surface of the base body (P-545.PP3) is pointing upwards. See
"Dimensions" (p. 22ff.) for details.
2. Place the holder into the aperture of the P-736.ZRx from above, so that the mounting
holes in the holder and the sample holder mounting plane (see Figure 6 above) of the
P-736.ZRx overlap.
3. Mount the holder with the M2 screws to the provided M2 threaded holes (see Figure 6
above). Pay attention to the specified torque range (p. 22) while doing so.

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 17 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Connecting the P-736.ZRx to the Controller
Requirements
You have read and understood the safety precautions (p. 7).
The controller is switched off, i.e., not connected to the power source.
You have connected the P-736.ZRx to the protective earth conductor (p. 12).
Connecting the P-736.ZRx to the controller
Connect the sub-D 9-pin connector of the P-736.ZRx to the corresponding sub-D 9-pin
socket of the controller.
Start-Up and Operation
Starting-Up and Operating the P-736.ZRx
INFORMATION
Systems are calibrated at the factory to achieve optimum performance. Replacing the system
components will cause a loss in performance.
Note the assignment of the stage axes to the controller channels which is given by the
calibration label of the piezo servo controller.
If the piezo servo controller or the stage has to be replaced, recalibrate the axis
displacement (see controller manual) or contact our customer service department (p. 18).
Requirements
You have read and understood the safety precautions (p. 7).
You have connected the P-736.ZRx to the protective earth conductor (p. 12) and to the
controller (p. 17).
Starting up and operating the P-736.ZRx
Follow the instructions in the user manual of the piezo controller used for start-up and
operation of the P-736.ZRx.
Discharging the P-736.ZRx
The P-736.ZRx must be discharged before demounting. Demounting is necessary e.g., before
cleaning or transporting the P-736.ZRx as well as for modifications.
Requirements
You have read and understood the safety precautions (p. 7).

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 18 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Discharging a P-736.ZRx connected to the controller
In closed-loop operation:
1. Switch off the servo mode on the controller.
2. Set the piezo voltage to 0 V on the controller.
In open-loop operation:
Set the piezo voltage to 0 V on the controller.
Discharging a P-736.ZRx not connected to the controller
Connect the P-736.ZRx to the switched-off PI controller for 10 seconds.
Maintenance
NOTICE
Damage due to improper maintenance!
The P-736.ZRx is maintenance-free and precisely aligned.
Only loosen screws according to the instructions in this manual.
Do not open the P-736.ZRx.
Cleaning the P-736.ZRx
Requirements
You have discharged (p. 17) the piezo actuators of the P-736.ZRx.
You have disconnected the P-736.ZRx from the controller.
Cleaning the P-736.ZRx
Clean the surface of the P-736.ZRx with a cloth that is lightly dampened with a mild
cleanser or disinfectant (e.g., alcohol or isopropyl alcohol).
Do not do any ultrasonic cleaning.
Customer Service
For inquiries and orders, contact your PI sales engineer or send us an email (info@pi.ws).
If you have questions concerning your system, have the following information ready:
Product and serial numbers of all products in the system
Firmware version of the controller (if present)
Version of the driver or the software (if present)
Operating system on the PC (if present)
If possible: Take photographs or make videos of your system that can be sent to our
customer service department if requested.
The latest versions of the user manuals are available for download on our website (www.pi.ws).

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 19 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Technical Data
Specifications
P-736.ZR1S
P-736.ZR2S
Unit
Tolerance
Active axes
Z
Z
Motion and positioning
Integrated sensor
Piezoresistive
Piezoresistive
Closed-loop travel
100
200
µm
Open-loop resolution
0.2
0.4
nm
typ.
Closed-loop resolution
0.4
0.7
nm
typ.
Mechanical properties
Settling time (10 % step width)
5
7
ms
Load capacity
500
500
g
max.
Drive properties
Piezo ceramic
PICMA® P-885
PICMA® P-885
Miscellaneous
Operating temperature range
15 to 40
15 to 40
°C
Material
Aluminum
Aluminum
Mass
550
550
g
±5 %
Cable length
1.5
1.5
m
±10 mm
Piezo controller
E-709 digital servo (included in delivery)
Communication interfaces
USB, RS-232, SPI
I/O connector
HD Sub-D 26-pin
1× analog input 0 to 10 V
1× sensor monitor 0 to 10 V
1× digital input (LVTTL, programmable)
1× analog output
5× digital outputs (LVTTL, 3× predefined, 2× programmable)
Command set
PI General Command Set (GCS)
User software
PIMikroMove
Software drivers
LabVIEW driver, dynamic libraries for Windows and Linux
Supported functionality
Wave generator, data recorder, auto zero, trigger I/O,
MATLAB, MetaMorph, µManager
Controller dimensions
160 mm × 96 mm × 33 mm

User Manual
P736T0002, valid for P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2017-06-21
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 20 / 25
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws
Maximum Ratings
P-736.ZRx piezo scanners are designed for the following maximum ratings:
Model Maximum operating
voltage
Maximum operating
frequency (unloaded)
Maximum power
consumption
P-736.ZR1
-30 to 130 V
100 Hz
20 W
P-736.ZR2 -30 to 130 V 100 Hz 20 W
Ambient Conditions and Classifications
The following ambient conditions and classifications must be observed for the P-736.ZRx:
Area of application For indoor use only
Maximum altitude 2000 m
Air pressure
1100 hPa to 0.1 hPa
Relative humidity Highest relative humidity 80 % for temperatures up to 31 °C
Decreasing linearly to 50 % relative humidity at 40 °C
Operating temperature*
15° C to 40 °C
Storage temperature 0°C to 70°C
Transport temperature -25 °C to 85 °C
Overvoltage category
II
Protection class I
Degree of pollution
1
Degree of protection
according to IEC 60529
IP20
*Specifications assured from 17 °C to 23 °C, performance may be reduced outside this range.
Other manuals for P-736.ZRxS PInano
1
Table of contents
Other PI Scanner manuals